KD-768 WRIST MEASUREMENT ELECTRONIC BLOOD PRESSURE MONITOR

K051847 · Kodon(Tianjin)Electronic&Electrical Apparatus Co., · DXN · Jan 5, 2006 · Cardiovascular

Device Facts

Record IDK051847
Device NameKD-768 WRIST MEASUREMENT ELECTRONIC BLOOD PRESSURE MONITOR
ApplicantKodon(Tianjin)Electronic&Electrical Apparatus Co.,
Product CodeDXN · Cardiovascular
Decision DateJan 5, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

KD-768 Wrist Measurement Automatic Electronic Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on anyone each time, except infants and unconscious people, with an air cuff buckled around one's wrist according to the instruction in the user's guide manual.

Device Story

Wrist-worn non-invasive blood pressure monitor; utilizes oscillometric measurement method with silicon integrated pressure sensor. Device features internal pump for automatic cuff inflation/deflation. Input: pressure oscillations from wrist cuff. Processing: electronic interface module analyzes signals to calculate systolic/diastolic blood pressure and pulse rate. Output: results displayed on LCD; memory storage for 120 measurements with timestamp. Used by medical professionals or patients at home. Provides objective blood pressure data to assist in health monitoring and clinical decision-making.

Clinical Evidence

Clinical testing performed to verify accuracy against ANSI/AAMI SP10-1992 standards. Results met or exceeded accuracy requirements specified in the standard.

Technological Characteristics

Oscillometric measurement; silicon integrated pressure sensor; internal pump for automatic cuff inflation/deflation; LCD interface; wrist cuff (6.1023-9.8425 inch circumference); memory for 120 readings. Standards: IEC 601-1, IEC 60602-2-30, GB9706.1-1995, ISO 1833-1977.

Indications for Use

Indicated for use by medical professionals or at home to measure diastolic, systolic blood pressure, and pulse rate in adults via a wrist-worn inflatable cuff (circumference 6.1023-9.8425 inches). Contraindicated for infants and unconscious individuals.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### JAN 5 2006 K05/847 # 510(k) Summary #### Identification of the submitter: | Submitter: | Kodon (Tianjin) Electronic & Electrical Apparatus<br>Co., LTD<br>No 31, Changjiang Road, Nankai District, Tianjin,<br>P.R. China, 300193 | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------| | Telephone number: | 86-22-6052 6082 | | Fax number: | 86-22-6052 6162 | | Contact: | Liu Yi | | Date of Application: | 07/05/2005 | ## Identification of the product: | Device proprietary Name: | KD-768 Wrist Measurement Electronic Blood<br>Pressure Monitor | |--------------------------|-------------------------------------------------------------------------------| | Common name: | Noninvasive blood pressure measurement<br>systems | | Classification name: | Noninvasive blood pressure measurement system<br>Class II per 21 CFR 870.1130 | # Marketed Devices to which equivalence is claimed: | Device | manufacture | 510(k) number | | |--------|--------------------------------------------------------|---------------|--| | KD-726 | Kodon (Tianjin) Electronic and<br>Electrical Co., Ltd. | | | ### Device description: KD-768 Wrist Measurement Blood Pressure Monitor is a Non-invasive blood pressure measurement system for only one person each time. Based on oscillometric and silicon integrate pressure sensor technology, this devise is used to monitor systolic, diastolic blood pressure and pulse rate which will be shown on a LCD with an electronic interface module. Buckling an air cuff around a wrist automatically inflated and released by an internal pump, the device can analyze the signals promptly and display the results and remember circularly for some sets of data. It can storage and show 120 times measuring result with the month, day and time of measuring. ### Intended use: KD-768 Wrist Measurement Automatic Electronic Blood Pressure Monitor is page 1 of 2 {1}------------------------------------------------ intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on anyone each time, except infants and unconscious people, with an air cuff buckled around one's wrist according to the instruction in the user's guide manual. ### Comparison of technological characteristics of new device to predicate devices: The design of KD-768 Wrist Measurement Electronic Blood Pressure Monitor utilizes Oscilliometric measurement method, with an electronic interface module and a silicon integrate pressure sensor and a air cuff automatically inflated and released by an internal pump over the wrist to obtain blood pressure signals in the same manner as the predicate device. #### Clinical Tests: Clinical tests were performed and complied the accuracy requirements of ANSI/AAMISP10-1992. The results meet or exceed the accuracy requirements of ANSI/AAMISP10-1992. #### Non-clinical Tests: tests coincide the following All non-clinical standards, including Electromagnetic Compatibility test. IEC601-1 (1988) Medical electrical equipment --- Part 1: General requirements for safety IEC601-1 (1988) Amendment 2 IEC 60602-2-30: 1995 Medical electrical equipment-part2: Particular requirements for the safety of automatic cycling indirect blood pressure monitoring equipment. GB9706.1-1995 ISO 1833-1977 page 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a black and white logo. The logo consists of a circle with text around the perimeter. Inside the circle is a stylized image of a bird or abstract shape with three curved lines. The text around the circle is difficult to read due to the image quality. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 5 2006 Kodon (Tianjin) Electronic and Electrical Apparatus Co., Ltd. c/o Mr. Liu Yi President No. 31 Changjiang Road Nankai District, Tianjin CHINA Re: K051847 Trade Name: Kodon KD-768 Wrist Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 05, 2005 Received: December 05, 2005 Dear Mr. Yi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Liu Yi Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bhimmarfor Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) number: Device name: KD-768 Wrist Measurement Automatic Electronic Blood Pressure Monitor #### Indications for use: The KD-768 Wrist Measurement Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 6.1023 inches to 9.8425 inches. Prescription Use AND/OR Over-The-counter Use YES (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) Bhummon 510(k) Number
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%