FORA DIAMOND PRIMA BLOOD GLUCOSE MONITORING SYSTEM, FORA DIAMOND MINI BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Taidoc Technology Corporation
Product Code
NBW · Clinical Chemistry
Decision Date
Feb 6, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
FORA Diamond Prima Blood Glucose Monitoring System: FORA Diamond Prima Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood from the finger. The system is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. This system is intended for single-patient use and should not be shared. The FORA Diamond Prima Blood Glucose Test Strips are for use with the FORA Diamond Prima Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. FORA Diamond Mini Blood Glucose Monitoring System: FORA Diamond Mini Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood from the finger. The system is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. This system is intended for single-patient use and should not be shared. The FORA Diamond Mini Blood Glucose Test Strips are for use with the FORA Diamond Mini Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Device Story
System consists of meter, test strips, and control solutions; measures glucose in fresh capillary whole blood via electrochemical biosensor technology. Input: 0.5 µL capillary blood applied to test strip. Operation: FAD-dependent glucose dehydrogenase (FAD-GDH) catalyzes glucose oxidation; electrons transferred to meter generate electrical current proportional to glucose concentration. Output: Digital glucose concentration readout on meter display. Used at home by patients with diabetes; single-patient use only. Healthcare providers use results to monitor diabetes control effectiveness. Benefits: Provides quantitative glucose data for self-management of diabetes.
Clinical Evidence
Bench testing only. System accuracy performance compared to predicate device demonstrated substantial equivalence.
Technological Characteristics
Electrochemical biosensor using glucose dehydrogenase. System includes meter, test strips, and control solutions. Designed for single-patient use.
Indications for Use
Indicated for people with diabetes mellitus for home self-testing of capillary whole blood glucose. Not for neonates, critically ill, dehydrated, shock, or hyperosmolar patients. Not for alternative site testing.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K113077 — BIOLAND G-423 GLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Aug 3, 2012
K122110 — ACON ON CALL VIVID PAL BLOOD GLUCOSE MONITORING SYSTEM · ACON Laboratories, Inc. · Nov 21, 2012
K120866 — DA01 BLOOD GLUCOSE MONITORING SYSTEM · Delbio Incorporation · Dec 11, 2012
K100322 — TD-4277 BLOOD GLUCOSE MONITORING SYSTEM, MODEL 4277 · Taidoc Technology Corporation · Jun 28, 2010
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY FOR ASSAY AND INTRUMENT TEMPLATE
A. 510(k) Number:
k111890
B. Purpose for Submission:
Clearance of new devices
C. Measurand:
Capillary whole blood glucose from the finger
D. Type of Test:
Whole blood glucose concentration through a quantitative amperometric assay (GDH-FAD)
E. Applicant:
Taidoc Technology Corporation
F. Proprietary and Established Names:
FORA Diamond Prima Blood Glucose Monitoring System
FORA Diamond Mini Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR: 862.1345, Blood Glucose Test System
2. Classification:
Class II
3. Product code:
NBW, LFR
{1}
4. Panel:
75 (clinical chemistry)
H. Intended Use:
1. Intended use(s):
See Indications for Use
2. Indication(s) for use:
FORA Diamond Prima Blood Glucose Monitoring System:
FORA Diamond Prima Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood from the finger. The system is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
This system is intended for single-patient use and should not be shared.
The FORA Diamond Prima Blood Glucose Test Strips are for use with the FORA Diamond Prima Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
FORA Diamond Mini Blood Glucose Monitoring System:
FORA Diamond Mini Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) in the quantitative measurement of glucose in fresh capillary whole blood from the finger. The system is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
This system is intended for single-patient use and should not be shared.
The FORA Diamond Mini Blood Glucose Test Strips are for use with the FORA Diamond Mini Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
3. Special conditions for use statement(s):
- For single person, over the counter use
- Not for use in the screening or diagnosis of diabetes
- Not for use in testing neonates
- Not for use on critically ill patients, dehydrated patients, patients in shock
{2}
or hyperosmolar patients
- Not for alternative site testing
4. Special instrument requirements:
The FORA Diamond Prima Blood Glucose Test Strips are must be used with the FORA Diamond Prima Blood Glucose Meter and the FORA Diamond Mini Blood Glucose Test Strips must be used with the FORA Diamond Mini Blood Glucose Meter.
