K023219 · Becton, Dickinson & CO · NBW · Dec 6, 2002 · Clinical Chemistry
Device Facts
Record ID
K023219
Device Name
BECTON DICKINSON DIABETES SOFTWARE
Applicant
Becton, Dickinson & CO
Product Code
NBW · Clinical Chemistry
Decision Date
Dec 6, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Intended Use
The BDTM Diabetes Software is intended for use as a data management tool for acceptance, transfer, display, storage, processing (e.g. averaging), reporting, and printing of patient blood glucose monitoring data. The device is intended for use with a BD Blood Glucose Monitoring System only.
Device Story
BD Diabetes Software functions as a data management system for blood glucose monitoring. Inputs include patient blood glucose data transferred from BD Blood Glucose Monitors via serial/USB cable or via manual entry. The software processes data through averaging and storage, producing reports and printouts for clinical review. Used in home and professional settings; operated by patients or healthcare providers. The software aids in diabetes management by organizing and displaying glucose trends, facilitating clinical decision-making based on historical data.
Clinical Evidence
No clinical data. Performance summary based on system, hardware, software, packaging, and electrical safety verification and validation testing.
Technological Characteristics
Microsoft Windows-based software application (Windows 95/NT 4.0 or later). Operates on Intel-compatible PCs. Connectivity via BD Interface serial or USB cable. Data management functions include averaging, storage, and reporting.
Indications for Use
Indicated for use as a data management tool for patients using BD Blood Glucose Monitoring Systems to accept, transfer, display, store, process, report, and print blood glucose monitoring data.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
WinGlucofacts (Bayer Diagnostics)
Related Devices
K070593 — TRACKRECORD DATA MANAGEMENT SOFTWARE · Home Diagnostics, Inc. · May 7, 2007
K984527 — IN TOUCH DIABETES MANAGEMENT SOFTWARE · Lifescan, Inc. · Apr 29, 1999
K080587 — EZ MANAGER MAX DIABETES MANAGEMENT SOFTWARE · Animas Corp. · May 15, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
K0233219
# 510K Summary of Safety and Effectiveness
#### 1. Submitted By:
John Schalago Manager, Regulatory Affairs
Becton Dickinson Consumer Healthcare 1 Becton Drive Franklin Lakes, NJ 07417-1883
Phone: 201-847-5663 201-848-0457 Fax:
#### 2. Device Name:
BDTM Diabetes Software Trade Name: Common Names: Blood Glucose Meter and Data Management System Classification Name: Glucose test system Classification: 862.1345
#### 3. Predicate Device:
WinGlucofacts
Manufactured by: Bayer Diagnostics
#### Device Description: 4.
BD™ Diabetes Software is data management software intended for use with BD Blood Glucose Monitoring Systems only. The software is designed to operate on a Windows/Intel compatible platform, and is available in home use and professional use options. Data may be downloaded from a BD Blood Glucose Monitor or manually entered.
BD™ Diabetes Software provides data management tools for acceptance, transfer, display, storage, processing (e.g. averaging), reporting, and printing of patient blood glucose monitoring data. In addition, the BDTM Diabetes Software has an online help feature.
{1}------------------------------------------------
## 510K Summary of Safety and Effectiveness (Continued)
### ડ. Intended Use:
The BDTM Diabetes Software is intended for use as a data management tool for acceptance, transfer, display, storage, processing (e.g. averaging), reporting, and printing of patient blood glucose monitoring data.
The device is intended for use with a BD Blood Glucose Monitoring System only.
#### Technological Characteristics: 6.
The BDTM Diabetes Software is a Microsoft Windows based software application for diabetes data management. The software allows for data transfer from a BD Blood Glucose Monitor via a BD Interface serial or USB cable. BD™ Diabetes Software is designed to operate an Intel compatible PC with Microsoft Windows 95 or later, and Microsoft NT version 4.0 or later operating system.
#### 7. Performance Summary:
BD™ Diabetes Software testing consisted of system, hardware, software, packaging, electrical safety, and end-user evaluation. System verification and validation activities demonstrate that BD™ Diabetes Software will perform as intended when used in accordance with device labeling. Electrical components have been tested and certified to conform to appropriate voluntary and mandatory standards.
The term "substantial equivalence" as used in this 510(k) Notification is limited to the definition of substantial equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E, under which a device can be marketed without pre-approval or classification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US patent Laws or their application by the courts.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA", which is arranged in a circular fashion around the left side of the logo. The logo is black and white.
Food and Drug Administratio 2098 Gaither Road Rockville MD 20850
Mr. John A. Schalago MS. RAC Regulatory Affairs Manager Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417
k023219 Re:
Trade/Device Name: Becton Dickinson Diabetes Software Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW; JQP Dated: September 18, 2002 Received: September 26, 2002
Dear Mr. Schalago:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{3}------------------------------------------------
### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Attachment XI .A.1
Page l of
510(k) Number (if known):
Device Name: Becton Dickinson Diabetes Software
Indications For Use:
The BDTM Diabetes Software is intended for use as a data management tool for acceptance, transfer, display, storage, processing (e.g. averaging), reporting, and printing of patient blood glucose monitoring data.
The device is intended for use with the BD Blood Glucose Monitoring Systems only.
im
(Division Sign-Off)
vision of Clinical Laboratory Devices
510(k) Number. X02309
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
Rx
RTC_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.