The Electronic Sphygmomanometer and Automatic Blood Pressure Monitor are intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.
Device Story
Battery-powered automatic non-invasive blood pressure monitors (CONTEC08A, CONTEC08C, ABPM50) measure systolic/diastolic blood pressure and pulse rate via oscillometric technique at the upper arm. Devices feature an inflatable cuff, pump, and MCU; data storage for measurement history; and physiological/technical over-limit prompts (e.g., low battery). Operated by patients or clinicians in home or medical settings. Users view results on integrated TFT LCD or segment code displays. Output data assists in monitoring blood pressure trends. Benefits include convenient, automated, non-invasive blood pressure tracking.
Clinical Evidence
Clinical validation conducted per ISO 81060-2. Study included 85 subjects. Results confirmed clinical accuracy (mean deviation and standard deviation) is not inferior to the predicate device. Compliance with ISO 81060-2 verified.
Technological Characteristics
Battery-powered (alkaline or 5V adapter) oscillometric blood pressure monitors. Components: LCD, keys, cuff (PVC/Nylon/Polyester), MCU, pump. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30, ISO 81060-2, ISO 10993-5/10. Connectivity: Data storage for review. User-selectable units (mmHg/kPa).
Indications for Use
Indicated for adult patients requiring non-invasive measurement of systolic/diastolic blood pressure and pulse rate at medical facilities or home.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Electronic Blood Pressure Monitor: Models AOJ-30A and AOJ-30B (K191180)
Related Devices
K191180 — Electronic Blood Pressure Monitor · Shenzhen AOJ Medical Technology Co., Ltd. · Aug 23, 2019
K110774 — ELECTRONIC SPHYGMOMANOMETER · Contec Medical System Co., Ltd. · May 13, 2011
K181154 — Med-link Wrist Digital Blood Pressure Monitor, Med-link Upper Arm Digital Blood Pressure Monitor · Shenzhen Med-Link Electronics Tech Co., Ltd. · Aug 2, 2019
K172972 — Wrist Type Automatic Blood Pressure Monitor · Shenzhen Jiacom Technology CO ., Ltd. · Mar 2, 2018
K222125 — Arm Blood Pressure Monitor, models AOJ-30A, AOJ-30B, AOJ-30C, AOJ-30D, AOJ-30E, AOJ-30F and AOJ-30G · Shenzhen AOJ Medical Technology Co., Ltd. · Dec 1, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 28, 2020
Contec Medical Systems Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, Beijing 102401 China
Re: K202757
Trade/Device Name: Electronic Sphygmomanometer, Automatic Blood Pressure Monitor CONTEC08A/CONTEC08C/ABPM50 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 24, 2021 Received: March 29, 2021
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K202757
#### Device Name
Trade Name: Electronic Sphygmomanometer, Automatic Blood Pressure Monitor Model: CONTEC08A/CONTEC08C/ABPM50
Indications for Use (Describe)
The Electronic Sphygmomanometer and Automatic Blood Pressure Monitor are intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# Tab #4 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K202757
- Date of Preparation 1.
03/19/2021
- 2. Sponsor
## Contec Medical Systems Co., Ltd.
No.112 Qinhuang West Street, Economic & Technical Development Zone, Qinhuangdao, Hebei Province, 066004,China
Establishment Registration Number: 3006979678 Contact Person: Xueyong Li Position: Quality Manager Tel: +86-355-8015490 Fax: +86-355-8015490 Email: lxy1011@163.com
- 3. Submission Correspondent
Mr. Ray Wang
Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401
Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Electronic Sphygmomanometer; Automatic Blood Pressure Monitor Common Name: system, measurement, blood-pressure, non-invasive Model(s): CONTEC08A/CONTEC08C/ABPM50
Regulatory Information: Classification Name: Noninvasive Blood Pressure Measurement System;
Classification: II: Product Code: DXN; Regulation Number: 21 CFR 870.1130; Review Panel: Cardiovascular ;
Indication For Use Statement:
The Electronic Sphygmomanometer and Automatic Blood Pressure Monitor are intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.
