ASSURE PRO BLOOD GLUCOSE MONITORING SYSTEM, MODEL 46001
K090332 · Arkray Factory USA, Inc. · NBW · Aug 14, 2009 · Clinical Chemistry
Device Facts
Record ID
K090332
Device Name
ASSURE PRO BLOOD GLUCOSE MONITORING SYSTEM, MODEL 46001
Applicant
Arkray Factory USA, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Aug 14, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Assure Pro Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. The Assure Pro Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
Device Story
System uses biosensor technology to measure glucose in fresh capillary whole blood from fingertips; requires manual code chip entry per test strip vial. Sample applied to test strip; glucose oxidase (GOD) and Hexammineruthenium (III) chloride reaction produces electric current proportional to glucose concentration. Meter converts current to glucose value. Used by patients at home or healthcare professionals in clinics. Output displayed on meter; aids monitoring of diabetes control effectiveness. Benefits include rapid, quantitative glucose monitoring for diabetes management.
Clinical Evidence
Method comparison study performed with 143 capillary samples (106 patient, 37 spiked/glycolyzed) ranging 28-495 mg/dL. Accuracy: For glucose <75 mg/dL, 100% within ±10 mg/dL. For glucose ≥75 mg/dL, 99% within ±15% and 100% within ±20%. Precision testing (within-run and day-to-day) showed CVs ranging 1.15%–5.67%. Bench testing included interference, altitude (up to 10,000 ft), and hematocrit (30-55%) studies.
Technological Characteristics
In vitro diagnostic glucose monitoring system. Uses test strips and meter for electrochemical measurement. Includes two levels of control solutions for quality verification. System is designed for capillary whole blood samples.
Indications for Use
Indicated for quantitative glucose measurement in fresh capillary whole blood from fingertips. For home (OTC) use by persons with diabetes or clinical use by healthcare professionals to monitor diabetes control. Contraindicated for neonates and not for diabetes diagnosis or screening.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K090653 — POCKETCHEM EZ BLOOD GLUCOSE MONITORING SYSTEM · Arkray Factory USA, Inc. · Oct 9, 2009
K053079 — ASSURE PRO BLOOD GLUCOSE MONITORING SYSTEM, MODEL 460001 · Hypoguard USA, Inc. · Mar 28, 2006
K051514 — ASSURE PRO BLOOD GLUCOSE MONITORING SYSTEM, MODEL 460002 · Hypoguard USA, Inc. · Oct 4, 2005
K063771 — GLUCOCARD X-METER BLOOD GLUCOSE MONITORING SYSTEM · Arkray Factory USA, Inc. · Aug 9, 2007
K092887 — EASYPLUS MINI MASTERDRIVER SELF MONITORING BLOOD GLUCOSE SYSTEM, MODEL EPS09009 · Eps Bio Technology Corp. · Oct 21, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K090332
#### 7 510(k) Summary
AUG 1 4 2009
| Submitter: | ARKRAY Factory USA, Inc.<br>5182 W. 76th Street<br>Minneapolis, MN 55439 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Hamid Idrissi<br>Regulatory Affairs Project Manager<br>ARKRAY Factory USA, Inc.<br>5182 W. 76th Street<br>Minneapolis, MN 55439<br>Phone: 952-646-3171<br>Fax: 952-646-3110<br>idrissih@ARKRAYusa.com. |
| Date Prepared: | February 6, 2009 |
| Trade Name: | Assure Pro Meter Test Strips |
| Classification: | Glucose test system, 21 CFR 862.1345 and 862.1660; Class II |
| Product Codes: | CGA, NBW, JJX |
| Predicate Device: | Assure Pro Meter (K053079) |
| Device Description: | The Assure Pro Blood Glucose Monitoring System consists of a meter,<br>test strips, and two levels of control solutions for use as an aid to<br>monitor the effectiveness of diabetes control. |
| Intended Use: | The Assure Pro Blood Glucose Monitoring System is intended for the<br>quantitative measurement of glucose in fresh capillary whole blood<br>samples drawn from the fingertips. Testing is done outside the body<br>( <i>In Vitro</i> diagnostic use). It is indicated for use at home (over the<br>counter [OTC]) by persons with diabetes, or in clinical settings by<br>healthcare professionals, as an aid to monitor the effectiveness of<br>diabetes control. |
| Functional and<br>Safety Testing: | A full array of in-house and clinical testing was done consistent with<br>relevant FDA guidance's for blood glucose monitoring systems.<br>Bench testing included evaluation of interferences, linearity, linear<br>range, altitude effects, control solution functionality, and analytical<br>precision.<br>Clinical testing included evaluation of accuracy for the finger stick. |
| Conclusion: | Labeling, bench testing results and clinical testing results support the<br>Indications for Use and the claim of substantial equivalence to the<br>predicate. |
・・
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
Arkray Factory USA. Inc. c/o Mr. Hamid Idrissi Regulatory Affairs Project Manager 5182 West 76th Street Minneapolis, MN 55439
AUG 1 4 2009
k090332 Re:
> Trade Name: Assure Pro Blood Glucose Monitoring System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW. CGA. JJX Dated: July 31, 2009 Received: August 4, 2009
Dear Mr. Idrissi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{2}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
G.C.H.
Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# 6- Indications for Use Statement
510(k) Number (if known):
## Device Name: Assure Pro Meter Blood Glucose Monitoring System
Indications For Use:
Assure Pro Blood Glucose Monitoring System:
The Assure Pro Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. The Assure Pro Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
#### Assure Pro Blood Glucose Meter:
The Assure Pro Blood Glucose Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.
### Assure Pro Blood Glucose Test Strips:
Assure Pro test strips are intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips when used with the Assure Pro Blood Glucose Meter. Testing is done outside the body (In Vitro diagnostic use). They are indicated for use in home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.
#### Assure Dose CONTROL:
For use with Assure Pro Blood Glucose Meter and Assure Pro Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results. Control solutions are available in two levels - Level 1 (Normal) and Level 2 (High).
Over-The-Counter Use Prescription Use AND/OR × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
rrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) vision Sign-Off
Office of in Vitro Diagnostic Device Evaluation and Sat
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.