ADVOCATE REDI-CODE+BMB-EA001S BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Broadmaster Biotech Corperation
Product Code
NBW · Clinical Chemistry
Decision Date
May 11, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Advocate Redi-Code BMB-EA001S Blood Glucose Monitoring System is intended for single patient home use only.
Device Story
System comprises blood glucose meter, test strips, and control solutions; utilizes amperometric method to measure glucose concentration in fresh capillary whole blood. User applies blood sample to test strip; meter measures current proportional to glucose level; displays quantitative result. Designed for home use by single patient; includes speaking function for audio output. Provides data for monitoring diabetes control programs. Differences from predicate include housing color/material and addition of speaking instructions; core measurement technology and algorithm remain identical.
Clinical Evidence
Bench testing performed including precision, linearity, accuracy, and interference studies. Clinical evaluation included lay user assessment and alternate site testing performance. Speaking function evaluated for user convenience. Testing confirmed performance meets requirements and matches predicate.
Technological Characteristics
Blood glucose monitoring meter; includes speaking function; validated for 3000 cleaning/disinfection cycles; software modified to support audio output; intended for single-patient home use.
Indications for Use
Indicated for single-patient home use for blood glucose monitoring.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Shepherd Blood Glucose Monitoring System (k102316)
Related Devices
K102049 — CLEVER CHOICE VOICE+ BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · May 27, 2011
K093635 — MODELS FORA V30 AND TD-4242 BLOOD GLUCOSE MONITORING SYSTEMS · Taidoc Technology Corporation · Feb 18, 2010
K113670 — ELEMENT BLOOD GLUCOSE MONITORING SYSTEM · Infopia Co, Ltd. · May 8, 2012
K112275 — FORA PREMIUM V10 BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Dec 15, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K12018 3
MAY 1 1 2012
# 510(k) Summary (Per 21 CFR 807.92)
- l. Submitter Information Company Name Address
| Contact Person | |
|----------------|--|
| Phone | |
| Fax | |
| Email | |
Date Prepared
- 2. Device Name Proprietary Name
Common Name Classification Number
Classification Panel Product Code Regulation Number
- 3. Predicate Device Proprietary Name
Common Name Manufacturer 510(k) Number
BroadMaster Biotech Corporation 7F, NO. 168-2 , Liancheng Rd , Zhonghe Dist , New Taipei City 235 , Taiwan ( R.O.C ) Roger Lai/Blanc Lin 886-987-111105/886-2-66375859 886-2-22428332 roger@broadmaster-biotech.com blanc@broadmaster-biotech.com 2012/1/20
ADVOCATE® Redi-Code+ BMB-EA001S Blood Glucose Monitoring System Blood Glucose Test System System, Test, Blood Glucose, Over the Counter 75, Clinical Chemistry NBW, CGA 21 CFR 862.1345 Glucose Test System
Glucose Shepherd Blood Glucose Monitoring System Blood Glucose Test System BroadMaster Biotech Corporation k102316
## 4. Device Description
The ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System consists f the ADVOCATE® Redi-Code BMB-EA004S Blood Glucose meter and the ADVOCATE® Redi-Code* BMB-BA006A Blood Glucose test strips, ADVOCATE® Redi-Code control solutions, lancing device, and commercially available sterilized lancets. This system utilizes amperometric method to generate a current. The size of the current is proportional to the amount of glucose
{1}------------------------------------------------
presented in the sample, providing a quantitative measure of glucose level in whole blood.
## Intended Use
ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) at home. It is used for quantitative measurement of glucose level in fresh capillary whole blood samples (from the finger, the palm, the forearm, the upper arm, the calf, and the thigh). The alternative site testing can be only used during steady-state blood glucose monitoring. The ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System is intended for use by a single person and should not be shared. In additional, it is intended for use at home as an aid in monitoring the effectiveness of diabetes control program. It should not be used for the diagnosis of diabetes, or for the testing of neonates.
The ADVOCATE® Redi-Code BMB-EA001S Blood Glucose Monitoring Systems consists of the ADVOCATE® Redi-Code+ BMB-EA001S Blood Glucose meter and the ADVOCATE® Redi-Code* BMB-BA006A Blood Glucose test strips. The ADVOCATE® Redi-Code+ BMB-EA001S meter are used only with ADVOCATE® Redi-Code BMB-BA006A Blood Glucose test strips to quantitatively measure glucose in fresh capillary whole blood samples drawn from finger tips, the palm, the forearm, the upper arm, the calf, and the thigh. The ADVOCATE® Redi-Code * BMB-EA001S Blood Glucose Meter also includes speaking functions but has not been validated for use by the visually impaired.
The ADVOCATE® Redi-Code* control solutions are for use with the ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System as a quality control check to verify the accuracy of blood glucose test results.
