Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5W

K222979 · Microlife Intellectual Property GmbH · DXN · May 19, 2023 · Cardiovascular

Device Facts

Record IDK222979
Device NameMicrolife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5W
ApplicantMicrolife Intellectual Property GmbH
Product CodeDXN · Cardiovascular
Decision DateMay 19, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

The Upper Arm Blood Pressure Monitor, Model BP3KV1-5W is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm for a circumference range from 22 to 52cm. The device is suitable for use by adults, including adults with conditions of diabetes, pregnancy, or pre-eclampsia. The device detects the appearance of irregular heartbeat during measurement and gives a warning signal with the reading once the irregular heartbeat is detected. The device can be used in connection with a smart phone via Bluetooth or with a personal computer (PC) via USB. The measurement data can be transferred to a smart phone running the Microlife Connected Health+ mobile software (App) or a PC running the Microlife BP Analyzer+ (BPA+) software.

Device Story

Automatic digital blood pressure monitor for home use; measures systolic/diastolic BP and pulse rate via oscillometric technique. Input: cuff pressure oscillations detected by semiconductor sensor. Processing: microprocessor converts pressure changes to electrical signals; FIR filter applied; oscillometric algorithm calculates BP/pulse. Output: BP/pulse readings on digital display; irregular heartbeat warning; data transfer via Bluetooth to mobile app or USB to PC. Features: MyCheck (qualitative comparison to 4-week average), MyBP (averaged readings from recent 3-7 days), 28-day average, cuff fit check. Used by adults for self-monitoring; aids in tracking BP trends for clinical management.

Clinical Evidence

Clinical validation performed on reference device BP3KT1-3N, which shares the same deflation valve, sensor, and oscillometric algorithm as the subject device. Study was a prospective, open-label, non-randomized, single-center validation following ANSI/AAMI/ISO 81060-2:2019 standards. Device passed all criteria for accuracy in adult populations.

Technological Characteristics

Oscillometric measurement; semiconductor pressure sensor; electromechanical solenoid valve; 4 AA batteries or 6V DC AC adapter; Bluetooth 4.2 and USB connectivity; nylon cuff fabric; microprocessor-based control; software-based signal processing (FIR filter). Tested to IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30, and ISO 10993 standards.

