TRANSTEK WRIST BLOOD PRESSURE MONITOR

K123669 · Zhongshan Transtek Electronics Co., Ltd. · DXN · Dec 21, 2012 · Cardiovascular

Device Facts

Record IDK123669
Device NameTRANSTEK WRIST BLOOD PRESSURE MONITOR
ApplicantZhongshan Transtek Electronics Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateDec 21, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

Transtek Wrist Blood Pressure Monitor TMB-1014-BT is a digital monitor intended for use in measuring blood pressure and heartbeat rate in adult patient population with wrist circumference ranging from 13.5 cm to 21.5 cm (about 5 1/4 - 8 1/2 inches). This device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. The Wrist Blood Pressure Monitor compares average blood pressure results to pre-established AHA (American Heart Association) hypertension guideline of 135/85 mmHz. Transtek Wrist Blood Pressure Monitor, TMB-1014-BT is not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.

Device Story

Wrist-worn oscillometric blood pressure monitor; uses inflatable cuff and electronic pressure sensor to detect pulse waves; microprocessor calculates systolic/diastolic pressure and heart rate; detects irregular heartbeats by comparing pulse wave intervals to mean interval (warning if >25% difference); displays results on LCD; includes Bluetooth module for wireless data transmission to external devices; intended for adult home use; provides blood pressure readings and irregular heartbeat alerts; aids in monitoring blood pressure trends relative to AHA guidelines; not diagnostic; requires physician consultation for hypertensive values.

Clinical Evidence

Bench testing only. No clinical data provided. Device conforms to AAMI/ANSI SP10, IEC60601-1, and IEC60601-1-2 standards.

Technological Characteristics

Oscillometric measurement; electronic pressure sensor; inflatable nylon cuff (13.5-21.5 cm); powered by 2 AAA alkaline batteries; Bluetooth connectivity; LCD display; microprocessor-based control; electromagnetic deflation valve. Conforms to ISO14971, AAMI/ANSI SP10, IEC60601-1, IEC60601-1-2, and FCC Part 15.

