K070017 · Polymer Technology Systems, Inc. · NBW · Jan 31, 2007 · Clinical Chemistry
Device Facts
Record ID
K070017
Device Name
PTS PANELS METABOLIC CHEMISTRY PANEL TEST STRIPS
Applicant
Polymer Technology Systems, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Jan 31, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Indications for Use
The PTS PANELS Metabolic Chemistry Panel Test Strips are intended for the quantitative determination of glucose, HDL cholesterol, and triglycerides in whole blood. The PTS PANELS Metabolic Chemistry Panel Test Strips are intended for use with the PTS PANELS readers. The PTS PANELS Metabolic Chemistry Panel Test Strips are intended for use by healthcare professionals and in point-of-care settings.
Device Story
PTS PANELS Metabolic Chemistry Panel Test Strips utilize dry chemistry technology to measure glucose, HDL cholesterol, and triglycerides from fingerstick whole blood samples. The device consolidates three separate tests into a single test strip, reducing total testing time to approximately two minutes. It is designed for use by medical professionals in clinical settings and by individuals in home environments. The test strips are used with a PTS reflectance photometer, which reads the colorimetric enzymatic reaction results. Calibration data is stored on an EEPROM chip packaged with the strips. The output provides quantitative measurements of glucose, HDL, and triglycerides, assisting healthcare providers in the diagnosis and management of metabolic and lipid-related disorders.
Clinical Evidence
Bench testing only. Performance evaluated through precision studies and comparative analysis against standard laboratory chemistry analyzers to ensure accuracy of the combined metabolic panel.
Technological Characteristics
Multi-assay test strip integrating glucose, HDL cholesterol, and triglyceride chemistries. Operates with PTS PANELS readers. Design control procedures per 21 CFR 820.30 applied. No change to fundamental scientific technology.
Indications for Use
Indicated for quantitative determination of glucose, HDL cholesterol, and triglycerides in whole blood. Intended for use by healthcare professionals in point-of-care settings for patients requiring metabolic monitoring.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K071507 — PTS PANELS CHOL+HDL+GLU PANEL TEST STRIPS · Polymer Technology Systems, Inc. · Sep 10, 2007
K022898 — PTS PANELS LIPID PANEL TEST STRIPS · Polymer Technology Systems, Inc. · Sep 24, 2002
K060617 — PTS PANELS HDL CHOLESTEROL TEST STRIPS · Polymer Technology Systems, Inc. · Apr 5, 2006
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
OIVD Review Memorandum
To: THE FILE
RE: DOCUMENT NUMBER k070017
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k013068 (PTS BioScanner Glucose Test Strips), k060617 (PTS PANELS HDL Cholesterol Test Strips), k991894 and k000586 (PTS BioScanner Triglycerides Test Strips).
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for combining all three assays into one test strip and changing name to PTS PANELS Metabolic Chemistry Panel Test Strips.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and performance characteristics. Studies include precision and comparison to chemistry test on laboratory analyzer.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.