K150299 · Philosys, Inc. · NBW · Nov 19, 2015 · Clinical Chemistry
Device Facts
Record ID
K150299
Device Name
Gmate SMART Blood Glucose Monitoring System
Applicant
Philosys, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Nov 19, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Gmate® SMART Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, thigh or calf. The Gmate® SMART Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Gmate® SMART Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. The Gmate® SMART should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternate site testing should be done only during steady state times (when glucose is not changing rapidly). The Gmate® Blood Glucose Test Strips are for use with the GMATE® SMART meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, thigh or calf. The Gmate™ SMART App is a component of the Gmate® SMART Blood Glucose Monitoring System and is intended to be used by people with diabetes at home as an aid to monitor and track the effectiveness of their diabetes management. The Gmate™ SMART App allows the user to view their glucose test results and store a lifetime of results. The user may e-mail their glucose test results to their healthcare provider to help them review, analyze, and evaluate their glucose test results to support an effective diabetes management program. The user can also graph and trend their glucose test results to provide an outlook of their diabetes management.
Device Story
System consists of glucose meter, test strips, and Gmate SMART App; meter plugs into headphone jack of smartphone (Apple iOS or Android). User inserts test strip, applies capillary blood; meter uses amperometric method with glucose oxidase enzyme to generate current proportional to glucose concentration. App performs 5-second countdown, converts signal to glucose value, displays result, and stores data. Used by patients at home for diabetes management; results can be emailed to healthcare providers for review. Benefits include tracking glucose trends and lifetime storage of results to support diabetes management programs.
Clinical Evidence
Clinical evaluation compared patient blood glucose results against YSI 2300 Glucose Analyzer reference. Study included 100 participants with diabetes and healthcare professionals. Testing assessed system accuracy, user performance, and alternative-site measurement. Results demonstrated the system met all reliability and performance requirements; no adverse events reported.
Technological Characteristics
Electrochemical glucose oxidase biosensor. Meter is a passive device powered by smartphone headphone jack. Connectivity via wired audio interface to iOS/Android devices. No internal batteries or calibration required. Performance validated against CLSI standards.
Indications for Use
Indicated for people with diabetes for self-testing of blood glucose levels using fresh capillary whole blood from fingertips, forearm, upper arm, palm, thigh, or calf. Not for diagnosis/screening of diabetes or neonatal use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Gmate® SMART Blood Glucose Monitoring System (k131230)
Related Devices
K131230 — GMATE SMART BLOOD GLUCOSE MONITORING SYSTEM, GMATE SMART APPLICATION · Philosys, Inc. · Aug 4, 2014
K150817 — Dario Blood Glucose Monitoring System · Labstyle Innovations, Ltd. · Dec 21, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 19, 2015
PHILOSY'S INC. LINDA CHAN MANAGING DIRECTOR 304 PARK AVENUE SOUTH, SUITE 218 NEW YORK NY 10010
Re: K150299
Trade/Device Name: The Gmate® SMART Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA, JOP Dated: November 11, 2015 Received: November 12, 2015
Dear Linda Chan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Katherine Serrano -S
For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) k150299
Device Name
The Gmate® SMART Blood Glucose Monitoring System
Indications for Use (Describe)
The Gmate® SMART Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, palm, thigh or calf. The Gmate® SMART Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
The Gmate® SMART Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. The Gmate® SMART should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternate site testing should be done only during steady state times (when glucose is not changing rapidly).
The Gmate® Blood Glucose Test Strips are with the GMATE® SMART meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, thigh or calf.
The Gmate™ SMART App is a component of the Gmate® SMART Blood Glucose Monitoring System and is intended to be used by people with diabetes at home as an aid to monitor and track the effectiveness of their diabetes management. The Gmate™ SMART App allows the user to view their glucose test results and store a lifetime of results. The user may e-mail their glucose test results to their healthcare provider to help them review, analyze, and evaluate their glucose test results to support an effective diabetes management program. The user can also graph and trend their glucose test results to provide an outlook of their diabetes management.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# philosys, Inc.