Indications for Use

Indicated for adult patients with wrist circumference 13.5-21.5 cm for non-invasive measurement of systolic/diastolic blood pressure and heartbeat rate. Detects irregular heartbeats. Not for diagnostic use.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K123669 ### TRANSTEK ### Section 5 - 510(k) Summary Date of Summary Preparation: 11/23/2012 ### DEC 2 1 2012 ### 1. Submitter's Identifications Submitter's Name: ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD. Address: Jin'an Road, Minzhong, Zhongshan City, Guangdong, China Contact Person: Lisa Li Contact Email Address: lisha1@transtek.cn Telephone: 086-760-88282982 ext. 876 Fax: 086-760-85339231 #### 2. Correspondent's Identifications Correspondent's Name: A03 Lab of BTS Address: No.1 Fanghua Street, Hi-tech Zone, Chengdu City, Sichuan, China Contact Person: Leo Wang Contact Email Address: leo.w@hibts.com Telephone: 086-28-86083300 Fax: 086-20-80727399 ### 3. Name of the Device Device Classification Name: System, Measurement, Blood-Pressure, Non-invasive Product Name: TRANSTEK Wrist Blood Pressure Monitor Trade Name: TRANSTEK Models: TMB-1014-BT Classification Panel: Cardiovascular Common/Usual Name: Automatic Blood Pressure Monitor Product Code: DXN Device Classification: Class II Contraindications: None. #### 4. The Predicate Devices TRANSTEK, Wrist Blood Pressure Monitor, Model TMB-1014, K122482 ### 5. Device Description Transtek Wrist Blood Pressure Monitor, TMB-1014-BT is designed to measure the systolic and diastolic blood pressure and heartbeat rate of an individual by using a non-invasive technique in which {1}------------------------------------------------ ### TRANSTEK an inflatable cuff is wrapped around the wrist. Measurement method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating heartbeat rate, which is a well-known technique in the market called the "Oscillometric method". Transtek Wrist Blood Pressure Monitor is single-mounted devices of the main unit and cuff unit. The preformed cuff unit, which is applicable to wrist circumference approximately between 13.5 and 21.5 cm, includes the inflatable bladder and nylon shell. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD. The subject device is powered by two AAA alkaline batteries. The device also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%. Transtek Wrist Blood Pressure Monitor TMB-1014-BT embeds a Bluetooth module that allows it to connect to nearby BT receiving end. The LCD of device displays results. And once measurement is over, the device will start transmission data by BT. Thus users can receive, and display/storage, measurement data from TMB-1014-BT unit through their end devices that embedded BT module. #### 6. Intended Use of Device Transtek Wrist Blood Pressure Monitor TMB-1014-BT is a digital monitor intended for use in measuring blood pressure and heartbeat rate in adult patient population with wrist circumference ranging from 13.5 cm to 21.5 cm (about 5 1/4 - 8 1/2 inches). This device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. The Wrist Blood Pressure Monitor compares average blood pressure results to pre-established AHA (American Heart Association) hypertension guideline of 135/85 mmHz. Transtek Wrist Blood Pressure Monitor, TMB-1014-BT is not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated. #### 7. Design Control Activities and Performance Tests Summary Design control activities for this modification were performed and bench tests have been done. Those performance tests, risk management, and design verification tests provide demonstration that the difference does not raise any new questions of safety and effectiveness. TMB-1014-BT conforms to the following standards: ISO14971, Risk management to medical devices {2}------------------------------------------------ ### TRANSTEK AAMI/ANSI SP10, Safety and performance characteristics IEC60601-1, Electrical safety; IEC60601-1-2, Electromagnetic compatibility FCC Part 15, EMI tests of FCC Radiation & RF rules and regulations Explanation: The cuff of TMB-1014-BT is identical to the predicate's cuff, and the new wireless function does not affect blood pressure monitor measurement function. Therefore we have not done the ISO10993 test and Clinical test. ### 8. Summary of Substantial Equivalence 8.1 Difference between proposed device and the predicate device The only difference between the two devices is that TMB-1014-BT add-on a Bluetooth data communication, what user option, which can transmit measurement results to those end devices that embedded a BT module. ### 8.2 Discussion The Transtek Wrist Blood Pressure Monitor TMB-1014-BT has identical indication for use, fundamental scientific technology, cuff, energy type, environmental specifications, dimensional specifications, performance specifications, and similar software/firmware, functions, labeling to the predicate device. The only function difference between TMB-1014-BT and the predicate device is that the TMB-1014-BT provides user an optional wireless (Bluetooth) data transmission. It is an add-on function that is entirely independent from the blood pressure monitor measurement function, which does not reply on the wireless connection to carry out a blood pressure measurement and heartbeat rate analysis and display its results. Thus the BT data transmission does not affect the safety and effectiveness of the wrist blood pressure monitor function. All required design control activities have been implemented and all applicable performance tests have been done according with demands of FDA guidance document "Non-Invasive Blood Pressure (NIBP) Monitor Guidance" FDA March 10, 1997. We found that the modified device does not create new significant risk. #### 9. Conclusions The Transtek Wrist Blood Pressure Monitor TMB-1014-BT is substantially equivalent to the predicate device TMB-1014 by having the identical indication for use, identical technologies, and an add-on function which does not impact the safety and effectiveness of the device. --- End of this section --- {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with three curved lines representing its body and wings. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ### DEC 2 1 2012 Zhongshan Tranktek Electronics Co., LTD. c/o Leo Wang A03 Lab of BTS No. 1 Fanghua Street Hi-Tech District Chengdu, Sichuan CHINA 610041 Re: K123669 Trade/Device Name: TRANSTEK Wrist Blood Pressure Monitor Model TMB-1014-BT Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN Dated: October 30, 2012 Received: November 29, 2012 Dear Mr. Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {4}------------------------------------------------ ### Page 2 - Mr. Leo Wang If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours, ## Owen P. Faris -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### ection 4 - Indications for ### 10(k) Number (if known): : 4 ### Device Name: 3 Transtek Wrist Blood Pressure Monitor Model: TMB-1014-BT - ・こ ndications for Use: ت Transtek Wrist Blood Pressure Monitor TMB-1014-B T is a digital monitor intended for use in measuring blood pressure and heartbeat rate in adult patient population with as in mounterine from 13.5 cm to 21.5 cm (about 5 1/4 - 8 1/2 mches). Wrist circumferice ranging from 13.5 cm to 21.5 cm (about 5 1/4 - 8 1/2 mches). This device detects t ives a warning signal with readings. - - Wrist Blood Pressure Monitor compares average blood pressure results established AHA (American Heart Association) hypertension guideline of 135/85 11 - 12 - 12 - 12 - 12 - 12 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - diagnostic device. Contact your physician if hypertensive values are indicated. 13. 15. 1990 05 25 0 it 21 CFR 801 Subpart D) 3 -> (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER NEEDED) A . . . . . . . . . . . ้า - จั : 上一 Concurrence of CDRH, Office of Device Evaluation (ODE Owen P. Faris -S 2012.12.21 13:15:47 -05'00' Indications for Use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...