EASY CHECK TD-4209 BLOOD GLUCOSE MONITORING SYSTEM; CLEVER CHECK TD-4222, TD-4225, TD-4226 BLOOD GLUCOSE MONITORING SYST

K062235 · Taidoc Technology Corporation · NBW · Sep 19, 2006 · Clinical Chemistry

Device Facts

Record IDK062235
Device NameEASY CHECK TD-4209 BLOOD GLUCOSE MONITORING SYSTEM; CLEVER CHECK TD-4222, TD-4225, TD-4226 BLOOD GLUCOSE MONITORING SYST
ApplicantTaidoc Technology Corporation
Product CodeNBW · Clinical Chemistry
Decision DateSep 19, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The easy check TD-4209/Clever Chek TD-4222/Clever Chek TD-4225/Clever Chek TD-4226 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The alternative site testing in the easy check TD-4209/Clever Chek TD-4222/Clever Chek TD-4225/Clever Chek TD-4226 Blood Glucose Monitoring System can be used only during steady-state blood glucose conditions.

Device Story

System comprises blood glucose meter, test strips, control solutions, and lancet device; measures glucose in capillary whole blood via amperometric assay. User applies blood sample to test strip; glucose oxidase enzyme reacts with glucose to produce electrical current proportional to concentration; meter converts current to glucose value. Used by healthcare professionals and patients at home for diabetes management. Provides quantitative results to aid in monitoring diabetes control programs. Requires code strip for batch calibration. Performance verified against YSI-2300 reference method; supports alternative site testing (palm, forearm, upper-arm, calf, thigh) during steady-state conditions.

Clinical Evidence

Bench testing only. Precision evaluated using control solutions and venous whole blood (CV% 1.60-3.98%). Linearity demonstrated across 20-600 mg/dL range (r=0.9896). Method comparison (N=120 finger; N=100 per AST site) against YSI-2300 reference method showed 95%+ of results within ISO 15197 accuracy requirements (±15 mg/dL for <75 mg/dL; ±20% for ≥75 mg/dL). Altitude and reducing substance interference studies confirmed equivalence to predicate.