2/6
{2}------------------------------------------------
| 5. Comparison to Predicate Device | | |
|---------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Item | Candidate device | Predicate device |
| | ADVOCATE® Redi-Code+<br>BMB-EA001S Blood Glucose<br>Monitoring System | k102316 |
| Similarities and Differences | | |
| Appearance | Image: ADVOCATE Redi-Code | Image: Bread Master |
| Similarities | | |
| Intended use | Same for both systems. | |
| Enzyme | Glucose Oxidase, same formula and strip design for both systems | |
| Test strips | Same, BMB-BA006A test strips are for the use with both systems | |
| Test sample (finger,<br>arm...) | Same for both systems, fresh capillary whole blood<br>(finger ,palm, forearm, upper arm, calf and thigh) | |
| Measuring range | Same for both systems, 20-600 mg/dL | |
| Hematocrit | Same for both systems, 20-60% | |
| Required sample volume | Same for both systems, 1.1μL | |
| Reaction time | Same for both systems, 5 seconds | |
| Coding function | Same for both systems, no coding | |
| Operation condition | 50°F~104°F<br>Below 85% R.H. | |
| MCU | TI MSP430 | |
| Algorithm (Blood<br>glucose concentration<br>calculation) | Same | |
| The core circuit for the | Same | |
| measurement | | |
| Power source | Two 1.5V AAA alkaline batteries | |
| Weight (without<br>batteries) | 53g | |
| Memory | 400 measurements | |
| Dimension(mm) | 64*95*29 | |
| LCD display | Image: LCD display | |
| General/ Pre-meal/<br>Post-meal selection | Same for both systems | |
| Precision | Same precision for both systems. Please refer to<br>Attachment 12.1 Precision Study | |
| Linearity | Same linearity for both systems. Please refer to<br>Attachment 12.2 Linearity Study | |
| Accuracy | Same accuracy for both systems. Please refer to<br>Attachment 13.1 System Accuracy Study | |
| Lay user evaluation | Attachment 13.3 Customer and<br>Alternate Site Testing<br>Performance Evaluation.<br>Speaking Function Evaluation<br>Report, indicating the speaking<br>function really increase the users'<br>convenience, is also attached. | Easy to be used. |
| Interference | Same interference as the table listed in k102316. When tested<br>following NCCLS guidelines, bilirubin, creatinine, methyldopa,<br>galactose, maltoase, xylose, salicylate, cholesterol,hemoglobinand<br>triglycerides at therapeutic concentrations do not significantly affect<br>glucose results. However, these levels of the following interferences<br>in blood may cause inaccurate test results:<br>Acetaminophen $\geq$ 12.5 mg/dL (Therapeutic level is 1.2-3.6 mg/dL)<br>Ascorbic acid $\geq$ 7.5 mg/dL(Therapeutic level is 0.4-2.1 mg/dL)<br>Dopamine $\geq$ 3 mg/dL (Therapeutic level is 0.04 mg/dL)<br>L-dopa $\geq$ 4 mg/dL (Therapeutic level is 0.02-0.3 mg/dL) | |
| Tolbutamide≥ 150 mg/dL (Therapeutic level is 3.6-7.2 mg/dL) | | |
| Uric acid ≥20 mg/dL (Normal level is 2.6-7.2 mg/dL) | | |
| Gentisic acid ≥ 25 mg/dL (Therapeutic level is 0.2-0.6 mg/dL) | | |
| Tolazamide ≥ 15 mg/dL (Therapeutic level is 2.0-2.5 mg/dL) | | |
| Mannose≥250 mg/dL(Therapeutic level is 1.15 mg/dL) | | |
| Ibuprofen≥ 50 mg/dL(Therapeutic level is 1.0-7.0 mg/dL) | | |
| Disinfectant Protocol | Same disinfectant protocol has been verified to work for both<br>systems | |
| Altitude | Same altitude, up to 10745 feet. | |
| EMC testing | IEC 61326-1:2005 | IEC 61326-1:2005 |
| | EN 61326-1:2006 | EN 61326-1:2006 |
| | IEC 61326-2-6:2005 | IEC 61326-2-6:2005 |
| | EN 61325-2-6:2006 | EN 61325-2-6:2006 |
| | (TUV Rheinland ,Registration<br>No.: AK 50221354 0001) | (TUV Rheinland ,Registration<br>No.: AK 50186220 0001) |
| | Glucose meter software | Blood glucose detection and data analysis algorithm is identical.<br>(See “Attachment 13.2 Method Comparison Report”) |
| Control solutions cleared<br>in 510(k) # | Same control solution level 1/2/3 | |
| | Difference | |
| User manual (mark<br>differences in red when<br>compared with the<br>predicate) | Please refer to the section marked in red in the user manual | |
| Readability assessment<br>for labeling material<br>(User manual; test strip<br>insert, control solutions<br>insert) | The only difference of the labeling material is the section regarding<br>the meter speaking function in the user manual. Per the evaluation<br>report from Arkansas it shows users have met no problem when<br>reading the material. | |
| Color and Material for<br>the housing and button | Red ABS | Blue ABS |
| Additional function--<br>English/Spanish<br>Speaking Instruction | Yes | No |
| Battery Life | Over 500 times | Over 1000 times |
{3}------------------------------------------------
.
