JAZZ BLOOD GLUCOSE MONITORING SYSTEM

K071393 · Agamatrix · NBW · Jul 20, 2007 · Clinical Chemistry

Device Facts

Record IDK071393
Device NameJAZZ BLOOD GLUCOSE MONITORING SYSTEM
ApplicantAgamatrix
Product CodeNBW · Clinical Chemistry
Decision DateJul 20, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

AgaMatrix JAZZ™ Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Blood Glucose Meter is intended for use with AgaMatrix JAZZTM Blood Glucose Test Strips for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Blood Glucose Test Strips are intended for use with AgaMatrix JAZZ™ Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Control Solutions are intended for use with the AgaMatrix JAZZ™ Meter and AgaMatrix JAZZ™ Test Strips as a quality control check to verify the accuracy of blood glucose test results.

Device Story

Portable, battery-operated blood glucose meter; uses test strips to measure glucose in fresh capillary whole blood samples; intended for OTC home use by diabetics or clinical use by professionals; provides quantitative glucose concentration output; aids in monitoring diabetes control effectiveness; system includes meter, lancing device, lancets, and control solutions; brand name change from Liberty Blood Glucose Monitoring System with minor user interface improvements.

Clinical Evidence

Bench testing only; no clinical data provided. Verification and validation activities performed per design control requirements to confirm that modifications met predetermined acceptance criteria.

Technological Characteristics

Portable, battery-operated glucose test system; in vitro diagnostic; includes meter, test strips, and control solutions; complies with ISO 15197:2003, ISO 14971:2000, IEC 61010-1, IEC 61010-2-101, and IEC 61000-4-3.

Indications for Use

Indicated for persons with diabetes for quantitative blood glucose monitoring using fresh capillary whole blood (finger, palm, forearm). Not for use with neonates.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER K071393 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k052762 and k063409 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for a name change from Liberty to JAZZ, user interface change - screen is larger, added meal tagging capability. Hardware changes were as follows - the meter shell, additional button added to turn the backlight on and off and batteries are now dedicated to meter function or backlight. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics. Labeling changes were made in association with the above modification in the calibration section, expected value section and performance characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. {1} 2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...