PRIMO NONINVASIVE BLOOD PRESSURE MEASUREMENT SYSTEM

K053185 · Medwave, Inc. · DXN · Jan 24, 2006 · Cardiovascular

Device Facts

Record IDK053185
Device NamePRIMO NONINVASIVE BLOOD PRESSURE MEASUREMENT SYSTEM
ApplicantMedwave, Inc.
Product CodeDXN · Cardiovascular
Decision DateJan 24, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

The Medwave, Inc. Primo™ Noninvasive Blood Pressure Measurement System is a hand held non-invasive blood pressure measurement system intended to be used on adult patients with wrist circumferences of 11 cm - 22 cm by trained medical personnel to measure systolic, diastolic blood pressure and pulse rate.

Device Story

Hand-held non-invasive blood pressure monitor; measures systolic/diastolic BP and pulse rate from radial artery at wrist. Device consists of hard plastic housing with user interface and pressure sensor. Operates by applying downward pressure to radial artery via sensor; uses 'sweep' technique (increasing pressure ~10mmHg/heartbeat, then decreasing) to generate pressure waveform. Microcomputer-controlled system performs curve fitting on beat amplitude vs. hold-down pressure to identify peak and filter artifacts. Used by trained medical personnel in clinical settings. Output displayed on device screen; assists clinicians in patient hemodynamic assessment.

Clinical Evidence

Bench testing only. Device performance validated against intra-arterial line pressure waveforms per AAMI standards. Results demonstrate mean difference of ±5mmHg or less with standard deviation of 8mmHg or less. Pulse measurement accuracy ±5 bpm or 10%. Biocompatibility testing per ISO 10993-1. Electrical safety and EMC testing per IEC 60601.

Technological Characteristics

Microcomputer-controlled, battery-powered, hand-held wrist sphygmomanometer. Uses proprietary pressure sensor for oscillometric radial artery waveform capture. Operating range: 10°C-40°C, 10-85% humidity. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601. Software-based signal processing.

