V20, V20a, AVSM3 SNF

K200434 · Mediana Co., Ltd. · DXN · Oct 9, 2020 · Cardiovascular

Device Facts

Record IDK200434
Device NameV20, V20a, AVSM3 SNF
ApplicantMediana Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateOct 9, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2
AttributesPediatric

Intended Use

The V20, V20a, AVSM3 SNF are intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO-), pulse rate (PR) and temperature (TEMP) in all areas of a hospital and hospital type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V20 and V20a are for adult, pediatric and neonatal patients and the AVSM3 SNF is for adult patients only. Also, the V20 is only suitable for single measurement but V20a and AVSM3 SNF are suitable for single as well as continuous measurement. Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers. Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor.

Device Story

Lightweight, compact vital signs monitor (V20, V20a, AVSM3 SNF) for hospital/clinical use; operates via AC mains or internal battery. Inputs: NIBP cuffs, SpO2 sensors, temperature probes. Transforms physiological signals into real-time vital sign data (systolic/diastolic/mean BP, SpO2, pulse rate, temperature) displayed on 8-inch TFT-LCD screen. Operated by trained clinicians (doctors, nurses, technicians) to monitor patient status. Output provides immediate clinical data for patient assessment; enables alarm notifications for abnormal vitals. Benefits include continuous or single-point monitoring to support clinical decision-making and patient safety in general care, ICU, OR, and sub-acute settings.

Clinical Evidence

Bench testing only. Testing included electrical safety, EMC, environmental testing (IEC60068-2-27, IEC60068-2-64, IEC60068-2-1/2-2/2-30), and ISTA 2A packaging tests. Performance of NIBP, SpO2, and temperature modules was validated via equivalence to predicate devices and licensed module performance data.

Technological Characteristics

Compact monitor (249x211x176mm, 3.1kg); 8-inch TFT-LCD display. Powered by AC (100-240VAC) or internal battery. NIBP, SpO2, and temperature sensing. Complies with ISO 13485 and 21CFR820. Bench testing per system V&V plan #MDR-EG160311-01.

Indications for Use

Indicated for monitoring NIBP, SpO2, pulse rate, and temperature in hospital and hospital-type facilities. V20/V20a: adult, pediatric, and neonatal patients. AVSM3 SNF: adult patients only. Users must be trained clinicians (technicians, doctors, nurses).

