The POPS!® one Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood taken from the finger. It is intended to be used by people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. The POPS!® one Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is not intended for the diagnosis of, or screening of diabetes. It is not intended for use on neonates.
Device Story
System comprises POPS! one meter, sterile sensor modules (lancet/strip), and mobile app. User obtains capillary blood from fingertip; applies to sensor module; meter performs electrochemical glucose dehydrogenase measurement; transmits results via encrypted Bluetooth to mobile app. App displays results, trends, and error messages. Intended for single-patient home use; aids diabetes management. Meter automatically codes; requires 0.5 μL sample. Benefits include integrated mobile display and trend monitoring. Not for clinical/assisted-use settings due to infection control risks.
Clinical Evidence
User study with 362 participants demonstrated accuracy against YSI 2300 comparator. For glucose ≥75 mg/dL, 99.7% of results were within ±20% of YSI values. For glucose <75 mg/dL, 100% were within ±15 mg/dL. Additional bench testing confirmed precision (CV 2.7-4.1%), linearity (R² ≥0.997), and performance across hematocrit (20-60%), altitude (up to 15,000 ft), and temperature/humidity ranges. No clinical data on diagnostic sensitivity/specificity provided as device is for monitoring only.
Indicated for people with diabetes mellitus (excluding neonates) for quantitative measurement of glucose in fresh capillary whole blood from the finger to monitor diabetes control program effectiveness. For single-patient, home use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
TD-4277 Blood Glucose Monitoring System (k100322)
Related Devices
K181316 — POGO Automatic Blood Glucose Monitoring System · Intuity Medical, Inc. · Jun 14, 2018
K152493 — POGO Automatic Blood Glucose Monitoring System · Intuity Medical, Inc. · Apr 25, 2016
K024347 — EXPRESS BLOOD GLUCOSE MONITORING SYSTEM · Hypoguard USA, Inc. · Mar 28, 2003
K150817 — Dario Blood Glucose Monitoring System · Labstyle Innovations, Ltd. · Dec 21, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 30, 2018
TaiDoc Technology Corporation Anne Kuo Regulatory Affairs Senior Specialist 6F. No.127. Wugong 2nd Rd., Wugu District New Taipei City, 24888 Taiwan
Re: K181588
Trade/Device Name: POPS! one Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: October 26, 2018 Received: October 31, 2018
Dear Anne Kuo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K181588
Device Name
POPS!® one Blood Glucose Monitoring System
#### Indications for Use (Describe)
The POPS!® one Blood Glucose Monitoring System is comprised of the POPS!® one blood glucose meter, the POPS!® one blood glucose sensor modules, and the POPS!® mobile application as the display component.
The POPS!® one Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood taken from the finger. It is intended to be used by people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program.
The POPS!® one Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is not intended for the diagnosis of, or screening of diabetes. It is not intended for use on neonates.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☐</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☑</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for TaiDoc, a company that specializes in medical devices. The logo consists of the word "TaiDoc" in gray, followed by three red hearts. Below the hearts is the website address "www.taidoc.com" in white text on a gray background. The logo is simple and clean, and the use of red hearts suggests a focus on health and well-being.
# 510(k) Summary
#### Assigned 510(k) number: K181588
In accordance with the requirements of 21 CFR 807.92, this summary is being provided to serve as basis for the substantial equivalence determination.
#### Submitter information
| Manufacturer | TaiDoc Technology Corporation |
|--------------------------------|-------------------------------------------------------------------------|
| Address | 6F, No. 127, Wugong 2nd Rd. Wugu Dist.<br>New Taipei City, Taiwan 24888 |
| Establishment Registration No. | 3004145393 |
| Date Prepared | November 30, 2018 |
| Correspondent | TaiDoc Technology Corporation |
| Correspondent Contact | Anne Kuo |
| Title | Regulatory Affairs Senior Specialist |
| Phone | +886-2-6625-8188 #1195 |
| E-mail | ra.cert@taidoc.com.tw<br>anne.kuo@taidoc.com.tw |
| Correspondent | POPS! Diabetes Care, Inc. |
| Correspondent Contact | Jennifer Englund |
| Title | Vice President, Clinical, Regulatory & Quality |
| Phone | 612-802-7606 |
| E-mail | jennifer.englund@popsdiabetes.com |
#### Proposed Device Information
| Proprietary name | POPS!® one Blood Glucose Monitoring System |
|----------------------|--------------------------------------------------|
| Common name | Blood Glucose Monitoring System |
| Product code | NBW, Blood Glucose Test System, Over-the-Counter |
| Classification panel | Clinical chemistry |
| Classification | Class II |
| Regulation citation | 21 CFR §862.1345, Glucose test system |
CONFIDENTIAL
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray letters, followed by three red hearts. Below the hearts is the website address "www.taidoc.com" in white letters on a gray background. The logo is simple and clean, with the hearts adding a touch of warmth and care.
