FUSION

K061557 · Medwave, Inc. · DXN · Sep 15, 2006 · Cardiovascular

Device Facts

Record IDK061557
Device NameFUSION
ApplicantMedwave, Inc.
Product CodeDXN · Cardiovascular
Decision DateSep 15, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

The Fusion™ System and accessories blood pressure measurement system is intended to be used on patients with wrist circumference of 11 cm - 22 cm by trained medical personnel to continually monitor systolic, diastolic and mean blood pressure and pulse rate. The Fusion System module option of Nellcor OxiMAX Option (K012891), is indicated for the continuous non-invasive monitoring of a patient's functional oxygen saturation of arterial hemoglobin (SpO2) by trained medical personnel. The Fusion System module option of Kendall Genius Thermometer (K920713) is indicated for the measuring of a patient's body temperature. This device is prescription use only.

Device Story

Fusion™ System is a non-invasive blood pressure monitor using a pressure sensor placed on the radial artery; eliminates occlusive inflatable cuffs. Device applies a 'sweep technique'—varying force on the radial artery—to generate a pressure waveform. Proprietary algorithms analyze waveforms to extract systolic, diastolic, and mean blood pressure and pulse rate approximately every 15 heartbeats. Operated by trained medical personnel in clinical settings. Output displayed on screen; visual/audible alarms trigger if measurements exceed user-defined limits. Optional modules include Nellcor OxiMAX for SpO2 and Kendall Genius for temperature. Data aids clinicians in therapeutic management by providing frequent, accurate, non-invasive hemodynamic monitoring. Benefits include patient comfort and continuous data availability without cuff-related complications.

Clinical Evidence

Bench testing only. Performance validated against intra-arterial line pressure waveforms. System meets AAMI standards for non-invasive blood pressure monitors, achieving a mean difference of ±5 mmHg or less with a standard deviation of 8 mmHg or less. Pulse measurement accuracy is ±5 bpm or 10%. Testing included 10,000 full-scale cycles, electrical safety, electromagnetic compatibility, and biocompatibility.

Technological Characteristics

Microcomputer-controlled non-invasive blood pressure monitor. Uses a pressure sensor for radial artery applanation tonometry. Features 'sweep' pressure application technique (~10mmHg/heartbeat). Connectivity includes optional pulse oximetry and thermometer modules; thermal printer output. Operating environment: 10°C-40°C, 10-90% humidity. Software-based signal processing and waveform analysis. Powered by AC/DC. Compliant with ANSI/AAMI SP10-2002 and IEC60601.