#### 5. Device Description
The proposed devices, ABPM50 Automaitc Blood Pressure Monitor, CONTEC08A and CONTEC08C Electronic Sphygmomanometers are battery driven automatic non-invasive Blood Pressure Monitor. They can automatically complete the inflation and BP measurement, which can measure systolic, and diastolic blood pressure as well as the pulse rate at upper arm within its claimed range and accuracy via the oscillometric technique. User can select the unit of the measurement: mmHg or kPa. ABPM50 and CONTEC08A and CONTEC08C can be only used on adult individuals.
The devices have the data storage function for data review including measurement time, systolic blood pressure, diastolic blood pressure and pulse rate.
ABPM50 and CONTEC08A have physiological over-limit prompt function which can be turned on or off by users. When the measurement results exceed the over-limit prompt limit, the physiological over-limit prompt function will be triggered. The over-limit can be set by users, and the low limit must be lower than the corresponding high limit.
In addition, all of the three proposed devices have technical over-limit prompt function, which will be triggered when the battery voltage is low, and this technical over-limit prompt function can not be cancelled unless being closed or the power replaced.
The main differences among the three (CONTEC08A / CONTEC08C / ABPM50) are as follows:
{5}------------------------------------------------
| model | CONTEC08A | CONTEC08C | ABPM50 |
|----------------------|--------------------------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Trade<br>Name | Electronic<br>Sphygmomanometer | Electronic<br>Sphygmomanometer | Automatic Blood Pressure Monitor |
| Over limit<br>prompt | yes | no | yes |
| display | TFT LCD display | Segment code screen<br>display | TFT LCD display |
| Storage<br>function | Three users, each user<br>stores 100 groups of<br>data | Three users, each user<br>stores 99 groups of data | General single blood pressure<br>measurement storage 300 groups of<br>data , automatic blood pressure<br>measurement (350 groups of data) |
| Power<br>Supply | four alkaline batteries<br>or 5V power adapter | four alkaline batteries or<br>5V power adapter | two alkaline batteries |
#### 6. Identification of Predicate Device Predicate Device: 510(k) Number: K191180
Product Name: Electronic Blood Pressure Monitor: Models AOJ-30A and AOJ-30B Manufacturer: Shenzhen AOJ Medical Technology Co., Ltd.
#### 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ハ IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- > ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11 Edition 2.0 2015-01 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
{6}------------------------------------------------
- > IEC 80601-2-30 Edition 1.1 2013-07, Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- ) ISO 81060-2 Third edition 2018-11 Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type [including: Amendment 1 (2020)]
- > ISO 10993-5 Third Edition 2009-06-01. Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
- > ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Biocompatibility)
- Clinical Test Conclusion 8.
Controlled human clinical studies were conducted on proposed device with predicates in accordance with ISO 81060-2, mean deviation, standard deviation has been evaluated per clinical validation. The clinical trial results verify that the clinical accuracy of the proposed device is not inferior to that of predicate device.
Total 85 subjects are included in each clinical study, the results of proposed device meet the performance parameters claimed in user manual, and the proposed device complies with ISO 81060-2.
