← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K211588 # Caretaker Platform, VitalStream Platform, CT5 (K211588) _Caretaker Medical · DXN · Aug 13, 2021 · Cardiovascular · SESE_ **Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K211588 ## Device Facts - **Applicant:** Caretaker Medical - **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md) - **Decision Date:** Aug 13, 2021 - **Decision:** SESE - **Submission Type:** Special - **Regulation:** 21 CFR 870.1130 - **Device Class:** Class 2 - **Review Panel:** Cardiovascular ## Intended Use The Caretaker Platform is intended for non-invasive, continuous or spot-check measurement of hemodynamic parameters. Parameters can be reported to an integrated screen and optionally to Remote Data Display Systems (RDDS) and/or the Caretaker Remote Monitor App via standard wireless or wired data transmission protocols. The device is intended for use by clinicians or other properly trained personnel on adult patients at rest. ## Device Story Caretaker Platform is a hemodynamic monitoring system using a finger cuff to capture arterial pulse signals. It employs Pulse Decomposition Analysis (PDA) to transform pulse data into hemodynamic parameters. The device operates via a physical monitor with a touchscreen interface, which can transmit data to Remote Data Display Systems (RDDS) or the Caretaker Remote Monitor App via wired/wireless protocols. Used in clinical settings by trained personnel, it provides continuous or spot-check measurements. The system requires calibration via manual/AAMI 81060 compliant devices or its internal self-calibration mode. Output is displayed for clinicians to monitor patient hemodynamics, aiding in clinical decision-making by providing non-invasive, continuous vital sign tracking without the risk of occluding blood flow associated with higher-pressure cuffs. ## Clinical Evidence Bench testing only. Performance was validated using the Caretaker Analyzer GUI (CA GUI) to compare PDA algorithm outputs between the subject device and the predicate. Testing confirmed that the subject device provides hemodynamic parameters consistent with AAMI SP-10 guidelines and the predicate device's performance. ## Technological Characteristics Hemodynamic monitoring system using finger cuff-based sensing. Employs Pulse Decomposition Analysis (PDA) algorithm. Features touchscreen interface, wireless/wired connectivity, and API support for Android, Windows, and Linux. Operates at coupling pressure below normal diastole. Calibrated via manual or AAMI 81060 compliant devices or internal self-calibration mode. ## Regulatory Identification A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body. ## Predicate Devices - Caretaker4 Physiological Monitor ([K151499](/device/K151499.md), [K163255](/device/K163255.md)) - Caretaker Monitor ([K181196](/device/K181196.md)) ## Related Devices - [K163255](/device/K163255.md) — CareTaker Continuous Non-lnvasive Blood Pressure Monitor, CareTaker Wireless Vital Signs Monitor, CareTaker Physiological Monitor · Caretaker Medical, LLC · Apr 21, 2017 - [K151499](/device/K151499.md) — CareTaker Continuous Non-Invasive Blood Pressure Monitor, CareTaker Wireless VItal Signs Monitor · Caretaker Medical, LLC · Mar 18, 2016 - [K213699](/device/K213699.md) — CareTaker Monitor App · Caretaker Medical, LLC · Mar 30, 2022 - [K160552](/device/K160552.md) — EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight System, EV1000 Clinical Platform · Edwards Lifesciences, LLC · Jun 1, 2016 - [K140312](/device/K140312.md) — EV1000 CLINICAL PLATFORM WITH CLEARSIGHT TM FINGER CUFF OR CLEARSIGHT TM SYSTEM · Edwards Lifesciences, LLC · Jun 13, 2014 ## Submission Summary (Full Text) {0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. August 13, 2021 Caretaker Medical Jeff Pompeo President & CEO 941 Glenwood Station Ln #301 Charlottesville, Virginia 22901 Re: K211588 Trade/Device Name: Caretaker Platform Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DRG Dated: May 12, 2021 Received: May 24, 2021 Dear Jeff Pompeo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211586 Device Name Caretaker Platform ### Indications for Use (Describe) The Caretaker Platform is intended for non-invasive, continuous or spot-check measurement of hemodynamic parameters. Parameters can be reported to an integrated screen and optionally to Remote Data Display Systems (RDDS) and/or the Caretaker Remote Monitor App via standard wireless or wired data transmission protocols. The device is intended for use by clinicians or other properly trained personnel on adult patients at rest. