VivaChek Link Plus Blood Glucose Monitoring System

K233058 · Vivachek Biotech (Hangzhou) Co., Ltd. · NBW · Dec 22, 2023 · Clinical Chemistry

Device Facts

Record IDK233058
Device NameVivaChek Link Plus Blood Glucose Monitoring System
ApplicantVivachek Biotech (Hangzhou) Co., Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateDec 22, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

VivaChek™ Link Plus Blood Glucose Monitoring System is comprised of VivaChek™ Link Plus Blood Glucose Meter and the VivaChek™ Ino Blood Glucose Test Strips. VivaChek™ Link Plus Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Device Story

System uses electrochemical amperometric biosensor to measure glucose concentration in fresh capillary whole blood. Blood sample drawn into test strip via capillary action; glucose reacts with glucose oxidase and mediator to generate electrons; resulting current correlates to glucose concentration. Device includes 4G module for data transmission. Used by patients with diabetes at home for self-monitoring. Output displayed on meter screen; helps patients track glucose levels to manage diabetes. Device is handheld, battery-operated, and intended for single-patient use.

Clinical Evidence

Lay user study (n=352) demonstrated 100% of results within ±20% of comparator (YSI 2300) across 48.1–314.5 mg/dL range. Accuracy at extreme glucose concentrations (n=105) showed 100% within ±20% for both <80 mg/dL and >250 mg/dL ranges. Bench testing validated precision, linearity, interference, hematocrit (20-70%), altitude (up to 10,413 ft), and environmental operating conditions.

Technological Characteristics

Electrochemical amperometric biosensor; glucose oxidase chemistry. Rechargeable 3.7V lithium polymer battery. 4G wireless connectivity. Dimensions: 100.9mm x 61.6mm x 23.7mm. Weight: ~85g. Operating range: 20-600 mg/dL; 20-70% hematocrit; 41-113°F. Complies with FCC Part 15, 2, 24, 27, and 47 CFR § 2.1093.

