ANDON HEALTH CARE MANAGEMENT SYSTEM SOFTWARE

K112738 · Andon Medical Co., Ltd. · NBW · Sep 30, 2011 · Clinical Chemistry

Device Facts

Record IDK112738
Device NameANDON HEALTH CARE MANAGEMENT SYSTEM SOFTWARE
ApplicantAndon Medical Co., Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateSep 30, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

Andon Health Care Management System Software is an optional software accessory for use with the Andon blood glucose meters with data management capacities, When used with one of these meters, Andon Health Care Management System Software transfers data from the device's memory into a computer for enhanced data management. Andon Health Care Management System Software is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating historical blood glucose test results and other biological statistics to support diabetes management. Andon Heath Care System Software is not intended to provide treatment decisions or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional.

Device Story

Andon Health Care Management System Software; manages data from blood glucose meters; modification adds compatibility with AG-6081 and AG-6951 blood glucose meters (k101307); software collects/processes glucose data; used by patients/clinicians for health management; fundamental scientific technology remains unchanged from predicate.

Clinical Evidence

Bench testing only; verification and validation activities performed per design control requirements; risk analysis conducted to assess modification impact; results demonstrated predetermined acceptance criteria were met.

Technological Characteristics

Software-based health management system; interfaces with blood glucose meters (AG-6081, AG-6951); data management functionality; design control procedures per 21 CFR 820.30.

