ANDON HEALTH CARE MANAGEMENT SYSTEM SOFTWARE, AG-608 SINGLE & AG-608 MULTI BLOOD GLUCOSE MONITORING SYSTEMS

K102678 · Andon Medical Co., Ltd. · NBW · Aug 26, 2011 · Clinical Chemistry

Device Facts

Record IDK102678
Device NameANDON HEALTH CARE MANAGEMENT SYSTEM SOFTWARE, AG-608 SINGLE & AG-608 MULTI BLOOD GLUCOSE MONITORING SYSTEMS
ApplicantAndon Medical Co., Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateAug 26, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Intended Use

Andon Health Care Management System Software is an optional software accessory for use with the Andon blood glucose meters with data management capacities. When used with one of these meters, Andon Health Care Management System Software transfers data from the device’s memory into a computer for enhanced data management. Andon Health Care Management System Software is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating about historical blood glucose test results and other biological statistics to support diabetes management. Andon Health Care Management System Software is not intended to provide treatment decisions or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional. The AG-608 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The AG-608 Single Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The AG-608 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-608 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. The AGS-1000 Single Blood Glucose Test Strips are for use with the AG-608 Single Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample. The AG-608 Multi Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The AG-608 MULTI Blood Glucose Monitoring System is intended to be used by healthcare professionals for multiple patients in a professional healthcare setting as an aid in monitoring the effectiveness of diabetes control. The AG-608 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. The AGS-1000 MULTI Blood Glucose Test Strips are for use with the AG-608 MULTI Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample. This system should only be used with single-use, auto-disabling lancing devices.

Device Story

System consists of Andon Health Care Management System Software and AG-608 blood glucose meters (Single/Multi). Meter inputs capillary whole blood via test strips; stores glucose results. Software connects to meter via USB cable; downloads stored data to PC. Software displays results in tabular/graphical formats for trend analysis; allows time setting and memory deletion. Used in home/clinical settings by patients or healthcare professionals. Output supports diabetes management communication; does not provide treatment/dosing recommendations. Infection control validated for single-patient (AG-608 Single) and multi-patient (AG-608 Multi) use via CaviWipes disinfection.

Clinical Evidence

No clinical trials performed. Evidence includes bench testing for data transfer accuracy (full memory download/rollover), usability study (n=40: 20 lay-users, 20 professionals), readability study (SMOG score ≤ 8th grade), and infection control/robustness studies (11,000 cleaning/disinfection cycles).

Technological Characteristics

Software-based data management system; Visual Basic programming; PC-based platform; USB connectivity; compatible with AG-608 series blood glucose meters; supports English and Spanish languages; generates coordinate graphs and tables.