Indications for Use

Indicated for persons with diabetes for quantitative measurement of glucose concentration in fresh capillary whole blood from fingertips to monitor diabetes control. Not for neonatal use, diagnosis, or screening. For single-patient, home-use, in vitro diagnostic self-testing.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT ## I Background Information: A 510(k) Number K233058 B Applicant VivaChek Biotech (Hangzhou) Co., Ltd. C Proprietary and Established Names VivaChek™ Link Plus Blood Glucose Monitoring System D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | NBW | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry | ## II Submission/Device Overview: A Purpose for Submission: New device B Measurand: Glucose in capillary whole blood from the fingertip C Type of Test: Quantitative amperometry (glucose oxidase). ## III Intended Use/Indications for Use: A Intended Use(s): Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} See Indications for Use below. ## B Indication(s) for Use: VivaChek™ Link Plus Blood Glucose Monitoring System is comprised of VivaChek™ Link Plus Blood Glucose Meter and the VivaChek™ Ino Blood Glucose Test Strips. VivaChek™ Link Plus Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared. ## C Special Conditions for Use Statement(s): OTC - Over The Counter - Not for use on critically ill patients, patients in shock or in a hyperglycemic-hyperosmolar state with or without ketosis. - Not for neonatal use. - Not for use on patient with severe dehydration. - Not for use on patient that is severely hypotensive. - Not for screening or diagnosis of diabetes mellitus. - Not for Alternative Site Testing (AST). - Do not use the system above 10413 ft (3174 meters) in altitude. - This meter is not intended for use in healthcare or assisted-use settings such as hospitals, physician offices, or long-term care facilities because it has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures. Use of this meter on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens. ## D Special Instrument Requirements: VivaChek™ Link Plus Blood Glucose Meter ## IV Device/System Characteristics: ### A Device Description: VivaChek™ Link Plus Blood Glucose Monitoring System, is designed to quantitatively measure the glucose concentration in fresh capillary whole blood from the fingertip. The system consists of: VivaChek™ Link Plus Blood Glucose Meter, VivaChek™ Ino Blood Glucose Test Strips, and the VivaChek™ Ino Control Solution with three levels (level 1, 2 and 3). The blood glucose meter contains a 3.7-volt lithium Ion Battery that can be charged using USB port, plastic housing and Buttons and a 4G module for data transmission. The materials needed but not provided include a single user lancing device (VivaChek lancing device) and sterile lancets (VivaChek lancets) both cleared under K220475, VivaChek™ Ino Control Solution and VivaChek™ Ino Blood Glucose Test Strips. K233058 - Page 2 of 12 {2} B Principle of Operation: VivaChek™ Link Plus Blood Glucose Monitoring System is designed to quantitatively measure the glucose concentration in fresh capillary whole blood. The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose oxidase and the mediator. Electrons are generated, producing a current that is proportional to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed. C Instrument Description Information: 1. Instrument Name: VivaChek™ Link Plus Blood Glucose Meter 2. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. 3. Specimen Sampling and Handling: The glucose system is intended to be used with capillary whole blood from the finger only. The whole blood sample is applied directly to the test strip by capillary action. 4. Calibration: The meter does not require calibration or coding by the user. The meter is automatically coded. 