{7}------------------------------------------------
K202757
# 9. Substantially Equivalent (SE) Comparison
| Item | Proposed Device<br>CONTEC08A/CONTEC08C/ABPM50 | | Predicate Device<br>AOJ-30A and AOJ-30B<br>K191180 | Remark | |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|------------|
| Manufacture | Contec Medical Systems Co., Ltd. | | Shenzhen AOJ Medical Technology<br>Co., Ltd. | -- | |
| Product Code | DXN | | DXN | SE | |
| Regulation Number | 21 CFR 870.1130 | | 21 CFR 870.1130 | SE | |
| Indication for Use | The Electronic Sphygmomanometer and Automatic Blood Pressure<br>Monitor are intended to measure the systolic and diastolic blood<br>pressure as well as the pulse rate of adult person via non-invasive<br>oscillometric technique at medical facilities or at home. | | The Electronic Blood Pressure Monitor is intended to<br>measure the systolic and diastolic blood pressure as well<br>as the pulse rate of adult person via non-invasive<br>oscillometric technique at medical facilities or at home. | SE | |
| Clinical Use | Medical Facilities and Home Use | | Medical Facilities and Home Use | SE | |
| Patient<br>Population | Adult | | Adult | SE | |
| Measurement Type | Upper arm | | Upper arm | SE | |
| Measurement<br>Principle | Oscillometric | | Oscillometric | SE | |
| Components | LCD / Key / Cuff / MCU / Pump / Batteries | | LCD / Key / Cuff / MCU / Pump / Batteries | SE | |
| Power Source | CONTEC08A/CONTEC08C | 4x1.5V(4 "AA" alkaline batteries);<br>5V adapter, optional | 4x1.5V | Analysis 1 | |
| | ABPM50 | 2x1.5V(2 "AA" alkaline batteries) | | | |
| Physical<br>Dimensions | CONTEC08A/CONTEC08C | 130(L)*110(W)*80(H) mm | Approx: 138 mm(Length)x120<br>mm(Width)x59 mm(Height) | Analysis 1 | |
| | ABPM50 | 128(L)*69(W)*36 mm(H) | | | |
| | CONTEC08A/CONTEC08C | 300 gram(without batteries) | | Approx: 483.8 g, excluding battery | Analysis 1 |
| Weight | ABPM50 | 240 gram(without batteries) | | | |
| Measurement | Blood Pressure | 10~270 mmHg | Blood Pressure | 30~255 mmHg | Analysis 2 |
| Range | Pulse Rate | 40~240BPM | Pulse Rate | 40-199 bpm | |
| Accuracy | Static Pressure | ±3 mmHg | Static Pressure | ±3 mmHg | SE |
| | Pulse | ±5% | Pulse | ±5% | |
| Arm Circumference | 22-32 cm and 32-43cm | | 22 cm~36 cm | | Analysis 2 |
| Patient Contact<br>Material | PVC+ Nylon+Ployester | | Cuff-Terylene | | Analysis 3 |
| Electrical safety | IEC 60601-1 | | IEC 60601-1 | | SE |
| | IEC 60601-1-11 | | IEC 60601-1-11 | | |
| EMC | IEC 60601-1-2 | | IEC 60601-1-2 | | SE |
| Performance test | IEC80601-2-30 | | IEC80601-2-30 | | SE |
| Clinical data | ISO81060-2 | | ISO81060-2 | | SE |
| Biocompatibility<br>test | Cytotoxicity | | Cytotoxicity | | SE |
| | Skin Sensitization | | Skin Sensitization | | |
| | Skin Irritation | | Skin Irritation | | |
| Operation<br>Environments | +5 °C~40 °C | | + 5°C~+40°C, | | Analysis 1 |
| | 15 %RH~85 %RH (Non-condensing) | | 15%RH~90%RH | | |
| | 700 hPa~1060 hPa | | 70 kPa~106 kPa | | |
| Storage<br>Environments | -20 °C~+55 °C; | | - 20°C~+ 55°C, | | Analysis 1 |
| | Relative humidity: ≤95 %; No corrosive gas and drafty. | | 10%RH~93%RH | | |
| | 700 hPa~1060 hPa | | 70 kPa~106 kPa | | |
Table 7-1 General Comparison
{8}------------------------------------------------
{9}------------------------------------------------
#### Analysis 1
The proposed device is different in Power Source, Waight, Operation Environments and Storage Environments from the predicate device. The differences is small, and the proosed device has passed the II 200601-1-1 test, the safety and performance of the proposed device can be guaranteed. By complying with non-clinical the proposed device is determined to be substantially equivalency with predicate device.
### Analysis 2
The proposed device is different in Measuremence from the predicate device. But the proposed device has passed the IEC8601-2-30 and ISO81069-2 test, the performance of the proposed device can be guaranted, by complying with not-clinical the proposed device is determined to be substantially equivalency with predicate device.
### Analysis 3
The proposed device is different in Patent Contact from the predicate device has passed the Biocompatibility tost (including Cytotoxicity , Skin Sensitzation , Skin Intritation ), we believe the effect the effectiveness and safety compared with the predicate device, the proposed device is determined to be substantially equivalency with predicate device.
### 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.