CARESENS BLOOD GLUCOSE MONITORING SYSTEM; CARESENS II METER, MODEL GM505C; CARESENS POP METER, MODEL GM505EA OR GM505EB
K080923 · I-Sens, Inc. · NBW · Oct 17, 2008 · Clinical Chemistry
Device Facts
Record ID
K080923
Device Name
CARESENS BLOOD GLUCOSE MONITORING SYSTEM; CARESENS II METER, MODEL GM505C; CARESENS POP METER, MODEL GM505EA OR GM505EB
Applicant
I-Sens, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Oct 17, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The CareSens II, CareSens POP Meters are used for the quantitative measurement of glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. CareSens Blood Glucose System is only for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip along with alternate sites such as forearm, palm, thigh and calf. CareSens Test Strips work with the CareSens II, CareSens POP meters to quantitatively measure glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management in the home or in clinical settings. CareSens Blood Glucose System is only for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip along with alternate sites such as forearm, palm, thigh and calf. CareSens Normal and Middle Control Solutions are a red liquid which is to be used to check that both the CareSens meters and CareSens test strips are working together properly. It contains a known range of glucose as specified on the vial.
Device Story
CareSens II and CareSens POP are electrochemical blood glucose monitoring systems. Device uses capillary action to draw blood into test strip chamber; glucose oxidase enzyme reacts with glucose in sample; resulting current measured by meter. System displays glucose concentration in 5 seconds. Used by patients at home or clinicians in clinical settings. Requires manual entry of batch-specific code number from test strip vial. Output used to monitor diabetes management effectiveness. Includes control solutions for performance verification. Not for diagnostic screening.
Clinical Evidence
Clinical data demonstrated performance of the CareSens system against laboratory glucose reference equipment. All predetermined acceptance criteria were satisfied, confirming substantial equivalence to the predicate device.
Technological Characteristics
In vitro diagnostic system; glucose oxidase enzyme; amperometric sensing; carbon electrodes; plasma-equivalent calibration; test range 20-600 mg/dL; 5-second test time; operating humidity 10-90%; supports mg/dL or mmol/L units; button-based coding system.
Indications for Use
Indicated for quantitative measurement of glucose in capillary whole blood to monitor diabetes management effectiveness in home or clinical settings. Suitable for use on fingertip, forearm, palm, thigh, and calf. For in vitro diagnostic use only.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K111371 — ON CALL CHOSEN BLOOD GLUCOSE MONITORING SYSTEM · ACON Laboratories, Inc. · Oct 6, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K080923
i-SENS, Inc. I Sens 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191
510(k) SUMMARY
(As required by 2I.CFR.807.92)
OCT 1 7 2008
| Type of 510(k): | Traditional |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Introduction: | According to the requirements of 21 CFR.807.92, the following<br>information provides sufficient data to understand the basis for a<br>determination of substantial equivalence. |
| Submitted By: | i-SENS, Inc.<br>465-6, Wolgye-dong, Nowon-gu, Seoul, 139-845, Korea<br>Tel.) ++82-2-916-6191<br>Fax) ++82-2-942-2514<br>www.i-sens.com |
| Contact Person: | Dr. Hyun Joon Oh, Phone: +82-33-903-0760. Fax +82-33-748-6191 |
| Date Summary,<br>Prepared: | March 18, 2008 |
| Device Name: | Trade name: CareSens Blood Glucose Monitoring System<br>Common Name: Blood Glucose Test System<br>Classification Name: Class II, 862.1345 Glucose Blood Tester |
| Predicate Device: | We claim substantial equivalence to the OneTouch® Ultra® from<br>LifeScan, Inc., |
| Device Description: | CareSens Blood Glucose Monitoring System is an in vitro diagnostic<br>device designed to measure the concentration of glucose in capillary<br>whole blood with CareSens Test Strips. The CareSens II model consists<br>of the CareSens II meter, CareSens test strips, CareSens control solution |
{1}------------------------------------------------
## 510(k) Summary, Continued
(Normal and Middle ranges), check strips, lancing device, lancets, user manual, quick reference guide and a logbook. CareSens POP model consists of the CareSens POP meter, CareSens test strips and lancing device, lancets, user guide, and quick guide.
