Dario Blood Glucose Monitoring System

K150817 · Labstyle Innovations, Ltd. · NBW · Dec 21, 2015 · Clinical Chemistry

Device Facts

Record IDK150817
Device NameDario Blood Glucose Monitoring System
ApplicantLabstyle Innovations, Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateDec 21, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The Dario Blood Glucose Monitoring System consists of the Dario Blood Glucose Test Strips. Dario Glucose Control Solutions and the Dario App as the display component of the Dario Blood Glucose Monitoring System. The Dario Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The Dario Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Dario Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. The Dario Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The Dario Blood Glucose Test Strips are for use with the Dario Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The Dario Glucose Control Solutions are for use with the Dario Blood Glucose Meter and the Dario Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

Device Story

Dario Blood Glucose Monitoring System (BGMS) measures glucose in fresh capillary whole blood. System includes meter, lancing device, test strips, control solutions, and mobile app. Meter connects to smart mobile device via 3.5mm audio jack for power and data transmission. User inserts test strip, applies blood sample; glucose oxidase reaction generates electrical current; meter processes current via internal algorithm; result displayed on mobile app. Used at home by single patient for diabetes management. Provides quantitative glucose readings to inform diabetes control decisions. Benefits include compact, integrated design and mobile-based tracking.

Clinical Evidence

Clinical performance evaluated in a study of 100 diabetic lay users. Participants performed fingerstick tests compared against YSI 2300 reference method. For glucose < 75 mg/dL, 100% of results were within ±15 mg/dL. For glucose ≥ 75 mg/dL, 98.9% of results were within ±20% of reference. Linear regression: y = 0.992x + 2.782, r² = 0.9967. Usability questionnaire showed 94% of users found the device easy to operate.

Technological Characteristics

Amperometric glucose oxidase biosensor. Powered via 3.5mm audio jack of smart mobile device. Dimensions/form factor: integrated meter/lancing device. Connectivity: wired to mobile device. Software: mobile app display. Standards: IEC 61010-1, EN 61010-2-101, EN 61326-1, EN 61326-2-6, IEC 60601-1-2, IEC 60068, ISO 15197:2003.

