K970629 · Suntech Medical Instruments, Inc. · DXN · Jan 2, 1998 · Cardiovascular
Device Facts
Record ID
K970629
Device Name
TANGO NON-INVASIVE BLOOD PRESSURE MONITOR
Applicant
Suntech Medical Instruments, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Jan 2, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Intended Use
The Tango Blood Pressure Monitor is intended for use as an aid or adjunct to ECG exercise stress testing, measuring and displaying an adult patients systolic and diastolic blood pressures. The Tango Monitor has two operating modes, Dimensional K-Sound (DKA) and Oscillometry. In the Oscillometric mode the Tango NIBP monitor measures: 1. Systolic Pressure (mmHg) 2. Diastolic Pressure (mmHg) 3. Pulse Rate (beats per minute) In the Dimensional K-Sound mode (DKA) the Tango NIBP monitor measures: 1. Systolic Pressure (mmHg) 2. Diastolic Pressure (mmHg) 3. Heart Rate (beats per minute) Depending on the mode the Monitor is in one of the following waveforms are displayed: 1. R-Wave trigger signal (DKA) 2. K-sound waveforms (DKA) 3. or Oscillometric pressure pulses (Oscillometric) Intended Patient Population The Tango NIBP monitor is only intended for use as a stress test monitor. The unit is not intended for use with children, neonates or geriatrics. The Tango monitor is limited to adult use, the age range for the unit is 18 and older. Arm sizes are limited to the following cuff sizes: - Small Adult: 18cm - 26 cm - Adult: 25 cm - 35 cm - Large Adult: 33 cm - 47 cm Intended Operators and Environment The Tango NIBP monitor is intended to be operated by a nurse, physician or clinician for stress/exercise tests. The Tango is only intended to be used in a hospital, clinic and stress test lab. There are no circumstances where the Tango NIBP monitor would be operated by a patient or outside of hospital, clinic or stress test lab. The Tango has been tested and meets the environmental operating conditions required by AAMI SP10. Refer to Appendix B1 for the AAMI SP10 test results.
Device Story
Non-invasive blood pressure monitor for ECG exercise stress testing; inputs include cuff pressure and ECG R-wave trigger signals. Operates in two modes: Dimensional K-Sound (DKA) and Oscillometry. DKA mode utilizes R-wave gating to isolate Korotkoff sounds during exercise; Oscillometric mode measures pressure pulses. Outputs include systolic/diastolic BP, heart/pulse rate, and waveforms (R-wave, K-sound, or oscillometric pulses). Operated by clinicians/nurses in hospitals or stress test labs. Provides real-time hemodynamic data to assist clinicians during stress testing; aids in monitoring patient response to exercise.
Clinical Evidence
Bench testing only. Device performance validated against AAMI SP10 environmental and accuracy requirements for non-invasive blood pressure monitors.
Technological Characteristics
Dual-mode NIBP monitor: Oscillometric and Dimensional K-Sound (DKA). DKA mode uses R-wave trigger signal for noise-gated Korotkoff sound detection. Compatible with specific cuff sizes (18-47 cm). Meets AAMI SP10 standards for environmental operating conditions.
Indications for Use
Indicated for adult patients (18+ years) undergoing ECG exercise stress testing in clinical/hospital settings to measure systolic/diastolic blood pressure and heart/pulse rate. Not for use on children, neonates, or geriatric patients.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Related Devices
K122401 — TANGO M2 BLOOD PRESSURE MONITOR · Suntech Medical, Inc. · Nov 2, 2012
K053209 — TANGO+ AUTOMATIC BLOOD PRESSURE AND OXYGEN SATURATION MEASUREMENT SYSTEM · Suntech Medical, Inc. · Jan 5, 2006
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Rockville MD 20857
JAN - 2 1998
Mr. David Gallick
Quality Assurance Manager
SunTech Medical Instruments, Inc.
8917 Glenwood Avenue
Raleigh, NC 27612
Re: K970629
Tango Non-Invasive Blood Pressure Monitor
Regulatory Class: II (Two)
Product Code: DXN
Dated: November 3, 1997
Received: November 4, 1997
Dear Mr. David Gallick:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}
Page 2 - Mr. David Gallick
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}
510(k) Number (if known): K970629
Device Name: Tango Non-invasive Blood Pressure Monitor
Indications For Use:
See Attached Sheet
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970629
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
{3}
Indications for Use: SunTech Medical Instruments
Tango Non-invasive Stress BP Test Monitor JAN - 1 1998
The Tango Blood Pressure Monitor is intended for use as an aid or adjunct to ECG exercise stress testing, measuring and displaying an adult patients systolic and diastolic blood pressures.
The Tango Monitor has two operating modes, Dimensional K-Sound (DKA) and Oscillometry.
In the *Oscillometric mode* the Tango NIBP monitor measures:
1. Systolic Pressure (mmHg)
2. Diastolic Pressure (mmHg)
3. Pulse Rate (beats per minute)
In the *Dimensional K-Sound mode (DKA)* the Tango NIBP monitor measures:
1. Systolic Pressure (mmHg)
2. Diastolic Pressure (mmHg)
3. Heart Rate (beats per minute)
Depending on the mode the Monitor is in one of the following waveforms are displayed:
1. R-Wave trigger signal (DKA)
2. K-sound waveforms (DKA)
3. or Oscillometric pressure pulses (Oscillometric)
### Intended Patient Population
The Tango NIBP monitor is only intended for use as a stress test monitor. The unit is not intended for use with children, neonates or geriatrics. The Tango monitor is limited to adult use, the age range for the unit is 18 and older. Arm sizes are limited to the following cuff sizes:
- Small Adult: 18cm - 26 cm
- Adult: 25 cm - 35 cm
- Large Adult: 33 cm - 47 cm
### Intended Operators and Environment
The Tango NIBP monitor is intended to be operated by a nurse, physician or clinician for stress/exercise tests. The Tango is only intended to be used in a hospital, clinic and stress test lab. There are no circumstances where the Tango NIBP monitor would be operated by a patient or outside of hospital, clinic or stress test lab.
The Tango has been tested and meets the environmental operating conditions required by AAMI SP10. Refer to Appendix B1 for the AAMI SP10 test results.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.