3TEST GLUCOSE MONITORING SYSTEM

K050224 · Neuero Engineering, Inc. · NBW · Jul 13, 2005 · Clinical Chemistry

Device Facts

Record IDK050224
Device Name3TEST GLUCOSE MONITORING SYSTEM
ApplicantNeuero Engineering, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateJul 13, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The 3test® glucose test strip is intended to measure the glucose in whole blood with the 3test® Glucose Monitoring System. It is suitable for a person with diabetes to monitor their blood glucose at home by themselves. The 3test® Glucose Monitoring System can also be used at clinical sites by nurses or professional people to test patient's glucose level in whole blood.

Device Story

3test® Glucose Monitoring System uses electrochemical biosensor technology; capillary action draws whole blood sample into test strip chamber. Meter processes sample; displays glucose result in 5 seconds. Features 100-result memory. Used at home by patients or in clinical settings by healthcare professionals. Provides quantitative glucose levels to assist in diabetes management.

Clinical Evidence

Pre-clinical and clinical data submitted per FDA Guidance for In Vitro Diagnostic Test Systems. No specific performance metrics (e.g., sensitivity, specificity) provided in summary.

Technological Characteristics

Electrochemical sensing using glucose oxidase and potassium ferricyanide. Hand-held meter with LCD display. Automatic calibration via lot-specific code chip. Powered by two 3V lithium batteries. Dimensions: 100x58x21 mm; Weight: 57g. Operating range: 57–104° F, 20-90% relative humidity. Sample volume: 2.0 µL. Test time: 10 seconds. Memory: 100 results.

