VANTAGE ABI

K103693 · Summit Doppler Systems, Inc. · DXN · Feb 15, 2011 · Cardiovascular

Device Facts

Record IDK103693
Device NameVANTAGE ABI
ApplicantSummit Doppler Systems, Inc.
Product CodeDXN · Cardiovascular
Decision DateFeb 15, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The Vantage ABI is a non-invasive device used to gauge the lower extremity arterial system using pneumatic volume plethysmography recording (PVR) and oscillometric systolic blood pressure to assist in the diagnosis of vascular disease.

Device Story

Vantage ABI is a non-invasive diagnostic system for peripheral arterial disease (PAD). It utilizes four pneumatic cuffs placed on the patient's limbs to record volume plethysmography (PVR) waveforms and measure oscillometric systolic blood pressure. The device processes these inputs to calculate the Ankle-Brachial Index (ABI). It is intended for clinical use to assist healthcare providers in diagnosing vascular disease. The system operates on internal battery or external line power. Output consists of PVR waveforms and pressure measurements, which the clinician uses to evaluate arterial health and inform clinical decision-making regarding vascular status.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and performance comparisons to predicate devices.

Technological Characteristics

Non-invasive diagnostic system; utilizes pneumatic cuffs for volume plethysmography (PVR) and oscillometric blood pressure measurement; powered by internal battery or external line supply; performs automated ABI calculation.

Indications for Use

Indicated for patients requiring assessment of the lower extremity arterial system to assist in the diagnosis of vascular disease, including peripheral arterial disease.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K103693 pg 1 of 1 ## FEB 1 5 2011 #### SAFETY AND EFFECTIVENESS SUMMARY Summit Doppler Systems, Inc. Vantage ABI Name and Address: Summit Doppler Systems, Inc. 4680 Table Mountain Dr. #150 Golden, CO 80403 Phone: (303) 423-7572 Fax: (303) 431-5994 Contact: Ken Jarrell - President Preparation Date: November 30, 2011 Device Name: Vantage ABI Common Name: Plethysmograph Classification: Class II: FR Number Product Code Cardio Vascular Monitoring Device 820.2780 IOM Indications for Use: The Vantage ABI is a non-invasive device used to gauge the lower extremity arterial system using pneumatic volume plethysmography recording (PVR) and oscillometric systolic blood pressure to assist in the diagnosis of vascular disease. Description: The Vantage ABI performs the ankle-brachial index examination to assist in the diagnosis of peripheral arterial disease. The system will be used to assess peripheral arterial disease using pneumatic volume plethysmography recording (PVR), estimate the systolic blood pressure for each limb and calculate the Ankle Brachial Index. The unit will record a PVR waveform at each ankle and measure pressures in each limb. Four cuffs will be connected. This unit can be powered from its internal battery or from an external line-powered supply. Substantial Equivalence: The Vantage ABI is substantially equivalent to the cleared devices shown below. BioMedix, PADnet K042616, Cleared 10/12/2004 Summit Doppler Systems, Vista AVS K063600, Cleared 12/22/2006 This device uses pneumatic volume plethysmography recording (PVR) and Technologies Summary: Ankle Brachial Index (ABI) to assist in the diagnosis of vascular disease in the same manner as the predicate devices. Based on comparisons of device features, materials, intended use and Conclusion: performance, and user instructions, the Vantage ABI is shown to be substantially equivalent to the commercially available and legally marketed device indicated above. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and body. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Summit Doppler Systems, Inc. c/o Ms. Dawn Tibodeau Responsible Third Party Official TUV SUD America, Inc. 1775 Old Highway 8 NW New Brighton, MN 55112-1891 FEB 1 5 201 Re: K103693 Trade Name: Vantage ABI Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: January 28, 2011 Received: January 31, 2011 #### Dear Ms. Tibodeau: . the same of the states of the We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. . . . . . . . . . . . . 11 - 1 - 1 - 1 - 1 . {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdaggov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number: K103693 Device Name: Vantage ABI The Vantage ABI is a non-invasive device used to gauge the lower extremity arterial system using pneumatic volume plethysmography recording (PVR) and oscillometric systolic blood pressure to assist in the diagnosis of vascular disease. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter-Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) W.M.A. (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K103693
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...