RIGHTEST Blood Glucose Monitoring System Max Tel

K240637 · Bionime Corporation · NBW · Nov 4, 2024 · Clinical Chemistry

Device Facts

Record IDK240637
Device NameRIGHTEST Blood Glucose Monitoring System Max Tel
ApplicantBionime Corporation
Product CodeNBW · Clinical Chemistry
Decision DateNov 4, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

RIGHTEST Blood Glucose Monitoring System Max Tel is intended to the quantitative measurement of glucose (sugar) in fresh capillary whole drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared. RIGHTEST Blood Glucose Monitoring System Max Tel is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to montor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steadystate times (when glucose is not changing rapidly). The RIGHTEST Blood Glucose Monitoring System Max Tel is comprised of the RIGHTEST Meter Max Tel and the RIGHTEST Blood Glucose Test Strip Max.

Device Story

System measures glucose in fresh capillary whole blood via amperometric electrochemical sensor. User applies blood to test strip; capillary action draws sample into strip tip. FAD-glucose dehydrogenase reacts with glucose and mediator to generate electrons; current produced correlates to glucose concentration. Meter displays result. Used at home by patients with diabetes. Device includes meter, test strips, control solution, lancing device, and lancets. Features LTE connectivity for data transmission. Healthcare providers review results to assess diabetes control effectiveness. Benefits include convenient, rapid self-monitoring of blood glucose levels.

Clinical Evidence

Bench testing only. Performance verified via extreme glucose studies using natural and modified blood samples to confirm accuracy per FDA SMBG OTC guidelines. Software safety, FCC compliance for LTE functionality, and cybersecurity evaluations were performed. All acceptance criteria met.

Technological Characteristics

Amperometric glucose dehydrogenase-FAD assay. Measuring range: 20-600 mg/dL. Connectivity: LTE network. Traceable to NIST SRM #917c. Standards: ISO 14971 (risk management), AAMI TIR57 (security), IEC 62304 (software lifecycle).

