On Call Express II Blood Glucose Monitoring System
K192175 · ACON Laboratories, Inc. · NBW · Oct 17, 2019 · Clinical Chemistry
Device Facts
Record ID
K192175
Device Name
On Call Express II Blood Glucose Monitoring System
Applicant
ACON Laboratories, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Oct 17, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The On Call® Express II Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips, forearm and palm by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. Alternative site testing should be done only during steady-state times (when blood glucose level is not changing rapidly). The On Call® Express II Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is for in vitro diagnostic use only. The On Call® Express II Blood Glucose Monitoring System is not intended for the diamosis of or diabetes, nor intended for use on neonates. The On Call® Express Blood Glucose Test Strips are used with the On Call® Express II Blood Glucose meter in the quantitative measurement of glucose in capillary blood from the fingertips, forearm and palm.
Device Story
System measures glucose in fresh capillary whole blood; utilizes electrochemical biosensor technology; intended for single-patient home use. User applies blood sample to test strip inserted into meter; meter performs electrochemical analysis; displays glucose concentration on screen. Device features 500-record memory; powered by two CR2032 3.0V batteries. Designed for patient self-monitoring of blood glucose levels to assist in diabetes management. Meter housing is textured white plastic; dimensions 81.5 mm x 48.3 mm x 16 mm; weight 47.4 g. Validated for 260 cleaning/disinfection cycles using DisCide Ultra Towelettes to ensure material integrity and viral inactivation.
Clinical Evidence
No clinical data provided; substantial equivalence supported by bench testing.
Technological Characteristics
Electrochemical glucose test system. Meter housing: textured white plastic. Dimensions: 81.5 mm x 48.3 mm x 16 mm. Weight: 47.4 g. Power: 2x CR2032 3.0V batteries. Memory: 500 records. Hematocrit range: 25-60%.
Indications for Use
Indicated for single-patient home use for the quantitative measurement of glucose in fresh capillary whole blood samples.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
On Call Express Blood Glucose Monitoring System (K132086)
Related Devices
K191038 — On Call Express Voice Blood Glucose Monitoring System · ACON Laboratories, Inc. · Jun 12, 2019
K192220 — ReliOn Premier Classic Blood Glucose Monitoring System · I-Sens, Inc. · Dec 13, 2019
K072854 — DIACHEX BASIC BLOOD GLUCOSE MONITORING SYSTEM, DIACHEX SUPERB BLOOD GLUCOSE MONITORING SYSTEM · Tyson Bioresearch, Inc. · Jun 30, 2008
K031843 — MODIFICATION TO EVENCARE BLOOD GLUCOSE MONITORING SYSTEM · Medline Industries, Inc. · Jul 3, 2003
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the text "ADMINISTRATION".
October 17, 2019
ACON Laboratories, Inc. Qiyi Xie Senior Officer, Regulatory & Clinical Affairs 10125 Mesa Rim Road San Diego, CA 92121
Re: K192175
Trade/Device Name: On Call Express II Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: September 16, 2019 Received: September 18, 2019
Dear Qiyi Xie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192175
Device Name
On Call® Express II Blood Glucose Monitoring System
## Indications for Use (Describe)
The On Call® Express II Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips, forearm and palm by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. Alternative site testing should be done only during steady-state times (when blood glucose level is not changing rapidly).
The On Call® Express II Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is for in vitro diagnostic use only.
The On Call® Express II Blood Glucose Monitoring System is not intended for the diamosis of or diabetes, nor intended for use on neonates. The On Call® Express Blood Glucose Test Strips are used with the On Call® Express II Blood Glucose meter in the quantitative measurement of glucose in capillary blood from the fingertips, forearm and palm.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.