Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DM

K211127 · Health & Life Co., Ltd. · DXN · Sep 8, 2021 · Cardiovascular

Device Facts

Record IDK211127
Device NameFull Automatic (NIBP) Blood Pressure Monitor, Model HL858DM
ApplicantHealth & Life Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateSep 8, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

This device uses the oscillometric method to automatically measure systolic and diastolic blood pressure as well as heart rate. The measurement position is at human being's arm. All values can be read out in the LCD panel. Measurement position is at human being's upper arm. The intended user of this over-the-counter device is adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) for home use. HL858DM features BP Category Indicator that will show the information with the readings on the user tracking their blood pressure level. HL858DM is equipped with an Advanced IHB detection feature to collect and analyze pulses. If the specific irregular heartbeats are detected and it may affect blood pressure reading with device will give the user a warning signal. The feature can inform the user that the measured blood pressure reading may be inaccurate once the specific irregular heartbeats are detected. Besides, the device features a built-in "Bluetooth Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure montor by means of DailyChek® application software with the paired Bluetooth-enabled device.

Device Story

Automatic upper-arm blood pressure monitor; uses oscillometric method during inflation to measure systolic/diastolic BP and heart rate. Features include BP category indicator, irregular heartbeat (IHB) detector, and Bluetooth connectivity. Operated by adult users in home environment. Device collects pulse signals; internal algorithms analyze pulses for IHB; if detected, warning signal alerts user to potential reading inaccuracy. Results displayed on LCD; data transmitted via Bluetooth to DailyChek® app for tracking, date/time synchronization, and battery status monitoring. Benefits include home-based BP monitoring and tracking for hypertension management.

Clinical Evidence

Clinical validation performed per ISO 81060-2:2018. Results demonstrated that predetermined acceptance criteria were fully met.

Technological Characteristics

Oscillometric measurement; ABS housing/keys; LCD display; Power: 4x AAA batteries or AC adapter (100-240V AC). Connectivity: Bluetooth. Standards: IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30, ISO 10993-1/5/10, IEC 62304, IEC 62366.

Indications for Use

Indicated for adults aged 18+ with arm circumference 9-17 inches (23-43 cm) for home use to measure systolic/diastolic blood pressure and heart rate via oscillometric method.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. September 8, 2021 Health & Life Co., Ltd Peggy Su RA/QA Div. Manager 9F, No. 186, Jian Yi Road, Zhonghe District New Taipei City, 23553 Taiwan Re: K211127 Trade/Device Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DM Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 6, 2021 Received: August 9, 2021 Dear Peggy Su: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Stephen C. Browning -S LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211127 #### Device Name Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DM Indications for Use (Describe) This device uses the oscillometric method to automatically measure systolic and diastolic blood pressure as well as heart rate. The measurement position is at human being's arm. All values can be read out in the LCD panel. Measurement position is at human being's upper arm. The intended user of this over-the-counter device is adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) for home use. HL858DM features BP Category Indicator that will show the information with the readings on the user tracking their blood pressure level. HL858DM is equipped with an Advanced IHB detection feature to collect and analyze pulses. If the specific irregular heartbeats are detected and it may affect blood pressure reading with device will give the user a warning signal. The feature can inform the user that the measured blood pressure reading may be inaccurate once the specific irregular heartbeats are detected. Besides, the device features a built-in "Bluetooth Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure montor by means of DailyChek® application software with the paired Bluetooth-enabled device. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------|-----------------------------------------------------------| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K211127 # PREMARKET NOTIFICATION # 510(k) SUMMARY (As Required By 21 CFR 807.92) This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. AUG 06 2021 Date: The assigned 510(k) number is: K211127 -------- #### 1. Submitter: Health & Life Co., Ltd. 9F. No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C TEL: +886-2-8227-1300 FAX: +886-2-8227-1301 Contact person: Peggy Su / RA & QA Div. Manager E-mail: Peggy Su@hlmt.com.tw Tel: 886-2-8227-1300 ext. 11203 Fax: 886-2-8227-1301 #### 2. Name of the Device: Trade Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DM Common Name: Blood Pressure Monitor Classification Name: Non-invasive Blood Pressure Measurement System Classification: Class II, 21 CFR 870.1130 Classification Panel: 74 Cardiovascular Product Code: DXN #### 3. Information for the 510(k) Cleared Device (Predicate Device): A. Full Automatic (NIBP) Blood Pressure Monitor, Model: HL858DK (K180240) {4}------------------------------------------------ # 4. Device Description: HL858DM automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use. The device has the Risk Category Indicator, Irregular Heartbeat Detector, Bluetooth data transmission functions. When the symbol Advanced IHB appears on screen indicates the specific heartbeat irregularity was detected during measurement. Additionally, after measurement, the BP Category Indicator feature will show the information with the readings on the screen for the user tracking their blood pressure level. Besides, HL858DM featured with an IHB feature can identify the specific irregular heartbeats that may cause deviated blood pressure reading. #### 5. Intended Use This device uses the oscillometric method to automatically measure systolic and diastolic blood pressure as well as heart rate. The measurement position is at human being's arm. All values can be read out in the LCD panel. Measurement position is at human being's upper arm. The intended user of this over-the-counter device is adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) for home use. HL858DM features BP Category Indicator that will show the information with the readings on the screen for the user tracking their blood pressure level. HL858DM is equipped with an Advanced IHB detection feature to collect and analyze pulses. If the specific irregular heartbeats are detected and it may affect blood pressure reading with deviation, the device will give the user a warning signal. The feature can inform the user that the measured blood pressure reading may be inaccurate once the specific irregular heartbeats are detected. {5}------------------------------------------------ Besides, the device features a built-in "Bluetooth Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device. # 6. Comparison of device to predicate device: # Product Specification Comparison Table of Subject Device HL858DM, and Predicate Device HL858DK (K180240) | Item | Predicate Device<br>HL858DK (K180240) | Subject Device<br>HL858DM | |-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------| | Method of<br>measurement | Oscillimetric | Same as left | | Measurement<br>Type | During inflation | During inflation | | Range of<br>measurement | Pressure 0- 300mmHg,<br>Pulse 40-199 Beats/minute | Same as left | | Accuracy | Pressure ± 3mmHg<br>Pulse ± 5% | Same as left | | Pressure<br>Changed Rate | 2~5mmHg/sec.<br>(from 90mmHg to 150mmHg) | Same as left | | Display | Liquid Crystal Digital | Same as left | | Power Supply | 4 × AAA/1.5V (LR03)<br>Alkaline batteries, or<br>AC Adapter (Model: SINPRO,<br>HPU15-102 Input: 100-240V<br>AC 47-63Hz / Output: 5.99V,<br>DC, 2A) | Same as left | | Storage/<br>Transportation<br>Environment | - 25 °C ~ 60 °C (- 13 °F~ 158 °F),<br>≤ 93% R.H. | Same as left | | Operating<br>Environment | 5 °C ~ 40 °C (41 °F~104 °F),<br>15% ~ 93% R.H. | Same as left | {6}------------------------------------------------ | | 700~1060hPa | | | |------------------------------------|----------------------------------------------------------------------|----------------|------------------------------------------------------------------------------------------| | Material | ABS housing and<br>ABS keys | | Same as left | | Sets of<br>memory | 3*40,total 120 | | Same as left | | Number of<br>Push Button | 5 keys control<br>(Start/Stop, Memory, Up, Down,<br>Time () button) | | Same as left | | Storage pouch | Yes | | Same as left | | Cuff size | Arm circumference approx.<br>23~43cm / 9~17 inch<br>(Universal cuff) | | Same as left | | Unit Weight | Approx. 320 ± 5g<br>(Excluding cuff and Batteries ) | | Same as left | | Risk/BP<br>Category<br>Indicator | Yes<br>(BP Category Indicator, Six Levels) | | Same as left | | Irregular<br>Heartbeat<br>Detector | Yes | | Same as left | | Bluetooth Data<br>Transmission | No | 1.<br>2.<br>3. | Measurement Data<br>Transmission<br>Date/Time<br>Synchronization<br>Battery Status Check | # Changes from the predicate devices HL858DK (K180240): - Add Data Link Function (Bluetooth Data Transmission) The subject device HL858DM adds Bluetooth technology function. It provides users an optional choice to log, track and store their measurement data. This feature doesn't directly affect the safety and effectiveness of blood pressure measurement. Relevant evaluations such as software verification and validation, risk management activities and usability engineering validation were performed accordingly, and the results show that the Bluetooth Data Transmission function of {7}------------------------------------------------ the subject device HL858DM could operate properly as intended. Please refer to Section 14. Substantial Equivalence Discussion for detail information. # 7. Discussion of Clinical Tests Performed: The subject device HL858DM is compliant to the standard of ISO 81060-2:2018 Noninvasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met. # 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows: The subject device was tested to evaluate its safety and effectiveness, including the followings: - a. EMC Test: IEC 60601-1-2 Edition 4:2014, Medical Electrical Equipment Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Tests # b. Safety Test: -IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance -IEC 60601-1-11:2015, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance- Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment # c. FCC Test: FCC 47 CFR Part 15, Subpart B, C # d. Biocompatibility Test: -ISO 10993-1:2018, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process -ISO 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity -ISO 10993-10:2010, Third Edition Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization - e. Reliability Test: {8}------------------------------------------------ IEC 80601-2-30 Edition1.1 2013-07 Medical electricalequipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. # f. Risk Assessment: ISO 14971:2007 Second Edition, Medical devices - Application of risk management to medical devices # g. Software Verification and Validation: -IEC 62304 Ed.1.0 (2006), Medical device software - Software life cycle processes, -IEC 60601-1-4 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1 #### h. Usability Validation: -IEC 62366:2015 Medical devices - Application of usability engineering to medical devices -IEC 60601-1-6:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability # 9. Conclusions: The subject device was tested and fulfilled the requirements of those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...