- Subpart B—Cardiovascular Diagnostic DevicesCFR Sub-Part
- DQDStethoscope, Electronic2Product Code
- DQECatheter, Oximeter, Fiber-Optic2Product Code
- DQGMicrosphere, Trace3Product Code
- DQKComputer, Diagnostic, Programmable2Product Code
- DQOCatheter, Intravascular, Diagnostic2Product Code
- DQPKit, Balloon Repair, Catheter3Product Code
- DQRCannula, Catheter2Product Code
- DQTOccluder, Catheter Tip2Product Code
- DQXWire, Guide, Catheter2Product Code
- DQYCatheter, Percutaneous2Product Code
- DRACatheter, Steerable2Product Code
- DRBStylet, Catheter2Product Code
- DRCTrocar2Product Code
- DREDilator, Vessel, For Percutaneous Catheterization2Product Code
- DRFCatheter, Electrode Recording, Or Probe, Electrode Recording2Product Code
- DSIDetector And Alarm, Arrhythmia2Product Code
- DSJAlarm, Blood-Pressure2Product Code
- DSKComputer, Blood-Pressure2Product Code
- DXGComputer, Diagnostic, Pre-Programmed, Single-Function2Product Code
- DXLInjector, Indicator2Product Code
- DXMDensitometer2Product Code
- DXNSystem, Measurement, Blood-Pressure, Non-Invasive2Product Code
- DXQBlood Pressure Cuff2Product Code
- DXWSystem, Phonocatheter, Intracavitary2Product Code
- DXXSystem, Catheter Control, Steerable2Product Code
- DYBIntroducer, Catheter2Product Code
- DYGCatheter, Flow Directed2Product Code
- JOQGenerator, Pulse, Pacemaker, External Programmable (For Electrophysiological Studies Only)2Product Code
- KRACatheter, Continuous Flush2Product Code
- KRBProbe, Thermodilution2Product Code
- KRKManometer, Blood-Pressure, Venous2Product Code
- LDEStethoscope, Manual1Product Code
- LITCatheter, Angioplasty, Peripheral, Transluminal2Product Code
- MAVSyringe, Balloon Inflation2Product Code
- MHXMonitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2Product Code
- MLDMonitor, St Segment With Alarm2Product Code
- MTDCatheter, Intracardiac Mapping, High-Density Array2Product Code
- MXDRecorder, Event, Implantable Cardiac, (With Arrhythmia Detection)2Product Code
- NFATemporary Coronary Saphenous Vein Bypass Graft For Embolic Protection2Product Code
- NKQGuidewire, Catheter, Reprocessed2Product Code
- NKRSystem, Catheter Control, Reprocessed2Product Code
- NKSCatheter, Steerable, Reprocessed2Product Code
- NKUInjector And Syringe, Angiographic, Balloon Inflation, Reprocessed2Product Code
- NKWActuator, Syringe, For Injector, Reprocessed2Product Code
- NLGCatheter, Intracardiac Mapping, High-Density, Reprocessed2Product Code
- NLHCatheter, Recording, Electrode, Reprocessed2Product Code
- NLICatheter, Angiography, Reprocessed2Product Code
- NMBCatheter, Oximeter, Fiber Optic, Reprocessed2Product Code
- NMKTrocar, Reprocessed2Product Code
- NMMCatheter, Angioplasty, Peripheral, Transluminal, Reprocessed2Product Code
- NPPReprocessed Blood Pressure Cuff2Product Code
- NTETemporary Carotid Catheter For Embolic Capture2Product Code
- NVMCatheter, Angioplasty, Peripheral, Transluminal, Dual-Balloon2Product Code
- OBICatheter, Pressure Monitoring, Cardiac2Product Code
- OBJCatheter, Ultrasound, Intravascular2Product Code
- OEDAntimicrobial Blood Pressure Cuff2Product Code
- OEQAngiography/Angioplasty Kit2Product Code
- OERAngioscopic Valvulotome Kit2Product Code
- OESCardiac Catheterization Kit2Product Code
- OEXCardiovascular Catheter Sheath Introducer Kit2Product Code
- OEZCardiovascular Procedure Kit2Product Code
- OFBCatheter Guide Wire Kit2Product Code
- OFCCatheter Introducer Kit2Product Code
- OFDCatheter Introducer Kit2Product Code
- OFECentral Venous Blood Pressure Kit2Product Code
- OMZPulmonary (Pulmonic) Valvuloplasty Catheters/Percutaneous Valvuloplasty Catheter2Product Code
