K110709 · Arkray Factory USA, Inc. · NBW · May 19, 2011 · Clinical Chemistry
Device Facts
Record ID
K110709
Device Name
ARK CARE DIABETES MANAGMENT SYSTEM
Applicant
Arkray Factory USA, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
May 19, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The ARKRAY Diabetes Management Software is an optional accessory for use with compatible blood glucose meters, such as ARKRAY Glucocard Vital Blood Glucose Meter with data management capabilities. The ARKRAY Diabetes Management Software transfers data from the meter's memory into a secured sever for enhanced data management. ARKRAY ARK Care™ Diabetes Management System is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating about historical blood glucose test results and other biological statistics to support improved diabetes management.
Device Story
ARK Care is a web-based diabetes management system; acts as interface between personal blood glucose monitors and secure health management database. Input: blood glucose data uploaded from compatible meters (ARKRAY, Roche, Bayer, LifeScan, Abbott) via USB cable; user-entered health data (weight, blood pressure, lab values, exercise). Processing: data storage, organization, and analysis. Output: logbooks, charts, graphs, and reports accessible via secure web portal. Used in home and clinical settings; operated by patients, family members, and healthcare professionals. Function: facilitates remote monitoring, data sharing, and communication between patients and care teams via password-protected portal. Benefit: improved diabetes treatment regulation and health outcomes through centralized data tracking and trending.
Clinical Evidence
Usability study conducted with 38 diabetic adult consumers and 6 healthcare professionals. Participants performed standardized tasks (enrollment, data upload, logbook entry, chart viewing). Results showed >99% pass rate for ease-of-use and performance questionnaires. 68% of personal users and 100% of professional users indicated intent to use the system for diabetes management. Pass/fail criteria exceeded 70-80% thresholds. Bench testing confirmed accurate data transmission from compatible meters, including memory capacity and rollover synchronization.
Technological Characteristics
Web-based diabetes management software; utilizes USB data transfer cable for meter connectivity. Compliant with HIPAA and HITECH standards for data security. Operates via internet-connected browser. No specific materials or ASTM standards cited for the hardware interface.
Indications for Use
Indicated for people with diabetes and their healthcare professionals in home and clinical settings to manage blood glucose data and other biological statistics.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
MyCare Team, MCT Diabetes (k073699)
Related Devices
K083263 — SYMCARE DIABETES MANAGEMENT PROGRAM · Symcare Personalized Health Solutions, Inc. · Mar 13, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a sequence of handwritten alphanumeric characters. The sequence appears to be "K110709". The characters are written in a simple, somewhat crude style, with varying stroke thicknesses.
MAY 1 9 2011 -
ARK Care™ 510(k) Summary
| Submitter: | ARKRAY Factory USA, Inc.<br>5182 W. 76th Street<br>Minneapolis, MN 55439 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tyler Foutch<br>Regulatory Affairs Project Manager<br>ARKRAY Factory USA, Inc.<br>5182 W. 76th Street<br>Minneapolis, MN 55439<br>Phone: 952-646-3171<br>Fax: 952-646-3110<br>fbutcht@ARKRAYusa.com |
| Date Prepared: | March 8, 2010 |
| Trade Name: | ARKRAY ARK Care Diabetes Management System™ |
| Classification: | Glucose test system, 21 CFR 862.1345; Class II |
| | Calculator/data processing module for clinical use, 21 CFR 862.2100;<br>Class I |
| Product Codes: | CGA, NBW, JQP. |
| Predicate Device: | MCT Diabetes™ (K073699). |
| Device Description: | ARK Care system serves as an interface between the software in<br>personal glucose monitoring devices and a general purpose health<br>management database to assist in the review, analysis and evaluation<br>of blood glucose test results. The secure system is compliant with<br>HIPAA and HITECH standards.<br><br>ARK Care is designed for home use and professional healthcare<br>settings. It is an accessory device to most manufactured model home-<br>use blood glucose monitors, including glucose monitoring devices by<br>ARKRAY, Roche, Bayer, LifeScan, and Abbott. The list of supported<br>devices is located at www.arkcare.net.<br><br>The purpose of the electronic diabetes management system is to help<br>users and healthcare teams manage blood glucose information to<br>better regulate diabetes treatments and control blood glucose for<br>better health outcomes. The ARK Care system holds a convenience<br>function, as the user can upload blood glucose data from a variety of<br>currently marketed blood glucose monitors into one location for<br>viewing. After the user transmits and stores blood glucose data to the<br>secure database, the ARK Care system allows family members and/or<br>healthcare professionals to view and monitor the user's data and<br>reports. Family and healthcare team members must receive<br>permission from the primary user and create a password protected<br>login before viewing data. ARK Care provides a safe communication<br>portal for the user, family, and healthcare team to send email-like |
| | entries, improving patient care and disease management. |
| | The user and approved healthcare team members can view the blood<br>glucose data in different formats such as logbooks, charts, and graphs.<br>The data can be viewed through selected time intervals and these<br>intervals can be compared over time to track disease management. |
| | The subject can also enter and track other health-related information<br>such as body weight, blood pressure, lab values, and exercise<br>activities. Groups of profiles can be queried for bulk reporting on data<br>related to tracking and trending of outcomes, supporting diabetes<br>disease management in individuals and in managed care<br>organizations. |
| Intended Use: | ARKRAY ARK Care Diabetes Management System™ is intended for<br>use in home and clinical settings via the internet to assist people with<br>diabetes and their healthcare professionals in uploading, storing,<br>analyzing, and communicating about historical blood glucose test<br>results and other biological statistics to support diabetes management. |
| Functional and<br>Safety Testing: | A full array of in-house testing was done consistent with relevant FDA<br>guidance's for blood glucose monitoring systems. Testing included<br>validation of the systems hardware (USB data transfer cable) and<br>software as well as consumer studies that demonstrated the systems<br>ability to be easily operated by in-home users. |
| Conclusion: | Labeling, validation testing results and consumer studies results<br>support the Indications for Use and the claim of substantial<br>equivalence to the predicate (K073699). |
{1}------------------------------------------------
.
・
.
:
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized emblem consisting of a series of curved lines, possibly representing a stylized caduceus or a similar symbol associated with health and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Arkray Factory Usa, Inc. c/o Tyler Foutch Project Manager Regulatory Affairs 5182 West 76th St., Minneapolis. MN 55439
MAY 1 9 2011
Re: k110709
Trade/Device Name: ARK Care™ Diabetes Management System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system. Regulatory Class: II Product Code: NBW, CGA, JQP Dated: March 11, 2011 Received: March 14,2011
Dear: Mr. Foutch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
## Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
CJC.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
:
Enclosure
{4}------------------------------------------------
## Indication for Use
510(k) Number (if known): k110709
## Device Name: ARK CARE™ DIABETES MANAGEMENT SYSTEM
Indication For Use:
The ARKRAY Diabetes Management Software is an optional accessory for use with compatible blood glucose meters, such as ARKRAY Glucocard Vital Blood Glucose Meter with data management capabilities. The ARKRAY Diabetes Management Software transfers data from the meter's memory into a secured sever for enhanced data management. ARKRAY ARK Care™ Diabetes Management System is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating about historical blood glucose test results and other biological statistics to support diabetes management.
Prescription Use (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use X . (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Diving Officer
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) = k110709
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.