I. Device Description:
FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System consist of three main components: the meter, test strips and control solutions. These products have been designed, tested, and proven to work together as a system to produce accurate blood glucose test results. Use only FORA test strips and FORA control solutions (cleared under k093724) with the FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System.
The blood glucose detection method and measurement is by an electrochemical biosensor technology using FAD-dependent glucose dehydrogenase (FAD-GDH).
The two blood glucose systems have the same technical components although the FORA Diamond Mini Blood Glucose is smaller in size and does not have the test strip ejector, ketone warning function and does not have the capacity to perform daily averaging of measurements. However, they use the same test strips.
J. Substantial Equivalence Information:
1. Predicate device name(s):
TD-4277 blood glucose monitoring system
2. Predicate 510(k) number(s):
k100322
3. Comparison with predicate:
| Item | TD-4277 Blood Glucose Monitoring System - Predicate | FORA Diamond Prima Blood Glucose Monitoring System | FORA Diamond Mini Blood Glucose Monitoring System |
| --- | --- | --- | --- |
{3}
| Intended Use/Indication s for Use | It is intended to be used for quantitative measurement of glucose in fresh capillary whole blood as an aid to monitor the effectiveness of diabetes control in people with diabetes. | Same | Same |
| --- | --- | --- | --- |
| Sample test time | 5 seconds | Same | Same |
| Detection method | Amperometry | Same | Same |
| Enzyme | Glucose dehydrogenase | Same | Same |
| Calibration Coding | No-coding: Code number is preinstalled and only one code number is assigned. User must choose and insert the correct test trip. | Same | Same |
| Memory | 1000 measurements | 450 measurements | 450 measurements |
| Test range | 20-600 mg/dL | Same | Same |
| Hematocrit range | 20-60% | Same | Same |
| Sample type | Fresh capillary and venous whole blood | Fresh capillary whole blood | Fresh capillary whole blood |
| Sample sites | Fingertip | Same | Same |
| Sample volume | 0.5 uL | Same | Same |
# K. Standard/Guidance Document Referenced (if applicable):
- ISO 14971:2007. Medical devices-Application of risk management to medical devices.
- ISO 15197. In vitro diagnostic test systems. Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
- EN 60601-1-1. Medical electrical equipment, Part 1-1. General requirements for safety. Safety requirements for medical electrical systems.
- EN 60601-1-2:2001 (A1:2006). Medical electrical equipment, Part 1-2. General requirements for basic safety and essential performance. Electromagnetic Compatibility.
- EN 61326-1:2006. Electrical equipment for measurement, control, and laboratory
{4}
use. EMC Requirements. General requirements.
- IEC/EN 61010-2-101:2002. Safety requirements for electrical equipment for measurement, control, and laboratory use, Part 2-101. Particular requirements for in vitro diagnostic (IVD) medical equipment.
L. Test Principle:
FORA Diamond Prima and FORA Diamond Mini Blood Glucose meters, in conjunction with the same test trips, utilize amperometric technology to quantitatively measure the glucose concentration in whole blood samples and in control solutions. A glucose dehydrogenase sensor based on the carbon electrode adopting the amperometric assay utilizes the enzyme glucose dehydrogenase to catalyze the formation of gluconolactone from the oxidation of glucose whereby two electrons are produced. Through the mechanism of the mediator, electrical current is generated and it is proportional to the quantity of glucose in the sample.
