G5 INFINITY BLODO GLUCOSE MONITORING SYSTEM WITH MODEL IGM, 0024A
K082201 · U. S. Diagnostics, Inc. · NBW · Feb 20, 2009 · Clinical Chemistry
Device Facts
Record ID
K082201
Device Name
G5 INFINITY BLODO GLUCOSE MONITORING SYSTEM WITH MODEL IGM, 0024A
Applicant
U. S. Diagnostics, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Feb 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Infinity ™ Meter device is used along with the Infinity ™ Test Strips is used for the quantitative measurement of glucose level in whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings. Infinity TM System is for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the dorsal palm, ventral palm, forearm, upper arm, calf and thigh.
Device Story
System measures capillary whole blood glucose via amperometric assay using glucose oxidase enzyme. User applies blood sample to single-use test strip inserted into G5 Infinity Meter; meter converts enzymatic reaction current into glucose concentration. Used in home by patients or clinical settings by professionals. Supports fingertip and alternate site testing (palm, forearm, upper arm, calf, thigh). Requires manual entry of batch-specific code number for calibration. Output displayed on meter screen; aids in diabetes management monitoring. Does not support wireless data transmission.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
In vitro diagnostic blood glucose monitor. Electrochemical sensing principle using dehydrogenase glucose reaction. Handheld meter form factor. System includes meter and test strips. No specific materials, software architecture, or connectivity standards disclosed.
Indications for Use
Indicated for quantitative blood glucose measurement in patients with diabetes to monitor management effectiveness. For home and clinical use. Testing sites: fingertip, dorsal/ventral palm, forearm, upper arm, calf, thigh. Contraindicated for neonates and not for diabetes screening or diagnosis.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K063212 — CLEVER CHEK TD-4231 AND TD-4223 BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Nov 22, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
KORZIO|
FEB 2.0 2009
## 510(k) SUMMARY
(As required by 21.CFR.807.92)
| Introduction: | According to the requirements of 21 CFR.807.92, the following information provides sufficient data to understand the basis for a determination of substantial equivalence. |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By: | US Diagnostics, Inc. 304 Park Avenue South Suite 218 New York, NY 10010 |
| Contact Person: | Edward Letko Phone: 917-402-5900 Fax: 212-202-5173 |
| Date Summary,<br>Prepared: | February 5, 2009 |
| Device Name: | Propriety Name: G5 Infinity™ Common Name: Blood Glucose Test System Classification Name: Class II, 862.1345 Glucose Blood Tester |
| Predicate Device: | We claim substantial equivalence to the LifeScan, Inc., OneTouch® Ultra®. |
| Device<br>Description: | The Infinity ™ Monitor is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood, which is used with the Infinity ™ Test Strips.<br>The test principle is:<br>This device is an in vitro diagnostic product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the dehydrogenase glucose that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. This reaction is measured by the Meter and displayed as your blood glucose result. |
Intended Use: The Infinity ™ Meter device is used along with the Infinity ™ Test Strips is used for the quantitative measurement of glucose level in whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings. Infinity TM System is for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the dorsal palm, ventral palm, forearm, upper arm, calf and thigh.
{1}------------------------------------------------
# 510(k) Summary, Continued
giving it an attractive, nearly painless alternative to the more painful fingertip site.
# Comparison to Predicate Device:
The US Diagnostics, Inc. Infinity ™ Module is substantially equivalent to the other products in commercial distribution intended for similar use. The most notable, it is substantially equivalent to the currently marketed item, the OneTouch® Ultra® by LifeScan, Inc.
The Infinity ™ Blood Glucose Monitoring System is substantially Conclusion: equivalent to the following predicate devices: K024194 - LifeScan, Inc. OneTouch® Ultra® K984261 -- LifeScan, Inc. SURESTEP® K021513 - Roche Diagnostics Corp. Accu-Chek Advantage
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread, a common symbol associated with the United States government. The seal is rendered in black and white.
Re:
#### Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
U.S. Diagnostics, Inc. c/o Jonathan Johnson 304 Park Avenue South Suite 218 New York, NY 10010-4301
FEB 2 0 2009
k082201 Trade Name: G5 Infinity Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Blood Glucose Test System, Glucose Oxidase Regulatory Class: Class II Product Codes: NBW, CGA, JJX Dated: February 9, 2009 Received: February 10, 2009
Dear Mr. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. .
If your device is classified (see above) into either class II (Special: Controls) or class III (BMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page - 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Corg C. He
Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{4}------------------------------------------------
# Indication for Use
### 510(k) Number (if known): K082201
Device Name: G5 InfinityTM Blood Glucose Monitoring System
Indication For Use:
The GS Infinity TM Meter device is used along with the G5 Infinity TM Test Strips for the quantitative measurement of glucose level in whole blood as an aid in monitoring the effectiveness of diabetes manage:nent in the home by patients with diabetes and in clinical settings by healthcare professionals. G5 Infinity TM System is for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the dorsal palm, ventral palm, forearm, upper arm, calf and thigh. The G5 Infinity™ Blood Glucose Monitoring System is not intended for screening or diagnosis of diabetes and also not for use with neonates.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use X (21 CFR Part 801 Subpart C)
(PLEASE.DONNOT WRITE BELOW THIS LINE; CONTINUE QN ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Ruta Chesler
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k082201
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.