Indications for Use

Indicated for self-testing of blood glucose levels in people with diabetes to monitor diabetes control. For in vitro diagnostic use with fresh capillary whole blood from the fingertip. Contraindicated for neonatal use, diagnosis or screening of diabetes, and testing of arterial or venous blood. Not for use by critically ill patients, or those in shock, dehydrated, or hyperosmolar states.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: K150817 B. Purpose for Submission: New Device C. Measurand: Glucose in fresh capillary whole blood obtained from the fingertip D. Type of Test: Quantitative, Amperometric Assay (Glucose Oxidase) E. Applicant: Labstyle Innovations Ltd. F. Proprietary and Established Names: Dario Blood Glucose Monitoring System G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1345, Glucose Test System 21 CFR § 862.1660; Quality Control Material (assayed and unassayed) 2. Classification: Class II, Class I (exempt) 3. Product code: NBW, CGA, JJX 4. Panel: {1} Clinical Chemistry (75) ## H. Intended Use: 1. **Intended use(s):** See Indications for Use below. 2. **Indication(s) for use:** The Dario Blood Glucose Monitoring System consists of the Dario Blood Glucose Meter, Dario Glucose Test Strips, Dario Glucose Control Solutions and the Dario App as the display component of the Dario Blood Glucose Monitoring System. The Dario Blood Glucose Monitoring System is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The Dario Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Dari Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home to monitor the effectiveness of diabetes control. The Dari Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The Dario Blood Glucose Test Strips are for use with the Dario Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The Dario™ Glucose Control Solutions are for use with the Dario™ Blood Glucose Meter and the Dario™ Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly. 3. **Special conditions for use statement(s):** - For Over-the-Counter use - Single-patient use only - Not for testing glucose levels of neonates - Not for the diagnosis of, or screening for diabetes - Not for testing glucose levels of arterial or venous blood - Not for testing glucose from sites other than the fingertip - Not for testing patients critically ill, in shock, dehydrated or hyperosmolar 4. **Special instrument requirements:** Dario Blood Glucose Meter Apple platform Smart Mobile Device, with iOS 6.1 or above {2} 3 I. Device Description: The Dario Blood Glucose Monitoring System consists of the Dario Blood Glucose meter (dongle), 25 Dario Test Strips housed in a disposable cartridge (sold separately), lancets (sold separately), Dario Control Solution Levels 1 and 2 (sold separately), Disposable Covers (for the mobile devices), User Manual and Quick User Guide, the Dario Application Software (needs to be downloaded from App Store or Google Play). The Dario BGMS can be used with Apple platform smart mobile devices, with operating system versions iOS 6.2 or above. The Dario Control solutions were previously cleared in k082683 as SD Check Gold and are renamed in this submission. J. Substantial Equivalence Information: 1. Predicate device name(s): Gmate SMART Blood Glucose Monitoring System; Philosys, Inc. 2. Predicate 510(k) number(s): K131230 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device Dario Blood Glucose Monitoring System | Predicate Gmate SMART Blood Glucose Monitoring System (K131230) | | Indication For Use | Intended for the quantitative measurement of glucose in fresh capillary blood samples by people with diabetes at home to monitor the effectiveness of diabetes control | Same | | Assay Detection Method | Amperometry | | | Enzyme | Glucose Oxidase | | | Measuring Range | 20-600 mg/dL | | | Hematocrit Range | 20-60% | Same | | Altitude | 10,000 feet | Same | | Coding | Not required | Same | | Display | Connected to smart-device or mobile device to display measurement results | Same | | Power Source | Powered by audio phone | Same | {3} | Similarities | | | | --- | --- | --- | | Item | Device Dario Blood Glucose Monitoring System | Predicate Gmate SMART Blood Glucose Monitoring System (K131230) | | | jack connected to smartphone or mobile device | | | Operating System | iOS | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Sample volume | 0.3 μL | 0.5 μL | | Test time | 6 sec | 5 sec | | Sample type | Fingertip | Fingertips, forearm, upper arm, palm, thigh or calf | | Temperature Range | 50-113°F (10-45°C) | 50-104°F (10-40°C) | | Dimensions LxWxH (mm) | 40.2 x 16.1 x 9.8 | 43 x 21 x 9 | | Weight (g) | 4.0 | 4.2 | | Software Application | Dario App | Gmate SMART App | # K. Standard/Guidance Document Referenced (if applicable): ISO 10993-1:2003, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. EN 13640, Stability Testing of In Vitro Diagnostic Reagent IEC 