YA HORNG BLOOD PRESSURE MONITOR, YA HORNG BLOOD PRESSURE MONITOR, BLUETOOTH TRANSMISSION

K121025 · Ya Horng Electronic Co., Ltd. · DXN · Aug 17, 2012 · Cardiovascular

Device Facts

Record IDK121025
Device NameYA HORNG BLOOD PRESSURE MONITOR, YA HORNG BLOOD PRESSURE MONITOR, BLUETOOTH TRANSMISSION
ApplicantYa Horng Electronic Co., Ltd.
Product CodeDXN · Cardiovascular
Decision DateAug 17, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

The YA HORNG Upper Arm Blood Pressure Monitor, model BP-700NW and Bluetooth Transmission BP-700W are noninvasive blood pressurement systems intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to be 9.0" for Arm type. ※ Optional model: BP-700W with Bluetooth module for the wireless communication function connects to the PC for record archiving and printing purposes.

Device Story

Device measures blood pressure via oscillometric method; senses pulse signal vibrations in closed air pipe system; microcomputer processes signal to determine systolic/diastolic pressure based on cuff oscillation amplitude. Used at home by adult patients; cuff wrapped around upper arm. BP-700W model includes Bluetooth module for wireless data transfer to PC for archiving/printing; BP-700NW lacks wireless function. Output displayed to user; assists in monitoring blood pressure trends. Benefits include convenient, non-invasive self-monitoring of cardiovascular health.

Clinical Evidence

Bench testing only. Compliance demonstrated with ANSI/AAMI SP 10:2002, ANSI/AAMI ISO 81060-2:2009, and EN 1060-4:2004 standards for blood pressure measurement accuracy. Electrical safety and EMC testing performed per IEC/EN 60601-1 and IEC/EN 60601-1-2. FCC/RF conformity verified per ANSI C63.4 and EN 300328.

Technological Characteristics

Oscillometric measurement; microcomputer-based signal processing. Materials include woven cotton sheeting. Connectivity via Bluetooth (BP-700W) or USB (predicate-based). Powered by internal electronics. Form factor: upper arm cuff (9.0" circumference). Standards: IEC/EN 60601-1, EN 1060-1/3/4, ANSI/AAMI SP 10, ANSI/AAMI ISO 81060-2, FCC Part 15.