The reaction principle of reagent depends on following reaction equation:
$$
\text{Glucose} + 2 \text{ferricyanide} = \text{gluconolactone} + 2 \text{ferrocyanide}
$$
$$
\text{electrode}
$$
$$
\text{ferrocyanide} = \text{ferricyanide} + \text{e-}
$$
GDH: glucose dehydrogenase
FAD: flavin adenine dinucleotide
The glucose biosensors recognize the glucose present in whole blood or control solutions by virtue of the specificity of the enzyme FAD dependent glucose dehydrogenase (GDH) present on the test strip. The electrons liberated by this reaction are transferred via a co-factor and mediator to the meter where they are read as a small electrical current. The magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The sponsor performed precision studies in accordance with ISO 15197 and CLSI EP-5A. Fresh venous whole blood adjusted to 5 glucose levels (hematocrit 45%) were used for within-day precision studies. Each concentration was tested 10 times on 10 meters, using 3 test strip lots, (100 total tests divided between 3 strip lots per blood glucose level). Results for each meter are summarized below:
5
{5}
Results of repeatability precision of FORA Diamond Prima
| | Interval 1 (30-50 mg/dL) | | | Interval 2 (51-110 mg/dL) | | | Interval 3(111-150 mg/dL) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot | PR110113 | TR100716 | TR100804 | PR110113 | TR100716 | TR100804 | PR110113 | TR100716 | TR100804 |
| Mean | 43.2 | 42.7 | 42.0 | 86.0 | 84.9 | 85.5 | 138.3 | 135.9 | 137.3 |
| SD | 1.51 | 1.53 | 1.60 | 2.95 | 2.98 | 2.89 | 4.81 | 4.37 | 4.43 |
| CV | 3.50% | 3.58% | 3.82% | 3.44% | 3.50% | 3.38% | 3.48% | 3.21% | 3.22% |
| Overall mean | | 42.6 | | | 85.5 | | | 137.2 | |
| SD | | 1.62 | | | 2.94 | | | 4.58 | |
| CV | | 3.81% | | | 3.43% | | | 3.34% | |
| | Interval 4 (151-250mg/dL) | | | Interval 5 (251-400 mg/dL) | | |
| --- | --- | --- | --- | --- | --- | --- |
| Lot | PR110113 | TR100716 | TR100804 | PR110113 | TR100716 | TR100804 |
| Mean | 225.1 | 223.0 | 219.7 | 358.6 | 363.4 | 364.2 |
| SD | 7.30 | 6.67 | 6.61 | 10.86 | 10.77 | 11.21 |
| CV | 3.24% | 2.99% | 3.01% | 3.03% | 2.96% | 3.08% |
| Overall mean | | 222.3 | | | 362.3 | |
| SD | | 7.15 | | | 11.14 | |
| CV | | 3.22% | | | 3.07% | |
Results of repeatability precision of FORA Diamond Mini
| | Interval 1 (30-50 mg/dL) | | | Interval 2 (51-110 mg/dL) | | | Interval 3(111-150 mg/dL) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot | PR110113 | TR100716 | TR100804 | PR110113 | TR100716 | TR100804 | PR110113 | TR100716 | TR100804 |
| Mean | 41.9 | 41.5 | 41.1 | 76.0 | 75.0 | 75.8 | 122.6 | 121.8 | 123.2 |
| SD | 1.61 | 1.54 | 1.43 | 2.51 | 2.25 | 2.25 | 3.94 | 3.86 | 3.80 |
| CV | 3.84% | 3.72% | 3.48% | 3.30% | 3.00% | 2.97% | 3.21% | 3.17% | 3.08% |
| Overall mean | | 41.4 | | | 75.6 | | | 122.6 | |
| SD | | 1.54 | | | 2.35 | | | 3.86 | |
| CV | | 3.72% | | | 3.10% | | | 3.15% | |
Table 3. (continued). Results of repeatability precision of FORA Diamond Mini
{6}
7
| | Interval 4 (151-250mg/dL) | | | Interval 5 (251-400 mg/dL) | | |
| --- | --- | --- | --- | --- | --- | --- |
| Lot | PR110113 | TR100716 | TR100804 | PR110113 | TR100716 | TR100804 |
| Mean | 203.9 | 204.1 | 203.0 | 332.2 | 333.4 | 332.0 |
| SD | 6.60 | 5.98 | 6.26 | 10.41 | 10.19 | 10.58 |
| CV | 3.24% | 2.93% | 3.08% | 3.13% | 3.06% | 3.19% |
| Overall mean | | 203.6 | | | 332.5 | |