Indications for Use

Indicated for adult individuals, including those with diabetes, pregnancy, or pre-eclampsia, for non-invasive measurement of systolic/diastolic blood pressure and pulse rate via upper arm cuff (22-52cm). Detects irregular heartbeats.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of the HHS logo is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 19, 2023 Microlife Intellectual Property GmbH % Vaibhav Rajal Official Correspondent for Microlife Intellectual Property GmbH mdi Consultants. Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021 Re: K222979 Trade/Device Name: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5W Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: September 26, 2022 Received: September 28, 2022 Dear Vaibhav Rajal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Kenneth J. Cavanaugh -S for Robert Kazmierski Acting Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K222979 #### Device Name Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3K V1-5W #### Indications for Use (Describe) The Upper Arm Blood Pressure Monitor, Model BP3KV1-5W is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm for a circumference range from 22 to 52cm. The device is suitable for use by adults with conditions of diabetes, pregnancy, or pre-eclampsia. The device detects the appearance of irregular heartbeat during measurement and gives a warning signal with the reading once the irregular heartbeat is detected. The device can be used in connection with a smart phone via Bluetooth or with a personal computer (PC) via USB. The measurement data can be transferred to a smart phone running the Microlife Connected Health+ mobile software (App) or a PC running the Microlife BP Analyzer+ (BPA+) software. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------|-----------------------------------------------| | <span></span> ☐ Prescription Use (Part 21 CFR 801 Subpart D) | ☒ Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Microlife Intellectual Property GmbH #### Espenstrasse 139 9443 Widnau Switzerland #### Tel: +41717277000 Fax: +41717277059 ### 510(k) SUMMARY The assigned 510(k) number is K222979 #### 1. Submitter's Identification: Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland Date Summary Prepared: May 18, 2023 - Contact: Mrs. Ariel Wang Global Regulatory Affairs & Quality management Director Microlife Intellectual Property GmbH, Switzerland Tel: 886-2-87971288 # 366 E-Mail: ariel.wanq@microlife.com.tw #### 2. Name of the Device: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5W Regulation Number: 21 CFR Part 870.1130 Requlation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: II Product Code: DXN #### 3. Information for the 510(k) Cleared Device (Predicate Device): #### Primary Predicate: - a. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MW1-4B, K153077, Microlife Intellectual Property GmbH. #### Reference Predicate: - b. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GX1-5X (BP A3 PC), K183469, Microlife Intellectual Property GmbH. {4}------------------------------------------------ ## 4. Device Description: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5W is designed to measure systolic and diastolic blood pressure, pulse rate of an individual with arm circumference sizes ranging from 22 -52 cm by using a non-invasive technique in which one inflatable cuff is wrapped around the single upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but using a semiconductor sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method". The device detects the appearance of irregular heartbeat during measurement, and the symbol " "\\\ "is displayed after the measurement. In addition, the device can be used in connection with smart mobile devices running the APP and via Bluetooth. The blood pressure monitor is a fully automatic digital blood pressure measuring device use by adults on the upper arm at home. ## 5. Indications for Use: The Upper Arm Blood Pressure Monitor. Model BP3KV1-5W is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm for a circumference range from 22 to 52cm. The device is suitable for use by adults, including adults with conditions of diabetes, preqnancy, or preeclampsia. The device detects the appearance of irregular heartbeat during measurement and gives a warning signal with the reading once the irregular heartbeat is detected. The device can be used in connection with a smart phone via Bluetooth or with a personal computer (PC) via USB. The measurement data can be transferred to a smart phone running the Microlife Connected Health+ mobile software (App) or a PC running the Microlife BP Analyzer+ (BPA+) software. ## 6. Comparison to the 510(k) Cleared Device (Predicate Device): Based on information from the comparison chart: The differences and changes are the microprocessor and firmware versions. The measurement algorithm and mechanism of operation, as well as the safety & essential performance of the devices remain identical before and after the changes. The subject (Modified) device BP3KV1-5W uses the same oscillometric method as the predicate device BP3GX1-5X/3MW1-4B with the same fundamental scientific technology to determine the systolic and diastolic blood pressure and pulse rate. Cuff is inflated automatically by pump and the pressures are transferred via tubing to a sensor in these units. The subject (Modified) device BP3KV1-5W and the predicate device BP3MW1-4B have traffic light function, IHD