5. Quality Control: Three levels of glucose control solutions are available for use with this system and can be purchased separately. Recommendations on when to test with control solutions are provided in the labeling. Acceptable ranges for each level of control solution are printed on the test strip vial label. The user is cautioned not to use the meter if the control result falls outside these ranges. The control solution readings are automatically marked by the meter as control results and are not included in the patient result averages. This medical device product has functions subject to FDA premarket review as well as functions that are not subject to FDA premarket review. For this application, if the product has functions that are not subject to FDA premarket review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the functions subject to FDA premarket review or they are included as a labeled positive impact that was considered in the assessment of the functions subject to FDA premarket review. K233058 - Page 3 of 12 {3} V Substantial Equivalence Information: A Predicate Device Name(s): VivaChek™ Ino Smart Blood Glucose Monitoring System B Predicate 510(k) Number(s): K173140 C Comparison with Predicate(s): | Device & Predicate Device(s): | K233058 | K173140 | | --- | --- | --- | | Device Trade Name | VivaChek™ Link Plus Blood Glucose Monitoring System | VivaChek Ino Smart Blood Glucose Monitoring System | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | For the quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. | Same | | Test Strip | VivaChek™ Ino Blood Glucose Test Strips | Same | | Strip Chemical Composition | Glucose oxidase | Same | | Sample Type | Fresh capillary whole blood | Same | | Measurement Range | 20-600 mg/dL | Same | | Sample Volume | 0.8 μL | Same | | Test Time | 5 seconds | Same | | General Device Characteristic Differences | | | | Battery Type | Rechargeable, 800 mAh, 3.7 Volt DC, lithium polymer battery | Rechargeable, 250mAh, 3.7 Volt DC, lithium polymer battery | | Data Transmission | 4G | Bluetooth | K233058 - Page 4 of 12 {4} VI Standards/Guidance Documents Referenced: FDA Guidance Document: Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use. Guidance for Industry and Food and Drug Administration Staff. Issued on September 29, 2020. VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Within-Run Precision (Repeatability) The sponsor performed a repeatability study using 3 lots of VivaChek™ Ino Blood Glucose Test Strips, 10 VivaChek™ Link Plus Blood Glucose Meters and venous whole blood adjusted to 5 glucose concentration levels (i.e., Level 1: 30 to 50 mg/dL, Level 2: 51 to 110 mg/dL, Level 3: 111 to 150 mg/dL, Level 4: 151 to 250 mg/dL, and Level 5: 251 to 400 mg/dL). Each sample was tested in replicates of 10 on each of 10 meters using 3 lots of test strips for a total of 300 tests per glucose concentration. Results are summarized below: | Glucose Level (mg/dL) | Strip Lot | n | Mean (mg/dL) | SD (mg/dL) | %CV | | --- | --- | --- | --- | --- | --- | | Level 1 (30 to 50) | 1 | 100 | 40.0 | 1.7 | 4.3% | | | 2 | 100 | 40.2 | 1.8 | 4.4% | | | 3 | 100 | 40.3 | 1.7 | 4.3% | | | Combined | 300 | 40.2 | 1.7 | 4.3% | | Level 2 (51 to 110) | 1 | 100 | 70.5 | 2.1 | 2.9% | | | 2 | 100 | 70.4 | 2.2 | 3.1% | | | 3 | 100 | 70.5 | 2.0 | 2.9% | | | Combined | 300 | 70.4 | 2.1 | 3.0% | | Level 3 (111 to 150) | 1 | 100 | 129.4 | 4.0 | 3.1% | | | 2 | 100 | 129.2 | 3.6 | 2.8% | | | 3 | 100 | 129.7 | 3.4 | 2.6% | | | Combined | 300 | 129.4 | 3.6 | 2.8% | | Level 4 (151 to 250) | 1 | 100 | 200.2 | 4.9 | 2.4% | | | 2 | 100 | 200.3 | 5.1 | 2.5% | | | 3 | 100 | 198.4 | 5.8 | 2.9% | | | Combined | 300 | 199.6 | 5.3 | 2.7% | | Level 5 (251 to 400) | 1 | 100 | 349.3 | 8.7 | 2.5% | | | 2 | 100 | 351.7 | 9.1 | 2.6% | | | 3 | 100 | 349.8 | 7.8 | 2.2% | | | Combined | 300 | 350.3 | 8.6 | 2.5% | K233058 - Page 5 of 12 {5} Intermediate Precision (Between Run) Intermediate precision study was performed using 3 lots of VivaChek™ Ino Blood Glucose Test Strips, 10 VivaChek™ Link Plus Blood Glucose Meters, and 5 levels of glucose control solutions (i.e., Level 1: 30 to 50 mg/dL, Level 2: 51 to 110 mg/dL, Level 3: 111 to 150 mg/dL, Level 4: 151 to 250 mg/dL, and Level 5: 251 to 400 mg/dL). Each sample level was measured once a day with each meter and each test strip lot for 10 days, for a total of 300 replicates per level. Results are summarized below: | Glucose Level (mg/dL) | Strip Lot | n | Mean (mg/dL) | SD (mg/dL) | %CV | | --- | --- | --- | --- | --- | --- | | Level 1 (30 to 50) | 1 | 100 | 40.0 | 1.5 | 3.7% | | | 2 | 100 | 39.9 | 1.4 | 3.6% | | | 3 | 100 | 40.3 | 1.4 | 3.6% | | | Combined | 300 | 40.0 | 1.5 | 3.7% | | Level 2 (51 to 110) | 1 | 100 | 69.8 | 1.8 | 2.6% | | | 2 | 100 | 70.4 | 1.7 | 2.4% | | | 3 | 100 | 69.8 | 1.8 | 2.6% | | | Combined | 300 | 70.1 | 1.8 | 2.6% | | Level 3 (111 to 150) | 1 | 100 | 130.1 | 3.2 | 2.5% | | | 2 | 100 | 129.7 | 3.3 | 2.6% | | | 3 | 100 | 130.4 | 3.1 | 2.4% | | | Combined | 300 | 130.1 | 3.2 | 2.5% | | Level 4 (151 to 250) | 1 | 100 | 199.1 | 5.1 | 2.5% | | | 2 | 100 | 200.1 | 4.8 | 2.4% | | | 3 | 100 | 198.9 | 4.9 | 2.4% | | | Combined | 300 | 199.5 | 4.9 | 2.5% | | Level 5 (251 to 400) | 1 | 100 | 349.3 | 8.5 | 2.4% | | | 2 | 100 | 350.2 | 8.7 | 2.5% | | | 3 | 100 | 349.6 | 6.6 | 2.4% | | | Combined | 300 | 349.9 | 8.6 | 2.4% | 2. Linearity: A linearity study was conducted with 2 VivaChek™ Link Plus Blood Glucose Meters, 3 lots of VivaChek™ Ino Blood Glucose Test Strips, and 11 blood glucose concentration levels (19, 76, 135, 195, 250, 310, 365, 428, 482, 540, and 600 mg/dL) as determined by the comparative method (YSI 2300 STAT PLUS Glucose and L-Lactate Analyzer). VivaChek™ Link Plus Blood Glucose Monitoring System results were compared with the values obtained from the comparative method. The results from the linear regression analysis are summarized below: | Test Strip Lot# | Slope | y-intercept | R²-value | | --- | --- | --- | --- | | Lot 1 | 0.9968 | 0.9922 | 0.9981 | | Lot 2 | 1.0006 | 1.5950 | 0.9980 | | Lot 3 | 0.9908 | 2.3351 | 0.9977 | K233058 - Page 6 of 12 {6} The results of the study support the sponsor’s claimed glucose measurement range of 20 to 600 mg/dL. If a sample is less than 20 mg/dL, the result is flagged by the meter as “LO”. If a sample result is more than 600 mg/dL, the result is flagged by the meter as “HI”. The low and high functions were validated and demonstrated to function as intended. 3. Analytical Specificity/Interference: Interference studies were performed with 10 VivaChek™ Link Plus Blood Glucose Meters and 3 lots of VivaChek™ Ino Blood Glucose Test Strips. The study was conducted using venous whole blood samples adjusted to 3 glucose levels (50-70, 110-130, and 225-270 mg/dL) as measured by the comparative method (YSI 2300 STAT PLUS Glucose and L-Lactate Analyzer). The adjusted blood glucose concentrations were separated into a control sample with no interferent added and a test sample containing one of 32 potentially interfering substances. No significant interference between the test and the control samples was defined by the sponsor as within ± 4.8 mg/dL at &lt; 75 mg/dL glucose and within ± 8% at &gt; 75 mg/dL glucose. The highest concentrations at which no significant interference t was observed are presented in the table below: | Test Substance | Maximum tested concentration with no significant interference | | --- | --- | | Acetaminophen | 20 mg/dL | | Ascorbic acid | 6 mg/dL | | Conjugated Bilirubin | 50 mg/dL | | Unconjugated Bilirubin | 40 mg/dL | | Cholesterol | 500 mg/dL | | Creatinine | 15 mg/dL | | Dopamine | 20 mg/dL | | EDTA | 200 mg/dL | | Galactose | 60 mg/dL | | Gentisic acid | 27.8 mg/dL | | Reduced Glutathione | 92.9 mg/dL | | Hemoglobin | 20000 mg/dL | | Heparin | 800 IU/dL | | Ibuprofen | 50 mg/dL | | L-Dopa | 3 mg/dL | | Maltose | 480 mg/dL | | Mannitol | 1800 mg/dL | | Methyldopa | 10.5 mg/dL | | Salicylic acid | 60 mg/dL | | Sodium | 180 mmol/L | | Tolbutamide | 100 mg/dL | | Tolazamide | 40 mg/dL | | Triglycerides | 3000 mg/dL | | Uric acid | 24 mg/dL | K233058 - Page 7 of 12 {7} | Test Substance | Maximum tested concentration with no significant interference | | --- | --- | | Xylose | 200 mg/dL | | Sorbitol | 0.09 mg/dL | | Lactose | 25 mg/dL | | Tetracycline | 1.5 mg/dL | | Xylitol | 0.09 mg/dL | | Lactitol | 0.09 mg/dL | | Isomalt | 0.09 mg/dL | | Maltitol | 0.09 mg/dL | The sponsor included the following statement in the labeling: - Do not test your blood during or soon after a xylose absorption test. Xylose in the blood can give inaccurate results with this meter. 