#### The test principle is:
CareSens Blood Glucose Monitoring System is an in vitro diagnostic product intended for the measurement of glucose concentration in human blood, based on the glucose oxidase method for glucose determination. Each test strip features an electrode containing the enzyme glucose oxidase. A blood sample is applied to the blood collection area at the tip of the strip and is automatically drawn into the reaction zone, where the glucose oxidase catalyzes the oxidation of glucose to produce gluconic acid. During the reaction, a mediator transfers electrons to the electrode surface and generates a current. The amount of the current is proportional to the amount of glucose present in the blood sample. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. This reaction is measured by the Meter and displayed as your blood glucose result.
#### Intended Use:
The CareScns II, CarcSens POP Meters are used for the quantitative measurement of glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. CareSens Blood Glucose System is only for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip along with alternate sites such as forcarm, palm, thigh and calf.
{2}------------------------------------------------
## 510(k) Summary, Continued
CareSens Test Strips work with the CareSens II, CareSens POP meters to quantitatively measure glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management in the home or in clinical settings. CareSens Blood Glucose System is only for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip along with alternate sites such as forearm, palm, thigh and calf.
CareSens Normal and Middle Control Solutions are a red liquid which is to be used to check that both the CareSens meters and CareSens test strips are working together properly. It contains a known range of glucose as specified on the vial.
Comparison to The i-SENS, Inc. CareSens Blood Glucose Monitoring System is Predicate Device: substantially equivalent to the other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed item, the OneTouch® Ultra® by LifeScan, Inc.
#### Predicate Device 510(k) number: K024194
| Similarities | | |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Item | Device | Predicate |
| Enzyme | Glucose Oxidase | |
| Measurement principle | Amperometric method | |
| Test principle | Glucose Oxidase chemical reaction. The instrument measures the<br>extent of current cause by presence of glucose in sample. | |
| Intended use | The test strips work with the device to quantitatively measure<br>glucose in whole blood. The test Strips are for <i>in vitro</i> (i.e., outside | |
{3}------------------------------------------------
I sens Tel. 82-2-916-6191
# 510(k) Summary, Continued
# Similarities:
| | the body) diagnostic use only. |
|---------------------------|--------------------------------|
| Coding System | Button |
| Sample | Fresh capillary whole blood |
| Electrode | Carbon |
| Calibration | Plasma-equivalent |
| Test Time | 5sec |
| Test Range(mg/dL) | 20~600 |
| Glucose unites | Either mg/dL or mmol/L |
| Checking the system | Control solution |
| Alternate Site Capability | Yes |
| Operating Humidity | 10~90% |
| Data demonstrating<br>Substantial<br>equivalence: | The clinical data demonstrates the performance of the CareSens Blood<br>Glucose Monitoring System well with the laboratory glucose reference<br>test equipment. All predetermined acceptance criteria were satisfied. The<br>data also demonstrate that the CareSens Blood Glucose Monitoring<br>System is substantially equivalent to the predicate device. |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | CareSens Blood Glucose Monitoring System is substantially equivalent<br>to the following predicate device system: K024194<br>- LifeScan, Inc. OneTouch® Ultra® Blood Glucose Monitoring System. |
.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
I-Sens, Inc. c/o Hyun Joon Oh Division Manager, Quality Assurance 465-6 Wolgye-Dong, Nowon-Gu Seoul, Republic of Korea 139-845
OCT 7 7 2008
k080923 Trade Name: CareSens Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW, CGA, JJX Dated: August 25, 2008 Received: August 25, 2008
Dear Hyun Joon Oh:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
## Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indication for Use
510(k) Number (k080923):
Device Name: CareSens Blood Glucose Monitoring System
Indication For Use: The CareSens II, CareSens POP Meters are used for the quantitative mcasurement of glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. CareSens Blood Glucose System is only for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip along with alternate sites (forearm, palm, thigh and calf)
CareScns Test Strips work with the CareSens II, CareSens POP meters to quantitatively measure glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management in at home or in clinical settings. CareSens Blood Glucose System is only for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip along with alternate sites (forearm, palm, thigh and calf)
CareSens Normal and Middle Control Solutions are a red liquid which is to be used to check that both the CareSens meters and CareSens test strips are working together properly. It contains a known range of glucose as specified on the vial.
Prescription Use (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use
(21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol Benson
Division Sign-off
Division Sign Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K080923
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.