K052671 · Hoca Medical Microsystem, Inc. · DXN · Feb 3, 2006 · Cardiovascular
Device Facts
Record ID
K052671
Device Name
HOCA EZ-CARE BP108, TENDER-CARE BP109
Applicant
Hoca Medical Microsystem, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Feb 3, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Intended Use
The device is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.25" - 7.75".
Device Story
Noninvasive blood pressure monitor; uses oscillometric method to measure systolic/diastolic pressure and pulse rate. Input: pressure oscillations sensed via wrist cuff and closed air pipe system. Processing: microcomputer automatically senses pulse signal characteristics; identifies systolic pressure as point where cuff oscillation amplitude begins to increase and diastolic pressure as point where amplitude stops decreasing. Output: digital display of blood pressure and pulse rate. Used at home by adult patients. Provides objective data for patient monitoring; assists in tracking blood pressure trends.
Clinical Evidence
Bench testing only. Compliance with EN 60601-1:1990 (electrical safety), EN 60601-1-2 (EMC), and ANSI/AAMI SP10-1992 (Electronic or Automated Sphygmomanometers).
Technological Characteristics
Oscillometric measurement principle; microcomputer-based signal processing; wrist-worn inflatable cuff (5.25"-7.75"); digital display. Standards: EN 60601-1, EN 60601-1-2, ANSI/AAMI SP10.
Indications for Use
Indicated for adults over age 18 for home-based noninvasive measurement of systolic/diastolic blood pressure and pulse rate via wrist cuff (5.25"-7.75" circumference).
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Automatic Digital Blood Pressure Monitor, APM BP108A (K040159)
Related Devices
K041400 — FORTIFIER DIGITAL WRIST BLOOD PRESSURE MONITOR, MODEL LF-01 · Fortifier International Co., Ltd. · Aug 24, 2004
K072860 — YA HORNG DIGITAL UPPER ARM BLOOD PRESSURE MONITOR, BP-100J, BP-110J; DIGITAL WRIST BLOOD PRESSURE MONITOR, BP-500,BP-500 · Ya Horng Electronic Co., Ltd. · Feb 25, 2008
K141416 — BLOOD PRESSURE MONITOR · Shenzhen Pump Medical System Co., Ltd. · Oct 9, 2014
K120673 — DIGITAL AUTOMATIC WRIST BLOOD PRESSURE MONITOR WBPM15 SERIES · Grandway Technology (Shenzhen) Limited · Feb 14, 2013
{0}------------------------------------------------
FEB S 2006
K052671
71/2
# HOCA Medical Microsystem, Inc.
No.336-2, Chien Kuo Road, Chu Nan Chen, Miaoli Shien, 35045, Taiwan Telephone (886) 37-481909 (886) 37- 484906 Fax # e-mail: neo@smarthoca.com Website : www.smarthoca.com
#### 510(k) SUMMARY " દા
HOCA Medical Microsystem, Inc. Submitter's Name:
> No.336-2, Chien Kuo Road, Miaoli Shien, 350, Taiwan Fax: 886-37-484906 Telephone: 886-37-481909 e-mail : neo@smarthoca.com
Date summary prepared:
Proprietary Name:
Device Name:
#### HOCA EZ-CARE BP108, TENDER-CARE BP109 NONINVASIVE BLOOD PRESSURE MEASUREMENT SYSTEM Common or Usual Name: Blood Pressure Monitor, Class II, Classification Name: 21 CFR 870.1130
September 19, 2005
Indications for Use:
The device is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.25" - 7.75".
### Description of the device:
HOCA EZ-CARE BP108 and TENDER-CARE BP109 use the Oscillometric method to measure the blood pressure. The Oscillometric method is adopted clinically to measure the blood pressure recently. It is not needed to use the stethoscope, as in the traditional measuring method, to monitor the Korotkov sound when deciding the systolic or diastolic pressure. The Oscillometric method senses the vibrating signal via the closed air pipe system and utilizes the microcomputer to automatically sense the characteristics of the pulse signal. Through simple calculation, clinically proven, the reading can reflect the accurate real blood pressure, and the systolic pressure is defined as the pressure when the cuff pressure oscillating amplitude begins to increase and the diastolic pressure as the pressure when the cuff pressure oscillating amplitude stops decreasing.
{1}------------------------------------------------
K052471
F2/2
Image /page/1/Picture/1 description: The image shows a logo with a stylized leaf shape in black and white. Inside the leaf, there is a white square with a black plus sign in the center. To the right of the leaf, there is a cursive word that appears to be "foca".
### HOCA Medical Microsystem, Inc.
No.336-2, Chien Kuo Road, Chu Nan Chen, Miaoli Shien, 35045, Taiwan (886) 37- 484906 Telephone (886) 37- 481909 Fax # e-mail: neo@smarthoca.com Website : www.smarthoca.com
Performance Testing:
Electric Safety Requirement Test Report of EN 60601-1:1990 and EMC test report of EN 60601-1-2 ( EN 55011:1991 and EN 61000-4-2:1995) ANSI/AAMI SP10-1992 Electronic or Automated Sphygmomanometers
Legally marketed device for substantial equivalence comparison:
Automatic Digital Blood Pressure Monitor, APM BP108A (K040159)
Summary for substantial equivalence comparison:
Same characteristics: intended use, technological characteristics, power supply, display, measuring range, accuracy, operating and storage environments. The new device HOCA BP108 are all the same as the predicate device APM BP108A.
Different characteristics: the new device HOCA BP109 has the different appearance for the "start key".
They are decided to be substantially equivalent.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 3 2006
HOCA Medical Microsystem, Inc. c/o Dr. Jen Ke-Min ROC Chinese-European Industrial Research Society No. 58 Fu-Chiun St. Hsin-Chu City Taiwan ROC
Re: K052671
Trade Name: HOCA Ez-Care BP108 and Tender-Care BP109 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: January 12, 2006 Received: January 18, 2006
Dear Dr. Ke-Min.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Dr. Jen Ke-Min
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bhimimor for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with a leaf-like shape on the left and the word "Coca" on the right. Inside the leaf shape, there is a plus sign. The leaf is shaded in black, and the word "Coca" is written in a cursive font.
# HOCA Medical Microsystem, Inc.
No.336-2, Chien Kuo Road, Chu Nan Chen, Miaoli Shien, 35045, Taiwan Telephone (886) 37-481909 (886) 37- 484906 Fax # e-mail: neo@smarthoca.com Website : www.smarthoca.com
### Indications for Use
510(k) Number:
K052671
Device Name: HOCA Medical Microsystem, Inc. HOCA EZ-CARE BP108 and TENDER-CARE BP109
. Indications for use:
The device is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.25" - 7.75".
Prescription Use
AND/OR
Over-The-Counter Use __
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummmmas
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K052671
Cl
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.