MICROLIFE WRIST WATCH BLOOD PREEURE MONITOR, BP3MO1-3P

K120430 · Microlife Intellectual Property GmbH · DXN · Aug 2, 2012 · Cardiovascular

Device Facts

Record IDK120430
Device NameMICROLIFE WRIST WATCH BLOOD PREEURE MONITOR, BP3MO1-3P
ApplicantMicrolife Intellectual Property GmbH
Product CodeDXN · Cardiovascular
Decision DateAug 2, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

The Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist for a circumference range from 13.5 to 21.5cm. The device detects the appearance of irregular heartbeat during measurement and gives a warning signal with the reading once the irregular heartbeat is detected.

Device Story

Wrist-worn blood pressure monitor; uses inflatable cuff and capacitor pressure sensor to detect cuff pressure oscillations; oscillometric method calculates systolic/diastolic blood pressure and pulse rate. Device features Irregular Heartbeat Detection (IHD) and traffic light function for visual feedback. Intended for adult self-use; provides readings on LCD screen. Benefits include convenient, non-invasive monitoring of blood pressure and pulse, with alerts for irregular heartbeats to facilitate patient awareness.

Clinical Evidence

Internal clinical testing performed to validate accuracy of blood pressure detection following integration of IMT technology. No repeated clinical testing per ANSI/AAMI SP10 deemed necessary due to technical identity with predicate device and lack of impact of added features on clinical accuracy.

Technological Characteristics

Non-invasive oscillometric wrist monitor; capacitor pressure sensor; automatic pump inflation; LCD display. Features: Irregular Heartbeat Detection (IHD), traffic light function. Software validation: moderate level of concern. Standards: ANSI/AAMI SP10:2008.

Indications for Use

Indicated for adult individuals (wrist circumference 13.5-21.5cm) for non-invasive measurement of systolic/diastolic blood pressure and pulse rate. Includes irregular heartbeat detection.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 2 2012 AUG Exhibit#1 ## 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510{k) number is: ## 1. Submitter's Identification: Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland Date Summary Prepared: May 2, 2012 Contact: Mr. Gerhard Frick Vice President of Technical and Service Microlife Intellectual Property GmbH, Switzerland Tel: +41 79 216 0070 E-Mail: gerhard.frick@microlife.ch ### 2. Name of the Device: Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P Requiation Number: 21 CFR Part 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: II Product Code: DXN ## 3. Information for the 510(k) Cleared Device (Predicate Device): a. Microlife Wrist Watch Blood Pressure Monitor, Model BP3MK1-3 (BP W100), K073398, Microlife Intellectual Property GmbH. b. OMRON Digital Wrist Blood Pressure Monitor, Model HEM-609N, K042505, Omron Healthcare, Inc. c. Microlife Upper Arm Blood Pressure Monitor BP3MC1-PC, K061471,Microlife Intellectual Property GmbH ## 4. Device Description: Microlife Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P is designed to measure systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses a capacitor pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and {1}------------------------------------------------ diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method". The device can detect electrical signals while inflating. And the device has Irreqular Heartbeat Detection (IHD) function. It detects the appearance of irreqular heartbeat during measurement and the irregular heart beat symbol "44 " is displayed on the LCD screen if any irreqular heart beat signal has been detected. ' In addition, the device has traffic light function. ## 5. Intended Use: The Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist for a circumference range from 13.5 to 21.5cm.. The device detects the appearance of irregular heartbeat during measurement and gives a warning signal with the reading once the irregular heartbeat is detected. #### 6. Comparison to the 510(k) Cleared Devices (Predicate Devices): The subject BP3MO1-3P and the predicate device model BP3MK1-3 (BP W100). use the same oscillometric method with the same fundamental scientific technology to determine the systolic and diastolic blood pressure and pulse rate. Wrist cuff is inflated automatically by pump and the pressures are transferred via tubing to a sensor in these two units. to The differences between the two models are the IMT technology, traffic light function and medication alarm function. The traffic light function is added to the subject device, while the medication alarm function is removed from subject device. But the differences do not affect the accuracy and normal use of this device because they use the same fundamental scientific technology based on the internal clinical test on BP3MO1-3P. The signal detection technology is same with what is used in the predicate device. Model HEM-609N, with 510(k) cleared number K042505. The subject BP3MO1-3P and the predicate BP3MC1-PC both have traffic light function. And the traffic light function has no impact on the clinical accuracy in terms of blood pressure detection. ## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: Testing information demonstrating safety and effectiveness of the Microlife Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance requirements. The following testing was conducted: {2}------------------------------------------------ - a. Reliability Test Storage test - Reliability Test Operating test ﻗ - Reliability Test Vibration test ﻥ - d. Reliability Test Drop test - e. Reliability Test Life test - EMC Test f. None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that Microlife Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P tested met all relevant requirements of the aforementioned tests. ## Discussion of Clinical Tests Performed: 8. The subject device Model BP3MO1-3P is from the technical point of view, identical to the predicate blood pressure monitor Model BP3MK1-3 (BP3MK1-3). The internal clinical test on BP3MO1-3P demonstrates that the clinical accuracy in terms of blood pressure detection and normal use of this device is not affected by change of signal detection technology to IMT technology. The other differences (traffic light function and medication alarm function) have no impact on the clinical accuracy in terms of blood pressure detection. Therefore repeated clinical test in accordance with the standard ANSI/AAMI SP10 is not necessary. ### 9. Software information: Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". ### 10. Conclusions: We have demonstrated that there are no significant differences between the Microlife Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P and the predicate devices, Model BP3MK1-3 (BP W100), Model HEM-609N, and model BP3MC1-PC in terms of safety and effectiveness based on electrical, mechanical and environmental test results per the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and the ANSI/AAMI Voluntary Standard, SP10: 2008. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three swooping lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. ## Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 > 2 2012 AUG Microlife Intellectual Property GmbH c/o Ms. Susan D. Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021 Re: K120430 > Trade/Device Names: Microlife Wrist Watch Blood Pressure Monitor, BP3MO1-3P Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: July 2, 2012 Received: July 3, 2012 Dear Ms. Goldstein-Falk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Susan D. Goldstein-Falk Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours Pram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use K120430 510(k) Number (if known): Device Name: Microlife Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P Indications For Use: The Microlife Wrist Watch Blood Pressure Monitor, Model BP3MO1-3P is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist for a circumference range from 13.5 to 21.5cm. The device detects the appearance of irregular heartbeat during measurement and gives a warning signal with the reading once the irregular heartbeat is detected. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use × (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |------------------------------------|---------| | Division of Cardiovascular Devices | | | 510(k) Number | K120430 | | Page | 1 of 1 | |------|--------| |------|--------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...