K052469 · Healthpia America Corp. · NBW · Jun 5, 2006 · Clinical Chemistry
Device Facts
Record ID
K052469
Device Name
GLUCOPACK
Applicant
Healthpia America Corp.
Product Code
NBW · Clinical Chemistry
Decision Date
Jun 5, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The GlucoPack™ Diabetes Monitoring System is used for the quantitative measurement of glucose level in whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings. GlucoPack™ System is for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertips only.
Device Story
GlucoPack™ is a blood glucose monitoring system consisting of a meter and test strips. It measures glucose levels in whole blood samples obtained from the fingertip. The device is intended for use by patients with diabetes at home or by healthcare professionals in clinical settings to monitor diabetes management effectiveness. The system performs in vitro diagnostic testing outside the body. Users apply a blood sample to the test strip, which is inserted into the meter; the meter processes the sample to provide a quantitative glucose reading. This output allows patients and clinicians to assess glycemic control and adjust diabetes management strategies accordingly.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison to legally marketed predicate devices.
Technological Characteristics
In vitro diagnostic blood glucose monitoring system. Uses test strips for whole blood glucose measurement. Class II device, product codes NBW and CGA. Regulation 21 CFR 862.1345.
Indications for Use
Indicated for quantitative blood glucose measurement in patients with diabetes mellitus. For home (OTC) or clinical use. Not for use on neonates. Testing site: fingertip.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
OneTouch Ultra (k024194)
Related Devices
K031388 — MODIFICATION TO HYPOGUARD ADVANCE BLOOD GLUCOSE MONITORING SYSTEM · Hypoguard USA, Inc. · May 22, 2003
K092887 — EASYPLUS MINI MASTERDRIVER SELF MONITORING BLOOD GLUCOSE SYSTEM, MODEL EPS09009 · Eps Bio Technology Corp. · Oct 21, 2009
K082417 — GLUCOCARD 01-MINI BLOOD GLUCOSE MONITORING SYSTEM, MODEL 73110 · Arkray Factory USA, Inc. · Oct 21, 2008
K020232 — HYPOGUARD ADVANCE BLOOD GLUCOSE MONITORING SYSTEM · Hypoguard USA, Inc. · Feb 12, 2002
K053312 — EASYMATE BLOOD GLUCOSE MONITORING SYSTEM · Bioptik Technology, Inc. · Mar 1, 2006
Submission Summary (Full Text)
{0}
1 of 14
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k052469
B. Purpose for Submission:
New device submission
C. Measurand:
Whole Blood Glucose
D. Type of Test:
Quantitative, utilizing glucose oxidase technology
E. Applicant:
HealthPia America Corp.
F. Proprietary and Established Names:
GlucoPack
G. Regulatory Information:
1. Regulation section:
21CFR Sec.- 862.1345-Glucose test system
2. Classification:
Class 2
3. Product code:
NBW - System, test, blood glucose, over the counter
CGA - Glucose oxidase, glucose
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The GlucoPack Diabetes Monitoring System is intended for use in the quantitative measurement of glucose in whole blood taken from the fingertip. Testing is done outside the body (in Vitro diagnostic use). It is indicated for use at home (over the counter (OTC)) by persons with diabetes mellitus, or in clinical
{1}
settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for use on neonates.
3. Special conditions for use statement(s):
Do not use plasma or serum or use sites other than the fingertip to obtain the blood sample. It is not intended for use on neonates. It is not intended to diagnose diabetes. Dehydration may lower test results. Inaccurate results may occur when individuals are in shock, in hypotensive conditions, hyperglycemic or hyperosmolar states, with or without ketosis,
4. Special instrument requirements:
GlucoPack™ Meter
I. Device Description:
GlucoPack™ Meter
Components
This glucose-meter is composed of a strip sensor analog circuit unit, step perception unit, micro controller unit, and communication unit.
- Strip sensor analog circuit unit
Once the blood sample is injected into the strip sensor, the electricity is on; voltage amplification is generated through the electric current-voltage converter. The signal amplified is filtered and sent to AD input unit of MCU and then digitalized.
- Step perception unit
When a person moves his or her body, the movement is captured and the counter in MCU counts it.