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. EF {3}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness - 1) 12 Mav 2021 Preparation Date: - 2) Submitted by: Caretaker Medical, LLC 941 Glenwood Station Ln; Suite 301 Charlottesville, Virginia 22901 User Fee Organization Number 397095 Owner/Operator #: 10054848 ### 3) Contact Person/Prepared by: Jeff Pompeo President & CEO Phone: 434-409-1945 Email: Jeff@caretakermedical.net #### 4) Device Identification: Trade Name: Common Name: Alternate Marketing Names: Caretaker Physiological Monitor System Caretaker Platform Vitalstream GT5 Classification: 21 CFR 870.1130. Product Code: DXN -System, Measurement, Blood-Pressure, Non-Invasive Device Class: II - Predicate Devices: Caretaker4 Physiological Monitor (K151499, K163255), Caretaker 5) Monitor (K181196) - Device Description: The Caretaker Platform is a hemodynamic monitoring system that non-6) invasively measures various parameters via a finger cuff based on the scientific method of Pulse Decomposition Analysis ("PDA"). The Caretaker physical device is calibrated using any manual or AAMI 81060 compliant blood pressure device or is automatically calibrated using its selfcalibration mode. The Caretaker Remote Monitor consists of an App that runs on Remote Data Display Systems (RDDS) and a set of APIs to for additional command and control functions developed on Android, Windows, or Linux systems. - 7) Intended Use: The Caretaker Platform is intended for non-invasive, continuous or spot-check measurement of hemodynamic parameters. Parameters can be reported to an integrated screen and optionally to Remote Data Display Systems (RDDS) and/or the Caretaker Remote Monitor App via standard wireless or wired data transmission protocols. The device is intended for use by clinicians or other properly trained personnel on adult patients at rest. #### 8) Comparison to Predicates: Summary of methodology Comparison bench-testing was performed to compare the current device to the predicate Caretaker4 since both use the Pulse Decomposition Analysis (PDA) method for the tracking of hemodynamic parameters. The comparison tests focused on the algorithmic performance 510(k) Summary Page 1 of 2 {4}------------------------------------------------ because the physiological sensing and electronic processing hardware that constitutes both device's front ends are the same. The bench tests are performed in the Caretaker Analyzer GUI (CA GUI), which is a customdesigned environment for the development and testing of the algorithms that analyze and process incoming arterial pulse signal data to produce vital signs. The CA GUI can perform comprehensive testing across platforms and versions, i.e. the performance of PDA algorithms, both in firmware and desktop software versions for both CT4 and DUT CT5, can be assessed by selecting appropriate menu items. ## Provision of continuous non-invasive hemodynamic parameters Both devices measure blood pressure within AAMI SP-10 guidelines. Both devices require calibration and are intended for use on adult patients administered by trained medical staff. Both systems are identical with regard to electronic circuitry, pneumatic layout and components and algorithmic pulse analysis. The new model CT5 has a touch screen interface which the predicate device does not have. Both systems operate at a coupling pressure significantly lower than normal diastole with a lower risk of occluding blood flow to the monitored digit, making the technology safer ## Provision of spot-check non-invasive hemodynamic parameters Both the predicate Caretaker and the present system use an oscillometric sweep to obtain pointin-time measurements within AAMI guidelines. The predicate device used this sweep solely for the purpose of self-calibration before switching to its relative change monitoring mode. In addition to the self-calibration, the present CT5 can also be set to periodically wake and use the same sweep to get multiple point-in-time measurement at isochronal periods set by the end user. ## Provision of use of software Both the predicate Caretaker and the present system use the same Caretaker Remote Monitor App software and APIs. The new CT5 device includes additional feature keys to set different configuration options for various hemodynamic parameters that were always on by default in the previous device. The feature keys are unique to each physical device serial number and are set at time of manufacture to create different models (all using the same hardware). #### 9) Conclusions: ## Provision of non-invasive hemodynamic parameters The Caretaker Platform is substantially equivalent to the predicate Caretaker device in regard to providing continuous, non-invasive hemodynamic monitoring. The spot-check mode of the present system is substantially equivalent to the self-calibration mode of the predicate. ## Provision of use of software The Caretaker Platform is substantially equivalent to the Caretaker Remote Monitor App and APIs in that the same software is used with the addition of the feature keys described above. --- **Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K211588](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K211588) **Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact). **Cite as:** FDA Device Explorer (Innolitics), https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K211588
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...