Technological Characteristics

Electrochemical glucose test system; capillary whole blood sample; handheld meter form factor; battery-powered; quantitative measurement.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood from finger, palm, forearm, upper-arm, calf, and thigh. Intended for healthcare professionals and patients with diabetes mellitus for monitoring diabetes control programs. Not for neonates, diagnosis, or screening. Alternative site testing restricted to steady-state blood glucose conditions.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k062235 B. Purpose for Submission: Change in volume, strip application port, read time, and addition of AST sites. C. Measurand: Glucose D. Type of Test: Blood Glucose Concentration through a Quantitative Amperometric Assay (Glucose Oxidase) E. Applicant: TaiDoc Technology Corporation F. Proprietary and Established Names: Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1345, Glucose Test System 2. Classification: Class II 3. Product code: NBW, CGA 4. Panel: 75 (Clinical Chemistry) H. Intended Use: 1. Intended use(s): See Indications for use. 2. Indication(s) for use: The Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. 3. Special conditions for use statement(s): The alternative site testing in the Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems can be used only during steady-state blood glucose conditions. {1} 4. Special instrument requirements: Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems I. Device Description: The Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems each consist of four main products: the blood glucose meter, test strips (including the "check &amp; code" strip), control solutions (2 levels of Taidoc control solution - cleared under k012430), and the lancet device (cleared under k833344). These products have been designed and tested to work together as a system to produce accurate blood glucose test results. The sponsor recommends that only TaiDoc test strips and control solutions be used with the blood glucose meters. The performance of the test strips is verified by the control solutions. The check &amp; code strip verifies the status of the meter. The meters were originally cleared under 510(k)s k042005 (Easy Check TD-4209, Clever Chek TD-4222) and k051854 (Clever Chek TD-4225 and Clever Chek TD-4226) and were themselves modifications of k041107. Model TD-4207 was cleared originally under 510(k) k042005, and then cleared for use with new strips and alternate testing under 510(k) k061181 with the trade name ACHTUNG. J. Substantial Equivalence Information: 1. Predicate device name(s): ACHTUNG TD-4207, Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems 2. Predicate 510(k) number(s): k042005, k061181 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Devices | Predicate (k061181) | | Detection Method | Amperometry | Amperometry | | Enzyme | Glucose Oxidase (Aspergillus niger) | Glucose Oxidase (Aspergillus niger) | | Test Range | 20 – 600 mg/dL | 20 – 600 mg/dL | | Temperature Range | 50-104°F, 10-40°C | 50-104°F, 10-40°C | | Humidity Range | Below 85% | Below 85% | | Warranty (meter) | 5 years | 5 years | | Open Use Time (strip) | 90 days | 90 days | | Coding | Code strip | Code strip | | Memory Capability | 450 measurements | 450 measurements | | Power | CR2032 3V lithium battery | CR2032 3V lithium battery | | Differences | | | | --- | --- | --- | | Item | Devices | Predicate (k061181) | | Mechanical Appearance | Varies depending on model | Rounded form | K. Standard/Guidance Document Referenced (if applicable): NCCLS EP9-A: Method Comparison and Bias Estimation Using Patient Samples ISO 15197: In vitro diagnostic test systems - Requirements for blood-glucose monitoring {2} systems for self-testing in managing diabetes mellitus. The results presented in this submission support that the devices conform to acceptance criteria in this ISO 15197. L. Test Principle: Once a whole blood sample is applied to the sample chamber of the test strip, glucose measurement commences. Glucose measurement is based on electrical potential caused by the reaction of glucose with the reagents contained on the strip's electrodes. The glucose in the sample is oxidized by the enzyme glucose oxidase, producing gluconolactone. The current resulting from this enzymatic reaction is measured and converted to glucose concentration by