FORE V10 BLOOD GLUCOSE MONITORING SYSTEM, MODEL TD-4244, FOREA V12 BLODO GLUCOSE MONITORING SYSTEM, MODEL TD-4247

K090404 · Taidoc Technology Corporation · NBW · Jun 23, 2009 · Clinical Chemistry

Device Facts

Record IDK090404
Device NameFORE V10 BLOOD GLUCOSE MONITORING SYSTEM, MODEL TD-4244, FOREA V12 BLODO GLUCOSE MONITORING SYSTEM, MODEL TD-4247
ApplicantTaidoc Technology Corporation
Product CodeNBW · Clinical Chemistry
Decision DateJun 23, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2

Device Story

Clever Chek TD-4227 Blood Glucose Monitoring System; modified version of previously cleared device (k072784). Modification involves integrated circuit upgrade: increased bit size CPU, increased ROM, increased RAM, and increased ADC channels. Device measures blood glucose levels; intended for use by patients for self-monitoring. Fundamental scientific technology remains unchanged from predicate. Design control activities, including risk analysis and verification/validation, performed to confirm safety and performance post-modification.

Clinical Evidence

Bench testing only; verification and validation activities performed per design control requirements to confirm that modifications to the integrated circuit did not adversely affect device performance.

Technological Characteristics

Blood glucose monitoring system; utilizes electrochemical sensing. Modification involves upgraded integrated circuit: increased bit size CPU, increased ROM, increased RAM, and increased ADC channels. Device is a modification of a previously cleared system; fundamental scientific technology remains unchanged.

Indications for Use

Indicated for blood glucose monitoring in patients with diabetes.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIVD Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER k090404 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) Taidoc Technology Corporation, Clever Chek TD-4227 Blood Glucose Monitoring System - k072784 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for: A. Change in an integrated circuit. The change upgrades several aspects of the device such as: - Increased bit size CPU - Increased ROM - Increased RAM - Increased channels for the ADC (analog to digital converter) Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and analytes. 4. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 5. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...