EBCARRYON BLOOD GLUCOSE MONITORING SYSTEM

K091765 · Visgeneer, Inc. · NBW · Jan 19, 2010 · Clinical Chemistry

Device Facts

Record IDK091765
Device NameEBCARRYON BLOOD GLUCOSE MONITORING SYSTEM
ApplicantVisgeneer, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateJan 19, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The eBcarryon Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The eBcarryon Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus and is not for use on neonates. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by person with diabetes, or in clinical setting by health care professionals, as an aid to monitor the effectiveness of diabetes control.

Device Story

System uses electrochemical biosensor technology; capillary action draws fresh capillary whole blood from finger into test strip chamber. Device measures glucose concentration; displays result in 10 seconds. Used at home by patients or in clinical settings by professionals to monitor diabetes control effectiveness. Provides quantitative glucose readings to assist in diabetes management.

Clinical Evidence

Pre-clinical and clinical data submitted in accordance with FDA Guidance for In Vitro Diagnostic Test Systems. No specific performance metrics (e.g., sensitivity, specificity) provided in summary.

Technological Characteristics

Electrochemical biosensor; carbon electrodes; glucose oxidase enzyme. Dimensions: 110x58x25 mm; weight: 63 g. Sample volume: 2.5 μL. Measurement time: 10 seconds. Range: 30-600 mg/dL. Connectivity: None. Calibration: Lot-specific code card. Standards: ISO 15197:2003, IEC/EN 60601-1:2001.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood from fingertip. For use by persons with diabetes (OTC) or healthcare professionals in clinical settings. Contraindicated for neonates, diagnosis/screening of diabetes, dehydrated/hypotensive/shock patients, hyperglycemic-hyperosmolar state, critically ill patients, or non-fingertip samples.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 5091765 Image /page/0/Picture/1 description: The image shows a small, pixelated black and white graphic. The graphic appears to be a stylized representation of a person or figure, possibly in a dynamic pose. Due to the low resolution and pixelation, specific details are difficult to discern, but the general shape suggests a human-like form with some artistic or abstract elements. ISGENEER INC. -2, No.83, Sec.2, Gongdao 561 Rd., Hsinchu City 300, Taiwan Fax: 886-5160161 Tel: 886-3-5160111 # Section III. Summary JAN 1 9 2010 This Summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of (Per 21 CFR 807.92) 2.1 General Information Establishment - Manufacturer: VISGENEER, INC. - Registration Number : 3005644014 . - 3F .- 2, No.83, Sec.2, Gongdao 5th Rd., Hsinchu City 300, . Address : - Taiwan - 886-3-5160111 Ext.3101 Phone : - 0 Fax : 886-3-5160161 - Contact Person: Evonne Chen 0 - April 13t , 2009 Date Prepared : 2.2 Name of Device | Trade Name: | eBcarryon Blood Glucose Monitoring System | |----------------------|-------------------------------------------------------------| | Common Name: | Blood Glucose Monitoring System | | Classification Name: | SYSTEM, TEST, BLOOD GLUCOSE,<br>OVER THE COUNTER, Class II, | 2.3 Predicate Device : Claim of Substantial Equivalence (SE) is eBsensor Blood Glucose Monitoring System, eB-G. (K062555) 2.4 Description of Device : Based on an electrochemical biosensor technology and the principle of capillary action, eBcarryon Blood Glucose Monitoring System only needs a small amount of blood . Capillary action at the end of the test strip draws the blood into the action chamber and your blood glucose result is precisely and displayed in 10 seconds. {1}------------------------------------------------ 3F .- 2, No.83, Sec.2, Gongdao 5th Rd., Hsinchu City 300, Taiwan Fax: 886-5160161 Tel: 886-3-5160111 ### 2.5 Intended Use: The eBcarryon Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The eBcarryon Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus and is not for use on neonates. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by person with diabetes, or in clinical setting by health care professionals, as an aid to monitor the effectiveness of diabetes control. ### 2.6 Synopsis of Test Methods and Results: Pre clinical and clinical data are employed upon submission of this 510(k) premarket notification according to the Guidance Document for In Vitro Diagnostic Test System ; Guidance for Industry and FDA Document provided by CDRH/FDA. - 2.7 Substantial Equivalence (SE): A claim of substantial equivalence is made to Visgeneer INC. - eBsensor Blood Glucose Monitoring System, eB-G (K062555). Both of them have the same working principle and technologies. The differences are electric voltage, Meter Dimension of meter, Memory Storage, and weight. There are no safety and effectiveness aspects arising from the subject device. They are substantially equivalent. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002 Visgeneer, Inc. c/o Ms. Evonne Chen Supervisor of Control Quality Dept. 3F-2, No.83, Sec.2, Gongdao 5th Road. Hsinchu City Taiwan 30070 JAN 1 9 2010 Re: k091765 > Trade name: eBcarryon Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, CGA Dated: December 22, 2009 Received: December 22, 2009 Dear Ms. Chen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ #### Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301.) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Signature Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image is a black and white graphic that appears to be a low-resolution scan or photograph of an object or scene. The details are not clear due to the image quality, but it seems to depict a textured or patterned surface with some darker areas that could be shadows or distinct features. The overall impression is that of a rough or uneven surface, possibly a close-up of a natural material or an abstract design. VISGENEER INC. 3F -2, No.83, Sec.2, Gongdao 5th Rd., Hsinchu City 300, Taiwan Tel: 886-3-5160111 - Fax: 886-5160161 # Indications for Use 510 (K) Number (If Known): KO9 17 65 Device Name: __eBcarryon Blood Glucose Monitoring System Indications for Use: The eBcarryon Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The eBcarryon Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus and is not for use on neonates. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by person with diabetes, or in clinical setting by health care professionals, as an aid to monitor the effectiveness of diabetes control. Prescription Use V and/or (21 CFR Part 801 Subpart D) Over the Counter Use V (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD). Carol Benam ! : on Sign-Off See of In Vitro Diagnostic Device mation and Safety K091765
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...