CLEVER CHEK TD-3213/ CLEVER CHEK TD-3215/ DR. T TD-3216/ CLEVER CHEK TD-3217/ CLEVER CHEK TD-3250 BLOOD GLUCOSE PLUS BLO
Applicant
Taidoc Technology Corporation
Product Code
NBW · Clinical Chemistry
Decision Date
Jan 12, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Intended Use
The Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217/ Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. The system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual, over age 16, at home. For the Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217 system, the blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 5.25”~7.75”. For Clever Chek TD-3250 system, the blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the arm. The cuff circumference is limited to 9.4”~13.8”.
Device Story
System combines quantitative amperometric glucose monitoring and non-invasive oscillometric blood pressure measurement. Glucose input: fresh capillary whole blood applied to test strip; glucose oxidase enzyme reaction generates electrical current proportional to glucose concentration. Blood pressure input: inflatable cuff (wrist or arm) measures systolic/diastolic pressure and pulse rate. Used at home by patients or by healthcare professionals. Output: digital display of glucose concentration, blood pressure, and pulse. Assists in monitoring diabetes control programs; not for diagnosis/screening. System includes meter, test strips, code strip, control solutions, and lancet device. Changes from predicate include strip volume, application port, read time, and addition of alternative site testing (AST) sites.
Clinical Evidence
Performance evaluated in laboratory and clinical settings by healthcare professionals and lay users. Studies confirmed suitability for intended use in glucose monitoring and blood pressure measurement. No specific sensitivity/specificity metrics provided in summary.
Technological Characteristics
System includes monitor, wrist/arm cuffs, and dry reagent test strips. Glucose sensing via electrochemical method; blood pressure via oscillometric method. Cuff sizes: 5.25"-7.75" (wrist) or 9.4"-13.8" (arm).
Indications for Use
Indicated for quantitative glucose measurement in fresh capillary whole blood (finger, palm, forearm, upper-arm, calf, thigh) and non-invasive blood pressure/pulse rate measurement in adults >16 years. Intended for home use by diabetics or professional use. Contraindicated for neonates; alternative site testing only during steady-state glucose conditions.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
ACHTUNG Blood Glucose Monitoring System (k061181)
Clever Chek TD-3213 (k042795)
Clever Chek TD-3250 Blood Glucose and Blood Pressure Measurement Systems (k061073)
Related Devices
K051936 — CLEVER CHECK TD-3215/DR. T TD-3216 BLOOD GLUCOSE AND BLOOD PRESSURE MEASUREMENT SYSTEMS · Taidoc Technology Corporation · Aug 8, 2005
K061073 — CLEVER CHECK TD-3250 BLOOD GLUCOSE AND BLOOD PRESSURE MEASUREMENT SYSTEM, MODEL TD-3250 · Taidoc Technology Corporation · Jun 2, 2006
K080014 — CLEVER CHEK TD-3250C BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTEM · Taidoc Technology Corporation · Jun 9, 2008
K093592 — ADVOCATE REDI-COD DUO BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTEM, MODEL TD-3223 · Taidoc Technology Corporation · Feb 25, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
# 510 (k) Summary
K062800
ട്ടിക്കുന്നു. 2007
| 1. | Company Information | |
|----|---------------------|-----------------------------------------------------------------------------------|
| | Company name | TaiDoc Technology Corporation |
| | Contact person | Pi-Shiou Li |
| | Address | 4F, No.88, Sec. 1, Kwang-Fu RD.,<br>San-Chung, Taipei County, Taiwan, R.O.C., 241 |
| | Phone | 886-2-6635-8080 |
| | E-mail | erica@taidoc.com |
| | FAX | 886-2-6635-5959 |
・
## 2. Device Identification
| Trade name | Clever Chek TD-3213 Blood Glucose plus Blood Pressure<br>Monitoring System<br>Clever Chek TD-3215 Blood Glucose plus Blood Pressure<br>Monitoring System<br>Dr. T TD-3216 Blood Glucose plus Blood Pressure<br>Monitoring System<br>Clever Chek TD-3217 Blood Glucose plus Blood Pressure<br>Monitoring System<br>Clever Chek TD-3250 Blood Glucose plus Blood Pressure<br>Monitoring System |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name | Blood Glucose and Blood Pressure Measurement System<br>Blood Glucose Test Strip |
| Classification name | Class II devices<br>21 CFR Section 862.1345, Glucose Test System<br>21 CFR Section 870.1130, Non-invasive Blood Pressure<br>Measurement System |
### 3. Predicate Device
| K061181 | ACHTUNG Blood Glucose Monitoring System by TaiDoc<br>Technology Corporation. |
|---------|--------------------------------------------------------------------------------------------------------------|
| K042795 | Clever Chek TD-3213 Blood Glucose and Blood Pressure<br>Measurement System by TaiDoc Technology Corporation. |
| K061073 | Clever Chek TD-3250 Blood Glucose and Blood Pressure<br>Measurement System by TaiDoc Technology Corporation. |
{1}------------------------------------------------
#### 4. Device Description
The Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217/ Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System consists of a monitor with wrist/arm cuff and test strips. The system utilizes an electrochemical method-based meter and dry reagent biosensor (test strips) for blood glucose testing. The size of the current is proportional to the amount of glucose present in the sample, providing a quantitative measurement of glucose in fresh whole blood and control solutions.
