TL-300 T-LINE NON-INVASIVE BLOOD PRESSURE MONITORING SYSTEM
K123446 · United States Gtm Medical Devices, Inc. · DXN · Dec 7, 2012 · Cardiovascular
Device Facts
Record ID
K123446
Device Name
TL-300 T-LINE NON-INVASIVE BLOOD PRESSURE MONITORING SYSTEM
Applicant
United States Gtm Medical Devices, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Dec 7, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1130
Device Class
Class 2
Intended Use
This device is intended for use by medically trained personnel in a clinical setting to continually monitor and display diastolic, systolic, and mean blood pressures and pulse rate. The device is intended for use on adult patients with a palpable pulse.
Device Story
TL-300 T-Line system provides continuous non-invasive blood pressure (NIBP) monitoring; utilizes applanation tonometry to measure arterial pressure waveforms; inputs include radial artery pulse signals; transforms signals into real-time systolic, diastolic, and mean blood pressure values and pulse rate; used in clinical settings by medically trained personnel; output displayed on monitor for clinician review; facilitates hemodynamic assessment and patient management; benefits include continuous monitoring without invasive arterial line placement.
Clinical Evidence
No clinical data provided; substantial equivalence established via bench testing and comparison of technological characteristics.
Technological Characteristics
Non-invasive blood pressure monitoring system; utilizes applanation tonometry for arterial pressure waveform acquisition; intended for continuous monitoring; class II device; product code DXN.
Indications for Use
Indicated for continuous monitoring and display of diastolic, systolic, mean blood pressure, and pulse rate in adult patients with a palpable pulse in clinical settings.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Related Devices
K121141 — TL-200 T-LINE NON-INVASIVE BLOOD PRESSURE MONITORING SYSTEM · Tensys Medical Systems, Inc. · May 15, 2012
K022783 — TL100 TENSMETER BLOOD PRESSURE MONITORING SYSTEM · Tensys Medical, Inc. · Nov 19, 2002
K020537 — TL10 TENSYMETER BLOOD PRESSURE MONITORING SYSTEM · Tensys Medical, Inc. · Aug 15, 2002
K102968 — INFINITY CNAP SMARTPOD · Draeger Medical Systems, Inc. · Feb 18, 2011
K231586 — Endophys Blood Pressure Monitor model BPM-30 · Endophys Technologies, LLC · Nov 3, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Tensys Medical Systems, Inc. c/o Dr. Jamie Sulley Chief Operating Officer 5825 Oberlin Drive, Suite 100 San Diego, CA 92121-3709
DEC 0 7 2012
Re: K123446
Trade/Device Name: TL-300 T-Line Non-invasive Blood Pressure Monitoring System Regulatory Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: November 6, 2012 Received: November 8, 2012
Dear Dr. Sulley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Dr. Jamie Sulley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Felipe Aguel
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Tensys Medical Systems Inc. -TL-300 Special 510(k)
## Appendix B
## Indications for Use Statement
## 510(k) Number (if known):
Device Name: TL-300 T-Line® Non-invasive Blood Pressure Monitoring System
## Indications for Use:
This device is intended for use by medically trained personnel in a clinical setting to continually monitor and display diastolic, systolic, and mean blood pressures and pulse rate. The device is intended for use on adult patients with a palpable pulse.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Felipe Aguel for BDZ
K123446
Page 1 of 1
(Posted November 13, 2003)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.