Indications for Use

Indicated for patients with wrist circumference 11-22 cm requiring continuous monitoring of systolic, diastolic, mean blood pressure, and pulse rate. Optional modules indicated for continuous SpO2 monitoring and body temperature measurement. For use by trained medical personnel; prescription only.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061557 # Section 2. SUMMARY AND CERTIFICATION # A. 510(K) SUMMARY SEP 1 5 2006 # A.1 Summary of Safety and Effectiveness In accordance with 21 CFR 807.92, the following information constitutes the Medwave, Inc. summary for the Vasotrac® APM205A System. | SUBMITTER'S NAME: | Medwave, Inc. | |---------------------|-------------------------------------------------| | ADDRESS: | 4382 Round Lake Road West<br>St. Paul, MN 55112 | | CONTACT PERSON: | Donna R. Lunak | | TELEPHONE NUMBER: | 651-639-1227 | | FAX NUMBER: | 651-639-1338 | | DATE OF SUBMISSION: | 05/22/06 | ### A.2 Identification Of Device Proprietary Name: Fusion™ Common Name: Patient Monitor Classification Status: Class II Non-invasive blood pressure measurement 870.1130 Module Option Oximeter Tyco Healthcare, Nellcor Oximax 870.2700 (K012891) Module Option Clinical Electronic Thermometer Tyco Healthcare, 880.2910 Kendall Genius Thermometer (K920713) ## A.3 Equivalent Devices Medwave, Inc. believes the Fusion™ System is substantially equivalent to Medwave, Inc. Vasotrac® APM2052A System (K011152). #### A.4 Description of the Device The Fusion™ System is a non-invasive blood pressure monitor that uses a pressure sensor placed directly on top of the radial artery. This sensor is noninvasive and eliminates the need for an occlusive inflatable cuff. This device is intended to be used on patients by trained medical personnel to continually monitor systolic, diastolic and mean blood pressure, and pulse rate. The Fusion™ System uses a patented "sweep technique" which applies a varying force on the radial artery. The counterpressure in the artery produces a signal which is digitized and used to calculate blood pressure parameters. page 1 of 6 {1}------------------------------------------------ ## Description of the Device (continued) The sensor measures the pulse at the point of maximum energy transfer; the wave shapes are analyzed by Medwave's proprietary algorithms. Parameters are extracted from the waveforms and a set of coefficients are applied to them yielding systolic, mean, and diastolic pressures approximately every 15 heart beats. These algorithms have been tested and validated in clinical studies by synchronized comparisons to arterial line pressure waveforms. Patient pressures can be monitored visually by viewing the screen and audibly entering limits into the Fusion™ System alarm menu. Patient measurements above or below the limits will be automatically brought to the attention of the caregiver through these visual and audible alarms. A Start/Stop key provides the operator with an option to cancel operation at anytime. The Fusion™ System can be ordered with additional module options: - · Tyco Healthcare Nellcor OxiMax Pulse Oximetry System (K012891) provides monitoring of a patient's functional oxygen saturation of arterial hemoglobin. - · Kendall Genius Thermometer (K920713) provides patient's body temperature readings. - · Seiko Thermal Printer SII providing hardcopy of readings selected to print. #### A.5 Intended Use The Fusion™ System and accessories blood pressure measurement system is intended to be used on patients with wrist circumference of 11 cm - 22 cm by trained medical personnel to continually monitor systolic, diastolic and mean blood pressure and pulse rate. The Fusion System with the Nellcor OxiMAX Option is indicated for the continuous non-invasive monitoring of a patient's functional oxygen saturation of arterial hemoglobin (SpO2) by trained medical personnel. The Fusion System will the Kendall Genius Thermometer (K920713) is indicated for the measuring of a patient's body temperature. This device is prescription use only. 061557 page 2 of 4 {2}------------------------------------------------ ## A.6 Technological Characteristics, Comparison to Predicate Device. Like the predicate device, the Fusion™ System is a non-invasive blood pressure monitor, which uses a pressure sensor placed on the wrist over the radial artery. This device is intended to be used on patients by trained medical personnel to continually monitor systolic, diastolic and mean blood pressure and pulse rate. The information from this device is intended to guide clinicians in the therapeutic management of their patients by: providing accurate and frequently updated blood pressure information in a safe, noninvasive, easily obtained, and comfortable manner. Both the devices measure the diastolic, systolic blood pressure and pulse rate from the wrist using oscillometric methods. Both the devices are microcomputer controlled. Like the Vasotrac® System (K011152), the Fusion™ System have a memory function that retains approximately 900 readings. Like the Vasotrac® System (K011152), the Fusion™ System display systolic, diastolic blood pressures ranging from 40 and 240 mmHg. Both systems have a blood pressure measurement accuracy of a mean difference of ± 5mmHg or less with a standard deviation of 8mmHg or less. The pulse measurement range is the same for both the Vasotrac® System (K011152) and the Fusion™ System, from 40 - 200 bpm. The accuracy of the pulse measurements are + 5 bpm or 10% of the measured pulse frequency. (Supported by study results in the supporting Clinical Data on file at Medwave, Inc.) Both Vasotrac® System (K011152) and the Fusion™ utilize the application of pressure to the artery (by the sensor); the counter pressure in the artery produces a pressure waveform. When maximum amplitude is achieved, mean blood pressure is calculated. The Vasotrac® System (K011152) and the Fusion™ System use a unique "sweep" technique for applying pressure to a the radial artery: downward pressure is applied by the sensor to the radial artery at a rate of ~10mmHg per heart beat increasing as the beat amplitude increases and decreasing rapidly when the beat amplitude begins to decrease. A curve fit is made using the amplitude of each beat versus the hold down pressure to form the bell shaped curve. This curve fit is used to determine the true peak that might occur between pulses as well as to filter out small variations due to artifacts or aberrancies. Both the Vasotrac® System (K01152) and the Fusion™ System utilize Medwave's proprietary algorithms in analyzing the