Indications for Use

Indicated for people with diabetes and their healthcare professionals in home and clinical settings to upload, store, analyze, and communicate historical blood glucose test results and biological statistics to support diabetes management.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## SEP 3 0 2011 # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92. ### 1.0 submitter's information | Name: | Andon Medical Co., Ltd. | |----------------------|--------------------------------------------------------------| | Address: | No.26 HangYu Road, Tianjin Airport Economic Area,<br>TIANJIN | | Phone number: | 86-22-8761 2426 | | Fax number: | 86-22-6052 6162 | | Contact: | Yi Liu | | Date of Application: | 9/10/2011 | ### 2.0 Device information Trade name: Andon Health Care Management System Software Common name: Data management software Classification name: Data management software ### 3.0 Classification Production code: NBW- Blood Glucose Monitoring System. Regulation number: 862.1345 Classification: II Panel: Clinical Chemistry ### 4.0 Predict device information Manufacturer: Andon Medical Co., Ltd. Andon Health Care Management System Software Device: 510(k) number: k102678 ### 5.0 Device description Andon Health Care Management System Software is an optional software accessory for use with AG-608 Single Blood Glucose Monitoring System and AG-608 Multi Blood Glucose Monitoring System, AG-6081 Single Blood Glucose Monitoring System and AG-6081 Multi Blood Glucose Monitoring System, AG-6951Single Blood Glucose Monitoring System and AG-6951 Multi Blood Glucose Monitoring System. When used with the above Blood Glucose Monitoring Systems, Andon Health Care {1}------------------------------------------------ Management System Software transfers data from the device's memory into a computer for enhanced data management. ### 6.0 Intended use Andon Health Care Management System Software is an optional software accessory for use with the Andon blood glucose meters with data management capacities, When used with one of these meters, Andon Health Care Management System Software transfers data from the device's memory into a computer for enhanced data management. Andon Health Care Management System Software is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating historical blood glucose test results and other biological statistics to support diabetes management. Andon Heath Care System Software is not intended to provide treatment decisions or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional. | item | Andon Health Care<br>Management<br>Software | Andon Health Care<br>Management<br>Software(K102678) | | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for use | The ANDON Health Care<br>Management System<br>Software is intended for use<br>in<br>home and clinical settings as<br>an aid for people with users<br>and their health<br>care professionals to review,<br>analyze and evaluate the<br>historical test results<br>to support health<br>management effectively. | The ANDON Health Care<br>Management System<br>Software is intended for use<br>in<br>home and clinical settings as<br>an aid for people with users<br>and their health<br>care professionals to review,<br>analyze and evaluate the<br>historical test results<br>to support health<br>management effectively. | | | | Installation method | Exe file | Exe file | | | Package Contents | N/A | N/A | | | Capable of deleting results | Delete all results in meter | Delete all results in meter | | | Language capabilities | English, Spanish | English, Spanish | | | Customizable schedule | N/A | N/A | | | Types of graphs etc. | Coordinates graph | Coordinates graph | | | | | | | | Auto-detect COM port | Yes | Yes | | | System components | PC, USB cable, meter | PC, USB cable, meter | | | Software platform | Microsoft | Microsoft | | | Hardware requirements | CPU: optimal at 1,2 GHz+<br>Main memory: optimal at 256 MB+ RAM<br>Disk space: optimal 200 MB+ free space - at least 100 MB<br>Graphic resolution starting from 1024 x 768, CD-ROM drive, USB interface | CPU: optimal at 1,2 GHz+<br>Main memory: optimal at 256 MB+ RAM<br>Disk space: optimal 200 MB+ free space - at least 100 MB<br>Graphic resolution starting from 1024 x 768, CD-ROM drive, USB interface | | | Technology | Visual Basic | Visual Basic | | | Performance specifications, including any testing | Read memories in meter.<br>Delete all memories in meter<br>Set time to meter<br>Draw table and graph<br>Print<br>Set the personal information | Delete all memories in meter<br>Set time to meter<br>Draw table and graph<br>Print<br>Set the personal information | | Meter Compatibility | AG-608 Single Blood Glucose Monitoring System<br>and AG-608 Multi Blood Glucose Monitoring System,<br>AG-6081 Single Blood Glucose Monitoring System<br>and AG-6081 Multi Blood Glucose Monitoring System,<br>AG-6951Single Blood Glucose Monitoring System<br>and AG-6951 Multi Blood Glucose Monitoring System | AG-608 Single Blood Glucose Monitoring System<br>and AG-608 Multi Blood Glucose Monitoring System | | ### 7.0 Summary comparing technological characteristics with predicate device {2}------------------------------------------------ 5- 3 {3}------------------------------------------------ ### 8.0 Performance summary Testing of Andon Health Care Management System Software included system test and unit test. ### 9.0 Comparison to the predict device and the conclusion Andon Health Care Management System Software is very similar with the predicted device Andon Health Care Management System Software(K102678), However, they are intended to use together with 4 additional BGMS: AG-6081 Single Blood Glucose Monitoring System and AG-6081 Multi Blood Glucose Monitoring System, AG-6951Single Blood Glucose Monitoring System and AG-6951 Multi Blood Glucose Monitoring System. However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo with a stylized bird in the center. The bird is depicted with abstract lines, suggesting movement or flight. Encircling the bird is text arranged in a circular pattern, likely the name of an organization or department. The text is in all caps and appears to be a government agency. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 Andon Medical Co., Ltd. c/o Yi Liu President No.26 HangYu Rd., Tianjin Airport Economic Area Tianjin, 300381 CH - CHINA SEP 30 2011 Re: k112738 Trade/Device Name: Andon Health Care Management System Software Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system. Regulatory Class: 11 Product Code: NBW, JQP Dated: September 20. 2011 Received: September 20. 2011 Dear: Mr. Liu: We have reviewed your Section 310(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ Page 2 - If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, signature Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indication for Use #### K112738 510(k) Number (if known): ### Device Name: Andon Health Care Management System Software ### Indication For Use: Andon Health Care Management System Software is an optional software accessory for use with the Andon blood glucose meters with data management capacities, When used with one of these meters, Andon Health Care Management System Software transfers data from the device's memory into a computer for enhanced data management. Andon Health Care Management System Software is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating historical blood glucose test results and other biological statistics to support diabetes management. Andon Heath Care System Software is not intended to provide treatment decisions or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional. Prescription Use Yes (21 CFR Part 801 Subpart D) And/Or Over the Counter Use Yes (21 CFR Part 801 Subpart C) Page 1 of 1 (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Ruttle Charles Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 112738 510(k)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...