ELECTRONIC BLOOD PRESSURE MONITOR
K102920 · Shenzhen Pango Electronic Co., Ltd. · DXN · Oct 15, 2010 · Cardiovascular
Device Facts
| Record ID | K102920 |
| Device Name | ELECTRONIC BLOOD PRESSURE MONITOR |
| Applicant | Shenzhen Pango Electronic Co., Ltd. |
| Product Code | DXN · Cardiovascular |
| Decision Date | Oct 15, 2010 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1130 |
| Device Class | Class 2 |
| Attributes | 3rd-Party Reviewed |
Intended Use
PG-800A Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist. It can be used at medical facilities or at home. The intended wrist circumference is 13.5-19.5 cm.
Device Story
PG-800A is a battery-powered, automatic, non-invasive wrist blood pressure monitor. Device uses oscillometric technique to measure systolic/diastolic blood pressure and pulse rate. Intended for adult use in clinical or home settings. User wraps inflatable cuff around wrist; device automatically inflates/deflates to capture pressure oscillations. Output displayed on screen; includes data storage and WHO blood pressure classification indicator. Healthcare providers or patients use output to monitor blood pressure trends and manage hypertension. Benefits include convenient, portable, non-invasive blood pressure tracking.
Clinical Evidence
Performance and clinical verification testing conducted per ANSI/AAMI SP10:2002+A1:2003+A2:2006. Biocompatibility testing performed per ISO 10993-5:2006 and ISO 10993-10:2002+AMD:2006. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
Battery-powered, automatic oscillometric wrist monitor. Materials: Nylon (cuff) and ABS (housing). Dimensions accommodate 13.5-19.5 cm wrist circumference. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ANSI/AAMI SP10 (performance), ISO 10993 (biocompatibility).
Indications for Use
Indicated for adult patients with wrist circumference 13.5-19.5 cm for non-invasive measurement of systolic/diastolic blood pressure and pulse rate via oscillometric technique. Suitable for use in medical facilities or home settings.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
- KD-795 Blood Pressure Monitor (K070826)
Related Devices
- K161845 — Electronic Blood Pressure Monitor · Shenzhen Pango Electronic Co., Ltd. · Jul 31, 2017
- K141683 — ELECTRONIC BLOOD PRESSURE MONITOR · Shenzhen Urion Technology Co., Ltd. · Apr 3, 2015
- K213503 — Wrist Blood Pressure Monitor, models AOJ-35A, AOJ-35B, AOJ-35C, AOJ-35D, AOJ-35E · Shenzhen AOJ Medical Technology Co., Ltd. · Feb 18, 2022
- K172972 — Wrist Type Automatic Blood Pressure Monitor · Shenzhen Jiacom Technology CO ., Ltd. · Mar 2, 2018
- K191180 — Electronic Blood Pressure Monitor · Shenzhen AOJ Medical Technology Co., Ltd. · Aug 23, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
## K102920
OCT 1 5 2010
## PANGAO®
·
Premarket Notification promission Promission
Traditional Sec. 510(L) Submission
PG-800A Ellectronic Blood Pressure Monitor
Sec. III 510(k) Summary
OCT 15 2016
IH-1
# Sec. III 510(k) Summary
| This 510(k) Summary is prepared per the request of 21 CFR 807.92. The Assigned 510(k) Number is: | |
|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Date | 20 JUL 2010 |
| Sponsor | Shenzhen Pango Electronic Co., Ltd |
| | No.25, 1st Industrial Park, Fenghuang Road, Xikeng Village, Henggang Town, |
| | Longgang District, Shenzhen, Guangdong, 518115, China |
| | Contact Person: Ms. Xiaoyun Yang, Vice General Manager |
| | T: +86-755-33825988 F: +86-755-33825989 E: sales@pan-go.com |
| Submission | MS. Diana Hong / MR. Lee Fu |
| Correspondent | Shanghai Mid-Link Consulting Co., Ltd |
| | P.O.BOX 237-023, Shanghai, 200237, China |
| Proposed Device | Electronic Blood Pressure Monitor, PG-800A |
| Classification | System, Measurement, Blood-pressure, Non-invasive DXN 21 CFR 870.1130 Class II |
| Intended Use | PG-800A Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood |
| | pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an |
| | inflatable cuff is wrapped around the wrist. It can be used at medical facilities or at home. The |
| | intended wrist circumference is 13.5-19.5 cm. |
| Device Description | The proposed device, PG-800A Electronic Blood Pressure Monitor, is a battery driven automatic |
| | non-invasive blood pressure monitor. It can measure systolic and diastolic blood pressure as well as |
| | the pulse rate of adult person at wrist within its claimed range and accuracy via the oscillometric |
| | technique. The device has the data storage function. It has an bar indicating function, which can |
| | indicate the WHO (World Health Organization) Blood Pressure Classification of the measured blood |
| | pressure by referencing Diastolic Blood Pressure issued at Journal of Hypertension 1999. Vol 17, No.2. |
| Testing | Electric Safety per IEC 60601-1:1988+A1:1991+A2:1995 / EMC per IEC 60601-1-2:2001+A1:2004 |
| | Performance and Clinical Verification Test per ANSI/AAMI SP10:2002+A1:2003+A2:2006 |
| | Biocompatibility per ISO 10993-5:2006 / ISO 10993-10:2002+AMD:2006 |
| Predicate Device | KD-795 Blood Pressure Monitor as cleared in K070826 |
| SE Conclusion | The proposed device, PG-800A Electronic Blood Pressure Monitor, measures the blood pressure via |
| | same principle, oscillometric, at the same site, wrist, as the predicate device. Both of them are |
| | driven by batteries. The patient contact materials of both are Nylon and ABS. Therefore, the |
| | proposed device, PG-800A Electronic Blood Pressure Monitor, is claimed to be Substantially |
| | Equivalent (SE) to the predicate device, KD795 Blood Pressure Monitor as cleared in K070826, in |
| | aspect of safety and effectiveness |
Shenzhen Pango Electronic Co., Ltd
Report SN: SHA01420100715-01
i
:
.
्
.
. •
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes above a wavy line, representing water. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shenzhen Pango Electronic Co., Ltd. c/o Mr. Ned Devine Underwriters Laboratories Inc. 333 Pfingsten Road Northbrook, IL 60062
OCT 1 5 2010
Re: K102920
Trade/Device Name: Electronic Blood Pressure Monitor, Model PG-800A Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: 74 DXN Dated: September 20, 2010 Received: October 1, 2010
Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Ned Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### PANGAO®
Premarket Notification Traditional Sec. 510(k) Submission PG-800A Electronic Blood Pressure Monitor Sec. II Indication for Use
1 5 2010 OCT
:
Sec. II Indication for Use Statement
510(k) Number:
Device Name: PG-800A Electronic Blood Pressure Monitor
Indications for Use:
PG-800A Electronic Blood Pressure Monitor is intended to measure the systelic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist.
It can be used at medical facilities or at home.
The intended wrist circumference is 13.5-19.5 cm.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
Page _ I_ of
510(k) Number k102920
Shenzhen Pango Electronic Co., Ltd Report SN: SHA01420100715-01
II-1
ﻋﻠﻰ ﺍ --ﺔ ﺍﻟﻘ