- OFFCentral Venous Catheter Tray2Product Code
- OFHDigital Angiography Tray2Product Code
- ORDOptical Coherence Tomography, Intravascular Catheter2Product Code
- OFJLead Introducer Kit2Product Code
- OFKPercutaneous Atrial Catheter Kit2Product Code
- OFLPercutaneous Sheath Introducer Kit2Product Code
- OFNWinged Intravenous Catheterization Kit2Product Code
- OGZCatheter, Intravascular, Plaque Morphology Evaluation2Product Code
- OWQReprocessed Intravascular Ultrasound Catheter2Product Code
- OZTBalloon Aortic Valvuloplasty2Product Code
- PDUCatheter For Crossing Total Occlusions2Product Code
- PFANon-Invasive Central Venous Manometer2Product Code
- PJACoronary Vascular Physiologic Simulation Software2Product Code
- PNEReprocessed Catheter Introducer2Product Code
- PNOCatheter, Percutaneous, Cutting/Scoring2Product Code
- PPNPercutaneous Catheter, Ultrasound2Product Code
- PQKPercutaneous Catheter For Creation Of An Arteriovenous Fistula For Hemodialysis Access2Product Code
- PTLWire, Guide, Catheter, Exempt2Product Code
- PTMSyringe, Balloon Inflation, Exempt2Product Code
- PUMTemporary Catheter, Embolic Protection, Transcatheter Intracardiac Procedures2Product Code
- QEKAngiographic Coronary Vascular Physiologic Simulation Software2Product Code
- QFJIntravascular Bleed Monitor2Product Code
- QJHReverse Central Venous Recanalization System2Product Code
- QMWCoronary Artery Disease Risk Indicator From Acoustic Heart Signals2Product Code
- QQIInterventional Cardiovascular Implant Simulation Software Device2Product Code
- QYFNeonatal Icu Continuous Non-Invasive Blood Pressure Monitor (Includes Alarms)2Product Code
- QYWHospital Cardiac Telemetry2Product Code
- QYXOutpatient Cardiac Telemetry2Product Code
- Subpart B—Diagnostic DevicesCFR Sub-Part
- Subpart C—Cardiovascular Monitoring DevicesCFR Sub-Part
- Subpart D—Cardiovascular Prosthetic DevicesCFR Sub-Part
- Subpart E—Cardiovascular Surgical DevicesCFR Sub-Part
- Subpart E—Surgical DevicesCFR Sub-Part
- Subpart F—Cardiovascular Therapeutic DevicesCFR Sub-Part
- Subpart F—Therapeutic DevicesCFR Sub-Part
- ChemistryReview Panel
- DentalReview Panel
- Ear, Nose, ThroatReview Panel
- Gastroenterology and UrologyReview Panel
- General HospitalReview Panel
- HematologyReview Panel
- ImmunologyReview Panel
- MicrobiologyReview Panel
- NeurologyReview Panel
- Obstetrical and GynecologicalReview Panel
- OphthalmicReview Panel
- OrthopedicReview Panel
- PathologyReview Panel
- Physical MedicineReview Panel
- RadiologyReview Panel
- General and Plastic SurgeryReview Panel
- ToxicologyReview Panel
- Medical GeneticsReview Panel
- UnknownReview Panel
Angiographic Coronary Vascular Physiologic Simulation Software
- Page Type
- Product Code
- Definition
- An angiographic coronary vascular physiologic simulation software device is intended to aid in the identification of functionally significant cardiovascular disease.
- Physical State
- Software
- Technical Method
- Performs perioperative analysis of angiographic imaging data to simulate blood flow in the coronary vasculature.
- Target Area
- Coronary Vasculature
- Regulation Medical Specialty
- Cardiovascular
- Review Panel
- Cardiovascular
- Submission Type
- 510(K)
- Device Classification
- Class 2
- Regulation Number
- 870.1415
- GMP Exempt?
- No
- Summary Malfunction Reporting
- Ineligible
- Implanted Device
- No
- Life-Sustain/Support Device
- No
- Third Party Review
- Not Third Party Eligible
- MAUDE Alerts
- View and Sign Up For MAUDE Alerts
CFR § 870.1415 Coronary vascular physiologic simulation software device
§ 870.1415 Coronary vascular physiologic simulation software device.