Indications for Use

Indicated for adult patients with wrist circumferences of 11-22 cm for measurement of systolic/diastolic blood pressure and pulse rate by trained medical personnel.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SECTION 2. SUMMARY AND CERTIFICATION ## A. 510(K) SUMMARY ## Summary of Safety and Effectiveness In accordance with 21 CFR 807.92, the following information constitutes the Medwave, Inc. summary for the Primo TM. | SUBMITTER'S NAME: | Medwave, Inc. | |---------------------|-------------------------------------------------| | ADDRESS: | 4382 Round Lake Road West<br>St. Paul, MN 55112 | | CONTACT PERSON: | Donna R. Lunak | | TELEPHONE NUMBER: | 651-639-1227 | | FAX NUMBER: | 651-639-1338 | | DATE OF SUBMISSION: | 11/14/05 | ### 1. Identification of device Proprietary Name: Primo TM Common Name: Wrist Sphygmomanometer Classification Status: Class II per regulations 870.1130 Product Codes ## 2. Equivalent devices Medwave, Inc. believes the Primo™ Noninvasive Blood Pressure Measurement System is substantially equivalent to Medwave, Inc. Vasotrax® Portable Wrist Noninvasive Blood Pressure Measurement System (K001898). ### 3. Description of the device The Primo™ Noninvasive Blood Pressure Measurement System is a hand-held non-invasive blood pressure measurement system that measure systolic, diastolic blood pressure and pulse rate from the user's wrist. The system is contained in hard plastic housings that contain a user interface panel and a wrist sensor. The characteristics of the arterial waveforms are recorded with a unique pressure sensor that is placed over the radial artery. This system provides a single reading of blood pressure, using a pressure sensor placed on the wrist over the radial artery. This sensor is noninvasive. {1}------------------------------------------------ # 053185 215 #### 4. Intended Use The Medwave, Inc. Primo™ Noninvasive Blood Pressure Measurement System is a hand held non-invasive blood pressure measurement system intended to be used on adult patients with wrist circumferences of 11 cm - 22 cm by trained medical personnel to measure systolic, diastolic blood pressure and pulse rate. #### 5. Technological characteristics, comparison to predicate device. Like the predicate device, the Primo™ Noninvasive Blood Pressure Measurement System measures the diastolic, systolic blood pressure and pulse rate from the wrist using oscillometric methods. Both the Primo ™ Noninvasive Blood Pressure Measurement System and the Vasotrax® Portable Wrist Noninvasive Blood Pressure Measurement System (K001898) are microcomputer controlled. Like the Vasotrax® Portable Wrist Noninvasive Blood Pressure Measurement System (K001898), the Primo™ Noninvasive Blood Pressure Measurement System display systolic, diastolic blood pressures ranging from 40 and 240 mmHg. Both systems have a blood pressure measurement accuracy of a mean difference of ± 5mmHg or less with a standard deviation of 8mmHg or less. The pulse measurement range is the same for both the Primo™ Noninvasive Blood Pressure Measurement Systcm and the Vasotrax® Portable Wrist Noninvasive Blood Pressure Measurement System (K001898), from 40 - 200 bpm. The accuracy of the pulse measurements are + 5 bpm or 10% of the measured pulse frequency. (Supported by study results in the supporting Clinical Data on file at Medwave, Inc.) Both Primo™ Noninvasive Blood Pressure Measurement System and the Vasotrax® System (K001898) utilize the application of pressure to the artery (by the sensor); the counter pressure in the artery produces a pressure waveform. When maximum amplitude is achieved, mean blood pressure is calculated. The Primo™ Noninvasive Blood Pressure Measurement System and the Vasotrax® System (K001898) use a unique "sweep" technique for applying pressure to a the radial artery: downward pressure is applied by the sensor to the radial artery at a rate of ~10mmHg per heart beat increasing as the beat amplitude increases and decreasing rapidly when the beat amplitude begins to decrease. A curve fit is made using the amplitude of each beat versus the hold down pressure to form the bell shaped curve. This curve fit is used to determine the true peak that might occur between pulses as well as to filter out small variations due to artifacts or aberrancies. {2}------------------------------------------------ K053135 p3/5 Technological characteristics, comparison to predicate device (continued) Both the Primo Noninvasive Blood Pressure Measurement System and the Vasotrax® (K001898) utilize Medwave's proprietary algorithms in analyzing the pressure waveforms to calculate the systolic and diastolic readings. Parameters are extracted from the waveforms and a set of coefficients is applied to them, yielding systolic and diastolic pressures. The algorithms have been tested against intra arterial line pressure waveforms and proven to meet industry standards set by the American Medical Instrumentation (AAMI), mean difference of + 5mmHg or less with a standard deviation of 8mmHg or less. (Supported by study results in Appendix H) The Primo™ Noninvasive Blood Pressure Measurement System, as well as the Vasotrax ® System (K001898) has a power switch and a display. The operating environment of 10°C - 40°C and 10% to 85% relative humidity. Both the Primo System and the Vasotrax ® Portable Wrist Noninvasive Blood