function, MAM function, Bluetooth function and DMT technology and the predicate device BP3GX1-5X have traffic light function, MAM function and DMT technology. {5}------------------------------------------------ The differences between these devices are: #### 1) Physical Dimension The physical dimension of the subject device BP3KV1-5W is 157.5 x 105 x 61.5mm, while predicate device BP3GX1-5X is 85 x 143 x 58mm and predicate device BP3MW1-4B is 110 x 120 x 85mm. The difference is caused because of their different appearance, but the difference does not raise any new safety and effectiveness questions. This has been tested and confirmed according to IEC 60601-1-2 EMC Test Report, IEC 60601-1, AAMIANSI ES60601-1 Safety Test Report and IEC 80601-2-30 Test Report. ### 2) Display The display of the Predicate device BP3MW1-4B is with touch pad; however, the display of the subject device BP3KV1-5W and the predicate device BP3GX1-5X is without touch pad. The functionality of the icons among the predicate devices (BP3MW1-4B and BP3GX1-5X) and subject device (BP3KV1-5W) are same, except that the subject device (BP3KV1-5W) has three additional icons representing the functions of Cuff fit check, My check and My BP. #### 3) Microprocessor The microprocessor of the subject device BP3KV1-5W is R5F11NMEAFB, whereas the microprocessor of the predicate device BP3GX1-5X is R5F2LA8CANFP and predicate device BP3MW1-4B is C38D59GF. The microprocessors have different connection pin specifications, but the computing architectures are the same, but the computing architectures are the same. This difference does not affect performance and accuracy which was evaluated in the performance testing. #### 4) Sensor The microprocessor of the subject device BP3KV1-5W is semiconductor sensor, whereas the microprocessor of the predicate device BP3GX1-5X and predicate device BP3MW1-4Bare capacitive sensor. Their calculation algorithms are the same. But the type of sensor is different. It does not affect performance and accuracy which was evaluated in the performance testing. #### 5) Cuff Fit Check Function The subject device BP3KV1-5W has the cuff fit check function, whereas the predicate device BP3GX1-5X/3MW1-4B does not have the function. The cuff fit check function is to check pulse strength during measurement, and indicates if pulse strength is OK. If pulse is too weak, nothing is indicated. This function is just a reference for properly wearing the cuff and does not affect the accuracy and efficacy of the use according to the clinical and essential performance testing, so it does not affect the safety or effectiveness. #### 6) 28 Days Average Function The subject device BP3KV1-5W has the 28 days average function. whereas the predicate device BP3GX1-5X/3MW1-4B does not have the function. The 28days average function is an indication of the blood pressure level of the most recent 4 weeks. This stands for the average of measurement values from the last 28 days, so this doesn't affect the clinical test. {6}------------------------------------------------ #### 7) MyCheck Function The subject device BP3KV1-5W has the MyCheck function, whereas the predicate device BP3GX1-5X/3MW1-4B does not have the function. The MyCheck function displays a qualitative comparison of the most recent measurement versus the 4-week measurement average (stored in memory), to provide an immediate, simplified indication whether the current measurement is similar, above or below the 4-week average, without additional operations. #### 8) MvBP Function The subject device BP3KV1-5W has the MyBP function, whereas the predicate device BP3GX1-5X/3MW1-4B does not have the function. The MyBP function shows an averaged value, which includes only readings taken in the morning or evening of the most recent 3 to 7 days for computation, per clinical guideline suggestion. The value is displayed only when sufficient number of readings meeting the criteria (below) have been obtained, to provide a more clinically relevant average. If data in the memory are insufficient, it is not displayed. This function is based on the measured data, so this doesn't affect the clinical test. Based upon the aforementioned information, the three devices are substantially equivalent. #### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: Testing information demonstrating safety and effectiveness of the Microlife Upper Arm Automatic Digital Blood Pressure Monitor. Model BP3KV1-5W in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance requirements. The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices: The following National and International Standards were utilized for testing the subject device: - 1. IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R) 2012 - 2. IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for safety and essential performance – Collateral standard: Electromagnetic Disturbances - Requirements And Tests. - 3. ISO 14971: 2007 Medical devices Application of risk management to medical devices. - 4. AAMI/ANSI//ISO 10993-1:2009/(R)2013, Biological evaluation of medical devices Part 1: Evaluation And Testing Within A Risk Management Process. - 5. AAMI / ANSI / ISO 10993-10:2010/(R)2014. Biological evaluation of medical devices Part 10: Tests for Irritation and Skin Sensitization - 6. AAMI/ANSI/IEC 80601-2-30 Medical electrical equipment Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers, 2018 - 7. AAMI/ANSI/ISO 10993-5:2009/(R)2014, Biological evaluation of medical devices