Indications for Use

Indicated for self-testing of blood glucose levels in people with diabetes at home to monitor diabetes control. Not for diagnosis, screening, or neonatal use. Alternative site testing (fingertips, forearm, palm) restricted to steady-state glucose conditions.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT ## I Background Information: A 510(k) Number K240637 B Applicant Bionime Corporation C Proprietary and Established Names RIGHTEST Blood Glucose Monitoring System Max Tel D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | NBW | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry | ## II Submission/Device Overview: A Purpose for Submission: Modification to an existing device to expand the claimed glucose measuring range (from 50-550 mg/dL to 20-600 mg/dL) and add Long Term Evolution (LTE) network capabilities. B Measurand: Glucose in capillary whole blood from the fingertips, forearm, or palm C Type of Test: Quantitative amperometric assay (glucose dehydrogenase-FAD) ## III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} B Indication(s) for Use: RIGHTEST Blood Glucose Monitoring System Max Tel is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared. RIGHTEST Blood Glucose Monitoring System Max Tel is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). The RIGHTEST Blood Glucose Monitoring System Max Tel is comprised of the RIGHTEST Meter Max Tel and the RIGHTEST Blood Glucose Test Strip Max. C Special Conditions for Use Statement(s): - OTC - Over The Counter - For in vitro diagnostic use only - For self-testing - Single-patient use only - Not for use on neonates - RIGHTEST Blood Glucose Monitoring System Max Tel can only use with capillary whole blood samples. - DO NOT use the results from alternative sites (palm, forearm) for insulin dose calculations. - DO NOT use the results from alternative site testing (palm, forearm) to calibrate Continuous Glucose Monitoring (CGM) devices. - Not for use on critically ill patients, severely hypotensive individuals, patients in shock, dehydrated patients, or in a hyperglycemic-hyperosmolar state with or without ketosis. - Do not use at altitudes greater than 10,000 feet (3,048 meters). - Severe dehydration and excessive water loss may cause inaccurately low results. - Not for screening or diagnosis of diabetes mellitus. - Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens. D Special Instrument Requirements: RIGHTEST Blood Glucose Meter Max Tel IV Device/System Characteristics: A Device Description: RIGHTEST Blood Glucose Monitoring System Max Tel, is designed to quantitatively measure the glucose concentration in fresh capillary whole blood from fingertip, palm or forearm. RIGHTEST Blood Glucose Monitoring System Max Tel consists of: RIGHTEST Blood Glucose Meter Max Tel, Blood Glucose Test Strip Max, RIGHTEST Control Solution GC700 (Level 1, K240637 - Page 2 of 8 {2} Level 2 and Level 4), RIGHTEST Lancing Device and Sterile Lancets (K221062). The RIGHTEST Blood Glucose Test Strip Max is the same as Test Strip Max cleared in K173638. The test strips, control solutions and lancing device can be purchased separately. ## B Principle of Operation: The RIGHTEST Blood Glucose Monitoring System Max Tel is designed to quantitatively measure the glucose concentration in fresh capillary whole blood. The glucose measurement is achieved by using the amperometric detection method that uses glucose dehydrogenase flavin-adenine dinucleotide (GDH-FAD) based chemistry. When a drop of blood is applied to the test strip it is pulled into the test strip through capillary action. Glucose in the sample reacts with test strip chemistry generating electrons and producing a current that is proportional to the glucose concentration in the sample. After the reaction time, the detected current is calculated by the meter and the resulting glucose concentration is displayed by the meter. The RIGHTEST Blood Glucose Monitoring System Max Tel system reports glucose results as plasma glucose. ## C Instrument Description Information: 1. Instrument Name: RIGHTEST Blood Glucose Meter Max Tel 2. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. 3. Specimen Sampling and Handling: Fresh capillary whole blood from user's fingertips, palm, or forearm. Samples are to be tested immediately upon collection. The whole blood sample is applied directly to the test strip by capillary action. 4. Calibration: No user calibration is required. The meter is automatically coded. 5. Quality Control: Three levels of control solution (Level 1, Level 2, Level 4) are for use to perform quality control testing to check whether the system is working properly. Instructions on when to perform a control test, details about the control solution, performing a control test, understanding out-of-range control results are provided in the labeling. User needs to enter the control solution mode (CS mode) on the meter to perform a quality control test. Control solution ranges are printed on the test strip vial label. This medical device product has functions subject to FDA premarket review as well as functions that are not subject to FDA premarket review. For this application, if the product has functions K240637 - Page 3 of 8 {3} that are not subject to FDA premarket review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the functions subject to FDA premarket review or they are included as a labeled positive impact that was considered in the assessment of the functions subject to FDA premarket review. K240637 - Page 4 of 8 V Substantial Equivalence Information: A Predicate Device Name(s): RIGHTEST Blood Glucose Monitoring