| SD | | 6.24 | | | 10.33 | |
| CV | | 3.06% | | | 3.11% | |
In addition to the study above, the sponsor also evaluated day-to-day precision using control samples with 3 different levels of glucose. Three lots of test strips and 10 meters were used in the study, with 1 test performed on each meter per day for 10 days, (100 total tests divided between 3 strip lots per control level). The results for both meters are summarized below
Results of intermediate precision of FORA Diamond Prima
| Control solution levels | Low (30-50 mg/dL) | | | Normal (96-144 mg/dL) | | | High (280-420 mg/dL) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot | PR1101 13 | TR1007 16 | TR1008 04 | PR1101 13 | TR1007 16 | TR1008 04 | PR1101 13 | TR1007 16 | TR1008 04 |
| Mean | 42.1 | 41.5 | 42.2 | 111.8 | 114.0 | 112.1 | 367.1 | 377.0 | 359.4 |
| SD | 1.78 | 1.70 | 1.74 | 4.09 | 3.99 | 4.04 | 12.58 | 14.19 | 12.50 |
| CV | 4.22% | 4.08% | 4.12% | 3.66% | 3.50% | 3.61% | 3.43% | 3.76% | 3.48% |
| Overall mean | | 42.0 | | | 112.6 | | | 367.0 | |
| SD | | 1.74 | | | 4.11 | | | 14.84 | |
| CV | | 4.15% | | | 3.65% | | | 4.04% | |
Results of intermediate precision of FORA Diamond Mini
| Control solution levels | Low (30-50 mg/dL) | | | Normal (96-144 mg/dL) | | | High (280-420 mg/dL) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot | PR1101 13 | TR1007 16 | TR1008 04 | PR1101 13 | TR1007 16 | TR1008 04 | PR1101 13 | TR1007 16 | TR1008 04 |
| Mean | 44.0 | 43.6 | 44.2 | 122.7 | 122.9 | 123.1 | 338.8 | 330.1 | 338.1 |
{7}
| SD | 1.65 | 1.70 | 1.62 | 4.30 | 4.03 | 4.26 | 12.07 | 12.19 | 11.70 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| CV | 3.75% | 3.89% | 3.66% | 3.51% | 3.28% | 3.46% | 3.56% | 3.69% | 3.46% |
| Overall mean | | 43.9 | | | 122.9 | | | 335.9 | |
| SD | | 1.65 | | | 4.16 | | | 12.44 | |
| CV | | 3.76% | | | 3.39% | | | 3.70% | |
# b. Linearity/assay reportable range:
The sponsor performed linearity studies using adjusted venous blood samples with 10 different glucose concentrations ranging from 10-20, 21-50, 51-80, 81-120, 121-200, 201-300, 301-400, 401-500, 501-600 and $601 - 700\mathrm{mg / dL}$ for the FORA Diamond Prima Blood Glucose and FORA Diamond Mini Blood Glucose Monitoring System a For each concentration, 10 consecutive tests (with 5 measurements per lot) by YSI-2300 and FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System respectively. The resulting data was compared and the linear regression analyses were as follows:
FORA Diamond Prima Blood Glucose Monitoring System
| | Slope | Slope 95% CI | Intercept | Intercept 95% CI | R² |
| --- | --- | --- | --- | --- | --- |
| Lot 1 | 1.0110 | 0.9983 -1.0237 | 0.1371 | -4.2103 - 4.4845 | 0.9981 |
| Lot 2 | 1.0036 | 0.9900 - 1.0172 | 0.2676 | -4.3760 - 4.9112 | 0.9978 |
| Lot 3 | 1.0142 | 1.0012 - 1.0272 | -1.5864 | -6.0285 - 2.8557 | 0.9981 |
FORA Diamond Mini Blood Glucose Monitoring System
| | Slope | Slope 95% CI | Intercept | Intercept 95% CI | R² |
| --- | --- | --- | --- | --- | --- |
| Lot 1 | 1.0127 | 0.9992 - 1.0262 | 0.7012 | -3.9170 - 5.3194 | 0.9979 |
| Lot 2 | 1.0009 | 0.9905 - 1.0114 | -2.3044 | -5.8797 - 1.2709 | 0.9987 |
| Lot 3 | 1.0113 | 0.9967 - 1.0259 | 2.4061 | -2.5886 - 7.4008 | 0.9975 |
The measuring range of $20 - 600\mathrm{mg / dL}$ for both Blood Glucose Monitoring Systems and have been shown to be linear.