61010-1:2001, Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General Requirements. EN 61326-1:2006, Electrical equipment for measurement, control and laboratory use – EMC requirements part 1: General requirements. EN 61326-2-6, Electrical equipment for measurement, control and laboratory use - EMC requirements Part 2-6: Particular requirements – In vitro diagnostic (IVD) medical equipment IEC 61010-2-101:2002, Safety requirements for electrical equipment for measurement, control and laboratory use – Particular requirements for In Vitro Diagnostic (IVD) Medical Equipment. IEC 60068-2-64:1993, Environmental testing Part 2: Test methods – Test Fh: vibration, broad-band random (digital control) and guidance. CLSI EP6-A:2005, Interference Testing in Clinical Chemistry, Approved Guideline {4} CLSI EP7-A2:2003, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. ## L. Test Principle: The Dario Blood Glucose Monitoring System measures glucose electrochemically. The glucose biosensor is capable of recognizing the glucose present in whole blood or control solutions by virtue of the glucose specificity of the enzyme glucose oxidase present on the glucose test strip. A voltage is applied and the glucose concentration is calculated by the meter from the electrical current. The final results are communicated to the smart mobile device through the audio jack. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: The sponsor performed within-run precision studies using venous whole blood spiked to five (5) different glucose concentration ranges (30 to 50, 51 to 110, 111 to 150, 151 to 250, and 251 to 400 mg/dL). Fifty (50) samples were prepared at each glucose level, and each sample was analyzed in duplicate for a total of 100 results per glucose level. Results were collected using three (3) test strip lots and 10 meters. Results are summarized below: Within-run precision for glucose: Test Strip Lot 1 | Glucose Level (mg/dL) | Number of tests | Mean (mg/dL) | SD (mg/dL) | CV (%) | | --- | --- | --- | --- | --- | | 30 to 50 | 100 | 48.3 | 1.1 | 2.4 | | 51 to 110 | 100 | 74.7 | 1.3 | 1.7 | | 111 to 150 | 100 | 172.2 | 1.8 | 1.4 | | 151 to 250 | 100 | 216.0 | 6.7 | 3.1 | | 251 to 400 | 100 | 344.2 | 5.6 | 1.6 | Test Strip Lot 2 | Glucose Level (mg/dL) | Number of tests | Mean (mg/dL) | SD (mg/dL) | CV (%) | | --- | --- | --- | --- | --- | | 30 to 50 | 100 | 47.6 | 1.2 | 2.6 | | 51 to 110 | 100 | 73.7 | 2.1 | 3.0 | | 111 to 150 | 100 | 125.4 | 2.6 | 2.1 | | 151 to 250 | 100 | 213.1 | 3.4 | 1.6 | | 251 to 400 | 100 | 339.2 | 6.7 | 2.0 | {5} Test Strip Lot 3 | Glucose Level (mg/dL) | Number of tests | Mean (mg/dL) | SD (mg/dL) | CV (%) | | --- | --- | --- | --- | --- | | 30 to 50 | 100 | 47.8 | 1.3 | 2.7 | | 51 to 110 | 100 | 73.7 | 1.6 | 2.2 | | 111 to 150 | 100 | 126.3 | 1.8 | 1.4 | | 151 to 250 | 100 | 214.4 | 5.1 | 2.4 | | 251 to 400 | 100 | 346.0 | 7.6 | 2.2 | Between-day precision was evaluated using three levels of glucose control solutions with concentrations 30 to 50, 96 to 144, and 280 to 420 mg/dL. Per day, each sample was measured with three (3) test strip lots and 10 meters. These tests were performed over 10 days, for a total of 300 results per glucose level. Results are summarized below: Between-day precision for glucose: Test Strip Lot 1 | Glucose Level (mg/dL) | Number of tests | Mean (mg/dL) | SD (mg/dL) | % CV | | --- | --- | --- | --- | --- | | 30 to 50 | 100 | 48.9 | 2.5 | 5.2 | | 96 to 144 | 100 | 103.0 | 3.2 | 3.1 | | 280 to 420 | 100 | 294.5 | 7.0 | 2.4 | Test Strip Lot 2 | Glucose Level (mg/dL) | Number of tests | Mean (mg/dL) | SD (mg/dL) | % CV | | --- | --- | --- | --- | --- | | 30 to 50 | 100 | 48.7 | 2.3 | 4.7 | | 96 to 144 | 100 | 104.8 | 3.2 | 3.0 | | 280 to 420 | 100 | 296.7 | 8.6 | 2.9 | Between-day precision for glucose: Test Strip Lot | Glucose Level (mg/dL) | Number of tests | Mean (mg/dL) | SD (mg/dL) | % CV | | --- | --- | --- | --- | --- | | 30 to 50 | 100 | 48.0 | 2.0 | 4.2 | | 96 to 144 | 100 | 103.9 | 2.5 | 2.4 | | 280 to 420 | 100 | 295.5 | 5.4 | 1.8 | {6} b. Linearity/assay reportable range: Linearity was evaluated in 2 studies: Linearity in Study #1 was evaluated using three (3) test strip lots and 11 mixed pools of venous blood with glucose concentrations at 16, 86, 157, 230, 301, 367, 438, 510, 578, 649 and 728 mg/dL, as measured by the YSI reference method. Linearity in Study #2 was evaluated at low concentrations, using three (3) test strip lots and six (6) mixed pools of venous blood with glucose concentrations at 16, 37, 57.5, 78.4, 99.2 and 118.8 mg/dL as measured by the YSI reference method. In both studies, each level was measured in replicates of five (5) with each of three (3) test strip lots and the values from the Dario™ Blood Glucose Monitoring System were compared with those obtained from the YSI-2300. Results from the combined regression analysis are as follows: $$ y = 1.008x - 1.459, r^2 = 0.998 $$ The results of the study support the sponsor’s claimed glucose measurement range of 20-600 mg/dL. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: The Dario Blood Glucose Monitoring System is traceable to NIST SRM 917b. The method comparison study was performed using the candidate device and YSI as the reference method (see Section 2.a. above) Test Strip Stability: Test strip stability was assessed in accelerated and real-time studies. The testing protocols and acceptance criteria were reviewed and found to be acceptable. The manufacturer claims shelf life stability of 24 months and an open-vial stability of 6 months at the recommended storage conditions of 36°F-90°F (2°C-32°C) and 10-90% relative humidity. The instructions state not to refrigerate open test strip cartridges or put test strips in a freezer. Control Solution Value Assignment and Stability: The Dario Glucose Control Solutions Level 1 and Level 2 were previously cleared (k082683). The control solutions offered are Level 1 (range 90-140 mg/dL) and Level 2 (range 170-240 mg/dL). The value assignment protocol and the stability protocols and acceptance criteria were reviewed under k082683. The control solutions are stable for 24 months and for 3 months after opening when stored at the recommended storage conditions of 46°F-86°F (8°C-30°C) and 10-90% humidity. d. Detection limit: See linearity study in Section M1b above. {7} e. Analytical specificity: Interference studies were performed by spiking venous blood with two levels of glucose concentrations (65 and 250 mg/dL). Each of these samples was divided into a test pool and a control pool and each potential endogenous and exogenous interfering substance was added to the test pool. Each substance was tested at three concentrations. Each sample was analyzed 10 times with the Dario meter and the % difference between the individual measurements and the reference (YSI) calculated. The sponsor defines no significant interference as ≤10% difference relative to YSI. Results are presented in the table below: | Potential Interfering Substances Tested | Concentration with no Significant Interference (mg/dL) | | --- | --- | | Acetaminophen | 6 mg/dL | | Acetyl-salicylic acid | 120 mg/dL | | Ascorbic Acid (Vit. C) | 4 mg/dL | | Bilirubin | 35 mg/dL | | Cholesterol | 506 mg/dL | | Creatinine | 30 mg/dL | | Dopamine | 2.5 mg/dL | | Ethanol | 400 mg/dL | | Fructose | 15 mg/dL | | Galactose | 60 mg/dL | | Gentisic Acid | 1.8 mg/dL | | Glutathione | 4.6 mg/d | | Hemoglobin | 200 mg/dL | | Heparin | 0.003 mg/dL (3000 U/L) | | Ibuprofen | 50 mg/dL | | Lactose | 25 mg/dL | | Levodopa | 4 mg/dL | | Maltose | 132mg/dL | | Maltotetraose | 120 mg/dL | | Maltotriose | 240 mg/dL | | Mannitol | 800 mg/dL | | Mannose | 5 mg/dL | | Methyl-Dopa | 2 mg/dL | | Sodium Fluoride | 200 mg/dL | | Sodium Salicylate | 63 mg/dL | | Sorbitol | 10 mg/dL | | Tetracycline | 5 mg/dL | | Tolazamide | 5 mg/dL | | Tolbutamide | 100 mg/dL | {8} | Triglyceride | 900 mg/dL | | --- | --- | | Urea | 500 mg/dL | | Uric Acid | 11 mg/dL | | Warfarin | 1 mg/dL | | Xylitol | 25 mg/dL | | Xylose | 50 mg/dL | The following limitations have been added to the labeling: If you are taking acetaminophen containing drugs (e.g. Tylenol) or Vitamin C (ascorbic acid) you may get inaccurate results with this system. If you have a disease or condition in which uric acid levels in your blood may be elevated (&gt;11 mg/dL), such as gout, you may get inaccurate results with this system. If you have very high levels of triglyceride (&gt;810mg/dL), you may get inaccurate results with this system. This system should not be used when undergoing xylose absorption tests. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: See lay user study in section M.3.c b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): To assess the performance of the Dario Blood Glucose Monitoring System in the {9} hands of the intended users the sponsor performed a study with 100 diabetic lay user participants. Participants obtained and tested their own fingerstick samples with the Dario BGMS. Blood glucose results from the Dario meter obtained by the lay user were compared to the YSI 2300 reference value. The samples ranged from 24 to 479 mg/dL as measured by YSI. The results are summarized in the tables below: ## Lay-user vs. YSI: For glucose concentrations &lt; 75 mg/dL | within ± 5 mg/dL | within ± 10 mg/dL | within ± 15 mg/dL | | --- | --- | --- | | 4/11 (36.4%) | 9/11 (81.8%) | 11/11 (100%) | For glucose concentrations ≥ 75 mg/dL | Within ± 5 % | within ± 10 % | within ± 15 % | within ± 20 % | | --- | --- | --- | --- | | 39/89 (43.8%) | 68/89 (76.4%) | 85/89 (95.5%) | 88/89 (98.9%) | Results of linear regression analysis: $$ y = 0.992x + 2.782, r^2 = 0.9967. $$ Subject usability was assessed using a questionnaire. 