New York Office 711 Third Avenue Suite 1502 New York, NY 10017 USA TEL: 855-GO-GMATE FAX: 800-939-8670
#### Seoul Office
635-5 Yeoksam-dong, Kangnam-ku, Seoul, 135-908, Korea TEL: +82-2-553-1421 FAX: +82-2-553-1423
over the
## 510(k) SUMMARY
(As required by 21.CFR.807.92)
| Manufacturer Name: | Philosys Co. Ltd.<br>827-3Sanpyeong-ri, Okgu-eup, Gunsan-ri<br>Jeollabuk-do, 573-907 Republic of Korea | | |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-------|
| | Phone: +82-63-453-1421<br>Fax: +82-63-453-1423<br>Contact: Byeong-Kyu Kim (Email: bkkim@philosys.com) | | |
| Submitted By: | Linda Chan<br>Philosys<br>711 Third Avenue<br>Suite 1502<br>New York, NY 10017 | | |
| Contact Person: | Linda Chan<br>Phone: 645-454-5414<br>Fax: 646-400-5240<br>Email: lchan@usdiagnostics.net | | |
| Date Summary, Prepared: | February 3, 2015 | | |
| Device Name: | Propriety Name:<br>Gmate® SMART Blood Glucose Monitoring System<br>Model Number: PG-101<br>Common Name: Blood Glucose Test System, Blood Glucose Monitoring System<br>Classification Name: Blood Glucose, Test, System, Glucose Oxidase, over counter | | |
| Product Code | Classification | Regulation Section | Panel |
| Product<br>Code | Classification | Regulation Section | Panel |
|-----------------|----------------|--------------------|----------------|
| CGA | II | 21 CFR 862.1345 | Chemistry (75) |
| NBW | II | 21 CFR 862.1345 | Chemistry (75) |
| JQP | I | 21 CFR 862.2100 | Chemistry (75) |
{4}------------------------------------------------
# sus, Inc.
New York Office 711 Third Avenue Suite 1502 New York, NY 10017 USA TEL: 855-GO-GMATE FAX: 800-939-8670
#### Seoul Office
635-5 Yeoksam-dong, Kangnam-ku, Seoul, 135-908, Korea TEL: +82-2-553-1421 FAX: +82-2-553-1423
#### 510(k) SUMMARY (As required by 21.CFR.807.92)
| Predicate Device: | We claim substantial equivalence to: 1) K131230- Gmate® SMART Blood Glucose Monitoring System |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description: | The Gmate® SMART Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood. The Gmate® SMART System is intended for self-testing outside the body (in vitro diagnostic use only) by people with diabetes at home as an aid to monitor the effectiveness of their diabetes management. The Gmate® SMART Blood Glucose Monitoring System consists of a glucose meter, test strips, and one control material (additional levels of control material are available upon request). The small Gmate® SMART meter does not require coding or calibration, no need of batteries, and no settings are required. The Gmate® SMART meter is powered on by plugging it into the headphone jack of the Apple/Android device. Insert the test strip into the meter, apply the blood or control solution to the strip and the meter will begin the 5 seconds count down of your test result. The Gmate™ SMART App converts the signal generated from the meter and test strip and displays the test result on the Apple/Android device. The Gmate® SMART Blood Glucose Monitoring System uses the smartphone technology, currently with Apple's iOS (with use of the Apple iPhone 3GS, iPhone 4, iPhone 4S, iPhone 5, iPod Touch 4th generation, iPad, and iPad2) and now with the following Android devices: Samsung Galaxy S3, S4, and S5; to view glucose test results. A simple download of the Gmate® SMART App, enables use of many functions. The Gmate™ SMART App is a component of the Gmate® SMART Blood Glucose Monitoring System and is intended to be used by people with diabetes at home as an aid to monitor and track the effectiveness of their diabetes management. The Gmate™ SMART App allows the user to view their glucose test results and store a lifetime of results. The user may e-mail their glucose test results to their healthcare provider or healthcare provider to help them review, analyze, and evaluate their glucose test results to support an effective diabetes management program. The user can also graph and trend their glucose test results to provide an outlook of their diabetes management. The test principle is: This device is an in vitro diagnostic only product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the glucose oxidase that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. The meter measures the current and calculates your blood glucose level. Combined with the Gmate™ SMART App, it displays the test result and stores them on your Apple/Android device. |