{8}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Test strips have been cleared under predicate k100322
d. Detection limit:
See linearity study
e. Analytical specificity:
The following protocol was applied to both FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System respectively.
The sponsor indicated that the interference study protocol was developed according to CLSI EP 7-A2. Venous blood was obtained from fasting subjects and collected in a heparin-Na (sodium heparin) vacutainer tube (Hct around $45\%$ ). Two concentrations of glucose were adjusted to $70\mathrm{mg/dL}$ (low) and $150\mathrm{mg/dL}$ (high) using the YSI 2300 as a reference instrument. The glucose samples were spiked with the potentially interfering compounds equivalent to the highest therapeutic dosage and toxic level (or ten times the highest therapeutic concentrations when toxic levels were not known), and tested on 3 lots of test strips. Bias was calculated as the mean percent difference in glucose reading between the test and control concentration groups. All samples tested showed $\%$ bias within $\pm 10\%$ between the test and the control groups. For both FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System, the sponsor claims no significant interference ( $\leq 10\%$ difference) for the substances and concentrations shown in the table below:
| Substance | Therapeutic / Physiologic Concentration Range (or Upper Limit) (mg/dL) | Concentration Tested (mg/dL) | Diamond Prima/Diamond mini |
| --- | --- | --- | --- |
| Acetylsalicylic Acid | 2 - 10 | 50 | No interference |
| Acyclovir | 0.23 - 0.31 | 3.1 | No interference |
| Allopurinol | 0.5 | 5 | No interference |
| Amitriptyline | 0.012 - 0.025 | 0.25 | No interference |
| Amoxicillin | 0.55 - 1.1 | 11 | No interference |
| Ampicillin | 0.5 | 5 | No interference |
| Aspirin (Salicylic Acid) | 10 - 30 | 60 | No interference |
| Atenolol | 0.1 - 0.2 | 10 | No interference |
| Bicarbonate | 244 (29 mM) | 336 (40 mM) | No interference |
| Bile Acids (Cholic Acid) | 0.7 | 6 | No interference |
{9}
| Caffeine | 0.3 - 1.5 | 10 | No interference |
| --- | --- | --- | --- |
| Calcium | 2.8 mM | 5 mM | No interference |
| Chloride | 108 mM | 140 mM | No interference |
| Cholesterol | 300 | 500 | No interference |
| Clonidine | 0.0001 - 0.0002 | 2 | No interference |
| Creatinine | 1.7 | 30 | No interference |
| Digoxin | 0.0001 - 0.00025 | 0.16 | No interference |
| Diphenhydramine | 0.01 - 0.1 | 1 | No interference |
| K2EDTA | 180 | 702 | No interference |
| K3EDTA | 175.5 | 702 | No interference |
| Enalapril | 0.012 - 0.015 | 0.15 | No interference |
| Ephedrine HCl | 1.8 | 50 | No interference |
| Erythromycin | 0.2 - 2.0 | 20 | No interference |
| Estrone | 0.0011 | 0.1 | No interference |
| Famotidine | 0.008 - 0.013 | 0.13 | No interference |
| Fluoxetine | 0.08 | 0.8 | No interference |
| Fructose | 7.5 | 1000 | No interference |
| Furosemide | 0.1 - 0.3 | 2 | No interference |
| Galactose | < 5 | 1000 | No interference |
| Gentisic Acid | 02 - 0.6 | 2 | No interference |