Part 5: Tests for In Vitro Cytotoxicity - 8. IEC 60601-1-11:2015 Medical Electrical Equipment Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical {7}------------------------------------------------ Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5W tested met all relevant requirements of the aforementioned tests #### 8. Discussion of Clinical Tests Performed: The proposed subject Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KV1-5W device uses the same deflation valve and sensor as BP3KT1-3N (European Device). Below is a comparison between the proposed BP3KV1-5W device and the BP3KT1-3N European Device: | Item | BP3KT1-3N European device | Proposed Subject BP3KV1-5W device | Similar or Different | |-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------| | Indications for Use | This oscillometric blood pressure<br>monitor is intended for measuring<br>non-invasive blood pressure in<br>people aged 12 years or older.<br><br>It is clinically validated in patients with<br>hypertension, hypotension, diabetes,<br>pregnancy, pre-eclampsia,<br>atherosclerosis, end-stage renal<br>disease, obesity and the elderly.<br><br>The device can detect an irregular<br>pulse suggestive of Atrial Fibrillation<br>(AF). Please note that the device is<br>not intended to diagnose AF. A<br>diagnosis of AF can only be<br>confirmed by ECG. The patient is<br>advised to see a physician.<br>Note: Content above from the CE<br>Mark version IFU.<br>This device is indicated for cuffs<br>covering the upper arm<br>circumference range from 17 to 52cm<br>(refer to<br>"Accessories"). | The Upper Arm Blood Pressure<br>Monitor, Model BP3KV1-5W is a<br>device intended to measure the<br>systolic and diastolic blood pressure<br>and pulse rate of an adult individual by<br>using a non-invasive oscillometric<br>technique in which an inflatable cuff is<br>wrapped around the upper arm for a<br>circumference range from 22 to 52cm.<br>The device is suitable for use by<br>adults, including adults with conditions<br>of diabetes, pregnancy, or<br>pre-eclampsia.<br>The device detects the appearance of<br>irregular heartbeat during<br>measurement and gives a warning<br>signal with the reading once the<br>irregular heartbeat is detected.<br>The device can be used in connection<br>with a smart phone via Bluetooth or<br>with a personal computer (PC) via<br>USB. The measurement data can be<br>transferred to a smart phone running<br>the Microlife Connected Health+<br>mobile software (App) or a PC running<br>the Microlife BP<br>Analyzer+ (BPA+) software | Different | | Device<br>Technology | Oscillometric method | Oscillometric method | √ | | Measuring<br>Location | Upper Arm | Upper Arm | √ | | Appearance (ID Design) | | | Different | | | | | | | Physical Dimension | 138 x 94.5 x 62.5 mm | 157.5mm x 105mm x 61.5mm | Different | | Microprocessor | HY16F198B<br>1. Hycon IC<br>2. Main clock frequency: 16 MHz<br>3. Sub clock frequency:32.768 kHz<br>4. System voltage level: 3.3 V<br>5. Pressure sensor type:<br>Semiconductor Sensor<br>6. The amount of MCU pins: 100<br>pins(package type) | R5F11NMEAFB<br>1. Renesas MCU core<br>2. Main clock frequency: 16 MHz<br>3. Sub clock frequency: 32.768 kHz<br>4. System voltage level: 3.3 V<br>5. Pressure sensor type:<br>Semiconductor Sensor<br>6. The amount of MCU pins: 80 pins | Different | | Sensor | Semiconductor sensor | Semiconductor sensor | √ | | Deflation Valve | Electromechanical solenoid valve | Electromechanical solenoid valve | √ | | Algorithm | 1. Blood pressure algorithm: Upper<br>arm for deflation measurement<br>2. Signal filter: FIR filter<br>3. Algorithm: Oscillometric<br>measurement<br>4. Irregular heart beat detection: Yes | 1. Blood pressure algorithm: Upper<br>arm for deflation measurement<br>2. Signal filter: FIR filter<br>3. Algorithm: Oscillometric<br>measurement<br>4. Irregular heart beat detection: Yes | √ | | Display | Digital liquid crystal display (Without<br>Touch pad)<br> | Digital liquid crystal display (Without<br>Touch pad)<br> | Different | | Pressure and<br>Pulse Rate<br>Accuracy | Pressure within ± 3 mmHg or 2% of<br>reading >200mmHg Pulse ± 5 % of<br>the reading | Pressure within ± 3 mmHg or 2% of<br>reading >200mmHg Pulse ± 5% of the<br>reading | √ | | Measuring<br>Range | SYS: 60-255 mmHg<br>DIA: 40-200 mmHg<br>Pulse: 40 to 199 beats per minute | SYS: 60-255 mmHg<br>DIA: 40-200 mmHg<br>Pulse: 40 to 199 beats per minute | √ | | Pressure<br>Resolution | 1mm Hg | 1mm Hg | √ | | Cuff Display<br>Range | O to 299mm Hg | O to 299mm Hg | √ | | Power Source | 4 AA batteries,<br>Or AC adapter 6 V DC 600 mA | 4 AA batteries,<br>Or AC adapter 6 V DC 600 mA | √ | | Low Battery<br>Voltage<br>Detection | YES | YES | √ | | Atrial Fibrillation<br>Detection | YES | NO | Different | | User | 1 | 2 | Different | | Last<br>Measurement<br>Recall | 1X99 sets | 2X99 sets | Different | | Beeper<br>Indication | NO | YES | Different | | Irregular<br>Heartbeat<br>Detection<br>Function | YES | YES | √ | | Traffic Light<br>Function | YES | YES | √ | | MAM Function | YES | YES | √ | | PC-link function | NO | YES | Different | | Blood Pressure<br>Analyzer<br>Software | NO | YES | Different | | Cuff Fit Check<br>Function | YES | YES | √ | | MyCheck<br>Function | NO | YES | Different | | MyBP Function | NO | YES | Different | | Display Backlight | NO | YES | Different | | AC/DC Adaptor<br>Compatible | YES | YES | √ | | 28 Day Memory<br>Day Average | NO | YES…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...