DIACHECK SMART/SUPERIOR BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Delbio Incorporation
Product Code
NBW · Clinical Chemistry
Decision Date
Feb 8, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The DiaCheck Superior Blood Glucose Monitoring System is intended for the rne Black our of glucose in fresh capillary whole blood samples from the fingertips by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The DiaCheck Superior Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. This system is intended for self testing outside the body (in vitro diagnostic use). This system should not be used for the screening or diagnosis of diabetes or for testing neonates.
Device Story
DiaCheck Superior Blood Glucose Monitoring System measures glucose levels in fresh capillary whole blood samples obtained from fingertips. Device intended for home use by individuals with diabetes to monitor glycemic control. System comprises blood glucose meter, test strips, and control solution. User applies blood sample to test strip inserted into meter; meter performs electrochemical measurement of glucose concentration. Results displayed on meter screen to assist users in managing diabetes. System designed for single-patient use; not for clinical screening, diagnosis, or neonatal testing. Control solution used to verify system accuracy and proper function of meter and strips.
Clinical Evidence
Bench testing only. Robustness studies demonstrated no change in performance or external materials after 260 cleaning and disinfection cycles using Super Sani-Cloth Germicidal Wipes (EPA #9480-4), simulating 5 years of single-patient use.
Technological Characteristics
Electrochemical glucose monitoring system; utilizes capillary whole blood samples; intended for in vitro diagnostic use.
Indications for Use
Indicated for single-patient use for the monitoring of blood glucose levels.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K103329 — DIATRUE PLUS BLOOD GLUCOSE MONITORING SYSTEM; DIATRUE PLUS BLOOD GLUCOSE TEST STRIP; DIATRUE GLUCOSE CONTROL SOLUTION · Delbio Incorporation · Mar 13, 2012
K122110 — ACON ON CALL VIVID PAL BLOOD GLUCOSE MONITORING SYSTEM · ACON Laboratories, Inc. · Nov 21, 2012
K031388 — MODIFICATION TO HYPOGUARD ADVANCE BLOOD GLUCOSE MONITORING SYSTEM · Hypoguard USA, Inc. · May 22, 2003
K192220 — ReliOn Premier Classic Blood Glucose Monitoring System · I-Sens, Inc. · Dec 13, 2019
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
OIVD Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER: k113343
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary) for the DiaCheck Smart and DiaCheck Superior Blood Glucose Monitoring Systems:
1. The name and 510(k) number of SUBMITTER'S previously cleared device: DiaCheck Premium Blood Glucose Monitoring System; k100806.
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials. (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for:
- Addition of one button for the Superior model and elimination of one button for the Smart model.
- Addition of strip ejection button to both Smart and Superior models
- Decreased memory size of Smart model
- Change in batteries from 2 AA to one 3V CR-2032 for both Smart and Superior models
- Change in meter size (in mm) from 90(L)x60(W)x25(H) to 96(L)x35(W)x20(H) for the Smart model and to 83(L)x61(W)x23(H) for the Superior model
- Software modification to both Smart and Superior models to modify the icon display on the LCD
- Change in coding procedure for the Smart model, from selection of correct code via meter display to insertion of code strip into port in the meter
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, and physical characteristics.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the
{1}
particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
The devices are intended for single-patient use only. Super Sani-Cloth Germicidal Wipes with EPA registration # 9480-4 were validated demonstrating complete inactivation of live virus for use with the meters. Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meters after 260 cleaning and disinfection cycles designed to simulate 5 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.