.
{4}------------------------------------------------
:
:
રાદ
・
. . . .
{5}------------------------------------------------
## 6. Performance Studies
The performance of the ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System was studied in the laboratory and in clinical settings. The studies have demonstrated that this system meets the performance requirements of its intended use.
## 7. Conclusion
The laboratory testing results, clinical testing results and labeling of ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System match the Indications for Use and support the claim of substantial equivalence to the predicate.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem that resembles an abstract representation of a human figure or a caduceus, a symbol often associated with medicine and healthcare.
10903 New Hampshire Avenue Silver Spring, MD 20993
Broadmaster Biotech Corperation c/o Roger Lai 7F, No. 168-2, Liancheng Rd Zhonghe Dist. New Taipei City 23553 Taiwan (R.O.C) China
MAY 1 1 2012
k120183 Re:
k120183
Trade Name: Advocate® Redi-Code+ BMB-EA001S Blood Glucose Monitoring
System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Codes: NBW, CGA, JJX Dated: April 10, 2012 Received: April 11, 2012
Dear Mr. Lai:
We have reviewed your Section 510(k) premarket notification of intent to market the device is We have reviewed your Section >>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>> referenced above and have detemmed the benefits substance of cadicate devices marketed in
indications for use stated in the enclosure) to legally marketed in
the enclosed in indications for use stated in the enclosure) to tegally manaded portuges of the Medical Device
interstate commerce prior to May 28, 1976, the enactment of the provisio interstate commerce prior to May 23, 1976, the eniace with the provisions of
Amendments, or to devices that have been receassified in accordance with the provisions of Amendments, or to devices that have been fectassined in actrequire approval of a premarket
the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of the Federal Food, Drug, and Cosmetic Act (Act (Act ) the device, subject to the general
approval application (PMA). You may, therefore, market the device, subject to the gen approval application (PMA). controls provisions of the Act. Ine general controls provisems of the entrols practice,
requirements for annual registration, listing of devices, good manufacturing practic requirements for annual registration, nomig and adulteration.
labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III
n affective and control and colditional controls. Existing major regulations af If your device is classified (see above) into eline) in (opermajor regulations affecting
(PMA), it may be subject to such additional controls. Existing major regulations (CF (PMA), it may be subject to such additional Colliols: Existing mayor roger roger of S00 to 895.
your device can be found in Title 21, Code of Federal Regulations (CFR), Part your device can be found in Title 21, Code of Federal Regulation (400),
In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA's issuance of a substantial equirements
mean that FDA has made a determination that your device with other requirements. mean that FDA has made a determination intributions administered by other Federal agencies.
of the Act or any Federal stations and regulations and not limited to: registratio of the Act or any Federal stantify and regulations administered of on the Act of registration
You must comply with all the Act's requirements, including, but not limited to: You must comply with all the Act S requirements, including of the more of the senses of the senses
and listing (21 CFR Part 807); Incience related adverse exemps) (21 CFR 803 and listing (21 CFR Part 807); labeling (21 CFR Pars overs) (21 CFR 803); and good
reporting (reporting of medical devices ast fark in the quality systems (QS) regulation reporting (reporting of medical device-related adverse evenis) (21 UN 000), in any former
manufacturing practice requirements as set forth in the quality systems (OS) regulat CFR Part 820).
{7}------------------------------------------------
#### Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours,
signature
Couriney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### 510(k) Number k120183
Device Name:
ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System
Indications For Use:
ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) at home. It is used for quantitative measurement of glucose level in fresh capillary whole blood samples (from the finger, the palm, the forearm, the upper arm, the calf, and the thigh). The alternative site testing can be only used during steady-state blood glucose monitoring. The ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System is intended for use by a single person and should not be shared. In additional, it is intended for use at home as an aid in monitoring the effectiveness of diabetes control program. It should not be used for the diagnosis of diabetes, or for the testing of neonates.
The ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring Systems consists of the ADVOCATE® Redi-Code+ BMB-EA001S Blood Glucose meter and the ADVOCATE® Redi-Code BMB-BA006A Blood Glucose test strips. The ADVOCATE® Redi-Code* BMB-EA001S meter are used only with ADVOCATE® Redi-Code+ BMB-BA006A Blood Glucose test strips to quantitatively measure glucose in fresh capillary whole blood samples drawn from finger tips, the palm, the forearm, the upper arm, the calf, and the thigh. The meter also included speaking functions but has not been validated for use by the visually impaired.
The ADVOCATE® Redi-Code* control solutions are for use with the ADVOCATE® Redi-Code* BMB-EA001S Blood Glucose Monitoring System as a quality control check to verify the accuracy of blood glucose test results.
Over-The-Counter Use X AND/OR Prescription Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Quto. Clule-
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) 1220183
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.