Indications for Use

Indicated for use with the 3test Glucose Monitoring System to measure glucose in capillary whole blood from the fingertip. Suitable for home use by persons with diabetes and for professional use at clinical sites. Not for use with neonates.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k050224 B. Purpose for Submission: Request for clearance of new device C. Measurand: Blood glucose, home-use glucose monitoring test D. Type of Test: Quantitative E. Applicant: Neuero Engineering Inc. F. Proprietary and Established Names: 3test Glucose Monitoring System G. Regulatory Information: 1. Regulation section: 21 CFR §862.1345, Blood Glucose Test System, Over-the-Counter 2. Classification: Class II 3. Product code: NBW, CGA 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): A glucose test system is intended to measure glucose quantitatively in blood and other bodily fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell tumors. 2. Indication(s) for use: "The 3test® Glucose test strip is intended to measure the glucose in whole blood with the 3test® Glucose Monitoring System. It is suitable for a person with diabetes to monitor their blood glucose at home by themselves. The 3test® Glucose Monitoring System can also be used at clinical sites by nurses or {1} professional people to test patient’s glucose level in whole blood. NOTE: a. the 3test is to be used with capillary whole blood from the fingertip b. the 3test is not for use with neonates c. the 3test meter is to be used with the 3test Blood Glucose Test Strip, and the 3test High and Low Glucose Control Solutions 3. Special conditions for use statement(s): This product is intended for over-the-counter and point-of-care use. 4. Special instrument requirements: None; this is a complete blood glucose monitoring system. I. Device Description: The 3test® Glucose Monitoring System consists of a hand-held blood glucose meter, test strips, and two levels of control materials. Each lot of test strips has a code chip containing lot-specific calibration information that the machine reads automatically. The meter is turned on by strip insertion; the user then supplies finger-tip blood or control solution to the strip and the meter makes an audible tone and starts the assay, which completes in ten seconds. The meter’s software converts the results read off the test strip into a plasma glucose concentration and displays the value on the meter’s LCD screen. Each test strip contains three different reaction areas: after the test is complete, the user breaks off the used reaction area with a provided tool and stores the strip for subsequent tests. J. Substantial Equivalence Information: 1. Predicate device name(s): Bayer Glucometer Elite Blood Glucose Meter and Test Strips 2. Predicate 510(k) number(s): k964630 (the Elite series has been subject to multiple 510(k) submissions since the claimed predicate including k043158, k020208, k990649, and k991242) 3. Comparison with predicate: {2} | Similarities | | | | --- | --- | --- | | Item | 3test | Ascensia ELITE | | Intended Use | Blood glucose monitoring for home and point-of-care | Same | | System Components | Meter, calibration code strip, test strip, check strip, battery, control solutions | Same | | Specimen | Capillary blood | Same, and approved for arterial and neonatal specimens | | Test Principle/ Enzyme/ Mediator | Electrochemical/ Glucose oxidase/ Potassium ferricyanide | Same/ same/ same | | Test Range | 20 – 600 mg/dL | 20 – 600 mg/dL | | Hematocrit Range | 30 – 55% | 20 – 60% | | Calibration | Automatic (code strip) | Same | | Stability | Closed strips and controls: 20 months Opened strips and controls: 3 months | Same | | Sample Volume | 2.0 ul | 2.0 ul | | Power Source | Two 3V lithium batteries | One 3V lithium battery | | Differences | | | | Item | 3test | Ascensia ELITE | | | | | | Test Time | 5 seconds | 30 seconds | | Operating Range | 57 – 104° F, relative humidity 20 - 90%, | 50 – 104° F, relative humidity 20 - 80% | | Memory Capability | 100 test results | 20 test results | | Size | 100x58x21 (mm) | 81x51x14 (mm) | | Weight | 57 grams | 50 grams | {3} K. Standard/Guidance Document Referenced (if applicable): | Area of Study | Reference Procedure | Procedure Title | | --- | --- | --- | | Method Comparison/Anticoagulant Studies | NCCLS EP9-T | Method Comparisons and Bias Estimations Using Patient Samples. | | Precision | NCCLS EP5-T2 | User Evaluation of Precision Performance of Clinical Chemistry Devices | | Linearity | NCCLS EP6-P | Evaluation of the Linearity of Quantitative Methods | | Interferences/Cross-Reactivity | NCCLS EP7-P | Interference Testing in Clinical Chemistry | | Guidance | CDRH Guidance | CDRH Review Criteria Portable Glucose Monitoring Devices Intended for Bedside Use in the Neonatal Nursery | | | NCCLS Guideline C30-A | Ancillary (bedside) blood glucose testing in acute and chronic care facilities. | L. Test Principle: The test is based on the release of electrical potential after a two-step reaction where glucose and ferricyanide, in the presence of glucose oxidase, are converted into gluconolactone and ferrocyanide. Ferrocyanide, when electrical current is applied, becomes ferricyanide and releases electrons; the increase in current measured after 10 seconds by the meter is proportional to the glucose concentration. As the rate of the chemical reaction is proportional to temperature, the meter also takes the temperature into account when calculating the glucose concentration. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision was determined at three POC (point-of-care) and three POL (physician office laboratories) by health-care professionals. Five levels of glucose-spiked blood (hydrolyzed overnight) were tested twenty times at each site, covering a range between $40\mathrm{mg/dL}$ and $400\mathrm{mg/dL}$. Within-day precision results are shown in the table below: {4} 5 3test Blood Glucose Monitoring System: Within-day Precision | Point of Care Studies | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Site 1 | YSI (mg/dL) | 41 | 75 | 127 | 203 | 311 | | | 3test Mean ± SD | 41 ± 2.4 | 77.1 ± 4.1 | 125 ± 7.0 | 206 ± 6.7 | 314 ± 11.7 | | | %CV | 5.9 | 5.4 | 5.6 | 3.3 | 3.7 | | | | | | | | | | Site 2 | YSI (mg/dL) | 48 | 98 | 142 | 221 | 372 | | | 3test Mean ± SD | 50 ± 2.7 | 100± 5.1 | 153 ± 6.5 | 220 ± 12.1 | 369 ± 11.8 | | | %CV | 5.4 | 5.1 | 4.3 | 5.5 | 3.2 | | | | | | | | | | Site 3 | YSI (mg/dL) | 45 | 91 | 135 | 186 | 294 | | | 3test Mean ± SD | 46 ± 3.0 | 96 ± 5.5 | 142 ± 7.8 | 194 ± 10.3 | 305 ± 13.6 | | | %CV | 6.5 | 5.7 | 5.5 | 5.3 | 4.5 | | | | | | | | | | Physician Office Laboratories | | | | | | | | Site 1 | YSI (mg/dL) | 47 | 82 | 142 | 196 | 309 | | | 3test Mean ± SD | 47 ± 2.7 | 82 ± 4.6 | 142 ± 5.3 | 199 ± 8.0 | 312 ± 12.1 | | | %CV | 5.7 | 5.6 | 3.7 | 4.0 | 3.9 | | | | | | | | | | Site 2 | YSI (mg/dL) | 50 | 104 | 127 | 213 | 262 | | | 3test Mean ± SD | 51 ± 2.4 | 106 ± 5.8 | 130 ± 6.2 | 216 ± 11.6 | 272 ± 10.8 | | | %CV | 4.7 | 5.5 | 4.7 | 5.4 | 4.0 | | | | | | | | | | Site 3 | YSI (mg/dL) | 43 | 75 | 147 | 226 | 325 | | | 3test Mean ± SD | 45 ± 2.6 | 77 ± 4.6 | 151 ± 6.7 | 230 ± 10.8 | 333 ± 11.6 | | | %CV | 5.8 | 6.0 | 4.5 | 4.7 | 3.5 | Between-day precision was tested at three POC sites and three POL sites by health-care professionals. Five levels of glucose-spiked blood (hydrolyzed overnight) were tested 10 times per day for five days at each site, covering a range between 50 mg/dL and 400 mg/dL. Precision was similar at both kinds of sites, so only POC results are summarized below: 3test Blood Glucose Monitoring System: Between-day Precision at 3 POC Sites | | | Glucose Concentration by YSI (mg/dL) | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 50 | 110 | 150 | 250 | 400 | | Site 1 | Mean 3test (mg/dL) | 52.6 | 115.3 | 154.3 | 254.3 | 408.7 | | | % CV | 5.7 | 5.4 | 3.9 | 3.3 | 2.9 | | Site 2 | Mean 3test (mg/dL) | 55.0 | 110.5 | 152.5 | 252.3 | 412.2 | | | % CV | 5.5 | 4.9 | 4.3 | 3.7 | 2.6 | | Site 3 | Mean 3test (mg/dL) | 52.6 | 114.5 | 156.0 | 252.5 | 406 | | | % CV | 6.6 | 5.8 | 3.4 | 3.5 | 3.0 | {5} The sponsor measured the difference in blood glucose values between the three reaction chambers on each strip by using blood samples of known value across the measurement range. Readings were collected for each reaction chamber on 15 strips and averaged by reaction chamber. Mean values between the three chambers varied 2.5% - 4.0%. This met the sponsor's acceptance criterion of ≤ 5% difference. b. Linearity/assay reportable range: Heparinized venous whole blood was allowed to glycolyze for 24 hours before testing then aliquoted and spiked with glucose. Twenty concentrations spanning 20 ~ 600 mg/dL were tested 8 times each by 3test (one lot of strips) and YSI and the mean value of each aliquot was calculated. The regression analysis showed a linear relationship between the 3test and the YSI method: y = 1.005x + 0.851, R = 0.999. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Control solutions are prepared by gravimetrically adding glucose to the control base solution. Value assignments for each lot of control are determined using 20 replicates of one test strip lot read on a 3test meter over 10 days to determine mean, standard deviation, and %CV. Real-time shelf life studies performed by the manufacturer indicate that unopened test strips have a 20 month life-span. Once the strip container is opened, the test strip has been shown to be stable for 3 months. d. Detection limit: The low and high detection limits for the 3test system have been set at 20 and 600 mg/dL glucose. Readings below or above these values will generate a "LO" or "HI" result respectively. e. Analytical specificity: Assay interferents were tested in a dose-response manner following NCCLS EP7-A guidelines. Aliquots of the blood were supplemented with glucose to a final concentration of 100 mg/dL and measured on an YSI analyzer. The interferent was prepared with an appropriate solvent, and spiked into the 100 mg/dL blood. A control pool was prepared by supplementing the blood with solvent minus the interferent. A series of four to five levels that included the maximum concentration of the substance that would be expected to be encountered in clinical practice were used