TL100 TENSMETER BLOOD PRESSURE MONITORING SYSTEM

K022783 · Tensys Medical, Inc. · DXN · Nov 19, 2002 · Cardiovascular

Device Facts

Record IDK022783
Device NameTL100 TENSMETER BLOOD PRESSURE MONITORING SYSTEM
ApplicantTensys Medical, Inc.
Product CodeDXN · Cardiovascular
Decision DateNov 19, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

This device is intended for use by medically trained personnel in a clinical setting to continually monitor and display diastolic, systolic, and mean blood pressures and pulse rate. The device is intended for use on adult patients with a palpable pulse.

Device Story

TL100 Tensymeter is a non-invasive blood pressure monitor; utilizes single-patient-use pressure sensor applied to wrist over radial artery; electronic interface module processes signals. Proprietary algorithms analyze radial artery pressure to derive diastolic, systolic, mean blood pressure, and pulse rate. Used by medically trained personnel in clinical settings. Output displayed as numerical values and pressure waveform; assists clinicians in continuous hemodynamic monitoring; enables real-time clinical decision-making for adult patients.

Clinical Evidence

Clinical testing performed to validate accuracy against AAMI Standard SP-10-1992. Results met or exceeded accuracy requirements for non-invasive blood pressure monitoring.

Technological Characteristics

Semiconductor pressure-sensing element; wrist-worn form factor; electronic interface module. Patient contact materials compliant with ISO 10993-1. Electrical safety per IEC 60601-1; electromagnetic compatibility per IEC 60601-1-2.

Indications for Use

Indicated for continuous monitoring of diastolic, systolic, mean blood pressure, and pulse rate in adult patients with a palpable pulse in a clinical setting.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KC022783 NOV 1 9 2002 # 510k Summary # 510k Summary for the TL100 Tensymeter Non-invasive Blood Pressure Monitor #### Submitter: Tensys Medical, Inc. 5825 Oberlin Drive Suite 100 San Diego, CA 92121 Phone 858-552-1941 858-552-1944 Fax #### Contact: Stuart L. Gallant, President #### Date prepared: #### Trade Name: TL100 Tensymeter Non-invasive Blood Pressure Monitor #### Common name: Non-invasive blood pressure measurement system. ## Classification: Class II per 21 CFR 870.1130 Non-invasive blood pressure measurement system. # Predicate Device: Tensys Medical, Inc. believes that the TL100 Tensymeter is substantially equivalent to the Tensys Medical Inc. TL10 Tensymeter(K020537) Image /page/0/Picture/17 description: The image shows the logo for Tensys Medical. The logo features the word "Tensys" in a bold, italicized font. Above the word "Tensys" is a line that resembles a heart rate monitor. Below the word "Tensys" is the word "medical" in a smaller font. {1}------------------------------------------------ ## Device Description: The TL100 Tensymeter is a non-invasive blood pressure monitor that utilizes a single patient use non-invasive pressure sensor placed on the wrist, over the radial artery, and an electronic interface module. The device uses proprietary algorithms to analyze the radial artery pressure and display diastolic, systolic, mean blood pressure, pulse rate and a pressure waveform. ## Intended Use: This device is intended for use by medically trained personnel in a clinical setting to continuously monitor and display diastolic, systolic, and mean blood pressures. The device is intended for use on adult patients with a palpable pulse. ## Comparison of Technological Characteristics of New Device to Predicate Device(s): The design of the TL100 Tensymeter utilizes a semiconductor pressure-sensing element applied to the wrist, over the radial artery, to obtain a pressure waveform in the same manner as the predicate device. # Clinical Tests: Clinical test were performed which show test results which meet or exceed the accuracy requirements of AAMI Standard SP-10 -1992 #### Non-Clinical Tests: Patient Contact Materials ISO 10993-1 Electrical Safety IEC 60601-1 ## Electromagnetic Compatibility IEC 60601-1-2 Image /page/1/Picture/13 description: The image shows the logo for Tensys Medical Inc. The logo is black and white and features the word "Tensys" in a bold, stylized font. Below the word "Tensys" is the word "Medical, Inc." in a smaller font. Above the word "Tensys" is a line that resembles a heart rate monitor. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 1 9 2002 Tensys Medical. Inc. c/o Mr. Stuart L. Gallant President and Chief Operating Officer 5825 Oberlin Drive, Suite 100 San Diego, CA 92121 Re: K022783 Trade Name: TL100 Tensymeter Blood Pressure Monitoring System Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: August 21, 2002 Received: August 22, 2002 Dear Mr. Gallant: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Mr. Stuart L. Gallant Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Dna Humulu for Bram D. Zuckerman, M.D. Bram D. Zuckerman, M. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Statement of Indications for Use 510(k) Number: Device Name: TL100 Tensymeter Blood Pressure Monitoring System ## Indications for Use: This device is intended for use by medically trained personnel in a clinical setting to continually monitor and display diastolic, systolic, and mean blood pressures and pulse rate. The device is intended for use on adult patients with a palpable pulse. Division of Cardiovascular & Respiratory Devices 510(k) Number K022783 DSF 11/1/02 Image /page/4/Picture/6 description: The image shows the logo for Tensys Medical. The logo features the word "Tensys" in a bold, sans-serif font, with a line above the word that resembles a heartbeat. Below the word "Tensys" are the words "medical inc.", in a smaller font.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...