4. Assay Reportable Range: 20 to 600 mg/dL glucose 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): The system is traceable to NIST SRM 917c glucose reference material. A method comparison study was performed using the candidate devices and YSI 2300 STAT PLUS Glucose and L-Lactate Analyzer as the comparative method (see section VII.C.3). Open-vial and closed-vial Stability: Test strip stability was assessed with accelerated studies and real-time stability studies. Protocols and acceptance criteria were reviewed and found acceptable. The labeling includes claims that the test strips are stable for 6 months after opening and 24 months unopened when stored between of 36 to 86°F (2 to 30°C) and 10 to 90% relative humidity. 6. Detection Limit: Please see linearity in Section VII.A.2 7. Assay Cut-Off: Not applicable. 8. Accuracy (Instrument): Not applicable. 9. Carry-Over: Not applicable. K233058 - Page 8 of 12 {8} K233058 - Page 9 of 12 # B Comparison Studies: 1. Method Comparison with Predicate Device: Please refer to lay user study below in section VII.C3. 2. Matrix Comparison: Not applicable. # C Clinical Studies: 1. Clinical Sensitivity: Not applicable. 2. Clinical Specificity: Not applicable. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Method Comparison/Lay User Performance Study: To assess the performance of the VivaChek™ Link Plus blood glucose monitoring system in the hands of lay users, the sponsor performed a study with 352 lay user participants who collected and tested their own fingertip capillary blood samples using only the instructions from the product labeling in English. The glucose concentrations in the samples ranged from 48.1 to 314.5 mg/dL, including 54 samples with glucose levels &lt;80 mg/dL and 66 samples with glucose level &gt;250 mg/dL, as measured by the comparator method (YSI 2300 analyzer). Results were analyzed by comparing blood glucose results obtained from the VivaChek™ Link Plus blood glucose meter by the lay user against the laboratory comparator value obtained by healthcare professionals. Results are summarized in the tables below: | VivaChek™ Link Plus Blood Glucose Monitoring System are summarized below: Accuracy Across the Entire Glucose Range: | | | | | --- | --- | --- | --- | | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% | | 243/352 (69.0 %) | 339/352 (96.3 %) | 352/352 (100 %) | 352/352 (100 %) | | Regression Analysis: | | --- | | y = 0.9965x + 0.3553; R² = 0.9942 | # Usability: At the end of the lay-user study, each participant was asked to complete a usability questionnaire regarding ease of understanding of information in the user manual and the ease {9} of use when performing a blood glucose test. The sponsor's analysis of the questionnaire responses demonstrated that the participants were satisfied with the ease of operation by following the instructions for use in the User's Manual and with the overall performance of the VivaChek™ Link Plus blood glucose monitoring system. ## Labeling Readability: The readability of the user manual and test strip package insert, were evaluated using a Flesch-Kincaid analysis and demonstrated that the readability was less than an 8th grade level. ## Accuracy at Extreme Glucose Study: An accuracy study was performed to evaluate the performance of VivaChek™ Link Plus Blood Glucose Monitoring System was evaluated with the 105 fingertip blood samples containing extreme glucose concentrations at the extreme lower and upper ends of the claimed glucose measuring range. Of the 105 blood samples, 53 samples were less than or equal to 80 mg/dL and 52 samples were greater than or equal to 250 mg/dL. Results on the candidate devices using 3 test strip lots were compared to the results obtained using the comparative method (YSI 2300 STAT PLUS Glucose and L-Lactate Analyzer) and are summarized below: | For glucose concentrations < 80 mg/dL: | | | | | --- | --- | --- | --- | | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% | | 29/53 (54.7 %) | 49/53 (92.5%) | 53/53 (100 %) | 53/53 (100 %) | | For glucose concentrations > 250 mg/dL: | | | | | --- | --- | --- | --- | | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% | | 33/52 (63.5 %) | 48/52 (92.3 %) | 52/52 (100 %) | 52/52 (100 %) | ## D Clinical Cut-Off: Not applicable. ## E Expected Values/Reference Range: The sponsor includes the following expected blood glucose values for people without diabetes in the labeling: | Time | Normal plasma glucose range for adults without diabetes, mg/dL | | --- | --- | | Before breakfast (fasting) | < 100 | | 2 hours after a