Indications for Use

Indicated for adults over age 18 for non-invasive measurement of systolic/diastolic blood pressure and pulse rate at home using an upper arm cuff.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # YA HORNG ELECTRONIC CO., LTD. No.35, Shalun, Anding Dist., Tainan City 745, Taiwan (R.O.C.) TEL: 886-6-5932201-8 FAX: 886-6-5930030 web:www.yahorng.com e-mail: rdsafety@yahomg.com ### SUMMARY OF SAFETY AND EFFECTIVENESS ( According to 21 CFR 807.92) AUG 17 2012 ### " 510(k) Summary for K121025 " Submitter's Name: YA HORNG Electronic Co., Ltd. > No. 35, Zsha Lun, Jon Zsha Village, Antin Address: District, Tainan, 74555, Taiwan, ROC 886-6-5932201 Telephone: > 886-6-5930030 FAX: Contact Person: Dr. Jen, Ke-Min Date Summary Prepared: March 23, 2012 > Upper Arm Blood Pressure Monitor, model Proprietary Name: BP-700NW and Bluetooth Transmission BP-700W BLOOD PRESSURE MONITOR Common Name: Classification Name: NON-INVASIVE BLOOD-PRESSURE MEASUREMENT SYSTEM ( per 21CFR section 870.1130) Class II (performance standards) Device Class: Specialty: CARDIOVASCULAR Product code: DXN O Legally Marketed ( Predicate ) Device : YA HORNG Digital Upper Arm Blood Pressure Monitor BP-700, BP-700T, BP-700U, BP-700B, BP-700TB, BP-700UB, and BP-700TUB (K090058) - A & D Medical UA-767BT Digital Blood Pressure Monitor (K040371) EI {1}------------------------------------------------ # YA HORNG ELECTRONIC CO., LTD. No.35, Shalun, Anding Dist., Tainan City 745,Taiwan (R.O.C.) TEL: 886-6-5932201-8 FAX: 886-6-5930030 web:www.yahorng.com e-mail: rdsafety@yahomg.com ### Description of the new device: (Same as the predicate devices) YA HORNG Upper Arm Blood Pressure Monitor, model BP-700NW and Bluetooth Transmission BP-700W use the Oscillometric method to measure the blood pressure. The Oscillometric method is adopted clinically to measure the blood pressure recently. It is not needed to use the stethoscope, as in the traditional measuring method, to monitor the Korotkov sound when deciding the systolic or diastolic pressure. The Oscillometric method senses the vibrating signal via the closed air pipe system and utilizes the microcomputer to automatically sense the characteristics of the pulse signal. Through simple calculation, the reading can reflect the accurate real blood pressure, and the systolic pressure is defined as the pressure when the cuff pressure oscillating amplitude begins to increase and the diastolic pressure as the pressure when the cuff pressure oscillating amplitude stops decreasing. ## Technological Characteristics of our new device compared to the predicate device: The technological characteristics of YA HORNG Upper Arm Blood Pressure Monitor, model BP-700NW and Bluetooth Transmission BP-700W are substantially equivalent to YA HORNG Digital Upper Arm Blood Pressure Monitor BP-700, BP-700T, BP-700U, BP-700B, BP-700TB, BP-700UB, BP-700TUB (K090058). There is the same Owner, YA HORNG Electronic Co., Ltd., which FDA owner number is 9040892. Especially, there are the same design specifications, the same form and intended to be used in the same manner that means the new devices are same as the predicate devices. Besides, the subject device BP-700W and the other predicate device A & D Medical UA-767BT Digital Blood Pressure Monitor (K040371) are also with the wireless communication function connect to the PC for record archiving and printing purposes. The major differences from the predicate devices and the subject devices as below: - The predicate devices: BP-700U, BP-700UB, and BP-700TUB (K090058), software . in the PC for record archiving and printing purposes. E2 {2}------------------------------------------------ # YA HORNG ELECTRONIC CO., No.35, Shatun, Anding Dist., Tainan City 745, Taiwan (R.O. TEL: 886-6-5932201~8 FAX: 886-6-5930030 web:www.yahorng.com e-mail: rdsafety@yahomg.com - The subject device BP-700W: replace the USB cable to transfer data of the BP-700W ● with Bluetooth module for the wireless communication function connect to PC for record archiving and printing purposes. - The predicate device: A & D Medical UA-767BT Digital Blood Pressure Monitor (K040371) and subject device BP-700W are also with the wireless communication function, and only has the different dimensions and the conditions of the operation and storage. Besides, they use the different size of cuffs. - The subject device BP-700NW is identical for the BP-700W; and only does not with the wireless communication function. Thus there are substantially equivalent. ## Test Summary: #### ELECTRIC SAFETY, EMC and FCC test reports, 1. | General safety | IEC/EN 60601-1:2007 | PASS | |----------------|-----------------------------------|------| | | EN 1060-1:2009, EN 1060-3:2009 | PASS | | EMC conformity | IEC/EN 60601-1-2: 2010 | PASS | | FCC conformity | ANSI C63.4: 2003 | PASS | | | FCC 47 part 15 subject B class B | PASS | | ERM conformity | EN 30148-1:2008, EN 30148-17:2009 | PASS | | RF conformity | EN 300328:2006 | PASS | - WOVEN COTTON SHEETING: (Same as the predicate devices K090058) 2. Uses the 510K Blood-Pressure Cuff: YA HORNG Blood-Pressure Cuff (K051539). #### PERFORMANCE & CLINICAL TEST 3. ANSI / AAMI SP 10: 2002 ANSI / AAMI ISO 81060-2:2009 EN 1060-4:2004 YA HORNG Electronic Co. Ltd. believes this information and referred document to be sufficient for the FDA to find our proposed device substantially equivalent to the predicate product and other products currently in distribution. Ke Zister Dr. Jen, Ke-Min official correspondent {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-W066-G609 Silver Spring MD 20993-0002 AUG 17 2012 Ya Horng Electronic Co., Ltd. c/o Dr. Ke-Min Jen ROC Chinese-European Industrial Research Society No. 58, Fu-Chiun St. Hsin-Chu City, 30067 Taiwan, ROC Re: K121025 > Trade/Device Names: 1) Ya Horng Upper Arm Blood Pressure Monitor BP-700NW; and 2) Ya Horng Upper Arm Blood Pressure Monitor, Bluetooth Transmission BP-700W Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: July 16, 2012 Received: July 24, 2012 Dear Dr. Jen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {4}------------------------------------------------ ### Page 2 - Dr. Ke-Min Jen If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. M. G. Killelea Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use #### K121025 510(k) Number: Device Name: YA HORNG ELECTRONIC CO., LTD. Upper Arm Blood Pressure Monitor BP-700NW, Upper Arm Blood Pressure Monitor, Bluetooth Transmission BP-700W #### . Indications for use: The YA HORNG Upper Arm Blood Pressure Monitor, model BP-700NW and Bluetooth Transmission BP-700W are noninvasive blood pressurement systems intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to be 9.0" for Arm type. ※ Optional model: BP-700W with Bluetooth module for the wireless communication function connects to the PC for record archiving and printing purposes. Prescription Use AND/OR Over-The-Counter Use √ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cmesf Hillehem Page sion Sian-Off) Division of Cardiovascular Devices 510(k) Number_V12025
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...