(a) Identification. A coronary vascular physiologic simulation software device is a prescription device that provides simulated functional assessment of blood flow in the coronary vascular system using data extracted from medical device imaging to solve algorithms and yield simulated metrics of physiologic information (e.g., blood flow, coronary flow reserve, fractional flow reserve, myocardial perfusion). A coronary vascular physiologic simulation software device is intended to generate results for use and review by a qualified clinician.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) Adequate software verification and validation based on comprehensive hazard analysis, with identification of appropriate mitigations, must be performed, including:
(i) Full characterization of the technical parameters of the software, including:
(A) Any proprietary algorithm(s) used to model the vascular anatomy; and
(B) Adequate description of the expected impact of all applicable image acquisition hardware features and characteristics on performance and any associated minimum specifications;
(ii) Adequate consideration of privacy and security issues in the system design; and
(iii) Adequate mitigation of the impact of failure of any subsystem components (e.g., signal detection and analysis, data storage, system communications and cybersecurity) with respect to incorrect patient reports and operator failures.
(2) Adequate non-clinical performance testing must be provided to demonstrate the validity of computational modeling methods for flow measurement; and
(3) Clinical data supporting the proposed intended use must be provided, including the following:
(i) Output measure(s) must be compared to a clinically acceptable method and must adequately represent the simulated measure(s) the device provides in an accurate and reproducible manner;
(ii) Clinical utility of the device measurement accuracy must be demonstrated by comparison to that of other available diagnostic tests (e.g., from literature analysis);
(iii) Statistical performance of the device within clinical risk strata (e.g., age, relevant comorbidities, disease stability) must be reported;
(iv) The dataset must be adequately representative of the intended use population for the device (e.g., patients, range of vessel sizes, imaging device models). Any selection criteria or limitations of the samples must be fully described and justified;
(v) Statistical methods must consider the predefined endpoints:
(A) Estimates of probabilities of incorrect results must be provided for each endpoint,
(B) Where multiple samples from the same patient are used, statistical analysis must not assume statistical independence without adequate justification, and
(C) The report must provide appropriate confidence intervals for each performance metric;
(vi) Sensitivity and specificity must be characterized across the range of available measurements;
(vii) Agreement of the simulated measure(s) with clinically acceptable measure(s) must be assessed across the full range of measurements;
(viii) Comparison of the measurement performance must be provided across the range of intended image acquisition hardware; and
(ix) If the device uses a cutoff threshold or operates across a spectrum of disease, it must be established prior to validation, and it must be justified as to how it was determined and clinically validated;
(4) Adequate validation must be performed and controls implemented to characterize and ensure consistency (i.e., repeatability and reproducibility) of measurement outputs:
(i) Acceptable incoming image quality control measures and the resulting image rejection rate for the clinical data must be specified, and
(ii) Data must be provided within the clinical validation study or using equivalent datasets demonstrating the consistency (i.e., repeatability and reproducibility) of the output that is representative of the range of data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment;
(A) Testing must be performed using multiple operators meeting planned qualification criteria and using the procedure that will be implemented in the production use of the device, and
(B) The factors (e.g., medical imaging dataset, operator) must be identified regarding which were held constant and which were varied during the evaluation, and a description must be provided for the computations and statistical analyses used to evaluate the data;
(5) Human factors evaluation and validation must be provided to demonstrate adequate performance of the user interface to allow for users to accurately measure intended parameters, particularly where parameter settings that have impact on measurements require significant user intervention; and
(6) Device labeling must be provided that adequately describes the following:
(i) The device's intended use, including the type of imaging data used, what the device measures and outputs to the user, whether the measure is qualitative or quantitative, the clinical indications for which it is to be used, and the specific population for which the device use is intended;
(ii) Appropriate warnings specifying the intended patient population, identifying anatomy and image acquisition factors that may impact measurement results, and providing cautionary guidance for interpretation of the provided measurements;
(iii) Key assumptions made in the calculation and determination of simulated measurements;
(iv) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance. Per-vessel clinical performance, including where applicable localized performance according to vessel and segment, must be included as well as a characterization of the measurement error across the expected range of measurement for key parameters based on the clinical data;
(v) A detailed description of the patients studied in the clinical validation (e.g., age, gender, race or ethnicity, clinical stability, current treatment regimen) as well as procedural details of the clinical study (e.g., scanner representation, calcium scores, use of beta-blockers or nitrates); and
(vi) Where significant human interface is necessary for accurate analysis, adequately detailed description of the analysis procedure using the device and any data features that could affect accuracy of results.
[80 FR 63673, Oct. 21, 2015]