the meter. The magnitude of the current is proportional to the concentration of glucose in the sample. M. Performance Characteristics (if/when applicable): Since the ACHTUNG meter has the same technological characteristics (with the exception of mechanical appearance) as the Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems, the ACHTUNG performance data was used to help support the addition of the test strips cleared in k061181 and the addition of a claim for alternate site testing in this submission. The sponsor also supplied performance data that supports that the Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems perform similarly to the predicate device (and to each other). 1. Analytical performance: a. Precision/Reproducibility: The sponsor evaluated the precision of the ACHTUNG meter (which is technologically similar to the Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 meters) and the modified strips in k061181. The sponsor evaluated replicate measurements of glucose control solutions and anticoagulated venous whole blood. In order to determine that the change in sample volume does not affect precision, the sponsor tested four volume levels (0.5, 0.6, 0.7, and 0.8 μL) of venous whole blood collected from 10 volunteers in k061181. The ten samples were pooled together, depleted of glucose, and then separated into three groups. Each group was spiked with a dextrose solution to within the desired concentration range: low 60-92 mg/dL, normal 109-165 mg/dL and high 259-389 mg/dL. Different testing volumes (0.5μL, 0.6μL, 0.7μL and 0.8 μL) of each sample were evaluated for repeatability. Since the volume claimed is 0.7 μL, only the 0.7 μL data from k061181 is presented below. 3 {3} | Patient | Low Control Level (60-92 mg/dL) | | | Mid Control Level (109-165 mg/dL) | | | High Control Level (259-389 mg/mL) | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Lot 1 | Lot 2 | Lot 3 | Lot 1 | Lot 2 | Lot 3 | Lot 1 | Lot 2 | Lot 3 | | 1 | 68 | 68 | 72 | 139 | 139 | 144 | 309 | 309 | 300 | | 2 | 68 | 68 | 65 | 132 | 132 | 141 | 310 | 310 | 297 | | 3 | 70 | 70 | 71 | 130 | 130 | 133 | 314 | 314 | 307 | | 4 | 68 | 68 | 71 | 136 | 136 | 137 | 308 | 308 | 300 | | 5 | 72 | 72 | 70 | 133 | 133 | 141 | 311 | 311 | 297 | | 6 | 72 | 72 | 72 | 133 | 133 | 134 | 312 | 312 | 313 | | 7 | 70 | 70 | 74 | 137 | 137 | 131 | 306 | 306 | 305 | | 8 | 69 | 69 | 71 | 135 | 135 | 134 | 313 | 313 | 302 | | 9 | 65 | 65 | 69 | 134 | 134 | 141 | 310 | 310 | 301 | | 10 | 64 | 64 | 74 | 137 | 137 | 145 | 305 | 305 | 305 | | Total CV(%) | 3.98 | | | 2.85 | | | 1.60 | | | To ensure that the meters perform similarly to the predicate device (and to each other), the strip cleared in k061181 was tested with the Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 meters. Precision testing was performed with 3 levels of TaiDoc control solutions with a single lot of strips with 25 measurements obtained for each meter at each control level. The average level for each control solution level is based on the observed result over a periods of 6 months. Each level was established as an average value first, which for low is $77\mathrm{mg / dL}$ , for normal is $135\mathrm{mg / dL}$ , and for high is $325\mathrm{mg / dL}$ . Then the range of each level was established from the average minus/plus $20\%$ (20% is from ISO15197, minimum acceptable accuracy), giving low ranges of 61 to $93\mathrm{mg / dL}$ , normal ranges from 108 to $162\mathrm{mg / dL}$ , and high ranges from 260 to $390\mathrm{mg / dL}$ . Results are summarized below. {4} | Low control solution | | | | | | --- | --- | --- | --- | --- | | meter | TD-4207 | TD-4209 | TD-4222 | TD-4225 | | Mean | 78.48 | 78.24 | 75.44 | 75.64 | | SD | 2.16 | 1.48 | 1.92 | 1.82 | | CV(%) | 2.76 | 1.89 | 2.54 | 2.41 | | Normal control solution | | | | | | --- | --- | --- | --- | --- | | meter | TD-4207 | TD-4209 | TD-4222 | TD-4225 | | Mean | 136.16 | 133.20 | 134.12 | 133.36 | | SD | 3.98 | 3.86 | 4.78 | 4.28 | | CV(%) | 2.92 | 2.90 | 3.56 | 3.21 | | High control solution | | | | | | --- | --- | --- | --- | --- | | meter | TD-4207 | TD-4209 | TD-4222 | TD-4225 | | Mean | 326.96 | 323.20 | 328.12 | 327.96 | | SD | 7.61 | 6.49 | 9.20 | 7.24 | | CV(%) | 2.33 | 2.01 | 2.80 | 2.21 | b. Linearity/assay reportable range: The linearity of the device was demonstrated by comparing 120 prepared whole blood samples on the