| Glyburide | 0.018 - 0.025 | 1.07 | No interference |
| Hemoglobin | 2.5 | 500 | No interference |
| Heparin (Li) | 35 - 100 U/dL | 6800 U/dL | No interference |
| Heparin (Na) | 35 - 100 U/dL | 6800 U/dL | No interference |
| Ibuprofen | 1 - 7 | 55 | No interference |
| Isomalt | N/A | 1000 | No interference |
| Lactose | < 0.5 | 1000 | No interference |
| Lactitol | N/A | 1000 | No interference |
| Lidocaine | 0.15 - 0.6 | 6 | No interference |
| Lipemic Samples (Triglycerides) | 3000 | 30 - 300 | No interference |
| Magnesium | 1.1 mM | 5 mM | No interference |
| Maltitol | N/A | 1000 | No interference |
| Maltose | N/A | 1000 | No interference |
| Metaproterenol | 0.00022 - 0.00130 | 1.81 | No interference |
| Metformin HCl | 0.5 - 4 | 50 | No interference |
| Metoprolol | 0.005 - 0.027 | 0.3 | No interference |
| Naproxen | 3-12 | 100 | No interference |
{10}
For both FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System the summary of drugs and concentrations in excess of $\pm 10\%$ bias is as follows:
| Substance | Limiting Concentration (mg/dL) | Therapeutic / Physiologic Concentration Range (or Upper Limit) (mg/dL) |
| --- | --- | --- |
| Acetaminophen | >5 | 0.45 - 3 |
| L-Dopa | >0.7 | 0.02 - 0.28 |
| Methyl-Dopa | >0.625 | 0.1 - 0.5 |
| Tolazamide | >6.25 | 1.6 |
| Mannose | >250 | 1.15 |
| Dopamine | >1.25 | 0.03 |
| Xylose | >6.25 | N/A |
| Pralidoxime Iodide | >5 | N/A |
| Methane | >5 | N/A |
| Methane | >250 | N/A |
| Methane | >1.25 | N/A |
| Methyldopa | >0.7 | 0.02 - 0.28 |
| Methyldopa | >0.625 | N/A |
| Methyldopa | >0.25 | N/A |
| Methyldopa | N/A | N/A |
{11}
The sponsor has the following limitations in their labeling of both FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System:
Xylose : Do not test blood glucose during or soon after an absorption test. Xylose can give falsely elevated results.
There is no significant interference $(\leq 10\%)$ in the presence of galactose, maltose, fructose or mannitol observed in blood glucose tests as demonstrated in studies up to $1,000~\mathrm{mg / dL}$
Lipemic Effects: Blood triglycerides up to $2000\mathrm{mg / dL}$ (22.8 mmol/L) do not affect the results significantly $(\leq 10\%)$ , but may affect results at higher levels.
f. Assay cut-off:
Not applicable
# 2. Comparison studies:
a. Method comparison with predicate device:
The sponsor conducted an accuracy study at three hospital sites. Trained professionals obtained 180 samples from diabetic and non-diabetic outpatient participants. The range of glucose values tested was $37 - 503\mathrm{mg / dL}$ for Diamond Prima and $37\sim 493\mathrm{mg / dL}$ for Diamond Mini BGMS. In order to obtain sufficient samples in the lowest and highest concentration intervals, 12 venous samples were obtained and either allowed to glycolyze to obtain low values or spiked with additional glucose. Samples that were $< 50\mathrm{mg / dL}$ and $>400\mathrm{mg / dL}$ were contrived samples and samples between 50 to $440\mathrm{mg / dL}$ were natural capillary samples from the fingertip.