for each interferent. The table below shows the effect of common interferents at the upper end of normal or therapeutic levels on 3test test levels: 6 {6} 7 # Interference at High-Normal or High Therapeutic Levels ## 3test System | Interferent | Upper End Therapeutic or Normal Range | Highest Concentration Tested with No Interference | | Interferent | Upper End Therapeutic or Normal Range | Highest Concentration Tested with No Interference | | --- | --- | --- | --- | --- | --- | --- | | Acetaminophen | 2 mg/dL | 8 mg/dL | | Tolazamide | 3 mg/dL | 100 mg/dL | | Ascorbic Acid | 2 mg/dL | 3 mg/dL | | Tolbutamide | 10 mg/dL | 100 mg/dL | | Bilirubin | 1.2 mg/dL | 20 mg/dL | | Triglycerides | 190 mg/dL | 3000 mg/dL | | Cholesterol | 300 mg/dL | 500 mg/dL | | Uric Acid | 7.7 mg/dL | 20 mg/dL | | Creatinine | 1.5 mg/dL | 30 mg/dL | | Citrate (K+) | 500 mg/dL | 1000 mg/dL | | Dopamine | N/A | 13 mg/dL | | EDTA (K+) | 150 mg/dL | 300 mg/dL | | Ibuprofen | 4.2 mg/dL | 40 mg/dL | | Fluoride | 500 mg/dL | 1000 mg/dL | | Methyl-Dopa | 0.75 mg/dL | 2.5 mg/dL | | Heparin | 4000 U/dL | 8000 U/dL | | Salicylates | 30 mg/dL | 50 mg/dL | | Oxalate (K+) | 400 mg/dL | 800 mg/dL | | Tetracycline | 0.4 mg/dL | 4 mg/dL | | Thymol | 140 mg/dL | Interfered at 140 mg/dL | The meter was tested at different altitudes to assess the effect of low oxygen levels on meter performance. No effect on performance was found when three different levels of blood were tested up to 8800 ft. Higher elevations were not tested. The sponsor presented data that supported using the test system between $14^{\circ}\mathrm{C}$ to $40^{\circ}\mathrm{C}$. ## Hematocrit Effect: The effect of sample hemoglobin variation on the 3test system was tested experimentally by preparing samples of known hematocrit (Hct) and spiking aliquots of these samples with three different levels of glucose. These samples were run on the 3test and YSI; there was less than an $\pm 11\%$ bias across the claimed range of $30\sim 55\%$ Hct. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: See clinical studies section below. b. Matrix comparison: Not applicable. This device is indicated for use with capillary blood only. The meter's software adjusts the whole-blood glucose reading to a plasma-equivalent reading. {7} 8 3. Clinical studies: Two separate clinical studies of the 3test system performance were performed. Results are presented in the Clinical Sensitivity section below. The consumer study was performed at three POC sites with a total of 150 lay-users. The lay-users ranged in age, education, and were about equally divided between males and females; type-2 diabetes was more common in the participant groups, although not all participants have diabetes. The native language of most of the participants was English. Each participant performed their own fingerstick and tested their blood using the instructions in the User's Guide. A trained technician then performed another fingerstick and tested the blood on the same meter. Capillary blood was collected and measured on a YSI analyzer. Another study was performed by two health care professionals (HCP) at three POC and three POL sites (n = 6 HCP). Forty patient volunteers participated at each site; six fingersticks were performed for testing on the 3test and six capillary tubes of blood were collected for testing on the YSI analyzer. a. Clinical Sensitivity: Consumer Study Results : 3test vs. YSI | Site | 3test v. YSI | r value | Sample Range (mg/dL, by YSI) | % Parkes Error Grid | | | | --- | --- | --- | --- | --- | --- | --- | | | | | | A | B | | | Consumer Results | | | | | | | | 1 | n = 47 | y = 1.001x + 0.558 | 0.987 | 66 – 250 | | | | 2 | n = 55 | y = 1.042x - 2.777 | 0.987 | 64 – 340 | | | | 3 | n = 48 | y = 0.999x - 0.388 | 0.981 | 65 – 351 | | | | Sum | n = 150 | y = 1.021x - 2.000 | 0.984 | 64 – 351 | 98.7% | 1.3% | | | | | | | | | | Technician Results | | | | | | | | 1 | n = 47 | y = 0.995x + 0.398 | 0.982 | | | | | 2 | n = 55 | y = 1.012x + 3.851 | 0.986 | | | | | 3 | n = 48 | y = 0.971x + 4.483 | 0.985 | | | | | Sum | n = 150 | y = 1.004x + 2.119 | 0.984 | | 98% | 2% | {8} HCP Study Results: 3test vs. YSI | Site (n = 40) | 3test v. YSI | r value | Sample Range (mg/dL, by YSI) | % ± 20% Reference value | % Parkes Error Grid | | | --- | --- | --- | --- | --- | --- | --- | | | | | | | A | B | | POC 1 | y =1.033x -0.32 | 0.994 | 65.7 – 344.7 | 97.5% | 97.5 | 2.5 | | POC 2 | y =1.028 – 1.45 | 0.989 | 63.7 – 327.2 | 100 | 97.5 | 2.5 | | POC 3 | y = 1.064 – 5.06 | 0.984 | 55.8 – 348.2 | 100 | 97.5 | 2.5 | | | | | | | | | | POL 1 | y = 1.025x – 2.08 | 0.989 | 64.2 – 323 | 100 | 100 | 0 | | POL 2 | y =1.006x + 2.58 | 0.989 | 64.8 – 346 | 100 | 97.5 | 2.5 | | POL 3 | y = 1.043 – 2.70 | 0.992 | 61 - 326 | 100 | 97.5 | 2.5 | b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: The normal fasting adult glucose range for a non-diabetic is 70 – 105 mg/dL. One to two hours after a meal, normal blood glucose levels should be less than 140 mg/dL. A medical professional should determine the range that is appropriate for diabetes patients. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...