The system is able to provide blood pressure measurement which adopts the "oscillometric method" as the measuring principle. It provides the systolic pressure, diastolic blood pressure and pulse rate on an adult individual, over age 16, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist (Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217) or in the arm (Clever Chek TD-3250). The cuff circumference is limited to 5.25" ~7.75" (13 ~20 cm) for wrist or 9.4"~13.8" (24~35 cm) for arm.
- 5. Intended Use
The Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217/ Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System is indicated for the quantitative measurement of glucose in fresh whole blood (capillary blood from the finger, the palm, the forearm, the upper-arm, the calf and the thigh) for self testing by persons with diabetes in the home or by healthcare professionals in healthcare facilities. Testing is done outside the body (in vitro diagnostic use).
The system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual, over age 16, at home.
For Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217 system, the blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 5.25" ~ 7.75".
For Clever Chek TD-3250 system, the blood pressure is measured by using
{2}------------------------------------------------
a technique in which an inflatable cuff is wrapped around the arm. The cuff circumference is limited to 9.4" ~ 13.8".
- 6. Comparison to Predicate Device
The Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217/ Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System has equivalent technological characteristics as the ACHTUNG Blood Glucose Monitoring System (K061181), the Clever Chek TD-3213 Blood Glucose and Blood Pressure Measurement System (K042795) and the the Clever Chek TD-3250 Blood Glucose and Blood Pressure Measurement System (K061073).
The system also contains the same intended use as the ACHTUNG Blood Glucose Monitoring System (K061181), the Clever Chek TD-3213 Blood Glucose and Blood Pressure Measurement System (K042795) and the the Clever Chek TD-3250 Blood Glucose and Blood Pressure Measurement System (K061073).
- 7. Performance Studies
The performance of the Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217/ Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System was studied in the laboratory and in clinical settings by healthcare professionals and lay users. The studies demonstrated that the above blood glucose plus blood pressure monitoring system is suitable for its intended use.
- 8. Conclusion
The Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217/ Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System demonstrates satisfactory performance and is suitable for their intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo of the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, symbolizing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 3 1 2007
Taidoc Technology Corporation 4F, 88, Sec.1, Kwang Fu Road San Chung, Taipei 241 China (Taiwan) ATTN: Ms. Pi-Shiou Li
Re: k062800
Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/Clever Chek TD-3217/ Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA, DXN Dated: December 28, 2006 Received: January 03, 2007
Dear Ms. Li:
This letter corrects our substantially equivalent letter of January 22, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number:
Device Name: Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216 Clever Chek TD-3217/Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System
Indications for Use:
The Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/Clever Chek TD-3217/ Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The alternative site testing in the systems can be used only during steady-state blood glucose conditions.
The system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual, over age 16, at home.
For Clever Chek TD-3213/ Clever Chek TD-3215/ Dr. T TD-3216/ Clever Chek TD-3217 system, the blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 5.25″ ~ 7.75″ .
For Clever Chek TD-3250 system, the blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the arm. The cuff circumference is limited to 9.4" ~ 13.8".
Prescription Use _ X _ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carl Benson
Division Sign-Off
iii
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of 1_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.