- Micro controller unit
This unit digitalizes the signal which is an input from the analog signal process circuit on the A/D input terminal, and counts the signal which is transmitted through step perception unit and transfers that signal to external display device.
- Communication unit
This unit communicates with an external display device and transmits data to the display device.
- External display device
This unit displays data and test results transmitted from the communication unit.
Features of product
- Type and extent of protection against the electric shock: internal power supply system device, type B device
- Safety device: power protection circuit, which is applied for safe use of the Lithium Polymer battery, has following functions;
- Over charge protect function: when the battery is overcharged by an external charger due to defective performance, this function shuts off the circuit and protects the device from overcharge.
- Over discharge protect function: this function protects the battery from
2 of 14
{2}
over discharge under the critical voltage.
- Over current shut off function: this function shuts off the over current flow as a result of the internal/external abnormal condition.
- Name and version of software:
- Name of software for blood glucose meter: IAR Workbench target descriptor for MSP 430 (C-based)
- Version: 1.26A/WIN
- Name of the software for external display:
- Brew (C-based)
- Version: 1.1
EasyGluco™ Test Strip Cleared under k043512
Two control solutions are sold separately from the kit. Controls previously cleared under k031501
Auto-Lancet Device, Greelan Lancets and the Check Strip for Calibration
# J. Substantial Equivalence Information:
1. Predicate device name(s):
LifeScan, Inc. OneTouch Ultra
2. Predicate 510(k) number(s):
k024194
3. Comparison with predicate:
The GlucoPack™ Blood Glucose Monitoring System provides the same glucose monitoring capability as the predicate device, the ONE TOUCH® Ultra®.
| | GlucoPack™ | ONE TOUCH® Ultra® |
| --- | --- | --- |
| Detection Method | Amperometry: current is generated by oxidation of reduced mediator. | Amperometry |
| Enzyme | Glucose Oxidase (Aspergillus niger) | Glucose Oxidase (Aspergillus niger) |
| Mediator | Potassium ferricyanide | Potassium ferricyanide |
| Electrode | Carbon electrode | Carbon electrode |
| Test range | 10 ~ 600 mg/dL | 20 ~ 600 mg/dL |
| Hematocrit Range | 30 ~55% | 30 ~ 55% |
| Test Time | 9 seconds | 5 seconds |
| Sample Volume | 3uL | 1uL |
| Temperature & Humidity range | 50 ~ 104°F
10 ~ 40°C
10 ~ 90% | 43 ~ 111°F
6 ~ 44°C
10 ~ 90% |
| Open use time | 3 months | 3 months |
| Coding | Button (C1 ~C40) | Button (C1 ~ C49) |
{3}
| | GlucoPack™ | ONE TOUCH® Ultra® |
| --- | --- | --- |
| Memory capability | Max 150 packet data on the external display device
Unlimited memory on the datacenter | 14-day average and last 150 tests in the memory |
| Power | Lithium-polymer DC 3.7V | 3V Li battery (CR2032) |
| Battery life | Running 5,000 tests | Running 1,000 tests |
| Size: LxWxH (mm) | 80.9x45x29.7 | 79x57x21 |
| Weight | 93g±1 | 42g (with battery) |
| Warranty | 1 year | 3 years |
| Software | GlucoPack :
IAR Workbench target descriptor for MSP430
The external display device : Brew Application | IN TOUCH® diabetes management software |
K. Standard/Guidance Document Referenced (if applicable):
1) Point-of-Care Blood Glucose Testing in Acute and Chronic care Facilities; Approved Guideline, 2nd Edition. CLSI Document C30-A2
2) Preliminary Evaluation of Quantitative Clinical Laboratory Methods; Approved Guideline. CLSI Document EP10-A
3) Evaluation of Matrix Effects; Approved Guideline, CLSI Document EP14-A2
4) Estimation of Total analytical Error for Clinical Laboratory Methods; Approved Guideline. CLSI Document EP21-A
5) User Demonstration of Performance for Precision and Trueness; Approved Guideline. CLSI Document EP15-A2
6) Interference Testing in Clinical Chemistry; Approved Guideline. CLSI Document EP7-A2
7) Evaluation of the Linearity of Quantitative Measurement Procedures; Approved Guideline, CLSI Document EP6-A
8) Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline. CLSI Document EP5-A2
9) Clinical Chemistry, 2nd Edition
10) MERCK INDEX, 11th Edition.