meal | <140 | K233058 - Page 10 of 12 {10} Reference: American Diabetes Association; Standards of Medical Care in Diabetes—2023 Abridged for Primary Care Providers. Clin Diabetes 2 January 2023; 41 (1): 4–31. F Other Supportive Instrument Performance Characteristics Data: 1. Hematocrit Study The effect of hematocrit on the VivaChek™ Link Plus Blood Glucose Monitoring System was evaluated with venous blood samples adjusted to hematocrit levels of 20%, 25%, 30%, 35%, 42%, 50%, 55%, 60%, 65%, and 70%. Each hematocrit level was adjusted to achieve five concentrations of glucose (30 to 50, 51 to 110, 111 to 150, 151 to 250, and 251 to 400 mg/dL). Each blood sample was tested using 10 VivaChek™ Link Plus Blood Glucose Meters with 3 lots of VivaChek™ Ino Blood Glucose Test Strips and tested on the YSI 2300 comparator method. The glucose values were compared to results obtained on the YSI 2300 comparator method, and to the normal hematocrit sample. The results support the claimed hematocrit range of 20% to 70% for the VivaChek™ Link Plus Blood Glucose Monitoring System. 2. System Operating Conditions Study The sponsor performed an operating condition study using venous whole blood samples adjusted to 3 glucose levels (60 to 70, 100 to 120, and 300 to 400 mg/dL). Testing was conducted under the following five temperature and relative humidity (RH) combinations: 113°F (45°C) and 41°F (5°C), at 10% and 90% RH each and at 73°F (23°C) at 40% RH. The candidate system results were compared to results from the YSI 2300 comparator method. The results support the claims in the labeling that the system can be used in conditions of 41–113°F (5-45°C) with relative humidity of 10 to 90%. 3. Altitude Effects The effect of altitude on the VivaChek™ Link Plus Blood Glucose Monitoring System was evaluated by testing whole blood samples at sea level (130 ft) and at 10,413 ft above sea level. The candidate system results were compared to those obtained with the comparator method (YSI 2300 analyzer). The results support the claims that the system functions as intended at altitudes up to the claimed altitude of 10,413 feet. 4. Sample Volume Study The sponsor conducted a sample volume study using venous blood samples prepared at three glucose range intervals at thee glucose levels (50-65, 110-120, and 200-250 mg/dL) that were tested at different sample volumes (0.6, 0.7, and 0.8 μL). The candidate system results were compared against the YSI 2300 comparator method measurements. Results support the claimed minimum sample volume of 0.8 μL. The sponsor also provided validation studies demonstrating that with blood volumes below 0.8 μL, the insufficient sample volume error message functioned as intended. 5. Flex Studies Intermittent sampling, sample perturbation, testing with used test strips, test strip early removal, drop/shock testing, and vibration testing was completed. The testing performed demonstrated that VivaChek™ Link Plus Blood Glucose Monitoring Systems are robust to intermittent sampling, sample perturbation, drop/shock, and vibration, and that an error message is returned to the user if a used test strip is inserted into the meter or test strip removal happens early. K233058 - Page 11 of 12 {11} K233058 - Page 12 of 12 6. Infection Control Studies The VivaChek™ Link Plus Blood Glucose Monitoring Systems is intended for single-patient use only. Disinfection efficacy studies were performed on the materials comprising the meters by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, Clorox Germicidal Wipes, 0.55% sodium hypochlorite (EPA Registration # 67619-12). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meter after 608 cleaning and disinfection cycles with the Clorox Germicidal Wipes. The robustness studies were designed to simulate 5 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures. 7. Electrical Safety and EMC Studies The sponsor provided documentation certifying that acceptable electrical safety and electromagnetic compatibility (EMC) testing had been performed and the system was found to be compliant. 8. Test Strip Lot Release Protocol The test strip lot release protocol and acceptance criteria were reviewed and found to be acceptable. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...