ACHTUNG TD-4207 Blood Glucose Monitoring System and a glucose reference method (YSI-2300) in k061181. The samples ranged in concentration from a low of approximately $23\mathrm{mg / dL}$ to a high of approximately 605 $\mathrm{mg / dL}$ (from YSI). Linear regression of the comparison data yielded the following relationship: ACHTUNG = (1.0052x YSI-2300) + 1.7264, r = 0.9896. The reportable range of the ACHTUNG TD-4207 Blood Glucose Monitoring System 20 - 600 mg/dL. c. Traceability, Stability, Expected values (controls, calibrators, or methods): The controls supplied with this device were previously cleared under k012430. The sponsor has shown traceability of the meter to a laboratory analyzer. d. Detection limit: $20\mathrm{mg / dL}$ . The sponsor supported this level with a linearity study (above). e. Analytical specificity: The specificity of the device was assessed in k041107. Given that the technology has not changed from k041107, the sponsor claims that elevated blood triglycerides and the following substances do not affect results: acetaminophen, dopa, methyldopa, L-dopa and tolbutamide occurring in expected blood concentrations. Due to the change in read time of the glucose oxidase reaction, the sponsor provided reducing substances studies along with an altitude study for the new strips. Reducing substances such as uric acid and ascorbic acid occurring in expected blood concentrations (0-3 mg/dL and 5-20 mg/dL respectively) was shown to be equivalent {5} to the predicate strip. The altitude study also showed that the percentage of mean difference between strips is within the acceptable range (from ISO15197, minimum acceptance accuracy: within $\pm 20\%$ when glucose concentration $\geq 75\mathrm{mg / dL}$ ) indicating that the new test strips are equivalent to the predicate strips at the same altitude (up to 10,744 feet). f. Assay cut-off: Not Applicable. # 2. Comparison studies: a. Method comparison with predicate device: The sponsor demonstrated in k061181 that the ACHTUNG TD-4207 Blood Glucose Monitoring System (as they share the same fundamental technology as the Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 meters) for finger stick is equivalent to a standard method (YSI-2300) and can be used on alternate sites (specifically the capillary blood from finger compared to the palm, forearm, upper arm, calf, and thigh). The sponsor assessed accuracy in k061181 by having 120 patients for the standard method compared to finger stick and 100 patients for each alternative anatomical testing site test the meter. For the finger stick against the standard method, samples ranged as follows: $10\%$ of samples were $20 - 50\mathrm{mg / dL}$ , $35\%$ of samples were $51 - 110\mathrm{mg / dL}$ , $30\%$ of samples were $111 - 150\mathrm{mg / dL}$ , $10\%$ of samples were $151 - 250\mathrm{mg / dL}$ , $10\%$ of samples were $251 - 400\mathrm{mg / dL}$ , and $5\%$ of samples were $401 - 600\mathrm{mg / dL}$ . For the AST sites, the samples ranged as follows: $40\%$ of samples were $51 - 110\mathrm{mg / dL}$ , $30\%$ of samples were $111 - 150\mathrm{mg / dL}$ , $20\%$ of samples were $151 - 250\mathrm{mg / dL}$ , and $10\%$ of samples were $251 - 400\mathrm{mg / dL}$ . All patients blood glucose levels were in a steady state for these studies. The studies from k061181 are summarized below: | | YSI vs. Finger | Finger vs. Palm | Finger vs. Forearm | Finger vs. Upper arm | Finger vs. Calf | Finger vs. Thigh | | --- | --- | --- | --- | --- | --- | --- | | N | 120 | 100 | 100 | 100 | 100 | 100 | | Slope | 1.0052 | 1.0057 | 0.8716 | 0.9102 | 0.9173 | 0.9842 | | Intercept | 1.7264 | 1.8744 | 9.1884 | 6.2178 | 5.3413 | 0.7822 | | r | 0.9948 | 0.9889 | 0.9839 | 0.9822 | 0.9822 | 0.9805 | Finger versus YSI and each site met the ISO 15197 requirement of ninety-five percent (95%) of the individual glucose results falling within $\pm 15\mathrm{mg / dL}$ of the results of the manufacturer's measurement procedure at glucose concentrations for samples $&lt; 75\mathrm{mg / dL}$ and within $\pm 20\%$ at glucose concentrations $\geq 75\mathrm{mg / dL}$ . These results from k061181 are summarized in the table below. | Site | Finger | Palm | Forearm | Upper arm | Calf | Thigh | | --- | --- | --- | --- | --- | --- | --- | | N | 120 | 100 | 100 | 100 | 100 | 100 | | Percentage That Met ISO Requirement | 100% (120/120) | 99% (99/100) | 