L. Test Principle:
The principle of the test relies on glucose in the blood sample to react to the electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. This reaction is measured by a glucose module integrated in the battery pack of the cell phone and the results of the blood glucose are displayed on the cell phone.
{4}
5 of 14
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Within-Run precision
The procedure was conducted by taking 4mL of blood that was treated with EDTA through a vacuum tube. Glucose was added to the 4 mL of blood to generate 5 different levels of glucose concentration for the test. Each of the samples was measured 50 times.
Between-Run precision
Three different control solutions of Low, Normal and High were measured twice a day, once in the morning (Run 1) and once in the afternoon (Run 2) during a month's time. For each Run, two measurement readings were taken in which the mean, standard deviation (SD), and coefficient variation percent (CV %) were calculated.
| | Mean (mg/dL) | SD | CV % |
| --- | --- | --- | --- |
| Within-run (venous blood) | 45.8 | 1.5 | 3.2 |
| | 99.6 | 2.5 | 2.5 |
| | 136.6 | 2.3 | 1.7 |
| | 196.8 | 3.3 | 1.7 |
| | 346.5 | 8.9 | 2.6 |
| Between-run/Total (control solution) | 56.6 | 2.1 | 3.7 |
| | 110.4 | 3.6 | 3.3 |
| | 300.2 | 9.7 | 3.2 |
b. Linearity/assay reportable range:
Samples with equally spaced concentrations were evaluated. The linear range is defined by the highest and lowest measured concentrations where the response is linear.
The GlucoPack phone used in this test can display below 10 mg/dL over 600 mg/dL for checking linear range. The planned GlucoPack phone for sale displays "Low" below 10 mg/dL, "Hi" over 600 mg/dL.
Each of the glucose levels were measured 5 times to test for precision. In order to evaluate the straight line for the Sensory Strip that was used, the following formula was used:
1st order polynomial, $y = ax + b$, 2nd order polynomial, $y = aX2 + bX + c$
In all dilution schemes, the applicant started with high and low concentration samples in which the concentrations meet or exceed the range of interest. For
{5}
the test, the highest and lowest glucose concentration used was $611.8\mathrm{mg / dL}$ and $8.8\mathrm{mg / dL}$ . If a strip sensor has an ideal linearity $(r^2 = 1)$ from lowest to highest concentration, the ideal concentration of level 2 mixed with $0.875\mathrm{L}$ and $0.125\mathrm{H}$ volume ratio is a $84\mathrm{mg / dL}$ $[(0.875*8.8 + 0.125*611.8) / (0.875 + 0.125)]$ .
| | GlucoPack™ | | | | | | YSI 2300 (Reference) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Dilution | Rep1 | Rep2 | Rep3 | Rep4 | Rep5 | Mean | | | Mean |
| 1 | 8 | 8 | 9 | 10 | 9 | 8.8 | 8 | 8 | 8.8 |
| 2 | 88 | 87 | 89 | 86 | 89 | 87.8 | 86 | 85 | 86.8 |
| 3 | 166 | 158 | 160 | 159 | 170 | 162.6 | 163 | 161 | 163.1 |
| 4 | 240 | 241 | 238 | 234 | 248 | 240.2 | 238 | 237 | 239.4 |
| 5 | 325 | 316 | 319 | 311 | 314 | 317 | 315 | 315 | 314.4 |
| 6 | 363 | 387 | 378 | 393 | 388 | 381.8 | 430 | 339 | 386.4 |
| 7 | 451 | 445 | 460 | 479 | 470 | 461 | 460 | 457 | 465.4 |
| 8 | 538 | 509 | 541 | 539 | 574 | 540.2 | 538 | 535 | 545.2 |
| 9 | 608 | 620 | 600 | 610 | 612 | 610 | 612 | 615 | 611.8 |
The Polynomial Evaluation of Linearity
| Dilution | Actual Mean | Predicted 1st order | Predicted 2nd order | Difference |
| --- | --- | --- | --- | --- |
| 1 | 8.8 | 12.1 | 10.3 | 1.8 |
| 2 | 86.8 | 87.1 | 86.7 | 0.5 |
| 3 | 163.1 | 162.1 | 162.7 | -0.5 |
| 4 | 239.4 | 237.2 | 238.3 | -1.1 |
| 5 | 314.4 | 312.2 | 313.5 | -1.3 |
| 6 | 386.4 | 387.2 | 388.3 | -1.1 |
| 7 | 465.4 | 462.2 | 462.7 | -0.5 |
| 8 | 545.2 | 537.2 | 536.7 | 0.5 |
| 9 | 611.8 | 612.2 | 610.4 | 1.8 |
The polynomial evaluation of linearity assumes that the data set is not linear. This approach assumes that the data points fall perfectly on a line or curve in the absence of random error. The method consists of two parts. The first part examines whether a nonlinear polynomial fits the data better than a linear one. The second part assesses whether the difference between the best-fitting nonlinear and linear polynomial is less than the amount of allowable bias for the method, which should be predefined.