Accuracy results for glucose concentration $< 75\mathrm{mg / dL}$
| | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| --- | --- | --- | --- |
| Diamond Prima | 15 / 20 (75.0 %) | 20 / 20 (100 %) | 20 / 20 (100 %) |
| Diamond Mini | 15 / 20 (75.0 %) | 20 / 20 (100 %) | 20 / 20 (100 %) |
Accuracy results for glucose concentration $\geq 75\mathrm{mg / dL}$
| | Within ± 5 % | Within ± 10 % | Within ± 15 % | Within ± 20 % |
| --- | --- | --- | --- | --- |
| Diamond Prima | 86 /160 (53.7 %) | 146 /160 (91.0 %) | 157 /160 (98.0 %) | 160 /160 (100 %) |
| Diamond | 71 /160 (44.3 | 148 /160 (92.5 | 158 /160 (99.0 | 160 /160 (100 |
{12}
| Mini | %) | %) | %) | %) |
| --- | --- | --- | --- | --- |
Linear regression for Diamond Prima (n= 180) is as follows:
$$
(y = 0.9766x + 5.7464, R^2 = 0.9785)
$$
Linear regression for Diamond Mini (n= 180) is as follows:
$$
(y = 0.9715x + 6.0535, R^2 = 0.9823)
$$
b. Matrix comparison:
Not applicable. Capillary whole blood is the only indicated sample matrix.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Consumer study:
A study was conducted to demonstrate that untrained lay users can correctly perform a glucose test using the DIAMOND Prima and DIAMOND Mini Blood Glucose Monitoring Systems and obtain accurate results. In all, 184 male and female subjects of varying demographics were recruited. Each subject was asked to perform a self test using only the provided English language labeling. A venous blood specimen from each subject was also obtained by a healthcare professional and tested on the YSI reference method.
The results are summarized in the tables below.
Difference distribution for glucose concentration $< 75\mathrm{mg/dL}$ of lay users versus YSI-2300
| | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15mg/dL |
| --- | --- | --- | --- |
| DIAMOND Prima | 12/23 (52.2 %) | 22/23 (95.7 %) | 23/23 (100 %) |
| DIAMOND Mini | 14/23 (60.9 %) | 21/23 (91.3 %) | 23/23 (100 %) |
Difference distribution for glucose concentration $\geq 75\mathrm{mg/dL}$
{13}
14
| | Within ± 5 % | Within ± 10 % | Within ± 15 % | Within ± 20 % |
| --- | --- | --- | --- | --- |
| DIAMOND Prima | 87/161 (54.0 %) | 137/161 (85.1 %) | 155/161 (96.3 %) | 161/161 (100 %) |
| DIAMOND Mini | 72/161 (44.7 %) | 127/161 (78.9 %) | 157/161 (97.5 %) | 161/161 (100 %) |
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
| Time of day | Normal plasma glucose range for people without diabetes |
| --- | --- |
| Fasting and before meal | Less than 100 mg/dL (5.6 mmol/L) |
| 2 hours after meals | Less than 140 mg/dL (7.8 mmol/L) |
Source: American Diabetes Association (2010). Clinical Practice Recommendations. Diabetes Care, 33 (Supplement 1): S1-100.
N. Instrument Name:
FORA Diamond Prima Blood Glucose Meter and
FORA Diamond Mini Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and requires 0.5 uL sample volume of capillary blood. Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☑
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☑ or No ☐
4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood from the finger only. The whole blood sample is applied directly to the test strip.
5. Calibration:
The calibration code for the vial of test strips should be selected or verified by the user from the available choices of code numbers programmed in the meter. Users are instructed where to find the calibration code information on the test strip vial label.
6. Quality Control:
The sponsor states that the system can be used only with FORA control solutions
{14}
(cleared under k093724). Recommendations on when to test the control materials are provided in the labeling. An acceptable range for each control level is printed on the test strip vial label. The user is cautioned not to use the meter if the control result falls outside these ranges
## P. Other Supportive Device and Instrument Information:
### 1. Hematocrit Study
The effect of different hematocrit levels was evaluated with one test strip lot and 10 FORA Diamond Prima Blood Glucose meters and 10 FORA Diamond Mini Blood Glucose Monitoring meters respectively. In both cases, blood samples at six glucose concentrations (20-30, 50-80, 100-150, 200-250, 350-400 and 550-600 mg/dL) were analyzed. The glucose samples were prepared from venous blood at 10 different hematocrit levels (20, 25, 30, 35, 40, 45, 50, 55, 60 and 65%). Glucose results for each concentration and hematocrit level were compared to samples tested on the YSI reference method.