System Max Tel B Predicate 510(k) Number(s): K231192 C Comparison with Predicate(s): | Device & Predicate Device(s): | K240637 | K231192 | | --- | --- | --- | | Device Trade Name | RIGHTEST Blood Glucose Monitoring System Max Tel | Same | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | For the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. | Same | | Measurement Technology | Glucose dehydrogenase-FAD | Same | | Sample Type | Fresh capillary whole blood | Same | | General Device Characteristic Differences | | | | Measuring Range | 20 - 600 mg/dL | 50 - 550 mg/dL | | Data Transmission | LTE network | N/A | {4} VI Standards/Guidance Documents Referenced: ISO14971-Third edition 2019-12: Medical Devices-Application of Risk Management to Medical Devices AAMI TIR57: 2016: Principles for medical device security - Risk management IEC- 62304 Edition 1.1 2015-06: Medical device software-Software life cycle processes VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Previously established in K231192. 2. Linearity: The linearity study provided in K231192 supports the sponsor’s claimed glucose measurement range of 20 - 600 mg/dL. The meter displays “LO” with glucose values below 20 mg/dL and “HI” with glucose values over 600 mg/dL. The LO and HI functions were validated and were demonstrated to function as intended. 3. Analytical Specificity/Interference: Previously established in K231192. The sponsor has included the following in the labeling: The sponsor has the following statements in their labeling: - If you have a condition, such as kidney disease or gout, that may cause your blood levels of uric acid to rise to more than 12 mg/dL, the results from your meter may not be correct. - If you are taking a high level of vitamin C (ascorbic acid level in your blood &gt; 3 mg/dL), your blood glucose results may not be reliable. If you are unsure, ask your doctor. - If you have a condition, such as jaundice, that may cause your blood levels of Conjugated Bilirubin to rise to more than 30 mg/dL, the results from your meter may not be correct. - Do not test your blood glucose during or soon after a xylose absorption test. Xylose in the blood can give inaccurate results with this meter. 4. Assay Reportable Range: Glucose 20 - 600 mg/dL. K240637 - Page 5 of 8 {5} 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): The system is traceable to NIST (National Institute of Standards and Technology) standard reference material NIST SRM #917c. A method comparison was performed using the candidate device and a YSI 2300 comparator method. Test strip stability protocols and acceptance criteria were previously reviewed in K231192 and found acceptable to support the labeling claims that the test strips are stable for 4 months after first being opened, and that closed vials are stable for 24 months when at the recommended storage temperatures 39°F - 86°F (4°C - 30°C) and 10 - 90% relative humidity. The labeling instructs the users not to freeze the test strips. 6. Detection Limit: The assay reportable range is 20 - 600 mg/dL. 7. Assay Cut-Off: Not applicable 8. Accuracy (Instrument): Not applicable 9. Carry-Over: Not applicable B Comparison Studies: 1. Method Comparison with Predicate Device: See Section VII.C.3 2. Matrix Comparison: Not applicable C Clinical Studies: 1. Clinical Sensitivity: Not applicable 2. Clinical Specificity: Not applicable 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): K240637 - Page 6 of 8 {6} Accuracy at Extreme Glucose Values: To supplement the previous accuracy performance studies conducted in K231192 and support the broader measuring range claim (from previously cleared 50-550 mg/dL in K231192 to 20-600 mg/dL) an additional extreme glucose study was performed using 113 capillary blood samples to assess the accuracy of the Rightest Blood Glucose Monitoring System Max Tel at the extreme glucose concentrations at the extreme lower and upper ends of the measuring range. Of the 113 samples, 72 were altered by glycolysis or spiking to achieve appropriate glucose concentrations to achieve 53 samples ranging from 20.4 to 77.4 mg/dL glucose and 60 samples with glucose concentrations ranging from 257 to 599 mg/dL (as measured by YSI 2300). Each sample was measured on the candidate device and results compared to results obtained on the YSI 2300 analyzer. Results are summarized in the table below: | For glucose concentrations < 80 mg/dL | | | | | --- | --- | --- | --- | | Within ±5% | Within ±10% | Within ±15% | Within ±20% | | 42/53 (79.2%) | 50/53 (94.3%) | 53/53 (100.0%) | 53/53 (100.0%) | | For glucose concentrations > 250 mg/dL | | | | | Within ±5% | Within ±10% | Within ±15% | Within ±20% | | 51/60 (85.0%) | 59/60 (98.3%) | 60/60 (100.0%) | 56/60 (100.0%) | D Clinical Cut-Off: Not applicable E Expected Values/Reference Range: The sponsor includes the following in the labeling for the expected blood glucose values for people without diabetes: &lt; 100 mg/dL fasting &lt;140 mg/dL 2 hours after a meal Reference: American Diabetes Association; Standards of Care in Diabetes—2023 Abridged for Primary Care Providers. Clin Diabetes 2 January 2023; 41 (1): 4–31. K240637 - Page 7 of 8 {7} F Other Supportive Instrument Performance Characteristics Data: Software and Cybersecurity: The sponsor provided software and cybersecurity documentation that was reviewed and found to be acceptable. Test Strip Lot Release: The test strip lot release protocol and acceptance criteria were reviewed and found to be acceptable. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K240637 - Page 8 of 8
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...