EOCENE SYSTEM

K090712 · Infopia Co, Ltd. · NBW · Oct 2, 2009 · Clinical Chemistry

Device Facts

Record IDK090712
Device NameEOCENE SYSTEM
ApplicantInfopia Co, Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateOct 2, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The GLUCOLAB™ Diabetes Monitoring System and EOCENE™ System is used for the quantitative measurement of glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings., including physician's office laboratories and point of care sites . The GLUCQLAB™ System and EOCENE'M System is not to be used for diagnosis, screening of diabetes or for neonatal use. The GLUCOLAB™ System and EOCENE™ System are for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the forearm, upper arm, palm, calf and thigh. GlucoLab™ control is used with GlucoLab™ Brand System to check that the meter and test strips are working together as a system and that you are performing the test correctly

Device Story

Eocene™ Glucose Telecommunication system acts as a data management accessory for GlucoLab™ or Eclipse™ Diabetes Monitoring Systems. The system consists of a cradle, DC 5V adapter, and phone line connection. It functions by transmitting blood glucose data from the monitoring meter to a database server via a phone line. Data is subsequently accessible to patients and healthcare professionals via the Eocene Secure website. The system is intended for use in home and clinical settings to facilitate remote monitoring of diabetes management effectiveness. By enabling regular data uploads, the system allows providers to review patient glucose trends, potentially improving clinical decision-making and diabetes management outcomes.

Clinical Evidence

Bench testing only; design control activities and verification/validation performed to assess impact of telecommunication modification; results met predetermined acceptance criteria.

Technological Characteristics

System includes Eocene cradle, DC 5V adapter, and phone line interface. Connectivity is via phone line to a remote database server. Software-based data transmission. No specific materials or energy standards cited.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood from the fingertip for home use by individuals with diabetes to monitor glycemic control. Not for neonatal use or diabetes diagnosis/screening.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary # EXHIBIT #1 109071 2 This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. The assigned 510(K) Number is not applicable. Date: March , 2009 OCT - 2 2009 #### 1. Submitter: | Submitted by: | Infopia Co.,ltd.<br>#1603, Dongil Technotown A Bldg.,<br>889-3, Kwanyang2-Dong, Dongan - Gu<br>Anyang, Kyunggi 431-716, Korea<br>Phone +82-31-423-6170<br>Fax<br>+82-31-423-6171 | |---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Bryan Oh<br>Phone: 1-321-267-9911<br>Fax: 1-321-267-5582 | #### 2. Device: | Propriety Name | Eocene™ Glucose Telecommunication system | |----------------------|--------------------------------------------------------| | | Blood glucose monitoring system | | Common Name | Data Management system; Accessory to Medical<br>Device | | Classification Name: | System, test, blood glucose, over the counter | | | Glucose Oxidase | | | Single (specified) analyte controls | | | Physiological signal transmitters and receivers | | | Class II | | Classification: | 21 CFR Part 862.1345, | | | 21 CFR Part 862.1660 | | | 21 CFR Part 870.2910 | | Product Code: | NBW, JQP | #### 3. Predicate Device: GlucoLab™ Blood Glucose Monitoring System(Infopia co., Ltd.) K051285 ・ {1}------------------------------------------------ #### Description: 4. Data transmitted by the Eocene™ system is available online for you or your designed healthcare professionals to review within the Eocene Secure website. To register your Eocene™ cradle for use with the Eocene systems, contact the dealer you purchased your device from or visit www.eocenesystems.com for more information. For best results. follow the daily testing frequency prescribed by your and upload your data at least once a week or as indicated by your provider. The Eocene™ system may be used with GlucoLab™ or Eclipse™ Diabetes Monitoring Systems. #### 5. Indications for use: The GLUCOLAB™ Diabetes Monitoring System and EOCENE " System is used for the quantitative measurement of glucose level in whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings., including physician's office laboratories and point of care sites . The GLUCOLAB " System and EOCENE "" System provide plasma equivalent results. The GLUCOLABTM System and EOCENE 100 System is not intended to be used with neonatal blood samples. The GLUCOLAB™ System and EOCENE™ System is for testing outside the body (in vitro diagnostic use onlv). Testing sites include the traditional fingertip testing along with alternate site testing on the forearm, upper arm, palm, calf and thigh. GlucoLab™ control is used with GlucoLab™ Brand System to check that the meter and test strips are working together as a system and that you are performing the test correctly. It is very important that you do control solution tests routinely to make sure you are getting accurate results. Control Solutions are sold separately. #### Comparison of Technological Characteristics with Predicate: 6. The GLUCOLAB™ Diabetes Monitoring System still operates under the same technological characteristics as the predicate device cleared under K051285. This Eocene™ System is used with the GLUCOLAB™ Diabetes Monitoring System. It is composed of the Eocene cradle, DC 5V Adaptor, and phone line. Confidential Infopia Co., Ltd. {2}------------------------------------------------ It allows to use the phone line to upload GlucoLab™ data to a database server. This does not affect to the safety and effectiveness of the GlucoLab data. ### Performance Data: 7. Non-clinical: Verification, validation and testing activities were conducted to establish the performance, functionality and reliability characteristics of the Eocene 100 Glucose Telecommunication. Testing involved the verification of software requirement specifications, product requirement specifications and user interface requirement specifications from the risk analysis. The device passed all of the tests based on pre-determined Pass/Fail criteria. #### 8. Conclusion The data from non clinical tests show that the use of the Eocene™ Glucose Telecommunication system is as safe and effective as the legally marketed predicate device (GlucoLab' ""). Therefore we conclude that the Eocene™ Giucose Telecommunication system is substantially equivalent to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread, which is a common symbol of the United States. The eagle is facing to the left and has its head turned towards the viewer. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Infopia Co., Ltd c/o Maria F Griffin, Official Correspondent mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021 OCT - 2 2009 Re: k090712 Trade/Device Name: Eocene System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system. Regulatory Class: II Product Code: NBW, JQP Dated: Sep 2, 2009 Received: Sep 3, 2009 Dear: Ms. Griffin, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. CA Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## EXHIBIT #3 # Indications for Use Page 1 of 1 510(k) Number (if known): K090712 Device Name : GLUCOLAB™ Diabetes Monitoring System and EOCENE™ System # Indications For Use: The GLUCOLAB™ Diabetes Monitoring System and EOCENE™ System is used for the quantitative measurement of glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings., including physician's office laboratories and point of care sites . The GLUCQLAB™ System and EOCENE'M System is not to be used for diagnosis, screening of diabetes or for neonatal use. The GLUCOLAB™ System and EOCENE™ System are for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the forearm, upper arm, palm, calf and thigh. GlucoLab™ control is used with GlucoLab™ Brand System to check that the meter and test strips are working together as a system and that you are performing the test correctly Prescription Use (Per 21 CFR 801 Subpart D) Over-The Counter Use × AND/OR (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off Division sign Office of In Vitro Diagnostic Device Evaluation and Safely 510(k) K090712
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...