Balloon Aortic Valvuloplasty

Page Type
Product Code
Definition
A balloon aortic valvuloplasty catheter is indicated for balloon valvuloplasty of the aortic valve.
Physical State
Generally, a balloon aortic valvuloplasty catheter is a coaxial catheter comprised of outer and inner layers. The catheter features a molded proximal end bifurcated with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon made of polyamide. Upon reaching a specified pressure, the waist will expand to the rated balloon diameter and dilate the valve to the rated diameter.
Technical Method
The catheter is inserted into the body and directed towards the aortic valve. Once the catheter is placed in the valve, the balloon is expanded which compresses and/or breaks the material causing stenosis.
Target Area
The aortic valve of the heart.
Regulation Medical Specialty
Cardiovascular
Review Panel
Cardiovascular
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
870.1255
GMP Exempt?
No
Summary Malfunction Reporting
Eligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 870.1255 Balloon aortic valvuloplasty catheter

§ 870.1255 Balloon aortic valvuloplasty catheter.

(a) Identification. A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) The device must be demonstrated to be biocompatible.

(2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components.

(3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal.

(4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis.

(5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.

[82 FR 34852, July 27, 2017]

Balloon Aortic Valvuloplasty

Page Type
Product Code
Definition
A balloon aortic valvuloplasty catheter is indicated for balloon valvuloplasty of the aortic valve.
Physical State
Generally, a balloon aortic valvuloplasty catheter is a coaxial catheter comprised of outer and inner layers. The catheter features a molded proximal end bifurcated with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon made of polyamide. Upon reaching a specified pressure, the waist will expand to the rated balloon diameter and dilate the valve to the rated diameter.
Technical Method
The catheter is inserted into the body and directed towards the aortic valve. Once the catheter is placed in the valve, the balloon is expanded which compresses and/or breaks the material causing stenosis.
Target Area
The aortic valve of the heart.
Regulation Medical Specialty
Cardiovascular
Review Panel
Cardiovascular
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
870.1255
GMP Exempt?
No
Summary Malfunction Reporting
Eligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 870.1255 Balloon aortic valvuloplasty catheter

§ 870.1255 Balloon aortic valvuloplasty catheter.

(a) Identification. A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) The device must be demonstrated to be biocompatible.

(2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components.

(3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal.

(4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis.

(5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.

[82 FR 34852, July 27, 2017]