CLEVER CHEK BLOOD GLUCOSE MONITORING SYSTEM; ACHTUNG BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Taidoc Technology Corporation
Product Code
NBW · Clinical Chemistry
Decision Date
Apr 24, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Achtung TD-4207 no code/Clever Chek TD-4209 one-code/Clever Chek TD-4222 no code/Clever Chek TD-4225 no code/Clever Chek TD-4226 one-code Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The alternative site testing in above systems can be used only during steady-state blood glucose conditions.
Device Story
Device modification involving integrated circuit upgrade; increased CPU bit size, ROM, RAM, and ADC channels. Modification maintains fundamental scientific technology of previously cleared device. Intended use and labeling remain unchanged. Design control activities, including risk analysis and verification/validation, performed to confirm performance meets acceptance criteria.
Clinical Evidence
No clinical data provided in the summary document; substantial equivalence established via bench testing and comparison to predicate technology.
Technological Characteristics
Integrated circuit hardware upgrade; increased CPU bit size, ROM, RAM, and ADC channels. Fundamental scientific technology unchanged from predicate.
Indications for Use
Indicated for quantitative glucose measurement in fresh capillary whole blood (finger, palm, forearm, upper-arm, calf, thigh) for healthcare professionals and patients with diabetes mellitus to monitor glycemic control. Contraindicated for neonates and for diabetes diagnosis or screening. Alternative site testing restricted to steady-state glucose conditions.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K073494 — CLEVER CHEK BLOOD GLUCOSE MONITORING SYSTEM, MODELS TD-4209, -4222, -4225 AND -4226 AND ACHTUNG BLOOD GLUCOSE MONITORING · Taidoc Technology Corporation · Apr 2, 2008
K042005 — ACHTUNG TD-4207, CLEVER CHEK TD-4209, CLEVER CHEK TD-4222 · Taidoc Technology Corporation · Aug 27, 2004
K082482 — FORA TD-4251/U-RIGHT TD-4252/U-RIGHT TD-4254 BLOOD GLUCOSE MONITORING SYSTEM, MODEL TD-4251/TD-4252A/TD-4252B/TD-4254 · Taidoc Technology Corporation · Jan 30, 2009
K083570 — U-RIGHT TD-4249/U-RIGHT TD-4250 BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Mar 11, 2009
K063212 — CLEVER CHEK TD-4231 AND TD-4223 BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Nov 22, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
APR 2 4 2009
TaiDoc Technology Corporation c/o Ms. Sophie Lin 6F, No. 127, Wugong 2nd Rd., Wugu Township Taipei County, 241 Taiwan, R.O.C.
Re: k090185
Trade name: Achtung TD-4207 no code, Clever Chek TD-4209 one-code, Clever Chek TD-4222 no code, Clever Chek TD-4225 no code, Clever Chek TD-4226 one-code Blood Glucose Monitoring Systems Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, CGA Dated: April 16, 2009 Received: April 16, 2009
Dear Ms. Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
## Page - 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
G.H.
Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
## Enclosure
{2}------------------------------------------------
## Attachment 2
## Indications for Use
510(k) Number:
Device Name: Achtung TD-4207 no code, Clever Chek TD-4209 one-code, Clever Chek TD-4222 no code, Clever Chek TD-4225 no code, Clever Chek TD-4226 one-code Blood Glucose Monitoring System
Indications for Use:
The Achtung TD-4207 no code/Clever Chek TD-4209 one-code/Clever Chek TD-4222 no code/Clever Chek TD-4225 no code/Clever Chek TD-4226 one-code Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The alternative site testing in above systems can be used only during steady-state blood glucose conditions.
| Prescription Use | |
|-----------------------------|--|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | X |
|------------------------|---|
| (21 CFR 807 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
| Division Sign-Off | |
|------------------------------------------------------------|--------|
| Office of In Vitro Diagnostic Device Evaluation and Safety | |
| 510(k) | K09185 |
| A2 | Page 1 of 1 |
|----|-------------|
|----|-------------|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.