pressure waveforms to calculate the systolic, mean and diastolic readings. Parameters are extracted from the waveforms and a set of coefficients is applied to them, yielding systolic, mean and diastolic pressures. The algorithms have been tested against intra arterial line pressure waveforms and proven to meet industry standards set by the American Medical Instrumentation (AAMI), mean difference of + 5mmHg or less with a standard deviation of 8mmHg or less. (Supported by study results in the Clinical Data on file at Medwave, Inc.) dl 1557 page 3 of 6 Page 9 of 46 Pages {3}------------------------------------------------ The Vasotrac® System (K011152) and the Fusion™ System software are identical with the exception of insignificant features incorporated. The Vasotrac System (K011152), as the Fusion™ System, has a power switch and a LCD display. The operating environment of 100C - 40°C and 10% to 90% relative humidity. The Fusion™ and the predicate device - the Vasotrac® APM205A Noninvasive Blood Pressure Measurement System (K011152) use the same non-invasive blood pressure measurement accessories. Any minor differences in the appearance, technology, or manufacture of the predicate device and the Fusion™ System do not raise any new questions of safety or effectiveness. Associated risks posed by the Fusion™ System are thought to be no more than those of well-designed automated cuff-based noninvasive blood pressure devices currently marketed in the interstate commerce are. Both the device and the sensor have been designed to minimize the risk to patients from burns, excessive pressure or sensor failure caused by either normal device use by patient and/or clinical abuse. # K061557 {4}------------------------------------------------ # A.7 Discussion of Performance Testing . An extensive collection of tests has been conducted and successfully completed. Summary follows: | Requirements | ANSI/AAMI SP10-2002 American<br>National Standards for Electronic or<br>Automated Sphygmomanometers | IEC60601 | Center For Devices And<br>Radiological Health<br>Noninvasive Blood<br>Pressure Monitor Guidance<br>Document | Other As<br>Listed | |---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|------------|-------------------------------------------------------------------------------------------------------------|---------------------| | Device Labeling | X | X | X | ---------- | | Outer Container Labeling | X | X | X | ---------- | | Information Manual | X | X | X | ---------- | | Component Labeling | X | X | X | ---------- | | Power System Labeling | X | ---------- | X | ---------- | | Storage Conditions -<br>20°C (-40°F) - 50°C<br>(122°F) | X | ---------- | ---------- | ---------- | | Operating Temperature<br>Conditions 10°C (50°F) -<br>40°C (104°F) | X | ---------- | ---------- | ---------- | | Operating Humidity<br>Conditions 15 - 90 percent<br>(noncondensing) | X | ---------- | ---------- | ---------- | | Operating Range in<br>Altitude Conditions -170<br>to 1700 meters<br>(-500 to 5000 feet),<br>referenced to sea level | X | ---------- | ---------- | ---------- | | Vibration and Shock | ---------- | ---------- | ---------- | NSTA | | Voltage Range | N/A | N/A | N/A | N/A | | Life test minimum of<br>10,000 full scale cycles | X | ---------- | ---------- | ---------- | | Maximum cuff pressure | N/A | N/A | N/A | N/A | | Cuff deflation | N/A | N/A | N/A | N/A | | Electrical safety | ---------- | X | ---------- | IEC601 | | Conductive components | ---------- | X | ---------- | ---------- | | Pressure indicator accuracy | X | ---------- | ---------- | ---------- | | Overall system efficacy | X | ---------- | ---------- | ---------- | | Auscultatory method as the<br>reference standard | N/A | N/A | N/A | N/A | | Intraarterial method as the<br>reference standard | X | ---------- | ---------- | ---------- | | Requirements | ANSI/AAMI SP10-2002 American<br>National Standards for Electronic or<br>Automated Sphygmomanometers | IEC60601 | Center For Devices And<br>Radiological Health<br>Noninvasive Blood<br>Pressure Monitor Guidance<br>Document | Other As<br>Listed | | Battery indicator | N/A | ---------- | ---------- | ---------- | | Requirements for devices<br>with manual inflation | N/A | N/A | N/A | N/A | | Comparison Testing | N/A | N/A | N/A | N/A | | Foreign Standards | N/A | N/A | N/A | N/A | | Software Testing | ---------- | ---------- | ---------- | Medwave<br>795-0000 | | Electromagnetic<br>Compatibility | ---------- | X | ---------- | ---------- | | Biocompatibility | ---------- | ---------- | X | ---------- | | Sterilization | N/A | N/A | N/A | N/A | | Packaging | N/A | N/A | N/A | N/A | | Shelf Life | N/A | N/A | N/A | N/A | 061557 page 5. {5}------------------------------------------------ ### A.8 Conclusion Based on extensive performance testing and a comparison to the predicate device, it is the conclusion of Medwave, Inc. that Fusion™ is substantially equivalent to devices already on the market (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness. page 6 of 6 K041557 {6}------------------------------------------------ #### Public Health Service ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 5 2006 Medwave, Inc. c/o Ms. Donna R. Lunak VP of Regulatory Affairs 4382 Round Lake Road West Arden Hills, MN 55112 Re: K061557 Trade Name: Fusion™ System Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: August 14, 2006 Received: August 23, 2006 ## Dear Ms. Lunak: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ Page 2 -- Ms. Donna R. Lunak Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bzimmernan for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # INDICATIONS FOR USE 510(k) Number (if known): Device Name: Fusion™ System Indications for Use: The Fusion™ System and accessories blood pressure measurement system is intended to be used on patients with wrist circumference of 11 cm - 22 cm by trained medical personnel to continually monitor systolic, diastolic and mean blood pressure and pulse rate. The Fusion System module option of Nellcor OxiMAX Option (K012891), is indicated for the continuous non-invasive monitoring of a patient's functional oxygen saturation of arterial hemoglobin (SpO2) by trained medical personnel. The Fusion System module option of Kendall Genius Thermometer (K920713) is indicated for the measuring of a patient's body temperature. This device is prescription use only. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Bhummana Division of Cardlovascular Devices 510(k) Number Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...