BLOOD PRESSURE MONITOR, MODEL M100
K032308 · Meditek Technology, Inc. · DXN · Nov 10, 2003 · Cardiovascular
Device Facts
| Record ID | K032308 |
| Device Name | BLOOD PRESSURE MONITOR, MODEL M100 |
| Applicant | Meditek Technology, Inc. |
| Product Code | DXN · Cardiovascular |
| Decision Date | Nov 10, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1130 |
| Device Class | Class 2 |
Intended Use
The M100 Blood Pressure Monitor is an electric device that provides a signal from which the systolic, diastolic, and rate of heart beats can be derived through the use of transducers placed on the surface of body. The signals of measurement can be stored in the memory of device, and then can be displayed on the screen of device and/or be transmitted out of device by using the standard signal transmission cable.
Device Story
Meditek M100 is a non-invasive electronic blood pressure monitor; uses oscillometric method to measure systolic/diastolic pressure and heart rate. Device consists of pneumatic pump, cuff, and microprocessor. Operation: cuff placed on wrist or arm; pump inflates cuff; device measures pressure oscillations during controlled deflation. Microprocessor software analyzes oscillation data to derive blood pressure and heart rate. Results displayed on LCD; data stored in internal memory; transmission via cable. Used in clinical or home settings; operated by patient or clinician. Provides objective physiological data to assist in cardiovascular monitoring.
Clinical Evidence
Bench testing only. Compliance with ANSI/AAMI SP 10-1992 (non-invasive blood pressure monitoring standard) and EN 60601 series (electrical safety and electromagnetic compatibility) was demonstrated. Software verification was performed in accordance with FDA software guidance.
Technological Characteristics
Oscillometric sensing principle; pneumatic pump-driven cuff inflation/deflation; microprocessor-based analysis. Form factor: wrist-worn monitor. Connectivity: signal transmission cable for data output. Standards: ANSI/AAMI SP 10-1992, EN 60601-1, EN 60601-1-1, EN 60601-1-2.
Indications for Use
Indicated for patients requiring non-invasive measurement of systolic/diastolic blood pressure and heart rate via wrist or arm-cuff oscillometric sensing.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
- Health care: HL-168 (K990807)
Related Devices
- K070473 — BLOOD PRESSURE MONITOR, MODELS ZSBP-OXX AND ZSBP-1XX · Zen Strong Medical Technology Co., Ltd. · May 23, 2007
- K110018 — FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR · Andon Health Co, Ltd. · Mar 15, 2011
- K222394 — Electronic Sphygmomanometers · Qingdao Yasee Medical Device Co., Ltd. · Mar 3, 2023
- K193387 — Arm Electronic Sphygmomanometer · Shenzhen Changkun Technology Co., Ltd. · Jan 27, 2021
- K994423 — AUTOMATED NON-INVASIVE BLOOD PRESSURE MONITOR, MODEL BPM-100 · Vsm Medtech , Ltd. · Mar 28, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
K032308
# NOV 1 0 2003
### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: ____________________
- 1. Submitter's Identifications: Meditek Technology Inc. 1F, No. 11, Lane 223, Sec. 3, Chung Ching N. Rd., Taipei, Taiwan, R.O.C.
Contact: Mr. Rodin Chang President
Date of Summary Preparation: September 25, 2002.
- 2. Name of the Device: Meditek Blood pressure Monitor, model M100.
- 3. Information of the 510(k) Cleared Device (Predicate Device): Health care: HL-168(K990807).
### 4. Device Description:
The Meditek Blood Pressure Monitor, model M100, is the non-invasive electronic medical device intended for the measuring of blood pressure (including systolic pressure and diastolic pressure) and the number of heart beats of patient by using the oscillometric method. According to the general principle of the oscillometric method, the cuff was placed on patient's wrist or arm. The cuff was inflated by using pneumatic pump to the estimated pressure and then was deflated at a certain deflation rate. During the deflation of pressure, the pressure and it's oscillation was measured. The measuring pressure and its oscillation are calculated and analysed by the software of the microprocessor according to the algorithm of oscillometric method.
After completing the measuring cycle, the systolic pressure, diastolic pressure, and heart beat rate are displayed on the LCD of device. All the measuring information and the time of measuring may be stored in the memories of device. Those stored memories may be transmitted out of device through the signal transmission cable.
Basically the blood pressure monitor M100 is the device measuring blood pressure at wrist.
### 5. Intended Use:
The M100 Blood Pressure Monitor is an electric device that provides a signal from which the systolic, diastolic, and rate of heart beats can be derived through the use of transducers placed on the surface of body. The signals of measurement can be stored in the memory of device, and then can be displayed on the screen of device and/or be transmitted out of device by using the standard signal transmission cable.
{1}------------------------------------------------
mediek technology mc.
- 6. Comparison to the 510(k) Cleared Device (Predicate Device):
The cleared device for the specification comparison of M100 is HL-168 (K990807).
- 7. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ANSI/AAMI, SP 10-1992, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
- 8. Conclusions
The Meditek Blood Pressure Monitor, including model M100, has the same intended use and technological characteristics as the cleared device of HL-168. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of a stylized eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 0 2003
Meditek Technology, Inc. c/o Mr. Tony C. S. Chang Official Correspondent 1F, No. 11, Lane 223, Sec. 3 Chung Ching N. Road Taipei Taiwan, R.O.C.
Re: K032308
Trade Name: Blood Pressure Monitor, Model M100 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Monitor Regulatory Class: Class II (two) Product Code: DXN Dated: November 3, 2003 Received: November 3, 2003
#### Dear Mr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Mr. Tony C. S. Chang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Uffaxmy
for Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
Device Name: Blood Pressure Monitor / M100 .
### Indications For Use:
The M100 of blood pressure monitor is an electronic device that provides a signal from which the systolic, diastolic and rate of heart beats can be derived through the use of transducers placed on the surface of the body. The signals of measurement can be stored in the memory of device, and then can be displayed on the screen of device and/or be transmitted out of device by using the standard signal transmission cable.
### PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
K032308
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | |
|---------------|--|
|---------------|--|
| Prescription Use (Per 21 CFR 801.109) | | OR | Over-The-Counter Use | <div> <br/> <br/> <br/> ✓ </div> |
|---------------------------------------|--|----|----------------------|-----------------------------------------|
|---------------------------------------|--|----|----------------------|-----------------------------------------|
K03230
And All and