100% (100/100) | 98% (98/100) | 98% (98/100) | 98% (98/100) | {6} The sponsor includes in the labeling for each meter instructions concerning conditions under which AST can and cannot be used. To ensure that the meters perform similarly to the predicate device, the Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 meters were compared against a standard laboratory method (YSI-2300). This accuracy test was performed using fresh capillary whole blood samples from the finger with a range of $28 - 571\mathrm{mg / dL}$ . Results are summarized below. | Comparison | N | Slope and y-intercept | r | | --- | --- | --- | --- | | TD-4207 vs. YSI-2300 | 120 | y=0.999+2.853 | 0.989 | | TD-4209 vs. YSI-2300 | 120 | y=0.997x+3.771 | 0.990 | | TD-4222 vs. YSI-2300 | 120 | y=0.995x-1.857 | 0.982 | | TD-4225 vs. YSI-2300 | 120 | y=1.005x-2.820 | 0.984 | Difference distribution for glucose concentration $&lt; 75\mathrm{mg / dL}$ | | Difference within ±5mg/dL | Difference within ±10mg/dL | Difference within ±15mg/dL | | --- | --- | --- | --- | | TD-4207 | 12/21(57%) | 20/21(95%) | 21/21(100%) | | TD-4209 | 13/21(62%) | 21/21(100%) | 21/21(100%) | | TD-4222 | 13/21(62%) | 18/21(86%) | 21/21(100%) | | TD-4225 | 10/21(48%) | 15/21(71%) | 21/21(100%) | 100% of the individual difference is within $\pm 15\mathrm{mg / dL}$ when glucose concentration is $&lt; 75\mathrm{mg / dL}$ . Difference distribution for glucose concentration $\geq 75\mathrm{mg / dL}$ | | Difference within ±5 % | Difference within ±10 % | Difference within ±15 % | Difference within ±20 % | | --- | --- | --- | --- | --- | | TD-4207 | 41/99(41%) | 77/99(78%) | 92/99(93%) | 97/99(98%) | | TD-4209 | 47/99(47%) | 74/99(75%) | 91/99(92%) | 95/99(96%) | | TD-4222 | 45/99(45%) | 67/99(68%) | 93/99(94%) | 95/99(96%) | | TD-4225 | 43/99(43%) | 71/99(71%) | 89/99(90%) | 95/99(96%) | 96% of the individual difference is within $\pm 20\%$ when glucose concentration is $\geq 75\mathrm{mg / dL}$ . Each meter met the ISO 15197 requirement of ninety-five percent (95%) of the individual glucose results falling within $\pm 15\mathrm{mg / dL}$ of the results of the manufacturer's measurement procedure at glucose concentrations for samples $&lt; 75$ mg/dL and within $\pm 20\%$ at glucose concentrations $\geq 75\mathrm{mg / dL}$ . # b. Matrix comparison: See above: Method comparison with predicate device {7} 8 3. Clinical studies: a. Clinical Sensitivity: Not Applicable. b. Clinical specificity: Not Applicable. c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not Applicable. 5. Expected values/Reference range: The sponsor included the following Expected Values for normal glucose levels in their strip labeling: | Status | Range (mg/dL) | Range (mmol/L) | | --- | --- | --- | | Before meals | 70-110 | 3.9-6.1 | | 2 hours after meals | <120 | <6.7 | Source: Krall, L.P. and Beaser, R.S.: Joslin Diabetes Manual. Philadelphia: Lea and Febiger (1989), 138. N. Instrument Name: Easy Check TD-4209, Clever Chek TD-4222, Clever Chek TD-4225 and Clever Chek TD-4226 Blood Glucose Monitoring Systems O. System Descriptions: 1. Modes of Operation: Each test strip is single use and must be replaced with a new strip for additional readings. 2. Software: FDA reviewed applicant’s Hazard Analysis and software development processes for this line of product types in k041107. See k041107 for more information. 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. 4. Specimen Sampling and Handling: This device is intended to be used with capillary whole blood from the finger, the palm, the forearm, the upper-arm, the calf, and the thigh only. Since the whole blood sample is applied directly to the test strip there are no special handling or storage issues. 5. Calibration: A code strip is provided with each batch of test strips to calibrate the meter for that batch. No further calibrations are required of the user. 6. Quality Control: The sponsor is providing a high and low glucose control solution with this device. When a test strip is inserted into the meter, the control mode can be activated. This prevents control results from being stored in the internal memory. An acceptable range for each {8} control level is printed on the test strip vial label. The user is referred to the troubleshooting section of the owner's manual if control results fall outside these ranges. P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...