IMPLANT INTEGRATION SYSTEMS DRILLS AND TAPS

K962394 · Implant Integration Systems, Inc. · DZE · Oct 11, 1996 · Dental

Device Facts

Record IDK962394
Device NameIMPLANT INTEGRATION SYSTEMS DRILLS AND TAPS
ApplicantImplant Integration Systems, Inc.
Product CodeDZE · Dental
Decision DateOct 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2

Intended Use

The Implant Integration Systems Surgical Drills are intended to be used in the placement of the “Dual Integrator” Dental Implant (K944225).

Device Story

Surgical drills used for site preparation during dental implant placement; specifically for Dual Integrator Dental Implant (K944225). Operated by dental surgeons in clinical settings. Drills facilitate bone preparation to receive implant. Benefit: enables precise surgical site creation for implant integration.

Clinical Evidence

No clinical data.

Technological Characteristics

Titanium nitride coated stainless steel surgical drills. Form factor: dental drill bits for bone preparation.

Indications for Use

Indicated for use in the surgical placement of the Dual Integrator dental implant system.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K962394 Implant Integration Systems, Inc. 6161 Clark Road, Suite 8 • Paradise, CA 95969 USA • (916) 872-1020 • Fax: (916) 877-7555 OCT 11 1996 Summary of Safety and Effectiveness Information The Implant Integration Systems Surgical Drills are intended to be used in the placement of the “Dual Integrator” Dental Implant (K944225). This product is similar in design, composition (titanium nitride coated stainless steel) and function to the Implant Innovations, Inc. Surgical Drills (K891615), the Impla-Med Corp. Titanium/Stainless Steel Dental Implant Drills (K902158) and the Steri-oss Drills and Taps (K915733). ![img-0.jpeg](img-0.jpeg) 6-19-96 )
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...