Arm Blood Pressure Monitor (ARM-30A+);Arm Blood Pressure Monitor (ARM-30G2);Arm Blood Pressure Monitor (ARM-30Q);Arm Blood Pressure Monitor (ARM-30E+);

K244000 · Shenzhen AOJ Medical Technology Co., Ltd. · DXN · Apr 30, 2025 · Cardiovascular

Device Facts

Record IDK244000
Device NameArm Blood Pressure Monitor (ARM-30A+);Arm Blood Pressure Monitor (ARM-30G2);Arm Blood Pressure Monitor (ARM-30Q);Arm Blood Pressure Monitor (ARM-30E+);
ApplicantShenzhen AOJ Medical Technology Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateApr 30, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.

Device Story

Battery-powered automatic non-invasive blood pressure monitor; uses oscillometric technique to measure systolic/diastolic blood pressure and pulse rate. Device inflates/deflates cuff wrapped around upper arm; processes pressure signals to derive hemodynamic parameters. Output displayed in mmHg or kPa; includes memory for storing measurements, time, and low battery alerts. Used in medical facilities or home environments by patients or clinicians. Provides objective BP/pulse data to assist in monitoring cardiovascular status. Effectiveness not established for pregnant or pre-eclamptic patients.

Clinical Evidence

Clinical accuracy study conducted per ISO 81060-2:2018 (including AMD1:2020) using the Same Arm Sequential Method. 28 models tested in 7 groups (n=92-100 per group; total subjects >600). Subjects aged 15-80 years. Primary endpoints: mean error and standard deviation of differences for systolic and diastolic pressure. Results met all ISO 81060-2 limits. No adverse effects reported.

Technological Characteristics

Oscillometric measurement principle. Battery-powered (4x AA or AAA). Cuff size 22-42 cm. Materials evaluated for biocompatibility per ISO 10993-1, -5, -10, -23. Standards: IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30, IEC 60601-1-2. Non-sterile. No wireless connectivity.

Indications for Use

Indicated for adult patients requiring non-invasive measurement of systolic/diastolic blood pressure and pulse rate. Not indicated for pregnant or pre-eclamptic patients.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 30, 2025 Shenzhen AOJ Medical Technology Co., LTD Jack Wang Deputy Chief Room 301&4F, Block A, Bldg A, ingfa Intelligent Manufacturing Park Xiaweiyuan, Gushu Community, Xixiang Str Bao'an District Shenzhen, Guangdong 518126 China Re: K244000 Trade/Device Name: Arm Blood Pressure Monitor (Models ARM-30A+, ARM-30G2, ARM-30Q, ARM-30E+, ARM-3010, ARM-3011, ARM-3012, ARM-3013, ARM-3110, ARM-30G+, ARM-30M, ARM-30S, ARM-30T, ARM-HA101, ARM-3040, ARM-30V, AMR-30W, SBM 15, BU 518, ARM-30U, ARM-3030, ARM-30Y, BU 520, ARM-3120, ARM-3020, ARM-3021, ARM-3022 and ARM-3023) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 28, 2025 Received: March 28, 2025 Dear Jack Wang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K244000 - Jack Wang Page 2 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical- {2} K244000 - Jack Wang Page 3 devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Stephen C. Browning -S LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K244000 | | | Device Name Arm Blood Pressure Monitor (Models ARM-30A+, ARM-30G2, ARM-30Q, ARM-30E+, ARM-3010, ARM-3011, ARM-3012, ARM-3013, ARM-3110, ARM-30G+, ARM-30M, ARM-30S, ARM-30T, ARM-HA101, ARM-3040, ARM-30V, AMR-30W, SBM 15, BU 518, ARM-30U, ARM-3030, ARM-30Y, BU 520, ARM-3120, ARM-3020, ARM-3021, ARM-3022 and ARM-3023) | | | Indications for Use (Describe) The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home. | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K244000 # 510(K) Summary Prepared in accordance with the requirements of 21 CFR Part 807.92 Prepared Date: 2025/02/26 ## 1. Submission sponsor Name: Shenzhen AOJ Medical Technology Co., Ltd. Address: Room 301&4F, Block A, Building A, Jingfa Intelligent Manufacturing Park, Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an District, 518126 Shenzhen, PEOPLE'S REPUBLIC OF CHINA Contact person: Jack Wang Title: Deputy Chief TEL: 86 755-27786026 ## 2. Subject Device Information | Trade/Device Name | Arm Blood Pressure Monitor | | --- | --- | | Model | models ARM-30A+,ARM-30G2,ARM-30Q,ARM-30E+, ARM-3010,ARM-3011,ARM-3012,ARM-3013,ARM-3110, ARM-30G+,ARM-30M,ARM-30S,ARM-30T,ARM-HA101, ARM-3040,ARM-30V,AMR-30W,SBM 15,BU 518,ARM-30U, ARM-3030,ARM-30Y,BU 520,ARM-3120,ARM-3020,ARM-3021, ARM-3022 and ARM-3023 | | Common Name | Automatic Blood Pressure Monitor | | Regulatory Class | Class II | | Product Code | DXN | | Submission type | Traditional 510(K) | ## 3. Predicate Device Manufacturer: Shenzhen AOJ Medical Technology Co., Ltd. Device name: Arm Blood Pressure Monitor, models AOJ-30A, AOJ-30B, AOJ-30C, AOJ-30D, AOJ-30E, AOJ-30F and AOJ-30G. 510(K) Number: K222125 ## 4. Device Description The Arm Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa. The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The device also has low voltage indication, which will be triggered when the battery is low. {5} K244000 The proposed device is intended to be used in medical facilities or at home. And the effectiveness of this sphygmomanometer has not been established in pregnant (including pre-eclamptic) patients. The product is provided non-sterile, and not to be sterilized by the user prior to use. All the models included in this submission follow the the same intended use, same measurement principle, same blood pressure core algorithm and similar product design. The main differences are appearance, Dimensions and some function which will not affect the safety and effectiveness of the device. # 5. Intended use & Indication for use The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home. # 6. Comparison to the Predicate Device | ITEM | Proposed Device Arm Blood Pressure Monitor ARM-30A+,ARM-30G2,ARM-30Q, ARM-30E+,ARM-3010,ARM-3011, ARM-3012,ARM-3013,ARM-3110, ARM-30G+,ARM-30M,ARM-30S, ARM-30T,ARM-HA101,ARM-3040, ARM-30V,AMR-30W,SBM 15, BU 518,ARM-30U,ARM-3030, ARM-30Y,BU 520,ARM-3120, ARM-3020,ARM-3021,ARM-3022, ARM-3023 / K244000 | Predicate Device Arm Blood Pressure Monitor AOJ-30A, AOJ-30B, AOJ-30C, AOJ-30D, AOJ-30E, AOJ-30F and AOJ-30G / K222125 | Comparison Result | | --- | --- | --- | --- | | Manufacturer | Shenzhen AOJ Medical Technology Co., Ltd. | Shenzhen AOJ Medical Technology Co., Ltd. | Same | | Intended Use/Indications for Use | The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home. | The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home. | Same | | Contraindications | No known contraindication existing | Not Known | Same | | Application scenario | Medical Facilities and Home Use | Medical Facilities and Home Use | Same | | Operational Specifications | | | | | Principle | Oscillometric | Oscillometric | Same | | Measurement Item | SYS, DYS, Pulse Rate | SYS, DYS, Pulse Rate | Same | | Patient population | Adult | Adult | Same | | Measurement site | Upper arm | Upper arm | Same | | Accuracy | ± 3 mmHg | ± 3 mmHg | Same | {6} K244000 | Heart rate measurement range | 40-199 bpm | 40-199 bpm | Same | | --- | --- | --- | --- | | Accuracy | ± 5% | ± 5% | Same | | Cuff size | 22 - 42 cm | 22 - 42 cm | Same | | Display | Blood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | Blood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | Same | | Auto shutdown | YES | YES | Same | | Operating environment | Temperature: 5°C~40°C Humidity: 15%-90% RH, Atmospheric pressure: 70 kPa -106 kPa | Temperature: 5°C~40°C Humidity: 15%-90% RH, Atmospheric pressure: 70 kPa -106 kPa | Same | | Storage environment | Ambient Temperature: -20°C to 55°C