The nonlinear 2nd fits the data better than a linear one, but the difference is lower than $1.8\mathrm{mg / dL}$ from $8.8\mathrm{mg / dl}$ to $611.8\mathrm{mg / dL}$ . The $\mathbb{R}^2$ of 1st order regression is a 0.9997
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
{6}
See k043512 and k031501
d. Detection limit:
10 – 600 mg/dL (linearity study above)
e. Analytical specificity:
Approximately 2 mL of blood was taken from 64 random diabetic individuals. The blood samples were treated with the EDTA vacuum tube and the glucose concentration was adjusted to < 50 ~ < 580 mg/dL by adding an adequate amount of the phosphate buffer (20mM with pH 7.4) that contains a different level of glucose. In order to adjust the Hematocrit value (30 ~ 55%), appropriate volumes of the centrifuged plasma was removed.
| Blood glucose range | number of samples |
| --- | --- |
| < 50 mg/dL | 8 |
| 51 ~ 110 mg/dL | 9 |
| 111 ~ 150 mg/dL | 11 |
| 151 ~ 250 mg/dL | 18 |
| 251 ~ 400 mg/dL | 24 |
| < 580 mg/dL | 10 |
| Total | 80 |
The Hematocrit level and glucose concentration in the blood was assessed by using the YSI2300 STAT PLUS.
The sponsor’s acceptance criteria are that all the data should be within +/- 30 % bias and that over 90% of data are within +/- 20%.
Bias between GlucoPack result and the corresponding comparison YSI result.
7 of 14
{7}

The % bias of the assay value of the GlucoPack system is relative to the YSI and does not have a negative or positive correlation to hematocrit level in this experiment. 99% of the data is within +/- 20% bias in the overall range of glucose and hematocrit.
## Interferences (see k043512):
- GlucoPack™ Test Strips are specific to D-glucose and do not react with other sugars which may be present in the blood.
- GlucoPack™ Test Strips do not interfere with the hematocrit at a normal range (30-55%) of blood glucose.
- Extreme levels in hematocrit may affect test results. Hematocrit levels less than 30% may cause falsely high readings. Hematocrit levels greater than 55% may cause falsely low readings.
- Blood samples that contain large amounts of ascorbic acid and uric acid may cause a slightly higher result than the actual glucose level.
- High concentrations of bilirubin, gentistic acid and other reducing substances in the blood may cause inaccurately high results.
- Lipemic samples; cholesterol up to 500mg/dL or triglycerides up to 3000 mg/dL do not significantly affect the results. Glucose values, however, in specimens beyond these levels should be interpreted with caution.
- Blood samples that contain a high concentration of dissolved oxygen may lower the test result.
8 of 14
{8}
- Dopamine treatment may increase the test result.
- Antiglycolysis may affect the test results.
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
The accuracy of the GlucoPack System was assessed by comparing blood Glucose levels on 410 diabetic patients obtained and tested by three healthcare professionals at three different clinical centers compared to matching venous plasma samples run on the Roche Hitachi 747 analyzer.