泰博科技股份有限公司 TaiDoc Technology Corp. 新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist., New Taipei City 24888. Taiwan
#### Predicate Device Information
| Manufacturer | TaiDoc Technology Corporation |
|------------------|-----------------------------------------|
| Proprietary Name | TD-4277 Blood Glucose Monitoring System |
| Common Name | Blood Glucose Monitoring System |
| 510(k) Number | K100322 |
#### Intended use
The POPS!" one Blood Glucose Monitoring System is comprised of the POPS!" one blood glucose meter, the POPS!® one blood glucose sensor modules, and the POPS!® mobile application as the display component.
The POPS!" one Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood taken from the finger. It is intended to be used by people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program.
The POPS!® one Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is not intended for the diagnosis of, or screening of diabetes. It is not intended for use on neonates.
#### Device Description
The POPS!® one Blood Glucose Monitoring System is designed to make blood glucose testing and diabetes management easier and more convenient. The traditional test kit is replaced by a low-profile meter that contains everything needed for testing and requires no assembly. The meter pairs with a mobile app installed on the user's phone via Bluetooth, allowing the meter to perform glucose testing and send the results to the app where they are communicated to the user and automatically stored.
The POPS!® one Blood Glucose Monitoring System contains the following:
- . POPS!® one blood glucose meter works directly with the POPS!® app installed on the user's mobile phone. The meter performs a blood glucose test and sends the result to the POPS!" app.
- . POPS!® one blood glucose sensor module is individually packed and sterile. It contains 3 test sites with separate foil covers; each containing a lancet and test strip. The user inserts the sensor module in the meter and replaces a used sensor module with a new once all 3 test sites have
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray letters, with three red hearts below it. Below the hearts is the website address "www.taidoc.com" in white letters on a gray background.
been used. The sensor module not only provides all components for performing a glucose test to the user in a convenient, self-contained format but it also provides a new lancet for each test; preventing lancet re-use.
- . POPS!® mobile application acts as the primary interface with users; enabling them to test, communicating blood glucose results and messages, and allowing them to monitor trends on their phone.
## Test principle
The blood glucose testing is based on the measurement of electrical current generated by the reaction of glucose with the reagent of the strip within the Sensor Module. The system measures the current, calculates the blood glucose level, and displays the result. The strength of the current produced by the reaction depends on the amount of glucose in the blood sample.
## Summary of Technological Characteristics and Comparison to the Predicate
The POPS!" one Blood Glucose Monitoring System is substantially equivalent to the predicate device, both in terms of intended use and technological characteristics. The primary differences between the proposed and predicate device are design changes to simplify the glucose testing experience and allow communication of results and messages to the user to occur via the POPS!® app.
The similarities and differences between the predicate and proposed devices are summarized in Table 1 and 2 below.
| Characteristic | Predicate device<br>TD-4277 | Proposed device<br>POPS!® one |
|---------------------|--------------------------------------|--------------------------------------|
| Operation principle | Electrochemical biosensor technology | Electrochemical biosensor technology |
| Detection method | Amperometric glucose biosensor | Amperometric glucose biosensor |
| Code calibration | No coding required | No coding required |
| Enzyme | Glucose dehydrogenase | Glucose dehydrogenase |
| Strip reaction time | 6 seconds | 6 seconds |
| Sample volume | 0.5 µL | 0.5 µL |
Table 1: Similarities between the Predicate and Proposed Devices
CONFIDENTIAL
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "TaiDoc" in gray font above three red hearts. The hearts are arranged in a horizontal row, with one heart directly below each of the first three letters of the word "TaiDoc". The hearts are all the same size and shape.
泰博科技股份有限公司 TaiDoc Technology Corp.
新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist., New Taipei City 24888, Taiwan
Tel : +886-2-6625-8188 Fax : +886-2-6625-0288
www.taidoc.com
| Characteristic | Predicate device<br>TD-4277 | Proposed device<br>POPS!® one |
|--------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Measurement range | 20~600 mg/dL | 20~600 mg/dL |
| Hematocrit range | 20~60% | 20~60% |
| Measurement unit | mg/dL | mg/dL |
| Measurement modes | Test Mode<br>AC (before meal)<br>PC (after meal)<br>Gen (not specified)<br>QC (quality control) | Test Mode<br>AC (before meal)<br>PC (after meal)<br>Gen (not specified)<br>QC (quality control) |
| Strip indication light | Yes | Yes |
| Meter storage/<br>transportation condition | -4 °F to 140°F, < 95% R.H. | -4 °F to 140°F, < 95% R.H. |
| Strip storage/<br>transportation condition | 35.6ºF to 89.6ºF, < 85% R.H. | 35.6ºF to 89.6ºF, < 85% R.H. |
Table 2: Differences between the Predicate and Proposed Devices
| Characteristic | Predicate device<br>TD-4277 | Proposed device<br>POPS!® one |
|---------------------|-----------------------------|-------------------------------|
| Physical Appearance | Image: TD-4277 device | Image: POPS! one device |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray, with three red hearts underneath it. The hearts are evenly spaced and aligned horizontally.
www.taidoc.com
泰博科技股份有限公司 TaiDoc Technology Corp.