Relative Humidity: 10-93% RH, Atmospheric pressure: 70 kPa -106 kPa | Ambient Temperature: -20°C to 55°C Relative Humidity: 10-93% RH, Atmospheric pressure: 70 kPa -106 kPa | Same | | Battery type | 1) 6Vdc (4*1.5V AAA batteries): ARM-30A+,ARM-30G2, ARM-30Q,ARM-30E+,ARM-3010, ARM-3011,ARM-3012,ARM-3013, ARM-3110,ARM-30G+,ARM-30M, ARM-30S,ARM-30T,ARM-HA101, ARM-3040,ARM-30V,AMR-30W, SBM 15,ARM-30U,ARM-3030, ARM-30Y,ARM-3120,ARM-3020, ARM-3021,ARM-3022,ARM-3023 2) 6Vdc (4*1.5V AA batteries): BU 518,BU 520 | 6Vdc (4*1.5VAAA batteries) | Different^{1} | | Weight | SBM 15: About 175g ARM-30V / ARM-30W: About 209g ARM-30Q: About 193g ARM-30A+: About 316g ARM-30G2: About 263g ARM-30G+: About 234g ARM-30M: About 280g ARM-HA101: About 290g ARM-3040: About 235g ARM-30Y:About 265g ARM-3030:About 266g BU 518: About 270g BU 520: About 286g ARM-30S / ARM-30T / ARM-30U: About 221g | AOJ-30A /AOJ-30B /AOJ-30G: (262 ±5)g without battery AOJ-30C: (316 ±5)g without battery AOJ-30D: (349 ±5)g without battery AOJ-30E: 220 g without battery AOJ-30F: 220 g without battery | Different^{2} | 3 / 6 {7} K244000 | | ARM-30E+ / ARM-3010 / ARM-3011 / ARM-3012 / ARM-3013 / ARM-3110: About 220g ARM-3120/ARM-3020/ ARM-3021/ ARM-3022/ ARM-3023:About 230g | | | | --- | --- | --- | --- | | Dimensions | ARM-30A+ / ARM-3030: 108mm×139mm×62mm ARM-30G2 / ARM-30G+: 130mm×100mm×60mm ARM-30Q: 100mm×100mm×59mm ARM-30M : 129mm×112mm×58.5mm ARM-HA101 : 130mm×112mm×58.3mm ARM-3040: 136mm×113mm×28mm ARM-30V / ARM-30W: 130mm×100mm×51mm SBM 15: 115.4mm×92mm×44.5mm ARM-30Y:140.8mm×110mm×61mm BU 518 / BU 520: 130mm×111.5mm×56mm ARM-30S / ARM-30T / ARM-30U: 130mm×100mm×49mm ARM-30E+ / ARM-3010 / ARM-3011 / ARM-3012 / ARM-3013 / ARM-3110: 118mm×98mm×62.5mm ARM-3120 /ARM-3020/ ARM-3021/ ARM-3022 / ARM-3023: 118mm×98mm×61mm | AOJ-30A /AOJ-30B /AOJ-30G: 127mm×93mm×73mm AOJ-30C: 108mm×139mm×62mm AOJ-30D:136mm×113mm×68mm AOJ-30E /AOJ-30F: 118mm×98mm×61mm | | | Patient Contacting | Surface-contacting, Less than 24 h | Surface-contacting, Less than 24 h | Same | | Biocompatibility evaluation | Cytotoxicity, skin sensitization and irritation | Cytotoxicity, skin sensitization and irritation | Same | | Electrical safety | IEC 60601-1 IEC 60601-1-11 ISO 80601-2-30 | IEC 60601-1 IEC 60601-1-11 ISO 80601-2-30 | Same | | EMC | IEC 60601-1-2 | IEC 60601-1-2 | Same | | Biocompatibility | ISO 10993-1 ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-1 ISO 10993-5 ISO 10993-10 | Different^{3} | | Data transmission | Not available | Has wireless function with Bluetooth and available for AOJ-30A and AOJ-30B | Different^{4} | ## Justification for the differences: 1) Different kind of power supply The subject device powered by internal battery (AA or AAA), which both belong to alkaline battery. It will not affect the safety and effective of subject device. The test results have been verified by verification testing. {8} K244000 2) Different Physical Specifications The subject and the predicate device are of different size but proximity. Moreover, such engineering design has been verified during the international standards, so such minor different ill not raise any safety and effectiveness questions. 3) Different Biocompatibility standard The subject device has been tested 10993-5,10993-10,10993-23 according to the update of the standard. 4) Different data transmission function The subject device has no data transmission function with Bluetooth. 7. Non-clinical Data Biocompatibility testing The biocompatibility evaluation for the Arm Blood Pressure Monitor were conducted in accordance with the International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests: - Cytotoxicity - Skin Sensitization - Skin Irritation Bench testing The device has been tested according to the following standards: - IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance - IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests - IEC 80601-2-30: Medical electrical equipment – Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. - IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment - FDA Guidance for Non-Automated Sphygmomanometer. 5 / 6 {9} K244000 # 8. Clinical data According to similar clinical characteristics, 28 submitted models were divided into 7 groups for clinical accuracy research. The clinical accuracy test report and data analysis followed the requirements of the ISO 81060-2 Third edition 2018-11 [Including AMD1:2020] Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type [Including: Amendment 1 (2020)]. The Same Arm Sequential Method was chosen for all studies. | Group No. | Test model | Number of Subjects (a minimum of 85 subjects) | Male (at least 30 %) | Female (at least 30 %) | Age range (>12 years old) | | --- | --- | --- | --- | --- | --- | | 1 | ARM-3030, ARM-30Y | 100 | 49 | 51 | 17 to 78 | | 2 | ARM-3120, ARM-3020, ARM-3021, ARM-3022, ARM-3023 | 100 | 54 | 46 | 15 to 80 | | 3 | ARM-30A+, ARM-30G2, ARM-30E+, ARM-30Q, ARM-3010, ARM-3011, ARM-3012, ARM-3013, ARM-3110 | 100 | 41 | 59 | 19 to 78 | | 4 | ARM-30M, ARM-30G+, ARM-HA101, ARM-30S, ARM-30T, ARM-30U | 100 | 54 | 46 | 17 to 71 | | 5 | ARM-3040 | 100 | 53 | 47 | 15 to 78 | | 6 | BU 520, BU 518 | 100 | 49 | 51 | 15 to 78 | | 7 | SBM 15, ARM-30V, ARM-30W | 92 | 47 | 45 | 15 to 70 | All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2 Third edition 2018-11 [Including AMD1:2020]. No adverse effect and/or complication is found in this study. # 9. Conclusion It is concluded from the non-clinical and clinical tests that demonstrate that the subject devices are as safe, as effective, and performs as well as the legally marketed predicate device identified above.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...