Indications for Use

Indicated for people with diabetes and their healthcare professionals in home and clinical settings to upload, store, analyze, and communicate historical blood glucose test results and biological statistics to support diabetes management. Not for treatment decisions or as a substitute for professional judgment.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102678 # 510(k) Summary AUG 26 2011 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92. # 1.0 submitter's information Andon Medical Co., Ltd. Name: No.04-23-3 AIRPORT INDUSTRIAL PARK, TIANJIN Address: Phone number: 86-22-8761 2426 Fax number: 86-22-6052 6162 Contact: Yi Liu Date of Application: 9/10/2010 # 2.0 Device information | Trade name: | Andon Health Care Management System Software | |----------------------|----------------------------------------------| | Common name: | Data management software | | Classification name: | Data management software | #### 3.0 Classification Production code: NBW- Blood Glucose Monitoring System. Requlation number: 862.1345 Classification: II Panel: Clinical Chemistry # 4.0 Predict device information Manufacturer: Taidoc Technology Corporation Device: Clever Chek Health Care System Software 510(k) number: k070941 # 5.0 Device description Andon Health Care Management System Software is an optional software accessory for use with AG-608 Single Blood Glucose Monitoring System and AG-608 Multi Blood Glucose Monitoring System. When used with AG-608 Single Blood Glucose Monitoring System and AG-608 Multi Blood Glucose Monitoring System, Andon Health Care Management System Software transfers data from the device's memory into a computer for enhanced data management. {1}------------------------------------------------ #### 6.0 Intended use Andon Health Care Management System Software is an optional software accessory for use with the Andon blood glucose meters with data management capacities, When used with one of these meters, Andon Health Care Management System Software transfers data from the device's memory into a computer for enhanced data management. Andon Health Care Management System Software is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating historical blood glucose test results and other biological statistics to support diabetes management. Andon Heath Care System Software is not intended to provide treatment decisions or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional. The AG-608 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The AG-608 MULTI Blood Glucose Monitoring System is intended to be used by healthcare professionals for multiple patients in a professional healthcare setting as an aid in monitoring the effectiveness of diabetes control. The AG-608 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. The AGS-1000 MULTI Blood Glucose Test Strips are for use with the AG-608 MULT1 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample. This system should only be used with single-use, auto-disabling lancing devices. The AG-608 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The AG-608 Single Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. {2}------------------------------------------------ The AG-608 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-608 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. The AGS-1000 Single Blood Glucose Test Strips are for use with the AG-608 Single Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample. | item | Andon | Clever | |-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for use | The ANDON Health Care<br>Management System<br>Software is intended for use<br>in<br>home and clinical settings as<br>an aid for people with users<br>and their health<br>care professionals to review,<br>analyze and evaluate the<br>historical test results<br>to support health<br>management effectively. | The CLEVER CHEK Health<br>Care System Software is<br>intended for use in<br>home and clinical settings as<br>an aid for people with users<br>and their health<br>care professionals to review,<br>analyze and evaluate the<br>historical test results<br>to support health<br>management effectively. | | Installation method | Exe file | Exe file | | Package Contents | N/A | N/A | | Capable of deleting results | Delete all results in meter | Delete all results in meter | | Language capabilities | English, Spanish | English, Spanish, Traditional<br>Chinese | | Customizable schedule | N/A | N/A | | Types of graphs etc. | Coordinates graph | Coordinates graph | | Auto-detect COM port | Yes | Yes | | System components | PC, USB cable, meter | PC, USB cable, meter | | Software platform | Microsoft | Windows XP, Windows<br>Vista, Windows 7 | | Hardware requirements | CPU: optimal at 1,2 GHz+ | Personal computer with 400<br>megahertz (MHz) or higher | # 7.0 Summary comparing technological characteristics with predicate device {3}------------------------------------------------ | | 256 MB+ RAM | processor clock | |------------------------------------------------------|-------------------------------|------------------------------| | | Disk space: optimal 200 | speed recommended; the | | | MB+ free space - at least 100 | software does not run on | | | MB | Apple computers. | | | Graphic resolution starting | Random access memory | | | from 1024 x 768, CD-ROM | (RAM) of 64 megabytes | | | drive, USB interface | (MB) or more | | | | Available hard disk space of | | | | 30 MB for running the | | | | program. | | | | Monitor with 1024 x 768 or | | | | higher resolution. | | Technology | Visual Basic | Visual Basic | | Performance specifications,<br>including any testing | Read memories in meter. | Read memories in meter. | | | Delete all memories in meter | Delete all memories in meter | | | Set time to meter | Set time to meter | | | Draw table and graph | Draw table and graph | | | Print | Print | | | Set the personal information | Set the personal information | . . {4}------------------------------------------------ #### 8.0 Performance summary Testing of Andon Health Care Management System Software included system test and unit test. #### 9.0 Comparison to the predict device and the conclusion Andon Health Care Management System Software is very similar with the predicted device Clever Chek Health Care System Software, However, they are intended to use together with the different meters, their port are also different. However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness. {5}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Andon Medical Co., Ltd c/o Mr. Yi Liu 1st Floor Building C No.3 Jin Ping Street Nankai District Tianjin, China 300190 Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 AUG 2 6 2001 Re: k102678 R Trade name: Andon Health Care Management System Software AG-608 Single Blood Glucose Monitoring System AG-608 Multi Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, JQP Dated: August 2, 2010 Received: August 2, 2010 Dear Mr. Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). Image /page/5/Picture/14 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, facing to the right. The eagle is composed of three curved lines that suggest feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle. The text is in all caps and is evenly spaced around the circle. {6}------------------------------------------------ Page 2 - If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, C.C. Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ Andon Health Care Management System Software FDA 510(k) Files # Indication for Use 510(k) Number (if known): Device Name: Andon Health Care Management System Software #### Indication For Use: Andon Health Care Management System Software is an optional software accessory for use with the Andon blood glucose meters with data management capacities, When used with one of these meters, Andon Health Care Management System Software transfers data from the device's memory into a computer for enhanced data management. Andon Health Care Management System Software is intended for use in home and clinical settings via the internet to assist people with diabetes and their healthcare professionals in uploading, storing, analyzing, and communicating historical blood glucose test results and other biological statistics to support diabetes management. Andon Heath Care System Software is not intended to provide treatment decisions or to be used as a substitute for professional healthcare judgment. All patient medical diagnoses and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional. Prescription Use Yes (21 CFR Part 801 Subpart D) Over the Counter Use Yes (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) And/Or Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Rute Chesler Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) k102678 Page 1 of 3 {8}------------------------------------------------ AG-608 Blood Glucose Monitoring System FDA 510(k) Files # Indication for Use 510(k) Number (if known): # Device Name: AG-608 MULTI Blood Glucose Monitoring System ### Indication For Use: The AG-608 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of qlucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The AG-608 MULTI Blood Glucose Monitoring System is intended to be used by healthcare professionals for multiple patients in a professional healthcare setting as an aid in monitoring the effectiveness of diabetes control. The AG-608 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. The AGS-1000 MULTI Blood Glucose Test Strips are for use with the AG-608 MULTI Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample. This system should only be used with single-use, auto-disabling lancing devices. Prescription Use J (21 CFR Part 801 Subpart D) Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) And/Or Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Autle C. Hesler Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K102678 Page 2 of 3 {9}------------------------------------------------ AG-608 Blood Glucose Monitoring System FDA 510(k) Files # Indication for Use K102678 510(k) Number (if known): Device Name: AG-608 Single Blood Glucose Monitoring System #### Indication For Use: The AG-608 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The AG-608 Single Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The AG-608 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-608 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. The AGS-1000 Single Blood Glucose Test Strips are for use with the AG-608 Single Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample. And/Or Prescription Use (21 CFR Part 801 Subpart D) Over the Counter Use √ (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Rute Chaler Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) Page 3 of 3
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...