{5}------------------------------------------------
## philosys, Inc.
New York Office 711 Third Avenue Suite 1502 New York, NY 10017 USA TEL: 855-GO-GMATE FAX: 800-939-8670
#### Seoul Office
635-5 Yeoksam-dong, Kangnam-ku, Seoul, 135-908, Korea TEL: +82-2-553-1421 FAX: +82-2-553-1423
### 510(k) SUMMARY
(As required by 21.CFR.807.92)
| Intended Use: | The Gmate® SMART Blood Glucose Monitoring System is intended for the<br>quantitative measurement of glucose (sugar) in fresh capillary whole blood<br>samples drawn from the fingertips, forearm, upper arm, palm, thigh or calf.<br>The Gmate® SMART Blood Glucose Monitoring System is intended to be<br>used by a single person and should not be shared. |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Gmate® SMART Blood Glucose Monitoring System is intended for self-<br>testing outside the body (in vitro diagnostic use) by people with diabetes at<br>home to monitor the effectiveness of diabetes control. The Gmate® SMART<br>should not be used for the diagnosis of or screening of diabetes or for neonatal<br>use. Alternate site testing should be done only during steady state times (when<br>glucose is not changing rapidly). |
| | The Gmate® Blood Glucose Test Strips are for use with the GMATE®<br>SMART meter to quantitatively measure glucose (sugar) in fresh capillary<br>whole blood samples drawn from the fingertips, forearm, upper arm, palm,<br>thigh or calf. |
| | The Gmate™ SMART App is a component of the Gmate® SMART Blood<br>Glucose Monitoring System and is intended to be used by people with diabetes<br>at home as an aid to monitor and track the effectiveness of their diabetes<br>management. The Gmate™ SMART App allows the user to view their glucose<br>test results and store a lifetime of results. The user may e-mail their glucose<br>test results to their healthcare provider to help them review, analyze, and<br>evaluate their glucose test results to support an effective diabetes management<br>program. The user can also graph and trend their glucose test results to provide<br>an outlook of their diabetes management. |
| Comparison to<br>Predicate Device: | The Gmate® SMART Blood Glucose Monitoring System (including meter,<br>control solutions, and test strips) is currently approved under K131230. Under<br>K131230, the Gmate® SMART was only compatible and approved to be used<br>with the Apple iOS devices. |
| | This new 510(k) is to update the previous submission with additional<br>compatible devices; the Android mobile smartphones. |
| | The studies and reports provided show that the Android devices with the<br>Gmate® SMART BGMS are as equivalent, safe, and effective as approved<br>under K131230. |
| Performance Data: | Non-clinical and clinical tests for the Gmate® SMART Blood Glucose System<br>were performed in accordance with FDA Guidance Documents and CLSI<br>reference standards. |
| | The clinical performance evaluation testing included system accuracy, user<br>performance, and alternative-site blood glucose measurement. |
{6}------------------------------------------------
New York Office 711 Third Avenue Suite 1502 New York, NY 10017 USA TEL: 855-GO-GMATE FAX: 800-939-8670
Seoul Office 635-5 Yeoksam-dong, Kangnam-ku, Seoul, 135-908, Korea TEL: +82-2-553-1421
FAX: +82-2-553-1423
### 510(k) SUMMARY
(As required by 21.CFR.807.92)
The non-clinical performance evaluation testing included precision, linearity, interference, hematocrit, altitude, and temperature/humidity. These evaluations were conducted to establish the performance, the functionality and the reliability characteristics of the Gmate® SMART Blood Glucose System.
The evaluations of the Gmate® SMART System was assessed by comparing blood glucose results obtained by patients with those obtained using the YSI 2300 Glucose Analyzer, a laboratory instrument. Glucose levels were measured on 100 persons with diabetes and healthcare professionals at a clinic center.
No adverse events occurred during the studies. The results demonstrated that the Gmate® SMART met all the reliability requirements and performance claims.
Based on the comparisons completed of the clinical and non-clinical tests performed, the devices passed all of the tests based on pre-determined Pass/Fail Criteria.
The Gmate SMART Blood Glucose Monitoring System is safe and effective Conclusion: and substantially equivalent to K131230 - Gmate® SMART Blood Glucose Monitoring System.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches