Arm Blood Pressure Monitor

K192023 · Truly Instrument Limited · DXN · Dec 23, 2019 · Cardiovascular

Device Facts

Record IDK192023
Device NameArm Blood Pressure Monitor
ApplicantTruly Instrument Limited
Product CodeDXN · Cardiovascular
Decision DateDec 23, 2019
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected. The devices' feature include Bluetooth function to transmit data to an external Bluetooth device with wireless communication

Device Story

Device measures systolic/diastolic blood pressure and pulse rate using oscillometric method; inflatable cuff wraps around upper arm; electronic pressure sensor converts cuff pressure changes to electrical signals; microprocessor analyzes signals to calculate blood pressure and pulse. Features irregular pulse rhythm detection (alerts if pulse interval variance >25%). Bluetooth 4.0 module transmits data (systolic, diastolic, pulse, date, time) to external devices. Intended for home use by adult patients. Healthcare providers review displayed results to inform clinical decisions. Benefits include convenient, non-invasive blood pressure monitoring and irregular heartbeat awareness.

Clinical Evidence

Clinical validation performed per ISO 81060-2:2013 (Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type). Results confirmed that mean difference and standard deviation requirements were fulfilled.

Technological Characteristics

Materials: ABS housing, nylon cuff shell. Sensing: Electronic pressure sensor. Energy: Polymer battery (DC 3.7V, 500mAh) or 4x 1.5V AA batteries. Connectivity: Bluetooth 4.0. Dimensions: Cuff fits 220-340mm arm circumference. Software: Microprocessor-based oscillometric analysis. Biocompatibility: ISO 10993-1, -5, -10 compliant.

Indications for Use

Indicated for adult individuals to measure systolic and diastolic blood pressure and pulse rate via non-invasive upper arm cuff. Includes irregular pulse rhythm detection and Bluetooth data transmission.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 23, 2019 Truly Instrument Limited % Max Wong Official Correspondent Truly (U.S.A) Inc 2620 Concord Avenue, Suite 106 Alhambra, California 91803 Re: K192023 Trade/Device Name: Arm Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 29, 2019 Received: November 29, 2019 Dear Max Wong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K192023 #### Device Name Arm Blood Pressure Monitor Indications for Use (Describe) Truly Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected. The devices' feature include Bluetooth function to transmit data to an external Bluetooth device with wireless communication Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 5-1 {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text 'K192023' in a simple, sans-serif font. The text is horizontally oriented and appears to be a code or identifier. The characters are uniformly sized and spaced, making the text easily readable. Image /page/3/Picture/1 description: The image shows the word "TRULY" in red, bold letters. To the right of the word is a circled R, indicating that the word is a registered trademark. A black line is located underneath the word. # 510(k) Summary V1.0 Date of Summary Preparation: May 21.2019 #### 1. Submitter's Identifications | Submitter's Name: | Truly Instrument Limited | |-------------------|---------------------------------------------------------------------------| | Address: | Site 2, Truly Industrial Area, Shanwei City,<br>Guangdong Province, China | | Contact Person: | Manager Su-HaiSen | | Telephone: | 86-0660-3363561 | | Fax: | 86-0660-3380377 | #### 2. Name of the Device | Device Classification Name: | System, Measurement, Blood-Pressure, Non-invasive | |-----------------------------|---------------------------------------------------| | Trade Name: | Arm Blood Pressure Monitor | | Models: | DB66-1, DB68 | | Classification Panel: | cardio-vascular | | Common/Usual Name: | Automatic Arm Blood Pressure Monitor | | Product Code: | DXN | | Device Classification: | Class II | | Contraindications : | N/A | #### 3. The Predicate Devices TRULY Instrument Limited. Truly Automatic Arm Bluetooth Blood Pressure Monitor DB Series, Model DB62 510 K Number : K161846 #### 4. Device Description Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method". The main components of the Truly Automatic Arm Blood Pressure Monitor DB series are the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 220 and 340 mm, includes the inflatable bladder and nylon shell. All models of the arm blood pressure monitor use a single size of cuff. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD. The devices also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the word "TRULY" in large, bold, red letters. To the right of the word, there is a circled "R" symbol, indicating that the word is a registered trademark. The word is slightly slanted to the right, giving it a dynamic appearance. The background is plain white. irregular pulse rhythm when the difference of the time intervals is over 25%. The devices embed a Bluetooth 4.0 Wireless network connections module that allows it to connect to nearby receiving end. Once measurement is over, the LCD of device displays results, and the device will start to send out data such as systolic, diastolic, pulse, date, time with Wireless method and protocol. ## 5. Intended use of device Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected. The devices' feature include Bluetooth function to transmit data to an external Bluetooth device with wireless communication . ## 6. Technological Characteristics of our new DEVICE COMPARED TO THE ## PREDICATE DEVICE: The technological characteristics of Automatic Arm Blood Pressure Monitor, models DB66-1, DB68 are substantially equivalent to Truly Automatic Arm Blood Pressure Monitor DB Series, Model DB62.( K161846). There is the same Owner, TRULY instrument Limited. Which FDA owner number is 9054344. In addition, there are the same design specifications, the same form and intended to be used in the same manner that means the new devices are same as the predicate devices. | Parameter | Predicate Devices DB62 | DB66-1 | DB68 | Explanation of the<br>differences | |------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-----------------------------------| | Intended use | Measuring systolic and diastolic blood<br>pressure and pulse rate of adult individual | Measuring systolic and diastolic blood<br>pressure and pulse rate of adult individual | Measuring systolic and diastolic blood<br>pressure and pulse rate of adult individual | Identical | | The trade name | Automatic Arm Bluetooth Blood Pressure<br>Monitor | Arm Blood Pressure Monitor | | Identical | | 510(K) Number | K161846 | | | | | Applicant | Truly Instrument Limited. | Truly Instrument Limited. | | Identical | | Regulatory Class | Class II | Class II | | Identical | | Panel | Cardiovascular | Cardiovascular | | Identical | | Product Code | DXN | DXN | | Identical | | Indications for<br>use | Truly Automatic Arm Bluetooth Blood<br>Pressure Monitor DB62 is intended to<br>measure the systolic and diastolic blood | Truly Automatic Arm Bluetooth Blood<br>Pressure Monitor DB62 is intended to<br>measure the systolic and diastolic blood | | Identical | #### Table-1: The comparison table Page 2 of 6 {5}------------------------------------------------ | TRULY®<br>r | | |-------------|--| |-------------|--| | Parameter | Predicate Devices DB62 | DB66-1 | DB68 | Explanation of the differences | |------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | pressure and pulse rate of an adult<br>individual by using a non-invasive<br>technique in which an inflatable cuff is<br>wrapped around the upper arm.<br>The devices' features include irregular<br>pulse rhythm detection during<br>measurement, and will display a alert<br>signal with the reading when irregular<br>heartbeat is detected.<br>The devices' feature include Bluetooth<br>function to transmit data to an external<br>Bluetooth device with wireless<br>communication .. Over-The-Counter Use | pressure and pulse rate of an adult<br>individual by using a non-invasive<br>technique in which an inflatable cuff is<br>wrapped around the upper arm.<br>The devices' features include irregular<br>pulse rhythm detection during<br>measurement, and will display a alert<br>signal with the reading when irregular<br>heartbeat is detected.<br>The devices' feature include Bluetooth<br>function to transmit data to an external<br>Bluetooth device with wireless<br>communication .. Over-The-Counter Use | | | | Target<br>Population | Adult | Adult | | Identical | | Anatomical sites | Upper Arm | Upper Arm | | Identical | | Where used<br>(hospital, home,<br>ambulance. etc) | Home | Home | | Identical | | Energy used and<br>/ or delivered | 4x 1.5V AA Battery | Polymer battery<br>DC3.7V,<br>500mAh | 4x 1.5V AA<br>Battery | Equivalent<br>The change in the<br>specification is<br>documented and<br>tested,does not<br>affect the intended<br>use or the<br>fundamental<br>scientific<br>technology. | | Human factors | Blood pressure | Blood pressure | | Identical | | performance | Measuring systolic and diastolic blood<br>pressure and pulse rate of adult individual,<br>Including irregular pulse rhythm detection | Measuring systolic and diastolic blood<br>pressure and pulse rate of adult<br>individual, Including irregular pulse<br>rhythm detection | | Identical | | biocompatibility | Cuff<br>According to ISO-10993 | Cuff<br>According to ISO-10993 | | Identical | | Compatibility with<br>the environment<br>and other devices | Operation Environment:<br>10°C~40°C,15%~90%RH<br>Storage Environment:<br>-20°C~60°C,10%~95%RH<br>Atmospheric:70KPa~106KPa | Operation Environment:<br>10°C~40°C,15%~90%RH<br>Storage Environment:<br>-20°C~60°C, 10%~95%RH<br>Atmospheric 70KPa~106KPa | | Identical | {6}------------------------------------------------ | Parameter | Predicate Devices DB62 | DB66-1 | DB68 | Explanation of the<br>differences | |-------------------|------------------------------------------------------|------------------------------------------------------|------|-----------------------------------| | Electrical safety | According to IEC60601-1-2<br>According to IEC60601-1 | According to IEC60601-1-2<br>According to IEC60601-1 | | Identical | | Sterilization | N/A | N/A | | Identical | | Chemical safety | N/A | N/A | | Identical | | sterility | N/A | N/A | | Identical | | Thermal safety | N/A | N/A | | Identical | | design | Refer to Table-2 | | | | | materials | Refer to Table-2 | | | | #### Table-2: The comparison table in Design and Materials | Parameter | Predicate Devices DB62 | DB66-1 | DB68 | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|---------------------------------| | Measurement algorithm<br>Method | Oscillometric method | No change ,all same | No change ,all same | | Measurement site of<br>body | Arm | No change ,all same | No change ,all same | | Max Cuff pressure | 300mmHg | No change ,all same | No change ,all same | | Bluetooth | BG312 | H04J08C | | | Irregular<br>heartbeat<br>detection | More than $\pm$ 25% to the<br>mean interval of pulse<br>intervals.<br>About the more detailed<br>description of the IH detection<br>algorithm, please refer to<br>"Software validation report<br>I-5. Algorithm description 4.<br>Determination method of<br>irregular heartbeat". | No change ,all same | No change ,all same | | Measurement Pressure<br>Range | 20 ~ 280 mmHg | No change ,all same | No change ,all same | | Measurement Pulse<br>Range | 40 ~ 195 beats/min | No change ,all same | No change ,all same | | Mesauring resolution | 1 mmHg | No change ,all same | No change ,all same | | Accuracy Pressure | $\pm$ 3mmHg | No change ,all same | No change ,all same | | Accuracy Pulse | $\pm$ 5% | No change ,all same | No change ,all same | | Cuff Deflation | Automatic deflation | No change ,all same | No change ,all same | | Pressurization Source | Automatic internal pump<br>DC6V | Automatic internal pump<br>DC3V | Automatic internal pump<br>DC6V | | Power Voltage | 4X 1.5V | Polymer battery<br>DC3.7V,500mAh | 4x 1.5V AA Battery | | Operating Environment | 10~40℃, 15~90%RH | No change ,all same | No change ,all same | | Storage Environment | -20°C~60°C,10%~95%RH | No change ,all same | No change ,all same | {7}------------------------------------------------ | Truly Instrument Limited | | | | |--------------------------|------------------------|----------------------------------------------|------| | Parameter | Predicate Devices DB62 | DB66-1 | DB68 | | Memory Size | 4 x 99 | 99 | 99 | | Display Type | Liquid crystal display | Liquid crystal display ,Only difference size | | | Cover | | Difference | | ## 7. Summary of Clinical study Testing Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the device. Compliant to the standard of ISO 81060-2: Second Edition 2013-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device. ## 8. Non-Clinical Tests Performed: - 8. Test Summary: 8-1. Electric Safety , EMC and FCC test reports. | General safety | IEC 60601-1:2005+A1:2012<br>Medical electrical equipment-<br>Part1: General requirements for safety and essential performance<br>IEC60601-1-11:2015<br>Medical electrical equipment-Part 1-11: General Requirement for<br>basic safety and essential performance- Collateral Standard:<br>Requirements for medical electrical systems used in the home<br>healthcare environment<br>EN60950-1:2013<br>Information technology equipment-Safety- Part1:General<br>requirements | |----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | EMC conformity | IEC 60601-1-2:2014<br>Medical Electrical Equipment - Part 1-2: General requirements for<br>safety - collateral standard: Electromagnetic compatibility -<br>Requirements and Tests | | FCC conformity | FCC 47 part 15 subject B class B | | ERM conformity | EN301489-1:2017;EN301489-17:2017 | | RF conformity | EN300328:2016 | | Health | EN62479:2010 | #### 8-2. Reliability & Clinical Test | IEC80601-2-30:20019<br>+A1:2013 | Medical electrical equipment -<br>Part 2-30: Particular requirements for the basic safety and essential<br>performance of automated non-invasive sphygmomanometer | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ISO 81060-2:2013 | Non-invasive sphygmomanometers – Part 2: Clinical validation | {8}------------------------------------------------ Image /page/8/Picture/1 description: The image contains the word "TRULY" in large, red, bold letters. To the right of the word is a small, red circle with the letter R inside, indicating a registered trademark. There is a black line underneath the word "TRULY". of automated measurement type #### 8.3 Biocompatibility Test | ISO10993-1:2009 | Biological evaluation of medical devices-Part 1: evaluation and<br>testing | |------------------|--------------------------------------------------------------------------------------------| | ISO10993-5:2009 | Biological evaluation of medical devices-Part 5: Test for<br>cytotoxicity in vitro methods | | ISO10993-10:2010 | Biological evaluation of medical devices-Part 10: Tests for irriation<br>and sensitization | #### 9. Conclusions The new subject devices of Arm Blood Pressure Monitor continue to follow principles of hardware and software design of the predicate device DB62(K161846), and the feature, safety, effectiveness are also as same as DB62., Thus, the subject devices are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...