Pressure Measurement System (K001898) are battery powered. Any minor differences in the appearance, technology, or manufacture of the Primo ™ Noninvasive Blood Pressure Measurement System device and the predicate device do not raise any new questions of safety or effectiveness. Associated risks posed by the Primo™ Noninvasive Blood Pressure Measurement System are thought to be no more than those of a well designed automated cuff-based noninvasive blood pressure devices currently marketed in the interstate commerce. Both the device and the sensor have been designed to minimize the risk to patients from excessive pressure or sensor failure caused by cither normal device use by the caregiver and/or clinical abuse. {3}------------------------------------------------ ## 6. Discussion of performance testing. An extensive collection of tests has been conducted and successfully completed. Summary follows: | Requirements | ANSI/AAMI SP10-2002<br>American National Standards<br>for Electronic or Automated<br>Sphygmomanometers | IEC60601 | Center For Devices<br>And Radiological<br>Health Noninvasive<br>Blood Pressure<br>Monitor Guidance<br>Document | Other<br>As<br>Listed | |----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|------------|----------------------------------------------------------------------------------------------------------------|----------------------------------| | Device Labeling | X | X | X | ---------- | | Outer Container<br>Labeling | X | X | X | ---------- | | Information Manual | X | X | X | ---------- | | Component Labeling | X | X | X | ---------- | | Power System Labeling | X | X | X | ---------- | | Storage Conditions --<br>20°C (-40°F) - 50°C<br>(122°F) | X | X | ---------- | ---------- | | Operating Temperature<br>Conditions 10°C<br>(50°F) - 40°C (104°F) | X | X | ---------- | ---------- | | Operating Humidity<br>Conditions 15 90<br>percent<br>(noncondensing) | X | X | ---------- | ---------- | | Operating Range in<br>Altitude Conditions -<br>170 to 1700 meters<br>(-500 to 5000 feet).<br>referenced to sea level | X | X | ---------- | ---------- | | Vibration and Shock | ---------- | ---------- | ---------- | NSTA | | Voltage Range | N/A | N/A | N/A | N/A | | Life test minimum of<br>10,000 full scale cycles | X | ---------- | ---------- | ---------- | | Maximum cuff<br>pressure | N/A | N/A | N/A | N/A | | Cuff deflation | N/A | N/A | N/A | N/A | | Electrical safety | ---------- | X | ---------- | IEC601 | | Conductive<br>components | ---------- | X | ---------- | ---------- | | Pressure indicator<br>accuracy | X | ---------- | ---------- | ---------- | | Requirements | ANSI/AAMI SP10-2002<br>American National Standards<br>for Electronic or Automated<br>Sphygmomanometers | IEC60601 | Center For Devices<br>And Radiological<br>Health Noninvasive<br>Blood Pressure<br>Monitor Guidance<br>Document | Other<br>As<br>Listed | | Auscultatory method as<br>the reference standard | N/A | N/A | N/A | N/A | | Intraarterial method as<br>the reference standard | X | ---------- | ---------- | ---------- | | Battery indicator | X | ---------- | ---------- | ---------- | | Requirements for<br>devices with manual<br>inflation | N/A | N/A | N/A | N/A | | Comparison Testing | N/A | N/A | N/A | N/A | | Foreign Standards | N/A | N/A | N/A | N/A | | Software Testing | ---------- | ---------- | ---------- | Medwave<br>795-0092,<br>795-0094 | | Electromagnetic<br>Compatibility | ---------- | X | ---------- | ---------- | | Biocompatibility | ---------- | ---------- | X | ISO10993<br>-1- | | Sterilization | N/A | N/A | N/A | N/A | | Packaging | N/A | N/A | N/A | N/A | | Shelf Life | N/A | N/A | N/A | N/A | {4}------------------------------------------------ ### 7. Conclusion Based on extensive performance testing and a comparison to the predicate device, it is the conclusion of Medwave, Inc. that the Primo™ noninvasive Blood Pressure Measurement System is substantially equivalent to devices already on the marked (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness. {5}------------------------------------------------ Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains an abstract image of an eagle. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged around the top of the circle. JAN 2 4 2006 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Medwave, Inc. c/o Ms. Donna R. Lunak Vice President, Regulatory Affairs 4382 Round Lake Road West Arden Hills, MN 55112 Re: K053185 Trade Name: Primo™ Noninvasive Blood Pressure Measurement System Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: November 14, 2005 Received: November 14, 2005 Dear Ms. Lunak: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Ms. Donna Lunak Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bhimmar for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ## Indications for Use 510(k) Number (if known): s 3/85 Device Name: Primo™ Indications For Use: The Primo™ Noninvasive Blood Pressure Measurement System is a hand held noninvasive blood pressure measurement system intended to be used on adult patients with wrist circumferences of 11 cm - 22 cm by trained medical personnel to measure systolic and diastolic blood pressure and pulse rate. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Bhumma SIGIT on of Cardlov scular Davices S . Jig Number Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...