94% of subjects reported the operation of the meter to be average, easy, or very easy. 90% of subjects reported being satisfied or very satisfied with the meter. Flesch-Kincaid readability assessment was conducted and the results demonstrated that the User Manual, quick start guide, test strip package insert and control solution package insert were written at an 8th grade level or less. ## 4. Clinical cut-off Not applicable ## 5. Expected values/Reference range: Expected glucose values without diabetes: | Status | Range | | --- | --- | | Before eating (FPG) | <100 mg/dL | | Two hours after meals | <140 mg/dL | American Diabetes Association: Standard of Medical Care in Diabetes 2015, Diabetes Care, vol.38, supplement 1, S14-S87, 2015. {10} 11 N. Instrument Name: Dario Blood Glucose Meter O. System Descriptions: 1. Modes of Operation: Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? Yes ☐ X ☐ or No ☐ Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? Yes ☐ X ☐ or No ☐ 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☐ X ☐ or No ☐ 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. 4. Specimen Sampling and Handling: The glucose test is intended to be used with capillary whole blood from the finger, palm, and forearm only. The whole blood sample is applied directly to the test strip by capillary action. 5. Calibration: There is no calibration required by the user for the DA12-B1 and DA12-B2 blood glucose meters. The meters are automatically coded. 6. Quality Control: Two levels of glucose control solutions are available with this system, but are sold separately. Recommendations on when to test the control materials are provided in the labeling. The user is cautioned not to use the meter if the control result falls outside these ranges. {11} P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: 1) Hematocrit Study: The effect of different hematocrit levels on the performance of the Dario Blood Glucose Monitoring System was evaluated using venous whole blood samples with hematocrit levels of 15, 20, 25, 30, 35, 40, 45, 50, 55, 60 and 65% spiked with glucose to achieve three (3) concentrations at 50-70, 100-200, and 250-300 mg/dL. Each sample was then tested using 10 Dario Meters and three (3) lots of test strips. The values were compared with those obtained from YSI 2300 analyzer. The % biases relative to YSI were acceptable within the claimed hematocrit range of 20 to 60%. 2) Altitude study: Capillary fingerstick samples with glucose concentrations ranging from 48 to 423.5 mg/dL were collected from 52 participants at 10,152 feet above sea level. Results were obtained from each participant sample in on the Dario for each of 3 test strip lots compared to additional fingerstick samples measured on the YSI. The results demonstrate acceptable bias to YSI to support the claims in the labeling that altitudes up to 10,000 feet (3048 meters) have no significant effect on blood glucose measurements from the Dario meter. 3) Sample volume study: The sponsor performed a study to support the claimed minimum sample volume for the Dario system. Blood samples with three levels of glucose (50-65, 100-120, and 200-250 mg/dL) were tested at six sample volumes (0.1, 0.2, 0.3, 0.4, 0.5, 0.75, 0.8, and 1.0 µL) and values obtained were compared to YSI values. Results support the claimed minimum sample volume of 0.3 µL. 4) Temperature and humidity studies: The sponsor performed temperature and humidity studies using venous blood samples at five glucose concentrations (approximately 60, 110, 200, 300, and 475 mg/dL) to evaluate temperatures ranging from 12±2°C to 43±2°C and relative humidity from 10% to 90%. Meter results were compared to YSI values. Six temperature and humidity combinations were tested including low temperature/low humidity, low temperature/high humidity, average temperature/low humidity, average temperature/high humidity, high temperature/low humidity, and high temperature/high humidity. The results support the claims in the labeling that the system can be used in conditions of 50 to 113°F (10 to 45°C) with relative humidity of 10 to 90%. 5) Infection Control Studies: The device is intended for single-patient use only. Super Sani-Cloth wipes with EPA registration #9480-4 were validated demonstrating complete inactivation of live hepatitis B virus for use with the test strips cartridge cap, Dario meter and Dario BGMS housing device. The sponsor also demonstrated that there was no change in performance or in the external materials of the test strips cartridge cap, Dario meter and Dario BGMS housing device after 156 cleaning and disinfection cycles (one cycle includes one cleaning wipe plus one disinfecting wipe) to simulate 3 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures. 12 {12} 6) EMC testing was certified and appropriate compliance certificates were provided. 7) Customer service is available Monday through Friday 9:00 am to 5:00 pm PST by calling 1-800-895-5921. Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: 1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 13
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...