Transtek Blood Pressure Monitor, Welch Allyn Remote Monitoring Blood Pressure Device, Welch Allyn Home Blood Pressure Device

K161886 · Guangdong Transtek Medical Electronics Co., Ltd. · DXN · Oct 14, 2016 · Cardiovascular

Device Facts

Record IDK161886
Device NameTranstek Blood Pressure Monitor, Welch Allyn Remote Monitoring Blood Pressure Device, Welch Allyn Home Blood Pressure Device
ApplicantGuangdong Transtek Medical Electronics Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateOct 14, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

Transtek Blood Pressure Monitor TMB-1591-A, Welch Allyn Blood Pressure Device RPM-BP100SBP and H-BP100SBP are digital monitors intended for use in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 15 cm to 54 cm (about 6-21 inches). These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings. Contraindications: This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator. These device are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.

Device Story

Non-invasive blood pressure monitor; uses inflatable arm cuff and electronic pressure sensor; operates via oscillometric method. Input: cuff pressure oscillations; Output: systolic/diastolic blood pressure, heart rate, and irregular heartbeat warning. Used in home/clinical settings by patients or clinicians. Microprocessor analyzes pressure signals to calculate vitals; detects irregular heartbeats by comparing pulse wave time intervals (warning if >25% difference). Includes Bluetooth module for automatic data transmission to mobile devices (iPhone/iPad) for storage/management. Powered by AA batteries or DC adapter. Benefits: enables convenient, automated monitoring of blood pressure and heart rate; provides early warning of potential irregular heartbeats.

Clinical Evidence

Clinical validation performed per AAMI/ANSI/ISO 81060-2:2013 to confirm accuracy of automated non-invasive blood pressure measurements across the specified arm circumference ranges. Bench testing included safety (AAMI/ANSI ES60601-1), EMC (IEC 60601-1-2), performance (IEC 80601-2-30), biocompatibility (ISO 10993), and wireless coexistence. Software verified per IEC 62304.

Technological Characteristics

Oscillometric non-invasive blood pressure measurement; electronic pressure sensor; microprocessor-based. Materials: ABS housing; biocompatible cuff materials (ISO 10993). Power: 4x AA batteries or 6V DC adapter. Connectivity: Bluetooth for data transmission. Dimensions: supports arm circumferences 15-54cm. Standards: IEC 80601-2-30, AAMI/ANSI ES60601-1, IEC 60601-1-11, IEC 60601-1-2.

Indications for Use

Indicated for adult patients with arm circumferences 15-54 cm for non-invasive measurement of blood pressure and heart rate. Contraindicated for patients with wearable or implantable electronic devices (e.g., pacemakers, defibrillators). Not for diagnostic use.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and features the department's emblem in the center. The emblem consists of a stylized human profile with three faces overlapping, representing the department's focus on health and human well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem in a circular fashion. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 14, 2016 Guangdong Transtek Medical Electronics Co., Ltd Ada Zhang Product Certification Supervisor Zone B, No.105, Dongli Road, Torch Development District Zhongshan, Guandong, China 528437 Re: K161886 Trade/Device Name: TRANSTEK Blood Pressure Monitor amd Welch Allyn Blood Pressure Device, Models: TMB-1591-A, RPM-BP100SBP and H-BP100SBP Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 5, 2016 Received: August 5, 2016 Dear Ada Zhang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Mude Yellm for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. #### 510(k) Number (if known) K161886 #### Device Name Transtek Blood Pressure Monitor, Welch Allyn Blood Pressure Device Models: TMB-1591-A, RPM-BP100SBP and H-BP100SBP #### Indications for Use (Describe) Indications of USe (Deschbe) Transtek Blood Pressure Monitor TMB-1591-A, Welch Allyn Blood Pressure Device RPM-BP100SBP and H-BP100SBP I rallstek blood fressure Momor se in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 15 cm to 54 cm (about 6-21 inches). These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings. Contraindications: This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator. These device are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the I he burden time for this Galloorch existing data sources, gather and maintain the data needed and complete time to review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Date of Summary Preparation: 10/13/2016 ### 1. Submitter's Identifications Submitter's Name: Guangdong Transtek Medical Electronics Co., Ltd Address: Zone A, No.105, Dongli Road, Torch Development District, Zhongshan, Guangdong, China 528437 Contact Person: Ada Zhang Contact Email Address: zhangjiapei@lifesense.com Telephone: 086-760-85166358 Fax: 086-760-85166358 ### 2. Name of the Device Device Classification Name: System, Measurement, Blood-Pressure, Non-invasive Product Name: TRANSTEK Blood Pressure Monitor, Welch Allyn Blood Pressure Device Trade Name: TRANSTEK, Welch Allyn Models: TMB-1591-A, RPM-BP100SBP, H-BP100SBP Classification Panel: Cardiovascular Common/Usual Name: Arm Blood Pressure Monitor Product Code: DXN Device Classification: Class II ### 3. The Predicate Devices TRANSTEK, Blood Pressure Monitor, Model TMB-1018-BT, K131395 ### 4. Device Description Blood Pressure Monitor TMB-1591-A, Welch Allyn Blood Pressure Transtek Device RPM-BP100SBP and H-BP100SBP are designed to measure the systolic and diastolic blood pressure and heartbeat rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the arm. The three devices are the same except brand name, model number and the layout of labeling. Measurement method to define systolic and diastolic pressure is similar to the ausculatory method but {4}------------------------------------------------ uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating heartbeat rate, which is a well-known technique in the market called the "Oscillometric method". These devices also compare the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%. These devices include the main unit and cuff unit. The main unit consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD. ABS is used to outer housing of the main unit. The three preformed cuff units, which are applicable to arm circumference between 15-24cm, 22-42cm and 40-54cm to meet different population's need. The subject device is powered by four AA alkaline batteries or by a DC 6V 1AA adapter. These devices embed a Bluetooth module that allows it to connect with nearby BT receiving terminal (such as iphone, ipad). Once measurement is over, the LCD of device displays results. And the device will start to transmit data to the pair-up BT terminal automatically. Thus users can display, manage and storage measurement data more easily. # 5. Intended Use of Device Transtek Blood Pressure Monitor TMB-1591-A, Welch Allyn Blood Pressure Device RPM-BP100SBP and H-BP100SBP are digital monitors intended for use in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 15 cm to 54 cm (about 6-21 inches). These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings. #### Contraindications: This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator. These device are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated. ### 6. Discussion of performance tests in design verification and design validation We conducted design verification and design validation activities to demonstrate that these proposed device met all design specifications as are Substantially Equivalent to the predicate device. The result {5}------------------------------------------------ show that TMB-1591-A, RPM-BP100SBP and H- BP100SBP conform to the following standards: a. Safetv Test: - AAMI/ANSI ES60601-1:2005+A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-1-11:2010, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance - Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment ### b. EMC Test: IEC 60601-1-2 Edition 3:2007-03, Medical Electrical Equipment - Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Tests ### c. Performance and Reliability test: IEC 80601-2-30 Edition1.1 2013-07 Medical electrical equipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. ### d. FCC Test: FCC 47 CFR Part 15, Subpart B ### e. Biocompatibility Test: - ISO 10993-1:2009, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process - ISO 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity - ISO 10993-10:2010, Third Edition Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization f. Risk Assessment: ISO 14971:2007 Second Edition, Medical devices - Application of risk management to medical devices ## g. Software Verification and Validation: -IEC 62304 Ed.1.0 (2006), Medical device software - Software life cycle processes, ## h. Usability Validation: -IEC 62366-1:2007+A1:2014 Medical devices - Application of usability engineering to medical devices -IEC 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability ## i. Wireless coexistence Radio-Frequency Wireless Technology in Medical Devices ## j. Clinical Validation AAMI/ANSI/ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. {6}------------------------------------------------ ### 7. Summary of Substantial Equivalence Transtek Blood Pressure Monitor TMB-1591-A, Welch Allyn Blood Pressure Device RPM-BP100SBP and H-BP100SBP have identical technology, energy type, cuff surface material and similar intended use, performance specifications, software/firmware and functions to the predicate device. The only significant difference between these proposed devices and the predicate device is that these proposed devices add two other intended cuffs which are applicable for population with arm circumference 15-24cm and 40-54cm. The additional cuffs size are appropriate to the claimed intended arm circumference per IEC 80601-2-30. The material of two additional cuffs is the same with original cuff. Thus there is no biocompatibility problem. The clinical performance and accuracy of these proposed devices have been demonstrated that meet the requirements of AAMI/ANSVISO 81060-2:2013. Thus the difference does not raise any new questions of safety and effectiveness. ### 8. Conclusions Transtek Blood Pressure Monitor TMB-1591-A, Welch Allyn Blood Pressure Device RPM-BP100SBP and H-BP100SBP are substantially equivalent to the predicate device TMB-1018-BT by having the identical technologies and similar intended use. Two added cuffs does not impact the safety and effectiveness of these proposed devices. --- End of this section ---
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...