A consumer study was conducted using 104 diabetics self tested with the GlucoPack System compared to matching venous plasma samples run on the Roche Hitachi 747 analyzer.
The sponsor's acceptance criteria are as follows:
Slope greater than 0.96 and less than 1.05
Correlation coefficient greater than 0.96

Obtained by Healthcare Professionals in Clinical Center
Slope 1.0081
y-intercept -1.976
Correlation coefficient(r) 0.9844
No. of samples 410
Range tested 65-433mg/dL
{9}
# C. Obtained by Lay diabetics

Slope 1.0069
y-intercept -0.8198
Correlation coefficient(r) 0.9854
No. of samples 104
Range tested 74-429 mg/dL
Using the Clark Error Grid Analysis the study done by the clinical centers resulted in $99.51\%$ of the tested specimens, in region marked A, had glucose values that deviated from the reference method of less than $20\%$ or are less than $70~\mathrm{mg / dL}$ by both methods.
{10}
Error Grid Analysis - Hitachi 747 Vs. GlucoPack™
# Obtained by Healthcare Professionals in Clinical Centers

# Summary of Professional Error Grid Analysis Results
| Zone | Number | % |
| --- | --- | --- |
| A | 408 | 99.51 |
| B | 2 | 0.49 |
| C | - | 0 |
| D | - | 0 |
| E | - | 0 |
| Total | 410 | 100 |
Zone A = clinically accurate, leading to correct treatment decisions
Zone B = deviation from the reference by more than 20%, leading to benign or no treatment.
Zone C = over-correcting of acceptable blood glucose levels.
Zone D = represent a potentially dangerous failure to detect and treat glucose levels outside the desired target range.
Zone E = represents results that would lead to an erroneous treatment.
The study done with the lay diabetics resulted in $99\%$ of the tested subjects, in region marked A, had glucose values that deviated from the reference method of less than $20\%$ or are less than $70 \mathrm{mg} / \mathrm{dL}$ by both methods.
{11}
Error Grid Analysis - Hitachi 747 Vs GlucoPack

Obtained by Lay Diabetics
Hitachi 747 Venous Plasma Glucose (mg/dL)
Summary of Consumer Error Grid Analysis Results
| Zone | Number | % |
| --- | --- | --- |
| A | 103 | 99 |
| B | 1 | 1 |
| C | - | 0 |
| D | - | 0 |
| E | - | 0 |
| Total | 104 | 100 |
Zone A = clinically accurate, leading to correct treatment decisions
Zone B = deviation from the reference by more than 20%, leading to benign or no treatment.
Zone C = over-correcting of acceptable blood glucose levels.
Zone D = represent a potentially dangerous failure to detect and treat glucose levels outside the desired target range.
Zone E = represents results that would lead to an erroneous treatment.
b. Matrix comparison: Not Applicable
3. Clinical studies:
a. Clinical Sensitivity: Not Applicable
b. Clinical specificity: Not Applicable
{12}
c. Other clinical supportive data (when a. and b. are not applicable): See the consumer method comparison study above.
4. Clinical cut-off: Not Applicable
5. Expected values/Reference range:
Expected blood glucose levels for people without diabetes:
| Time | Range (mg/dL) | Range (mmol/L) |
| --- | --- | --- |
| Before Breakfast: | 70-105 | 3.9-5.8 |
| Before lunch or dinner: | 70-110 | 3.9-6.1 |
| 1 hour after meals: | Less than 160 | Less than 8.9 |
| 2 hours after meals: | Less than 120 | Less than 6.7 |
| Between 2 and 4 AM: | Greater than 70 | Greater than 3.9 |
N. Instrument Name:
HealthPia GlucoPack™ Meter
O. System Descriptions:
1. Modes of Operation:
Manual fingerstick
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
FDA has reviewed applicant’s Hazard Analysis and software development processes for this device has determined the device was developed under good software lifecycle processes.
3. Specimen Identification:
No sample identification
4. Specimen Sampling and Handling:
Individual Fingerstick
13 of 14
{13}
5. Calibration:
Utilizes a Check Strip for Calibration
6. Quality Control:
Two control materials are available for use with the meter. The meter does not have a quality control data analysis software application.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14 of 14
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.