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. October 9, 2020 Mediana Co., Ltd. % Charlie Mack Principal Engineer International Regulatory Consultants 2950 E Lindrick Drive Chandler, Arizona 85249 Re: K200434 Trade/Device Name: V20, V20a, AVSM3 SNF Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DQA, FLL Dated: August 30, 2020 Received: September 10, 2020 Dear Charlie Mack: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200434 Device Name V20, V20a, AVSM3 SNF #### Indications for Use (Describe) The V20, V20a, AVSM3 SNF are intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO2), pulse rate (PR) and temperature (TEMP) in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V20 and V20a are for adult, pediatric and neonatal patients and the AVSM3 SNF is for adult patients only. Also, the V20 is only suitable for single measurement but V20a and AVSM3 SNF are suitable for single as well as continuous measurement. Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital-type facilities include ambulance, physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub-acute care centers. Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatent's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor. | Type of Use ( <i>Select one or both, as applicable</i> ) | | |----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. #### This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary ### Submitter Mediana Co., Ltd. 132, Donghwagongdan-ro, Munmak-eup, Wonju-si, Gangwon-do, Republic of Korea Tel) +82-33-742-5400 Fax) +82-33-742-5483 Company Contact: Dana Mun Date Summary Prepared: January 10th 2020 ### Device Name | Trade Name: | V20, V20a, AVSM3 SNF | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Vital Signs Monitor | | Classification Name: | NIBP measurement system (21CFR870.1130), also contains<br>Non-invasive pulse oximetry, SpO2 (21CFR870.2700) and<br>clinical electronic thermometer (21CFR880.2910) | | Classification: | Class II | | Product Code: | DXN, DQA, FLL | ### Predicate Devices (Legally Marketed Devices) The predicate device for the V20, V20a, AVSM3 SNF are: - Mediana Co., Ltd. Vital Signs Monitors, V10 and AVSM2 . Cleared by FDA through 510(k) No. K170497 - . Covidien, Nellcor™ Bedside Respiratory Patient Monitoring System Cleared by FDA through 510(k) No. K141518 - . SunTech Cardiac Monitor, LifeWindow LW8 Lite Cleared by FDA through 510(k) No. K183687 - . Covidien, Filac 3000 Thermometer Cleared by FDA through 510(k) No. K003313 - . Exergen Termometer Cleared by FDA through 510(k) No. K011291 # Device Description The V20, V20a, AVSM3 SNF are intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO2), pulse rate (PR) and temperature (TEMP) in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V20 and V20a are for adult, pediatric and neonatal patients and the AVSM3 SNF is for adult patients only. Also, the V20 is only suitable for single measurement but V20a and AVSM3 SNF are suitable for single as well as continuous measurement. The Mediana V20, V20a, AVSM3 SNF vital signs monitors are a lightweight and compact device (249 x 211 x 176 (mm) (W×H×D) and 3.1 kg for Standard configuration) powered by AC mains (100-240VAC, 50Hz/60Hz) and also powered by internal battery. The monitor provides patient data and monitoring status on 8″ TFT-LCD screen. {4}------------------------------------------------ ### Intended Use The V20, V20a, AVSM3 SNF are intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO-), pulse rate (PR) and temperature (TEMP) in all areas of a hospital and hospital type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V20 and V20a are for adult, pediatric and neonatal patients and the AVSM3 SNF is for adult patients only. Also, the V20 is only suitable for single measurement but V20a and AVSM3 SNF are suitable for single as well as continuous measurement. - Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers. - Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These dinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor. ### Summary of Technical Characteristics of the Device Compared to the Predicate Devices (Legally Marketed Devices) The subject device (Mediana vital signs monitor, Model V20, V20a, AVSM3 SNF, is substantially equivalent to Mediana vital signs monitor V10, Covidien Nellcor™ Bedside Respiratory Patient Monitoring System, SunTech Cardiac Monitor LifeWindow LW8 Lite, Covidien Filac 3000 Thermometer, Exergen Thermometer. They have the same technological characteristics and material, and they are comparable in kev safety and effectiveness features and software design and have the same intended uses and basic operation modes as the predicate device. - The Non-Invasive Blood Pressure (NIBP) measurement specifications and performance are equivalent to Mediana Co., Ltd. vital signs monitors V10 and SunTech LifeWindow LW8 Lite. The clinical studies which were carried out by SunTech and AND companies were used to support different NIBP modules of the subject device under their original 510(k) submissions, and that Mediana has licensed the two NIBP modules for use in the subject device. - The Pulse rate specifications and performance derived from either Non-Invasive Blood Pressure (NIBP) or Pulse Oximetry (SpO2) are equivalent to Mediana Co., Ltd. vital signs monitors V10, SunTech LifeWindow LW8 Lite and Covidien Nellcor™ Bedside Respiratory Patient Monitoring System. - The Pulse Oximetry (SpO₂) specifications and performance are equivalent to Mediana V10 vital signs monitor and Covidien Nellcor™ Bedside Respiratory Patient Monitoring System. {5}------------------------------------------------ - The Temperature specifications and performance are equivalent to Covidien Genius 2 thermometer, Covidien Filac 3000 thermometer and Exergen thermometer. #### Summary of Performance Testing The Mediana vital signs monitor, Model V20, V20a, AVSM3 SNF substantially has been tested in accordance with the system V & V plan #MDR-EG160311-01 included with the submission using production equivalent units prior to release to market. The bench testing (electrical safety testing, EMC testing, IEC60068-2-27, IEC60068-2-64, IEC60068-2-1/2-2/2-30 and ISTA 2A test) has been performed and the results all tests demonstrate that the subject device is substantially equivalent to the predicated device. A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Mediana design control procedure. Mediana's quality system confirms to 21CFR820, ISO 13485 certified by BSI. #### Conclusions . As stated above, the V20, V20a, AVSM3 SNF all comply with the appropriate medical device guidance and standards and are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...