新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist., New Taipei City 24888, Taiwan
Tel : +886-2-6625-8188 Fax : +886-2-6625-0288
| Characteristic | Predicate device<br>TD-4277 | Proposed device<br>POPS!® one |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | TD-4277 Blood Glucose Monitoring<br>System is intended for use in the<br>quantitative measurement of glucose<br>in fresh capillary whole blood from<br>the finger. It is intended for use by<br>healthcare professionals and people<br>with diabetes mellitus at home as an<br>aid in monitoring the effectiveness of<br>diabetes control program. It is not<br>intended for the diagnosis of or<br>screening for diabetes mellitus, and is<br>not intended for use on neonates.<br>Professionals may use the test strips<br>to test capillary and venous blood<br>samples, but lay user may not test<br>venous blood samples. | The POPS!® one Blood Glucose<br>Monitoring System is intended for<br>use outside the body (in vitro<br>diagnostic use) in the quantitative<br>measurement of glucose in fresh<br>capillary whole blood taken from the<br>finger. It is intended to be used by<br>people with diabetes mellitus at home<br>as an aid in monitoring the<br>effectiveness of a diabetes control<br>program.<br>The POPS!® one Blood Glucose<br>Monitoring System is intended to be<br>used by a single patient and should<br>not be shared. It is not intended for<br>the diagnosis of, or screening of<br>diabetes. It is not intended for use on<br>neonates. |
| Mode of operation | User initiates test on meter, inserts<br>test strip into meter, lances finger<br>with lancing device, and places blood<br>sample on test strip. If test result<br>tracking is desired, user must do so<br>manually. | User initiates test within app, lances<br>finger with integrated lancet and<br>places blood sample on the test strip.<br>Recording of test results is done<br>wirelessly and automatically in the<br>app. |
| Characteristic | Predicate device<br>TD-4277 | Proposed device<br>POPS!® one |
| Strip preparation | Strips are stored in a<br>humidity-controlled vial (25-50 strips per vial), user removes one strip from vial and inserts in meter for testing. | Individual strips are sealed in the sensor module with a foil seal to control humidity. User peels back the foil seal for the test site prior to testing, keeping the other strips covered/secure until use. This improves storage integrity of strips and minimizes user interaction with strips. |
| Lancet | Lancet in lancing device is changed out by user resulting in frequent re-use of lancets (increased infection risk and callous formation). | A new sterile lancet is provided in the sensor module with each test site, so users have a new lancet for each test. |
| Software design | Meter software controls the measurement and calculation of blood glucose and displays result on the meter. | Meter software controls the measurement and calculation of blood glucose and transmits the information to the POPS!® app for display and storage of results. |
| Meter Size (mm) | 96 (L) x 61 (W) x 26 (H) | 43 (L) x 39 (W) x 24 (H) |
| Meter Weight (g) | 67.2 g | 13 g |
| Power source | 2 x 1.5V AAA batteries | 1 CR2032 battery |
| User interface | LCD and buttons | Mobile application |
| Display of test results | LCD screen | Mobile application |
| Transmission function | USB | Encrypted and secure Bluetooth |
| Error messages | Error number displayed on LCD screen | Error number translated into text and displayed on App |
| PCB layout | circuit designed for one test site | circuit designed for three test sites |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "TaiDoc" in gray font above three red hearts. The hearts are evenly spaced and aligned horizontally. The word "TaiDoc" is in a sans-serif font and is larger than the hearts. The overall impression is one of health and well-being.
泰博科技股份有限公司 TaiDoc Technology Corp.
新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist., New Taipei City 24888, Taiwan
Tel : +886-2-6625-8188 Fax : +886-2-6625-0288
www.taidoc.com
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for TaiDoc. The word "TaiDoc" is written in gray at the top of the image. Below the word are three red hearts. At the bottom of the image is the website address, "www.taidoc.com", written in white on a gray background.
#### Summary of Testing
Non-clinical and clinical studies were conducted to test, verify and validate the performance of the proposed device according to FDA Guidance issued on October 11, 2016: Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use. Results from these studies show that all performance criteria were met.
Non-Clinical Testing Summary: Design verification and validation testing was performed to ensure that the POPS!® one System met design specifications and requirements. Testing activities included electrical/mechanical safety tests, functional performance tests (precision, linearity, interference, flex studies) as well as disinfection, cleaning, robustness, sterilization, shelf life and transit studies. Software validation was performed for this moderate level of concern device per FDA Guidance Content of Premarket Submissions for Software Contained in Medical Devices.
Clinical Testing Summary: A user evaluation confirmed the system accuracy, operation according to design, and ease of use to support the intended use as described in the proposed labeling.
#### Conclusion
Based on the information provided in this submission, the POPS!® one Blood Glucose Monitoring System has been shown to be substantially equivalent.to the TD-4277 Blood Glucose Monitoring System.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.