The data supports the sponsor's claim that hematocrit in the range of 20%-60% does not significantly interfere (more than ±15%) with glucose measurements using both test system.
### 2. Altitude:
A study was conducted to evaluate the effect of altitude up to 15,000 feet on performance of the FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System using a glove box system which simulates 4 elevations from sea level to 15,000 feet (4500 meters). Venous whole blood samples at five glucose concentrations ranging from 50-600 mg/dL were tested at four altitudes using three lots of test strips and 4 FORA Diamond Prima Blood Glucose meters and FORA Diamond Mini Blood Glucose meters respectively. Each sample was also evaluated by the YSI method. At altitudes up to 15,000 feet, test results were within +/-10% of YSI values for both meters.
### 3. Test system operating conditions:
Studies were performed using four FORA Diamond Prima Blood Glucose meters and four FORA Diamond Mini Blood Glucose meters respectively with three lots of test strips, and three venous whole blood samples with glucose concentrations of 64.2, 124 and 318 mg/dL. Testing was performed at various conditions in the claimed conditions 50-104°F (10-40°C) and at a relative humidity from 20-85% and results compared to the reference YSI. There were no significant differences in glucose concentrations across the temperature and humidity ranges tested. Results demonstrated that the test system can be used at temperatures from 50-104°F (10-40°C) and at a relative humidity from 20-85%.
### 4. User performance study:
For the user performance study summarized under consumer study above, the participants were asked to complete a questionnaire to evaluate the ease of use of the device and the clarity of the English language labeling. Overall the users indicated that they could successfully perform the test and that the user manual was written clearly.
### 4. Readability assessment:
{15}
The sponsor performed a readability assessment of the labeling and states that the user manual, strip insert, and control insert are written at 8th grade level or below based on SMOG analysis.
## 5. EMC testing
EMC testing was evaluated and certified by Electronics Testing Center Taiwan and a letter of attestation was issued and attached to file.
## 6. Sample volume study
A sample volume study was performed to verify the test strip sample volume requirement of 0.5 µL for both FORA Diamond Prima Blood Glucose Monitoring System and FORA Diamond Mini Blood Glucose Monitoring System. Three samples with glucose concentrations approximately 50, 150, and 350 mg/dL were evaluated with three lots of test strips with 5 different FORA Diamond Prima and FORA Diamond Mini meters respectively. Each of the three blood samples was applied to the strips at samples volumes 0.4, to 1.5 µL. Protocols and acceptance criteria were provided and found to be acceptable. The sponsor concluded that sample volumes ≥ 0.5 µL produced accurate results and sample volumes < 0.5 µL produced inaccurate results for both meters. The labeling provides instructions and graphics to assist the user in obtaining and applying an adequate sample volume in both cases.
## 7. Infection Control
The devices are intended for single-patient use only. Disinfection studies were performed on these meters by an outside commercial testing company to determine the robustness of the meter to the recommended cleaning and disinfection protocol, and its effectiveness in preventing the spread of blood borne pathogens, particularly Duck hepatitis B virus (HBV). Micro-Kill Plus™ disposable wipes (EPA Reg. No: 59894-10-37549) were validated, demonstrating complete inactivation of live virus for use with the meter. The sponsor also demonstrated that there was no change in performance or in the external materials of the meter and lancing device after 5,000 cleaning and disinfection cycles designed to simulate 5 years of use by lay users. Labeling has been reviewed for adequate instructions in validated cleaning and disinfection procedures.
## 8. Software validation
Software validation verification has been reviewed and the information the sponsor provided was deemed to be sufficient.
## Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.